(29 days)
The GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® HDx system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The 1.5T and 3.0T Signa® HDx MR systems are a modification to the previously cleared MR systems K041476 which utilizes a superconducting magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 1.5T and 3.0T Signa® HDx Magnetic Resonance System features a superconducting magnet operating at either 1.5T, or 3.0T. The data acquisition system supports 1, 4, 8, 12, 16, 32 independent receive channels and multiple independent coil elements per channel during a single acquisition series. Additionally, the system architecture is designed for expansion in 16 channel increments. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anafomical applications, short scan times, and multinuclear spectroscopy. The 1.5T Signa® HDx MR system is also available in a mobile configuration.
The GE 1.5T and 3.0T Signa® HDx MR Systems are modifications of previously cleared MR systems (K041476). The provided documentation does not contain specific acceptance criteria or a detailed study explicitly proving the device meets those criteria in the way a clinical performance study for an AI algorithm would. Instead, the submission focuses on demonstrating substantial equivalence to the predicate devices through technical modifications and compliance with general performance and safety standards.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not outline specific, quantitative acceptance criteria for image quality or diagnostic performance of the 1.5T and 3.0T Signa® HDx MR Systems in the way a new medical image analysis algorithm might. Instead, the primary "acceptance criterion" is conceptual: substantial equivalence to the predicate devices (Signa® Excite 1.5T MR system and Signa® Excite 3.0T MR system) in terms of safety and effectiveness, while offering improved technical capabilities.
The reported "device performance" is described in terms of these improvements and compliance with standards, rather than specific metrics like sensitivity or specificity.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Substantial equivalence to predicate devices (K041476) | The 1.5T and 3.0T Signa® HDx MR Systems are considered substantially equivalent to the currently marketed Signa® Excite 1.5T and 3.0T MR systems, respectively. The systems maintain the same intended use. |
| Improved technical capabilities over predicate | Main differences: Change to receiver chain architecture with thirty-two independent receive channels (and future expansion in 16-channel increments).Overall system improvements: Simplified User Interface, single 23" Liquid Crystal Display, improved multi-channel surface coil connectivity, improved image reconstruction architecture (Volume Recon Engine - VRE). |
| Compliance with relevant performance standards | Evaluated to appropriate NEMA performance standards. |
| Compliance with relevant safety standards | Evaluated to IEC 60601-1 International Medical Equipment Safety standard and IEC 60061-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis. |
| No new potential hazards introduced | It is the opinion of GE that usage of the 1.5T Signa® HDx Magnetic Resonance System and the 3.0T Signa® HDx Magnetic Resonance System does not result in any new potential hazards. |
| Ability to produce high-resolution, high signal-to-noise ratio images, and short scan times for diagnostic use across various body parts | The Indications for Use state the system is "designed to support high resolution, high signal-to-noise ratio, and short scan times" and is "indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body...". The context implies that the improvements contribute to meeting these objectives effectively for diagnostic interpretation by a trained physician. |
2. Sample Size Used for the Test Set and Data Provenance
The submission does not describe a specific "test set" in the context of a clinical performance study using patient data or images to evaluate the device's diagnostic accuracy. The evaluation appears to be primarily technical, comparing the new system's specifications and performance against established standards (NEMA, IEC) and the predicate device. Therefore, information regarding "data provenance" (country of origin, retrospective/prospective) is not provided.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Since no test set for diagnostic accuracy is described, there is no mention of experts used to establish ground truth or their qualifications. The evaluation is focused on technical equivalence and safety, not on a clinical ground truth for diagnostic performance.
4. Adjudication Method for the Test Set
Given the absence of a test set for diagnostic performance, no adjudication method is described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No MRMC comparative effectiveness study is mentioned. The submission makes no claims about human readers' improvement with or without AI assistance, as it is for an MR system itself, not an AI-assisted diagnostic tool.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This question is not applicable as the submission concerns an MR imaging system, not a standalone algorithm.
7. The Type of Ground Truth Used
For the purpose of this 510(k), the "ground truth" is primarily:
- Predicate device specifications and performance: The fundamental aspect is proving that the new device is functionally equivalent or improved without introducing new safety concerns compared to the already cleared predicate.
- Compliance with recognized standards: NEMA performance standards, IEC 60601-1 (safety), and IEC 60061-2-33 (particular requirements for MR safety).
There is no mention of ground truth established through expert consensus, pathology, or outcomes data in the context of a diagnostic performance study.
8. The Sample Size for the Training Set
Not applicable. The submission does not describe an AI/ML algorithm that would require a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. No training set or associated ground truth establishment is described.
In summary, this 510(k) submission for the GE 1.5T and 3.0T Signa® HDx MR Systems focuses on demonstrating substantial equivalence to existing predicate devices based on technical modifications, safety compliance, and adherence to recognized performance standards, rather than a clinical performance study with defined acceptance criteria and ground truth for diagnostic accuracy.
{0}------------------------------------------------
GE Healthcare Technologies
SEP 2 1 2005
P.O. Box 414, Milwaukee, WI 53201
510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE Healthcare TechnologiesPO Box 414Milwaukee, WI 53201 |
|---|---|
| Contact Person: | Larry A. Kroger Ph.D.Manager, Regulatory Programs |
| Telephone: | 262- 544-3894 |
| Fax: | 262- 548-4768 |
| Date Prepared: | August 22, 2005 |
Device Name:
GE 1.5T Signa® HDx MR System, and GE 3.0T Signa® HDx MR System Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH
Marketed Device:
The 1.5T Signa® HDx MR System is substantially equivalent to the currently marketed Signa® Excite 1.5T MR system (K041476) with the main differences being a change to the receive chain architecture that includes thirty two independent receive channels, and allows for future expansion in 16 channel increments.
The 3.0T Signa® HDx MR System is substantially equivalent to the currently marketed Signa Excite 3.0T MR system (K041476) with the main differences being a change to the receive chain architecture that includes thirty two independent receive channels, and allows for future expansion in 16 channel increments.
Device Description:
The 1.5T and 3.0T Signa® HDx MR systems are a modification to the previously cleared MR systems K041476 which utilizes a superconducting magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 1.5T and 3.0T Signa® HDx Magnetic Resonance System features a superconducting magnet operating at either 1.5T, or 3.0T. The data acquisition system supports 1, 4, 8, 12, 16, 32 independent receive channels and multiple independent coil elements per channel during a single acquisition series. Additionally, the system architecture is designed for expansion in 16 channel increments. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anafomical applications, short scan times, and multinuclear spectroscopy. The 1.5T Signa® HDx MR system is also available in a mobile configuration.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo is a circular emblem with the letters 'GE' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is black and white.
P.O. Box 414, Milwaukee, WI 53201
Indications for Use:
The GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR svstem is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and obligue images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® HDx system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
Comparison with Predicate Device:
The 1.5T and 3.0T Signa® HDx MR Systems are a modification of the previously cleared MR systems K041476 with the main differences being the receive chain architecture that includes a thirty two independent receive channels, and allows for future expansion in 16 channel increments. The overall system has been improved with a simplified User Interface and a single 23" Liquid Crystal Display, improved multi-channel surface coil connectivity, and an improved image reconstruction architecture known as the Volume Recon Engine (VRE).
Summary of Studies:
The 1.5T and 3.0T Signa® HDx Magnetic Resonance Systems were evaluated to the appropriate NEMA performance standards as well as the IEC 60601-1 International Medical Equipment Safety standard and IEC 60061-2-33 Particular Requirements for Safety of Macnetic Resonance Equipment for Medical Diagnosis. The 1.5T Signa® HDx Magnetic Resonance System is comparable to the currently marketed Signa® Excite 1.5T Magnetic Resonance System. The 3.0T Signa® HDx Magnetic Resonance System is comparable to the currently marketed Signa® Excite 3.0T Magnetic Resonance System.
Conclusion:
It is the opinion of GE that the 1.5T Signa® HDx Magnetic Resonance System is substantially equivalent to the Signa® Excite 1.5T Magnetic Resonance System. Usage of the 1.5T Signa HDx Magnetic Resonance System does not result in any new potential hazards.
It is the opinion of GE that the 3.0T Signa® HDx Magnetic Resonance System is substantially equivalent to the Signa® Excite 3.0T Magnetic Resonance System. Usage of the 3.0T Signa HDx Magnetic Resonance System does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird in flight. The bird is composed of three curved lines that suggest movement and feathers. The logo is encircled by text that reads "DEPARTMENT OF HUMAN". The text is arranged in a circular fashion around the bird, with the words "DEPARTMENT" and "HUMAN" being the most prominent.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 1 2005
Larry A. Kroger, Ph.D.. Senior Regulatory Programs Manager GE Healthcare Technologies P.O. Box 414, W-400 MILWAUKEE WI 53201
Re: K052293
Trade/Device Name: GE 1.5T and 3.0T Signa® HDx MR Systems Regulation Number: 21 CFR $892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: August 22, 2005 Received: August 23, 2005
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactnent date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and the entire logo is presented in black against a white background.
GE Healthcare Technologies
P.O. Box 414, Milwaukee, WI 53201
Indications for Use
Kos2193 510(k) Number (if known): _
Device Name:
GE 1.5T Signa® HDx MR System, and GE 3.0T Signa® HDx MR System
Indications for Use:
The GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® HDx system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diaqnosis.
Prescription Use × (Part 21 CFR 80.1 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancyc brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 1 of 1
N/A