K Number
K052459

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-01-11

(126 days)

Product Code
Regulation Number
870.4290
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medtronic model 6248VAL adjustable valve is intended to minimize blood loss during percutaneous catheter procedures used to place cardiac leads and implant accessories such as guide wires and subselecting catheters.

Device Description

The 6248VAL Adjustable Valve is a single use valve designed to reduce blood loss during percutaneous catheter procedures. The valve sealing diameter is adjustable so it may be tightened or loosened to provide an appropriate seal around devices such as leads, guidewires, and catheters, which are passed through the main valve port. Blood flow through the valve is reduced as the valve is closed.

AI/ML Overview

Here's an analysis of the provided text regarding the 6248VAL Adjustable Valve:

It's important to note that this document is a 510(k) summary for a medical device (an adjustable valve), not an AI/ML-driven device. Therefore, many of the typical acceptance criteria and study components for AI/ML devices, such as performance metrics like sensitivity/specificity, sample sizes for test sets, ground truth establishment by experts, MRMC studies, or training sets, are not applicable here.

The summary focuses on demonstrating "substantial equivalence" to predicate devices through traditional engineering and biocompatibility testing.


Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Visual requirementsMet
Mechanical requirementsMet
Performance requirementsMet
Biocompatibility consistent with ISO 10993-1:2003Conducted
Sterilization process validated (gamma radiation)Will be sterilized using a validated gamma radiation process

Study Details (as inferable from the document)

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified. The document states "Device verification testing was performed to demonstrate the valve meets established performance criteria." This typically involves a defined number of units or test samples, but the specific quantity is not reported.
  • Data Provenance: Not specified, but implied to be internal laboratory testing conducted by Medtronic (the manufacturer). This would be prospective testing of newly manufactured devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. For a physical device like a valve, "ground truth" is typically defined by engineering specifications, material standards, and functional requirements. Expert medical consensus or pathology is not relevant for this type of testing.

4. Adjudication method for the test set:

  • Not applicable. This concept pertains to resolving discrepancies in interpretations, usually for image-based or diagnostic AI/ML systems. For mechanical testing, results are typically objective measurements against a defined specification.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is a physical device, not an AI/ML diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This is a physical device, not an algorithm.

7. The type of ground truth used:

  • Engineering Specifications and Performance Standards: The "ground truth" for this device's performance would be the pre-defined engineering design specifications, functional requirements (e.g., ability to reduce blood loss, seal effectively), and compliance with relevant standards (like ISO 10993-1 for biocompatibility).

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device; there is no "training set."

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/ML device; there is no "training set."

Overall Conclusion from the Document:

The 510(k) summary for the 6248VAL Adjustable Valve demonstrates substantial equivalence to predicate devices based on:

  • Similar intended use.
  • Equivalent design, function, materials, and method of operation.
  • Successful completion of visual, mechanical, and performance testing, indicating all design and performance requirements were met.
  • Biocompatibility testing consistent with ISO 10993-1.
  • Validated sterilization process.

The document does not provide the detailed raw data or specific quantitative results from these tests, but rather states that the criteria were "met" and "conducted," which is typical for a 510(k) summary demonstrating substantial equivalence for a non-AI/ML device.

{0}------------------------------------------------

510(K) SUMMARY

Date Prepared:September 06, 2005
Date Revised:December 28, 2005
Submitter:Medtronic, Inc.Cardiac Rhythm Management7000 Central Avenue NEMinneapolis, MN 55432
Contact:Michelle NivalaRegulatory Affairs Specialist
Telephone:(763) 505-7863
Fax:(763) 505-7877
E-Mail:michelle.d.nivala@medtronic.com
Proprietary Name:6248VAL Adjustable Valve
Common Name:Cardiopulmonary bypass adaptor.stopcock, manifold, or fitting.
Device Classification:Class II, 21 CFR 870.4290
Product Code:DTL

Summary of Substantial Equivalence

The intended use, design, function, materials and method of operation, of the 6248VAL Adjustable Valve are substantially equivalent with regard to these features, to the following predicate devices:

  • Medtronic Adjustable Hemostasis Valve (model 6218VAL) K012083 and K012130 cleared August 28, 2001.
  • · Guidant Rapido Cut-A-Way Rotating Hemostasis Valve -- K031688 cleared July 2, 2003
  • · Angeion Y-Adaptor with Touhy-Borst Valve K895580 cleared April 30, 1990

Image /page/0/Picture/10 description: The image shows the Medtronic logo with the word "Confidential" to the right of it. The Medtronic logo consists of a symbol on the left and the word "Medtronic" in bold, black letters. The symbol appears to be a stylized representation of the human body. The word "Confidential" is in a smaller, lighter font than the Medtronic name.

{1}------------------------------------------------

Device Description

The 6248VAL Adjustable Valve is a single use valve designed to reduce blood loss during percutaneous catheter procedures. The valve sealing diameter is adjustable so it may be tightened or loosened to provide an appropriate seal around devices such as leads, guidewires, and catheters, which are passed through the main valve port. Blood flow through the valve is reduced as the valve is closed.

Indications for Use

The Medtronic model 6248VAL adjustable valve is intended to minimize blood loss during percutaneous catheter procedures used to place cardiac leads and implant accessories such as guide wires and subselecting catheters.

Technological Characteristics

Mechanical method of operation, sealing mechanism and packaging materials are substantially equivalent to the predicate devices referenced.

Summary of Testing

Device verification testing was performed to demonstrate the valve meets established performance criteria and to support equivalency to the referenced predicate devices. Visual, mechanical and performance testing was completed. All design and performance requirements were met.

Biocompatibility testing consistent with ISO 10993-1: 2003. "Biological Evaluation of Medical devices - Part 1: Evaluation and Testing" was also conducted.

The 6248VAL Adjustable Valve will be sterilized using a validated gamma radiation process.

Conclusion

Medtronic considers the 6248VAL Adjustable Valve to be substantially equivalent to legally marketed predicate devices through the data and information presented. No safety or effectiveness issues were identified.

Image /page/1/Picture/14 description: The image shows the Medtronic logo with the word "Confidential" to the right of the logo. The Medtronic logo consists of a circular emblem with stylized figures inside, followed by the company name in bold, sans-serif font. The word "Confidential" is in a smaller, regular font, indicating the sensitive nature of the document or information associated with the logo.

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines that suggest the shape of a bird's head and wings.

JAN 1 1 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medtronic, Inc. c/o Ms. Michelle Nivala Regulatory Affairs Specialist 7000 Central Avenue NE Minneapolis, MN 55432

Re: K052459 6248VAL Adjustable Valve Regulation Number: 21 CFR 870.4290 Regulation Name: Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting Regulatory Class: Class II (Two) Product Code: DTL Dated: December 5, 2005 Received: December 6, 2005

Dear Ms. Nivala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Michelle Nivala

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Dina R. Vicines

Image /page/3/Picture/5 description: The image shows a signature. The signature is written in cursive and appears to be a stylized version of the letters 'Sa'. The signature is written in black ink and is on a white background. The signature is simple and easy to read.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): K052459

Device Name: 6248VAL Adjustable Valve

Indications for Use: The Medtronic model 6248VAL adjustable valve is intended to minimize blood loss during percutaneous catheter procedures used to place cardiac leads and implant accessories such as guide wires and subselecting catheters.

Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Duma R. Vechner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_K 052459

Image /page/4/Picture/17 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left, the word "Medtronic" in bold black letters in the center, and the word "Confidential" in smaller black letters on the right. The circular graphic appears to depict a human figure.

§ 870.4290 Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.

(a)
Identification. A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.