(16 days)
The E-2320 PA S is intended to be used in displaying and viewing digital images for review by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.
E-2320 PA S is a display for medical viewing. It consists of 2 components: E-2320 PA S is a 20.1″ grayscale display. NioWatch is user-friendly software that allows to optimize the display for DICOM-compliant viewing.
The provided document is a 510(k) summary for the Barco E-2320 PA S medical flat panel display. This type of regulatory submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with specific acceptance criteria and performance data for the device itself.
Therefore, the document does not contain the detailed information required to fill out a table of acceptance criteria and reported device performance in the way you've requested for a device that performs a diagnostic task. The study described is a comparison against a predicate device to establish substantial equivalence.
Here's an assessment based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not specify quantitative acceptance criteria or detailed performance metrics for the E-2320 PA S display in a format that would typically be found for a diagnostic algorithm. The acceptance criteria for this submission are related to demonstrating substantial equivalence to the predicate device, K051902 (Barco E 2621), primarily based on technical characteristics, general function, application, and intended use.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable. This is a display device, not a diagnostic algorithm that processes a test set of medical images for performance evaluation in the traditional sense. The "study" here is a technical comparison for regulatory purposes.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. Ground truth is not established for image interpretation by clinicians in this submission. The "ground truth" for substantial equivalence would be the technical specifications and performance of the predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a display device, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable in the context of image interpretation. The "ground truth" for substantial equivalence is the technical specifications and performance data of the predicate device (Barco E 2621).
8. The sample size for the training set
Not applicable. This device does not use a "training set" in the context of machine learning.
9. How the ground truth for the training set was established
Not applicable.
Summary of the "study" for this device:
The "study" presented in this 510(k) submission is a declaration of substantial equivalence to an existing legally marketed device (predicate device).
- Predicate Device: Barco E 2621 (510(k) number: K051902)
- Comparison: The E-2320 PA S is compared to the E 2621.
- Key Finding: "The new and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application and intended use. Any difference between both devices does not affect safety or efficacy."
- Differences noted: The E-2320 PA S has a different LCD panel with a smaller screen size compared to the predicate device. The NioWatch software and other components are the same.
- Conclusion: The FDA reviewed the information and determined the device is substantially equivalent to the predicate device.
To reiterate, this document describes a regulatory pathway for a display monitor, not a diagnostic algorithm. Therefore, the detailed performance metrics and study designs typically associated with AI or diagnostic device evaluations are not present.
{0}------------------------------------------------
K05229
510(k) Summary In accordance with 21 CFR 807.92
1. Date of preparation
September 09, 2005
SEP 3 0 2005
2. Company information
BarcoView 35 President Kennedypark B-8500 Kortrijk, Belgium Tel. +32-(0)56-233-211 Fax +32-(0)56-233-457
3. Contact person
Lieven De Wandel Official correspondent
4. Device information
- Trade name: E-2320 PA S .
- . Common name: Medical flat panel display
- . Classification name: System, Image Processing
- . Classification number: 21 CFR 892.2050 / Procode 90LLZ
5. Predicate device
- . Name: E 2621
- . 510(k) number: K051902
- . Manufacturer: Barco NV
6. Device description
E-2320 PA S is a display for medical viewing. It consists of 2 components: E-2320 PA S is a 20.1″ grayscale display. NioWatch is user-friendly software that allows to optimize the display for DICOM-compliant viewing.
7. Intended use
"The E-2320 PA S is intended to be used in displaying and viewing digital images for review by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.
{1}------------------------------------------------
8. Summary of technological characteristics
The device consists of two components:
- One 2-megapixel flat panel display (E-2320 PA S) .
- . NioWatch software
The flat panel display has a resolution of 1600x1200 pixels. It can be used in landscape and portrait mode.
The NioWatch software allows to set the display test patterns, calibrate the display and view additional display and display controller information.
Compared to the predicate device, the E-2320 PA S display has a different LCD panel with a smaller screen size. The other components of the system are the same.
The device does not come into contact with the patient. It does not control any life sustaining devices either.
9. Conclusion:
The Barco E-2320 PA S is substantially equivalent to the predicate device, E 2621.
The new and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application and intended use.
Any difference between both devices does not affect safety or efficacy.
The 510(k) Pre-Market Notification for the Barco E-2320 PA S contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three abstract human figures, represented by curved lines, facing to the right. The figures are stacked vertically, with the top figure being the largest and the bottom figure being the smallest.
Public Health Service
SEP 3 0 2005
Food and Drug Administrati 9200 Corporate Boulevard Rockville MD 20850
Mr. Lieven De Wandel Official Correspondent BarcoView - Medical Imaging Systems 35 President Kennedypark 8500 Kortijk BELGIUM
Re: K052529 Trade/Device Name: E-2320 PA S Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 9, 2005 Received: September 15, 2005
Dear Mr. De Wandel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807) vilaling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality of your (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
0(k) Number (if known): _ڀيڪي 529
Device Name: E-2320 PA S
Indications for Use:
*The E-2320 PA S is intended to be used in displaying digital images for review t trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.
Prescription Use _XX
(Part 21 CFR 801 Subpart D) AND/OR
(21 CFR 801 Subpart C)
Over-The-Counter Use _
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel R. Leggson
(Division Sian-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
N/A