K Number
K052546

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-02-03

(140 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AEGIS Anterior Lumbar Plate System is indicated for use as an anteriorly placed supplemental fixation device via the lateral or anterolateral surgical approach above the bifurcation of the great vessel or via the anterior surgical approach, below the bifurcation of the great vessels.

The device is intended as a temporary fixation device until fusion is achieved. The AEGIS Anterior Lumbar Plate System is intended for anterior lumbar (L1 - S1) fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Device Description

The AEGIS Anterior Lumbar Plate System consists of an assortment of plates and screws.

The AEGIS Anterior Lumbar Plate System also contains Class 1 manual surgical instruments and cases that are considered exempt from premarket notification.

AI/ML Overview

This document describes the regulatory submission for the AEGIS Anterior Lumbar Plate System, a medical device. It does not contain information about a study proving the device meets acceptance criteria in the context of AI/ML performance.

The provided text is a 510(k) summary for a spinal implant system, which focuses on device description, intended use, materials, and substantial equivalence to predicate devices, rather than performance as would be measured for AI/ML systems. Therefore, most of the requested information regarding AI/ML study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies is not applicable to this document.

However, I can extract the information provided about the performance data and the type of evaluation conducted for this medical device:

1. A table of acceptance criteria and the reported device performance:

The document states: "Performance data were submitted to characterize the AEGIS Anterior Lumbar Plate System components."

Without the actual performance data or explicit acceptance criteria from the 510(k) submission, a detailed table cannot be created. For a mechanical implant like this, "acceptance criteria" generally refer to specific mechanical testing standards (e.g., fatigue strength, torsional rigidity) that the device must meet, and "reported device performance" would be the results of those tests. The 510(k) process primarily relies on demonstrating substantial equivalence to predicate devices, often through mechanical testing to show similar performance characteristics.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

Not applicable. This is a medical device (spinal implant) approval, not an AI/ML diagnostic or predictive device. The "test set" in this context would refer to mechanical test specimens, and not a dataset of patient information. The document does not specify the number of test specimens or the provenance of the material testing data, but it is implied to be laboratory-based testing of the device components.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not applicable. "Ground truth" in the context of this device's approval would relate to established engineering principles, material science standards, and the performance of predicate devices, not expert consensus on diagnostic interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. Adjudication methods are relevant for subjective evaluations, typically in clinical studies or AI/ML ground truth establishment, neither of which are detailed in this 510(k) summary for a mechanical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI-assisted diagnostic or therapeutic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For a mechanical device like this, the "ground truth" for demonstrating performance usually comes from:

  • Mechanical testing standards: Compliance with established ASTM or ISO standards for spinal implants (e.g., fatigue, static strength).
  • Material properties: Verification that the materials used (ASTM F-136 implant grade titanium alloy) meet published specifications.
  • Predicate device comparison: Performance is often compared against legally marketed predicate devices to establish substantial equivalence.

8. The sample size for the training set:

Not applicable. This is not an AI/ML device requiring a training set.

9. How the ground truth for the training set was established:

Not applicable. This is not an AI/ML device requiring a training set.

{0}------------------------------------------------

AEGIS Anterior Lumbar Plate System

510(k) Summary IX.

? ੋਵਿੰਗ ਤੇ

SUBMITTER: DePuy Spine, Inc. 325 Paramount Drive Raynham, MA 02780

  • CONTACT PERSON: Mary Gray Phone: (508) 828-3649 Fax: (508) 828-3797
  • DATE PREPARED: January 12, 2006
  • CLASSIFICATION NAME: Spinal Intervertebral Body Fixation Orthosis 8888.3060
  • AEGIS Anterior Lumbar Plate System PROPRIETARY NAME:
  • DePuy AcroMed M-2 Anterior Plate System, K972718 PREDICATE DEVICES: Medtronic Sofamor Danek PYRAMID Anterior Plate Fixation System, K013665
  • DEVICE DESCRIPTION: The AEGIS Anterior Lumbar Plate System consists of an assortment of plates and screws.

The AEGIS Anterior Lumbar Plate System also contains Class 1 manual surgical instruments and cases that are considered exempt from premarket notification.

  • INTENDED USE: The indications for use for the devices described in this submission are as follows:
    The AEGIS Anterior Lumbar Plate System is indicated for use as an anteriorly placed supplemental fixation device via the lateral or anterolateral surgical approach above the bifurcation of the great vessel or via the anterior surgical approach, below the bifurcation of the great vessels.

{1}------------------------------------------------

AEGIS Anterior Lumbar Plate System

The device is intended as a temporary fixation device until fusion is achieved. The AEGIS Anterior Lumbar Plate System is intended for anterior lumbar (L1 - S1) fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

MATERIALS:Manufactured from ASTM F-136 implant grade titanium alloy.
------------------------------------------------------------------------

PERFORMANCE DATA:

Performance data were submitted to characterize the AEGIS Anterior Lumbar Plate System components.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a stylized emblem. The emblem consists of a triple-line design, resembling a stylized human form, which is a common symbol associated with HHS.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 3 2006

Ms. Mary Gray Senior Regulatory Affairs Associate Depuy Spine 325 Paramount Drive Raynham, MA 02767

Re: K052546

Trade Name: AEGIS Anterior Lumbar Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation System Regulatory Class: II Product Code: KWQ Dated: January 12, 2006 Received: January 13, 2006

Dear Ms. Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have letermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require accordailed with the provisions of the Pourmay, therefore, market the device, subject to approval of a provisions of the Act. The general controls provisions of the Act include the general construction, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be rryour do nec to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I oter a requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Noah Bartsch

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

A Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K052546

AEGIS Anterior Lumbar Plate System Device Name:

Indications For Use:

The AEGIS Anterior Lumbar Plate System is indicated for use as an anteriorly placed supplemental fixation device via the lateral or anterolateral surgical approach above the bifurcation of the great vessel or via the anterior surgical approach, below the bifurcation of the great vessels.

The device is intended as a temporary fixation device until fusion is achieved . The AEGIS Anterior Lumbar Plate System is intended for anterior lumbar (1.1 - S1) fixation for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with dcgcneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEIEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

2

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of 1

510(k) Number K052546

N/A