K Number
K052353

Validate with FDA (Live)

Device Name
Z-CHAIR
Date Cleared
2005-09-09

(11 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Pride Mobility Products Corporation Z-Chair Powered Wheelchair, is to provide mobility to persons limited to a seated position that have the capability of operating a Powered Wheelchair.

Device Description

The Z-Chair is a compact battery-operated power wheelchair featuring rear-wheel drive technology, front casters, rear anti-tip wheels and is designed for, but not limited to Pride providers / retailers and their consumers. As a motorized wheelchair, the Z-Chair offers 2 motors for operational purposes, electronic regenerative disc brakes, suspension, off-board battery charger, a fully programmable 50-amp controller, and a 12-volt, 30-amp hour battery pack. Accessories include a lap belt, cup holder, and zippered accessory pouch. The Z-Chair is designed with ultimate safety, stability, performance, and portability in mind. The main feature of the Z-Chair is that it can be disassembled into 4 parts: the rear section, the front section, the battery pack, and the seat. This allows for ease of transport. The Z-Chair has a lightweight foldable seat, and a single-post, quick-release seat mount.

AI/ML Overview

The provided document is a 510(k) summary for the Pride Mobility Products Corporation Z-Chair Power Wheelchair. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. This document does NOT describe a study involving acceptance criteria and device performance in the context of a multi-reader multi-case (MRMC) comparative effectiveness study, or a standalone algorithm performance study.

Therefore, I cannot extract the information required in your request regarding acceptance criteria, device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for a medical device algorithm.

The document explicitly states:

  • Discussion of Clinical Testing Performed: N/A

This indicates that no clinical studies were performed to assess the performance of the device in a human-in-the-loop setting or as a standalone algorithm. The device described is a physical power wheelchair, not an AI/software device.

However, I can provide information based on the non-clinical testing discussed in the document, framed as demonstrating compliance with applicable testing standards, which serve as a form of "acceptance criteria" for this type of physical device.


Summary of Non-Clinical Testing and Compliance (Interpreted as "Acceptance Criteria" for a Physical Device)

The Z-Chair Power Wheelchair underwent non-clinical testing to demonstrate compliance with various ANSI/RESNA and other standards. These standards implicitly define the "acceptance criteria" for the safety, performance, and functionality of the wheelchair.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard & Parameter)Reported Device Performance (Compliance)
ANSI/RESNA WC/01: Static StabilityComplied
ANSI/RESNA WC/02: Dynamic StabilityComplied
ANSI/RESNA WC/03: Effectiveness of BrakesComplied
ANSI/RESNA WC/05: Overall Dimensions, Mass & Turning SpaceComplied (presumably within specified limits)
ANSI/RESNA WC/08: Static, Impact and Fatigue StrengthsComplied
ANSI/RESNA WC/09: Climatic TestsComplied
ANSI/RESNA WC/10: Obstacle ClimbingComplied
ANSI/RESNA WC/15: Documentation and LabelingComplied
ANSI/RESNA WC Vol. 2-1998 Section 21: Electromagnetic CompatibilityComplied
ANSI/RESNA WC/93: Maximum overall DimensionsComplied (presumably within specified limits)
CAL 117: Flammability TestingComplied

2. Sample Size for Test Set and Data Provenance

  • Test Set Sample Size: Not applicable in the context of an algorithm's test set. For a physical product, testing is typically performed on one or more representative units (prototypes or production models). The document does not specify the number of units tested.
  • Data Provenance: Not applicable in the traditional sense of retrospective/prospective clinical data or country of origin. The data provenance is from laboratory/engineering testing against specified performance standards.

3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

  • Not applicable. This document pertains to the physical performance testing of a wheelchair, not the assessment of diagnostic accuracy by medical experts. "Ground truth" here is defined by objective engineering measurements and the pass/fail criteria of the specified standards.

4. Adjudication Method for the Test Set

  • Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus in diagnostic studies. For physical product testing, results are typically determined by adherence to established test protocols and instrument readings or visual inspections, with criteria defined by the testing standards themselves.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

  • No, an MRMC comparative effectiveness study was not done. The device is a physical power wheelchair, not an AI diagnostic/analytic tool.

6. Standalone (Algorithm Only) Performance Study

  • No, a standalone performance study was not done. The device is a physical power wheelchair, not an AI algorithm.

7. Type of Ground Truth Used

  • For this physical device, the "ground truth" for the non-clinical testing is implicitly defined by the objective pass/fail criteria and measurement tolerances set forth in each of the referenced ANSI/RESNA and CAL 117 standards. For example, for static stability, the ground truth would be whether the wheelchair remains stable under specified conditions as defined by ANSI/RESNA WC/01.

8. Sample Size for Training Set

  • Not applicable. This is not an AI/machine learning device that requires a training set. The design and engineering of the Z-Chair would be based on general engineering principles and design specifications, not a "training set" of data.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. There is no training set for this physical device.

{0}------------------------------------------------

KOS2JS3

SEP - 9 2005

Image /page/0/Picture/1 description: The image shows the logo for Pride Mobility Products Corp. The word "Pride" is written in a bold, stylized font, with a horizontal line running through the top half of the letters. Below the word "Pride", the words "Mobility Products Corp." are written in a smaller, less bold font. The logo is black and white.

Research & Development 182 Susquehanna Ave Exeter, PA 18643 570-655-5574 FAX 655-2990 www.pridemobility.com

Exhibit 1

510(k) Summary Pride Mobility Products Corporation Z-Chair Power Wheelchair

Submitter's Name & Address:

Pride Mobility Products Corporation 182 Susquehanna Avenue Exeter, Pa. 18643 Phone: (570) 655-5574 Facsimile: (570) 655-2990

Contact Person:

Thomas Schappert Official Correspondent

Date Prepared:

08-02-05

Name of Device and Proprietary Name:

Z-Chair / Pride Mobility

Common or Usual Name:

Powered Wheelchair Base Unit

Classification Name:

Physical Medicine / Wheelchair, Powered

Product Code:

ITI

Comparison to Predicate Devices:

The Z-Chair is substantially equivalent to the Pride Mobility Quantum Blast (K011993) when comparing configuration, maneuverability, structure. The performance whor companing companisms of the drive mechanisms are similar to achieve the same intended use function that enables the user to maintain optimum stability without hindering performance. Both utilize rear-wheel drive technology with rear drive wheels, front casters, and rear anti-tip wheels. The key functional change between the Quantum In one out of and the Z-Chair is the Z-Chair's ability to disassemble into four pieces. The reason behind this is to make the product more portable for users who need to travel with the device. The device can essentially be taken apart and loaded into the trunk of a vehicle or stored in a compact space.

{1}------------------------------------------------

K052353

Device Description:

Device Description.
The Z-Chair is a compact battery-operated power wheelchair featuring rear-wheel drive The Z-Chair is a compact battery oportuled power to programmable 50-amp technology, from casters, rear and is the is designed for, but not limited to Pride r enriry & Giles Corp. providers / retailers and their consumers.

As a motorized wheelchair, the Z-Chair offers 2 motors for operational purposes, electronic regenerative disc brakes, suspension, off-board battery charger, a fully electronic regenerative diso brance, battery pack. Accessories include a lap belt, programmable continent cup holder, and zippered accessory pouch.

The Z-Chair is designed with ultimate safety, stability, performance, and portability in The Z-Chair is designed with allinder calbity, than be disassembled into 4 parts the mind. The main feature of the Echants to the seat. This allows for ease of real section, the front section, the battery path, and the shas a lightweight foldable seat, and a single-post, quick-release seat mount.

Intended Use:

The intended use of the Pride Mobility Products Corp. Z-Chair Powered Wheelchair is to The intended use of the Pride Mobility a seated position that have the capability of operating a powered wheelchair.

Non-Clinical Testing

Compliance to applicable Testing Standards is as follows: ANSI/RESNA WC/01 Determination of Static Stability ANSI/RESNA WC/02 Determination of Dynamic Stability ANSI/RESNA WC/03 Effectiveness of Brakes ANSI/RESNA WC/05 Overall Dimensions, Mass & Turning Space ANSI/RESNA WC/08 Test methods for Static, Impact and Fatigue Strengths ANSI/RESNA WC/09 Climatic Tests ANSI/RESNA WC/10 Obstacle Climbing ANSI/RESNA WC/15 Documentation and Labeling ANSI/RESNA WC Vol. 2-1998 Section 21 - Requirements and Test Methods for Electromagnetic Compatibility. ANSI/RESNA WC/93 Maximum overall Dimensions CAL 117 - Flammability Testing

Discussion of Clinical Testing Performed: N/A

Conclusions:

The Z-Chair has the same intended use and similar technological characteristics as the Quantum Blast (K011993), moreover, the non-clinical testing and the predicate Quarkan - Dido. (10 110 110 110 110 11 their technological characteristics do companions any new questions of safety or effectiveness. Thus, the Z-Chair is not Taloo Jany Tren on the predicate device (Blast). The Z-Chair has passed all the necessary testing procedures and is considered to be safe for user operation.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized graphic of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle graphic.

SEP - 9 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Thomas Schappert Official Correspondent Pride Mobility Products Corporation 182 Susquehanna Avenue Exeter, Pennsylvania 18643

Re: K052353

Trade/Device Name: Z-Chair - Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: August 19, 2005 Received: August 29, 2005

Dear Mr. Schappert:

We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave reviewed your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreatives of the enactment date of the Medical Device Amendments, or to conniner of the they 20, 2017 - ) in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, aterers, which of the Act include requirements for annual registration, listing of general controls provision practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or sabyer is one of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Thomas Schappert

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and w you to ough finding of substantial equivalence of your device to a legally prematication. The areaults in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you deems of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Neil R. Oyler

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K

Device Name: Z-Chair - Powered Wheelchair

Indications for Use:

The intended use of the Pride Mobility Products Corporation Z-Chair Powered Wheelchair, is to provide mobility to persons limited to a seated position that have the capability of operating a Powered Wheelchair.

AND / OR Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of GDRH, Office of Device Evaluation (ODE) ATTIN NIY (Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number_Kos2333

Page 1 of 1

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).