K Number
K052406

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-11-07

(67 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Uni-Thread Spinal System is a pedicle screw system indicated for skeletally mature patients who:

  • have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra;
  • receive fusions using autogenous bone graft only;
  • have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and
  • have the device removed after the development of a solid fusion.

In addition, the Uni-Thread Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

  • Degenerative spondylolisthesis with objective evidence of neurologic impairment
  • fracture
  • dislocation
  • scoliosis
  • kyphosis
  • spinal tumor
  • failed previous fusion (pseudoarthrosis)
Device Description

The Uni-Threaad Spinal System instrumentation consists of screws, rods, and connectors that can be assembled in a variety of constructs. The instrumentation is combined and placed to treat a variety of spinal disorders, with the constructs varying according to the nature of the spinal pathology. The Uni-Thread System components are available in titanium alloy manufactured according to ASTM-F136.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Uni-Thread™ Spinal System. This document focuses on establishing substantial equivalence to previously cleared devices rather than presenting data from a clinical study with specific acceptance criteria and performance metrics typically associated with AI/ML devices or novel diagnostic tools.

Therefore, the requested information regarding acceptance criteria, device performance, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, or ground truth establishment is not available in the provided text. The submission relies on demonstrating similarity in design, material, and intended use to predicate devices, supported by a "beginning in" (likely meaning 'benchmark' or 'comparison') study to show equivalence of the construct design to currently marketed systems.

Here's a breakdown of why each requested point cannot be addressed from the given document:

  1. A table of acceptance criteria and the reported device performance: This document does not specify quantitative acceptance criteria for device performance or report specific performance metrics (e.g., sensitivity, specificity, accuracy) akin to what would be expected for an AI/ML or diagnostic device. The basis for clearance is "substantial equivalence" to predicate devices.

  2. Sample sized used for the test set and the data provenance: Not applicable. No "test set" in the context of AI/ML or diagnostic performance evaluation is mentioned. The "beginning in a was performed to demonstrate the equivalence of the construct design" likely refers to mechanical or engineering testing, but details are absent.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. There is no mention of human experts establishing ground truth for a test set.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No test set requiring expert adjudication is described.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a spinal implant, not an AI-assisted diagnostic tool, so an MRMC study is irrelevant to its clearance.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. As a physical medical device (spinal system), the concept of "standalone algorithm performance" does not apply.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of diagnostic or AI performance. The "ground truth" for this device would be its mechanical and material properties meeting established standards, and its design being equivalent to predicate devices.

  8. The sample size for the training set: Not applicable. There is no mention of a training set as this is not an AI/ML device.

  9. How the ground truth for the training set was established: Not applicable. There is no mention of a training set.

In summary, the provided document relates to the clearance of a physical medical device (a spinal implant) based on its substantial equivalence to existing devices, not on the performance metrics of an AI/ML system.

{0}------------------------------------------------

K052406

510(k) SUMMARY OF SAFETY AND EFFECTIVENE

NAME OF FIRM:SpineVision, Inc.3003 Summit Blvd., Suite 1400Atlanta, GA 30319Phone: 404-460-5077
510(K) CONTACT:Lynnette WhitakerVice President, Regulatory Affairs574-269-9776
TRADE NAME:Uni-Thread™ Spinal System
COMMON NAME:Rod and Screw Spinal Instrumentation
CLASSIFICATION:888.3070 Pedicle Screw Spinal System
DEVICE PRODUCT CODE:Product code: 87 MNH, MNI
SUBSTANTIALLYEQUIVALENT DEVICES:K013301 and K021510, Uni-Thread Spinal System,SpineVision, Inc.

DEVICE DESCRIPTION AND INTENDED USE:

The Uni-Threaad Spinal System instrumentation consists of screws, rods, and connectors that can be assembled in a variety of constructs. The instrumentation is combined and placed to treat a variety of spinal disorders, with the constructs varying according to the nature of the spinal pathology. The Uni-Thread System components are available in titanium alloy manufactured according to ASTM-F136.

INDICATIONS FOR USE

The Uni-Thread Spinal System is a pedicle screw system indicated for skeletally mature patients who:

  • . have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra;
  • . receive fusions using autogenous bone graft only;
  • have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and .
  • . have the device removed after the development of a solid fusion.

{1}------------------------------------------------

In addition, the Uni-Thread Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

  • Degenerative spondylolisthesis with objective evidence of neurologic impairment .
  • fracture ●
  • dislocation .
  • scoliosis .
  • kyphosis ●
  • spinal tumor .
  • failed previous fusion (pseudoarthrosis) .

BASIS OF SUBSTANTIAL EQUIVALENCE:

DAOX OF DUDD FAN FITT Firead Spinal System are identical in design, material, and intended use to other spinal instrumentation system that have been cleared by FDA for posterior spinal use. to other beginn in a was performed to demonstrate the equivalence of the construct design to currently marketed spinal systems.

00005 Rege 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" arranged in a circular pattern around the symbol. The logo is black and white and appears to be a simple, clean design.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 7 2005

Lynnette Whitaker Vice President, Regulatory Affairs Spinevision, Inc. 3003 Summit Blvd., Suite 1400 Atlanta, Georgia 30319

K052406 Re:

Trade/Device Name: Uni-Thread™ Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: August 31, 2005 Received: September 2, 2005

Dear Ms. Whitaker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 -- Ms. Lynnette Whitaker

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

So Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K052406

Device Name: Uni-Thread™ Spinal System, Additional Components

Indications for Use:

The Uni-Thread Spinal System is a pedicle screw system indicated for skeletally mature patients who:

  • have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra; .
  • receive fusions using autogenous bone graft only; ●
  • have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and .
  • have the device removed after the development of a solid fusion. .

In addition, the Uni-Thread Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

  • Degenerative spondylolisthesis with objective evidence of neurologic impairment .
  • fracture ●
  • dislocation ●
  • scoliosis ●
  • kyphosis ●
  • spinal tumor �
  • failed previous fusion (pseudoarthrosis)

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of

510(k) Number K052406

র্ম Post र्मः

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.