(143 days)
The CardioSoft PC ECG is intended for use in acquiring, processing, recording, archiving and displaying electrocardiographic information. It can provide serial comparison of ECG information to facilitate review of current and previous records and is intended for use by trained operators. The system offers no diagnostic opinion to the user, but may be used by a physician to render his or her own medical opinion.
The CardioSoft PC ECG is an electrocardiograph that detects signals associated with cardiac activity and produces an ECG graphical record. The device operates on a personal computer through a USB hardware module. The device can acquire, process, record, archive and display 3-, 6-, or 12-lead ECG and vector cardiograms. The device provides portable ECG with important advantages compared to traditional machines.
The presented document is a 510(k) summary for the CardioSoft PC ECG device, a substantially equivalent submission to the predicate device CARDIAX (K924544). The document does not describe specific acceptance criteria or a dedicated study to prove the device meets performance criteria through clinical data or comparative effectiveness. Instead, it relies on demonstrating substantial equivalence to a predicate device and compliance with recognized standards.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document explicitly states: "The FDA has not established performance standards for this device." Therefore, there are no specific FDA-mandated acceptance criteria or direct reported device performance metrics (e.g., sensitivity, specificity, accuracy) derived from a clinical study for this 510(k) submission.
The "performance" section instead focuses on demonstrating equivalence through non-clinical testing and adherence to quality assurance measures.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| None explicitly defined by FDA in this document. | Non-clinical testing demonstrated "equivalence with the predicate device CARDIAX and the results show similar performance with this device." |
| Compliance with established voluntary/consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25, IEC 60601-2-27, ECS53, EC38, UL 2601-1, ANSI/UL 2601-1 ad CAN/CSA C22.2) | Device "complies with the following recognized voluntary and consensus standards". |
| Conformance to product development procedures and plans | "Assured by the application of system tests, design reviews, and product verification and validation testing performed prior to product release." |
| Quality assurance measures applied during development | Requirements specification reviews, Code inspections, Software and hardware evaluation and testing, Safety testing, Environment testing, Final validation. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not describe a clinical "test set" with a specific sample size. The performance assessment relies on non-clinical testing and comparison to a predicate device. Therefore, no information on data provenance (country of origin, retrospective/prospective) for a clinical test set is provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. As no clinical test set or ground truth establishment process for diagnostic performance is described, there's no mention of experts or their qualifications in this context.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no clinical test set described in the submission, therefore no adjudication method is mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. The document makes no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. The device, an electrocardiograph, "offers no diagnostic opinion to the user" and there is no AI component mentioned that would assist human readers in a diagnostic setting.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The device is an "electrocardiograph that detects signals associated with cardiac activity and produces an ECG graphical record." It is not an algorithm providing diagnostic output without human interpretation. It generates a record "that may be utilized only by a qualified healthcare professional for review and interpretation when based upon patient history, medical examination, and other clinical findings." Therefore, a standalone algorithm-only performance study as typically understood for AI-driven diagnostic tools was not conducted or described. The "performance" refers to the device's ability to accurately acquire, process, record, archive, and display ECG information.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Given the nature of the device (ECG acquisition and display) and the absence of a clinical diagnostic performance study, the concept of "ground truth" as typically applied to diagnostic algorithms (e.g., expert consensus for disease presence or absence, pathology results) is not relevant or mentioned in this submission. The "ground truth" for demonstrating performance would likely involve the accuracy of signal acquisition, processing, and display compared to known electrical signals or established ECG benchmarks.
8. The sample size for the training set
Not applicable. The CardioSoft PC ECG is described as an upgrade/modification of an existing device, preserving the "software processes of the predicate device." It's not a machine learning or AI-driven device that typically requires a 'training set' in the conventional sense for a 510(k). The focus is on hardware and interface improvements while maintaining functional equivalence.
9. How the ground truth for the training set was established
Not applicable, as there is no mention of a training set for a machine learning model.
{0}------------------------------------------------
Corporate Office: 1776 Yorktown Ste LL30 Houston, Texas 77056
Image /page/0/Picture/2 description: The image shows the CardioSoft logo. The logo consists of a circular shape that is black on the bottom and white on the top. The word "CardioSoft" is written in a simple font below the circular shape.
(713) 623-4009 Phone: (713) 623.4003 Fax:
JAN 1 3 2006
510(K) SUMMARY
Applicant's Name:
CardioSoft, LP 1776 Yorktown Street, Suite LL30 Houston, Texas 77056
Tel: 713-623-4009 Tel: 713-623-4009 corporate@CardioSoft.com
Contact Person: Saeed Moradi, Vice-President
Summary Prepared: July 15, 2005
CardioSoft® PC ECG Proprietary Name:
Classification Name: Electrocardiograph
Classification Name: & 870.2340, Electrocardiograph Product Code: DPS, Class II
Establishment Registration No .: 3003786840
Predicate Device(s): The CardioSoft PC ECG is substantially equivalent to the CARDIAX device, K924544 (Imed / Fairlake Marketing Group) [predicated on the MICROMED QRS-CARD], cleared by prior 510(K) Premarket Notification, in terms of intended use, indications for use, technological characteristics, performance and user interface. The predicate is a Class II device.
- Standards: The CardioSoft PC ECG complies with the following recognized voluntary and consensus standards: IEC 60601-1:1988 and amendments, IEC 60601-1-2:2001, IEC 60601-2-25:1999 and amendment, IEC 60601-2-27:1994, ECS53:1995, EC38:1998, UL 2601-1, ANSI/UL 2601-1 ad CAN/CSA C22.2
The device also complies with 21 CFR Parts 801 and 820 for labeling and GMP, and with 21 CFR 898 for electrode lead wires and patient cables.
000009
{1}------------------------------------------------
CARDIOSOFT PC ECG
510(K) SUMMARY
510(K) SUMMARY
1.214
The CardioSoft PC ECG is intended for use in acquiring,
processing, recording, archiving and displaying
electrocardiographic information. It can provide serial
electrocardiographic information. It can provide serial comparison of ECG information to facilitate review of current and previous records and is intended for use by trained operators. The system offers no diagnostic opinion to the user, but may be used by a physician to render his or her own medical opinion.
The CardioSoft PC ECG is an electrocardiograph that Device Description: detects signals associated with cardiac activity and produces an ECG graphical record. The device operates on a personal computer through a USB hardware module. The device can acquire, process, record, archive and display 3-, 6-, or 12-lead ECG and vector cardiograms. The device provides portable ECG with important advantages compared to traditional machines.
The system requires a USB connection and Windows® operating system and is intended for use by trained operators. It is intended for use by or on the order of a physician or other qualified health care professional.
ECGs are routinely used to diagnose cardiac abnormalities, determine response todrug therapy, and reveal changes in heart function. The ECG record data may be utilized only by a qualified healthcare professional for review and interpretation when based upon patient history, medical examination, and other clinical findings.
Technology:
Intended Use /
Indications for Use
The CardioSoft PC ECG employs the same functional technology as the predicate device. The only differences are technological improvements made with respect to
{2}------------------------------------------------
510(K) SUMMARY
USB interface, greater frequency response, and use with current Windows® operating systems.
Both devices utilize preamplifiers to collect electrocardiographic information from skin electrodes. The preamplifiers provide electrical isolation of the patient and the device. There are differences in two components that provide electrical isolation. The proposed device uses a different isolation DC/DC converter (made by Recom) and optocoupler (made by Hewlett Packard) than the predicate device. The proposed device utilizes an integrated USB module and provides improved frequency response. Both devices are in compliance with electromagnetic compatibility conformance and electrical and thermal safety compliance by the applicable standards. No energy is delivery to the patient. The enhancements have no effect on safety or performance.
The software processes of the predicate device have been fully preserved in the proposed device and have been adapted for use with the latest versions of the Windows® operating system. Generated data is identical, however the proposed device allows the user to more aptly configure screen display and printer output to a desired format. The enhancements have no effect on safety or performance.
{3}------------------------------------------------
510(K) SUMMARY
Performance:
The FDA has not established performance standards for this device. A listing of voluntary or consensus standards is provided in SECTION O - APPENDIX: Attach 5 -CARDIOSOFT PC ECG (USB version) STANDARDS TEST REPORT.
Non-clinical testing has been performed demonstrating equivalence with the predicate device CARDIAX and the results show similar performance with this device. Conformance to the product development procedures and plans have been assured by the application of system tests, design reviews, and product verification and validation testing performed prior to product release.
The following quality assurance measures were applied to the development of the CardioSoft PC ECG :
- Requirements specification reviews o
- Code inspections o
- Software and hardware evaluation and testing 0
- Safety testing o
- Environment testing 0
- Final validation O
Conclusion:
The CardioSoft PC ECG is an upgrade modification to the predicate device CARDIAX of like design and manufacture. This premarket notification submission demonstrates that the CardioSoft PC ECG is substantially equivalent because this device has the same basic intended use and the differences in technological characteristics do not raise new questions of safety or effectiveness. The CardioSoft PC ECG is as safe, as effective, and performs as well as the predicate device.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 3 2006
CardioSoft, LP c/o Mr. Saeed Moradi Vice-President 1776 Yorktown Street, Suite LL30 Houston, Texas 77056
Re: K052295
Trade Name: CardioSoft PC ECG Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: October 24, 2005 Received: November 18, 2005
Dear Mr. Moradi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Saeed Moradi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product reastill allow you to begin marketing your device as described in your Section 510(k) I mis lotter with and w yourse by a finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Bhiminimo for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
K052295 510(k) # (if known):
Device Name: CARDIOSOFT PC ECG
Indications for Use:
The CardioSoft PC ECG is intended for use in acquiring, processing, recording, archiving and displaying electrocardiographic information. It can provide serial comparison of ECG information to facilitate review of current and previous records and is intended for use by trained operators. The system offers no diagnostic opinion to the user, but may be used by a physician to render his or her own medical opinion.
Prescription Use x (21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Brimmer
Division sign-Off
Division of Cardiovascular Devices
510(k) Number Kns2295
000007
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).