K Number
K052229
Manufacturer
Date Cleared
2006-10-13

(423 days)

Product Code
Regulation Number
876.4270
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Coloplast Ostomy Rod is used for loop ostomy surgery. The ostomy rod is a device that is placed through a loop of the colon brought out through the abdominal wall to temporarily keep it from slipping back through the surgical opening during the loop (colostomy or ileostomy) ostomy procedure.

Device Description

The Coloplast Ostomy Rod is a 90mm white HDPE rod with a fixed T at one end and a separate T piece that can be snaoped into the open end of the rod once it is in place. Holes in the ends of the T allow for suturing. The device is packaged in a laminated polvethylene and cas-permeable paper pouch. The product is sterilized by irradiation.

AI/ML Overview

The provided text describes a 510(k) summary for the Coloplast Ostomy Rod, which is a medical device. This type of regulatory submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria through a clinical study with detailed performance metrics.

Therefore, much of the requested information regarding acceptance criteria, specific study designs, sample sizes, ground truth establishment, and multi-reader multi-case studies for AI algorithms is not applicable to this document, as it describes a traditional medical device's regulatory submission, not an AI/ML-based device.

However, I can extract the relevant information that is present in the provided text.

Here's a breakdown of the information that can be extracted and why other parts are not applicable:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Tensile StrengthTested and monitored through ongoing testing. (Specific numerical criteria or results are not provided).
Flexural StrengthTested. "Flexural strength and flexibility were compared with results for the Coloplast Ostomy Rod and the Conva-Tec Sure-Fit System giving similar results both well above the estimated effect from the colon." (Specific numerical criteria or results are not provided).
Biocompatibility (Cytotoxicity, Irritation, Sensitization)Tested according to ISO 10993-01 Biological Evaluation of Medical Devices. "All criteria for acceptance were met." (Specific criteria or results are not provided).
Leachable Profiles (Stability over time)Tested for two lots (one at baseline, one near expiration). "There was no significant change associated with time." (Specific numerical criteria or test results are not provided).
SterilityLabeled as sterile by irradiation. (Implies meeting sterility requirements, but no specific test results are detailed in this section).
Material Biocompatibility (HDPE)"This is not the same material used for the ConvaTec Sure-Fit System, but is widely used in the medical industry and is shown to meet biocompatibility requirements." (Implies material acceptance based on general industry use and prior biocompatibility testing for the material itself, not necessarily for the final device in this context).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable: The document describes non-clinical performance data (material properties, biocompatibility, leachables) rather than a clinical study or a test set for an AI/ML algorithm. Therefore, "test set" in the context of an AI/ML device is not relevant here. The tests mentioned (tensile, flexural, biocompatibility, leachables) are done on the device itself and likely involve a small number of samples (e.g., a few units from different lots) for engineering and biological safety testing. No country of origin for "data" is mentioned, as it's product testing, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable: There is no "ground truth" in the clinical sense mentioned in this document, as it's not a diagnostic device or an AI algorithm requiring expert consensus. The "ground truth" for the material and performance tests would be established by scientific and engineering standards for device functionality and safety, executed by qualified lab personnel.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable: No clinical test set or subjective assessment requiring adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable: This is not an AI/ML device, so no MRMC study, human reader improvement, or AI assistance is relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable: This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable: As mentioned, for this physical medical device, "ground truth" relates to objective engineering and biological safety standards (e.g., force measurements meeting a minimum threshold, biological assays showing no toxicity).

8. The sample size for the training set

  • Not Applicable: This is not an AI/ML device.

9. How the ground truth for the training set was established

  • Not Applicable: This is not an AI/ML device.

{0}------------------------------------------------

page 1(3)

K052229

510(k) Summary

.

:新闻


.

.

Coloplast Ostomy Rod

1. Submitters name, address, phone and fax numbersColoplast Corp1975 West Oak CircleMarietta, GA 30062 - 2249U.S.A.Tel.: (770) 281 8400Fax: (770) 281 8500
2. Contact person at Coloplast CorpElizabeth BootsBA, MTQuality Assurance Vice PresidentColoplast Corp, Skin Care Division1940 Commerce DriveP.O. Box 8300N. Mankato, MN 56003 - 8300 U.S.A.Tel.: (507) 386 4362Cell: (507) 351 6750Fax: (507) 345 3291Email: usbb@coloplast.com
3. Date that the 510 (k) summary was preparedSeptember 29, 2006

and the comments of the comments of

{1}------------------------------------------------

4. Name of the medical device (trade, common and classification
name)and the first to the state of the states12 11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Trade nameColoplast Ostomy Rod

Loop Ostomy Rod Common Name

Classification name Rod. Colostomy

  1. Legally marketed device to which substantial equivalence is claimed

ConvaTec Sur-Fit System K811240

  1. Description of the device

The Coloplast Ostomy Rod is a 90mm white HDPE rod with a fixed T at one end and a separate T piece that can be snaoped into the open end of the rod once it is in place. Holes in the ends of the T allow for suturing. The device is packaged in a laminated polvethylene and cas-permeable paper pouch. The product is sterilized by irradiation.

7. Intended use of the device of

Coloplast Ostomy Rod is used for loop ostomy surgery. The ostomy rod is a device that is placed through a loop of the colon brought out through the abdominal wall to temporarily keep it from slipping back through the surgical opening during the loop (colostomy or ileostomy) ostomy procedure.

  1. Technological characteristics companson to the predicate device

The ConvaTec Sur-Fit System has a size range of 65-90 mm that incluces the 90mm size for the Coloplast Ostomy Rod.

The ConvaTec Sur-Fit System has a fixed triangular shaped in a T at one end and an adjustable position triangular T at the other end to allow the T to be swiveled after insertion. It may be sutured through the triangular shapes, but does not need to be.

The Coloplast Ostomy Rod has a fixed T on one end a second T piece that can be snapped onto the opposite end of the rod once it is in place. It may be sutured through holes in the ends of the T arms, but does not need to be. Both are produced by injection molding,

The Coloplast Ostomy Rod and the ConvaTec Sure-Fit System are labeled as sterile by irradiation and for single use only.

Flexural strength and flexibility were compared with results for the Coloplast Ostomy Rod and the Conva-Tec Sure-Fit System giving similar results both well above the estimated effect from the colon.

The Coloplast Ostomy Rod is made from HDPE. This is not the same material used for the ConvaTec Sure-Fit System, but is widely used in the medical industry and is shown to meet biocompatibility requirements.

{2}------------------------------------------------

9. Non-clinical performance data

The Coloplast Ostomy Rod has been tested for tensile strength, flexural strength, The Colopiast Ostonly Rod has been testour testour wonitored through ongoing testing.

The Coloplast Ostomy Rod has been tested for biocompatibility, cytotoxicity, Trie Colopiast Ostomy Nod has bron 10993-01 Biological Evaluation of il htation and Sensitization do roquired by 100 - 100 all critieria for acceptance were met.

Leachable profiles were tested for two lots of ostomy rods one at baseline and one near expiration and there was no significant change associated with time.

  1. Conclusion 1 3 % ( 1 % 1 % 1

The Coloplast Ostomy Rod is similar in size, construction and design to the ConvaTec Sur-Fit System. The intended uses and classifications are the same.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in a simple, sans-serif font.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Ms. Elizabeth Boots Vice President, Quality Assurance Coloplast Corporation 1940 Commerce Drive N MANKATO MN 56002-8300

OCT 1 3 2006

Re: K052229

Trade/Device Name: Coloplast Ostomy Rod, Model 12814 Regulation Number: 21 CFR §876.4270 Regulation Name: Colostomy rod Regulatory Class: II Product Code: EZP Dated: September 29, 2006 Received: October 3, 2006

Dear Ms. Boots:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval); it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/10 description: The image shows a circular logo with the text "FDA Centennial" and the years "1906-2006". The letters "FDA" are prominently displayed in the center of the logo. The logo is surrounded by a circular border with additional text, though the specific words are difficult to discern due to the image quality. The overall design commemorates the centennial anniversary of the FDA.

Protecting and Promoting Public Health

{4}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (31 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy Cbrogdon

Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Statement of Indications for Use

510(k) Number (if known): Not known

Device Name: Coloplast Ostomy Rod

Indications for Use:

Coloplast Ostomy Rod is for use in loop ostomy surgery. The ostomy rod is a device that is placed through a loop of the colon brought out through the abdominal wall to temporarily keep it from slipping back through the surgical opening during the loop (colostomy) ostomy surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

ﻟﻤﺴﺎ Prescription Use_ (Per 21 CFR 801.109

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

David br. Heymann
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 51U(k) Number

§ 876.4270 Colostomy rod.

(a)
Identification. A colostomy rod is a device used during the loop colostomy procedure. A loop of colon is surgically brought out through the abdominal wall and the stiff colostomy rod is placed through the loop temporarily to keep the colon from slipping back through the surgical opening.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.