K Number
K052109

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-08-11

(7 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Small Organ (breast, thyroid, testicles, prostate)

Device Description

The device referenced in this Submission is a highly portable, software-controlled, diagnostic ultrasound system with accessories. This Submission does not include any technological or feature changes from the previously cleared SonoSite devices or transducers.

AI/ML Overview

The provided text is a 510(k) Summary of Safety and Effectiveness for a diagnostic ultrasound system. It addresses an application to add a new clinical application ("Small Organ") to an existing device.

Here's an analysis of the acceptance criteria and study information, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes the device's compliance with various safety and performance standards for ultrasound systems and medical devices. However, it does not specify quantitative performance acceptance criteria (e.g., sensitivity, specificity, resolution) for the imaging quality relevant to the new "Small Organ" application, nor does it provide a table of reported device performance against such criteria.

Instead, the acceptance criteria are implicitly that the device continues to meet the safety and effectiveness standards of its predicate device and applicable recognized consensus standards, with the new clinical application being demonstrated as substantially equivalent.

Acceptance Criteria (Implicit)Reported Device Performance
Conformance to applicable medical device safety standards (acoustic output, biocompatibility, cleaning/disinfection, thermal, electrical, mechanical safety).The SonoSite Hand-Carried Ultrasound System (C1 Series) was evaluated for academic output, biocompatibility, cleaning and disinfection effectiveness, as well as thermal, electrical, and mechanical safety, and was found to conform to applicable medical device safety standards (listed in Section 4). These reports were included in previous predicate submissions.
Substantial equivalence in safety and effectiveness to the predicate device (Advanced Technology Laboratories HDI 5000 Ultrasound System (K011224)) for the "Small Organ" application.The applicant believes that the testing and analysis (referencing previously submitted reports and the fact that the added indication is not novel) demonstrates that the SonoSite Hand-Carried Ultrasound System (C1 Series) and the "Small Organ" indication on the C60/5-2 5.0-2.0 MHz Curved Array transducer are substantially equivalent with respect to safety and effectiveness to the identified predicate device. No new clinical testing was required as the added indication is not a novel indication and the anatomical site is amenable to current transducer and post-processing ultrasound technology.

2. Sample Size Used for the Test Set and Data Provenance:

The document explicitly states: "No additional clinical testing is required, as the added indication for use is not a novel indication as shown by the predicate devices in Section 3."

This indicates there was no specific clinical test set used for this particular submission to evaluate the "Small Organ" application. The data provenance is effectively based on the previously cleared predicate device and the existing SonoSite system.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Since no new clinical test set was required and no clinical study was performed for this submission, there were no experts specifically engaged to establish ground truth for a test set related to the "Small Organ" application. The determination of substantial equivalence relies on existing regulatory clearances and technical specifications.

4. Adjudication Method for the Test Set:

Not applicable, as no new clinical test set was used for this submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

Not applicable. This submission is for an ultrasound system, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This is for a diagnostic ultrasound system, not a standalone algorithm.

7. The Type of Ground Truth Used:

The ground truth for this submission is based on the established safety and effectiveness of the predicate device (Advanced Technology Laboratories HDI 5000 Ultrasound System (K011224)) which already included "Small Organ" as a clinical application. The argument is that the SonoSite system, with its existing cleared technologies, is capable of performing this application equivalently. The document implies that the "anatomical site is amenable to current transducer and post-processing ultrasound technology available with the SonoSite Hand-Carried Ultrasound System (C1 Series) and predicate devices."

8. The Sample Size for the Training Set:

Not applicable. This submission does not involve a machine learning or AI model that requires a training set. The device is a traditional ultrasound system.

9. How the Ground Truth for the Training Set Was Established:

Not applicable.

{0}------------------------------------------------

AUG 1 1 2005

SonoSite, Inc. 21919 30th Drive SE WA 98021-3904 USA

Telephone:1.425.951.1200 imile:1.425.951.1201

www.sonosite.com

1052109

510(K) Summary of Safety and Effectiveness

This summary of safety and effectiveness is provided as part of this Pre-market Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1) Submitter's name, address, telephone number, contact person:

SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904

Corresponding Official:Daina L. GrahamVice President, Regulatory Affairs and Quality Assurance
E-mail:Daina.Graham@sonosite.com
Telephone:(425) 951-1275
Facsimile:(425) 951-1201
Date prepared:July 29, 2005
    1. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:

Common/ Usual Name

Diagnostic Ultrasound System with Accessories

Proprietary Name

SonoSite Hand-Carried Ultrasound System (C1 Series) (subject to change)

Classification Names

NameFR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

510(K) Summary Of Safety And Effectiveness

{1}------------------------------------------------

3) ldentification of the predicate or legally marketed device:

This 510(k) adds the clinical application "Small Organ (breast, thyroid, testicles, prostate)" to the indications for use of the transducer C60/5-2 5.0-2.0 MHz Curved Arrav on the SonoSite Hand-Carried Ultrasound System (C1 Series) (K010374 and K014116). No other change, including technological change was made to this system. A predicate device that has "Small Organ" as a clinical application or indication is Advanced Technology Laboratories HDI 5000 Ultrasound System (K011224).

Device Description: 4)

The device referenced in this Submission is a highly portable, software-controlled, diagnostic ultrasound system with accessories. This Submission does not include any technological or feature changes from the previously cleared SonoSite devices or transducers.

By this Submission, the clinical application "Small Organ (breast, thyroid, testicles, prostate)" is being added to previously cleared indications for use to the following transducer:

SystemTransducerTransducerTypeFrequencyRange
SonoSite Hand-Carried Ultrasound SystemC60/5-2Curved Array5.0 - 2.0 MHz

FDA Recognized Consensus Standards that SonoSite Hand-Carried Ultrasound System (C1 Series) is designed to comply with, as applicable to its features, are listed below:

Applicable Standards

Reference No.Title
AIUMAIUM Medical Ultrasound Safety, American Institute of Ultrasound in Medicine(1994)
ANSI/AAMI EC53ECG Cables and Electrodes except for sections 4.4 and 4.5.9 (1995)
AAMI/ANSI/ISO 10993-1Biological evaluation of medical devices - Part 1: Evaluation and testing (ISO10993-1:1997)
AAMI/ANSI/ISO 30993-4Biological evaluation of medical devices - Part 4: Selection of tests forinteractions with blood (ISO 10993-4:1992)
AAMI/ANSI/ISO 10993-5Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity(ISO 10993-5:1999)
AAMI/ANSI/ISO 10993-10Biological evaluation of medical devices - Part 10: tests for irritation andsensitization (ISO 10993-10:1995)
AAMI/ANSI/ISO 10993-11Biological evaluation of medical devices - Part 11: Tests for systematic toxicity(ISO 10993-11:1993)

{2}------------------------------------------------

Reference No.Title
AAMI/ANSI/ISO 10993-12Biological evaluation of medical devices - Part 12: sample preparation andreference materials (ISO 10993-12:1996)
EN 980 A1Graphical symbols for use in the labeling of medical devices (2003)
IEC 60601-1Medical electrical equipment. Part 1: General requirements for safety - IEC 601-1:1988
IEC 60601-1/A1Medical electrical equipment. Part 1: General requirements for safety - IEC 601-1:1988/A1:1991
IEC 60601-1/A2Medical electrical equipment. Part 1: General requirements for safety - IEC 601-1:1988/A2:1995 + corrigendum June 1995
IEC 60601-1-1Medical electrical equipment. Part 1: General requirements for safety - 1.Collateral standard: Safety requirements for medical electrical systems - IEC601-1-1:2000
IEC 60601-1-2Medical electrical equipment - Part 1: General requirements for safety; 2.collateral standard: electromagnetic compatibility; requirements and tests - IEC60601-1-2:200
IEC 60601-1-4Medical electrical equipment - Part 1: General requirements for safety - 4.Collateral standard: Programmable electrical medical systems - IEC 60601-1-4:1996 Amendment A1
IEC 60601-2-25Medical Electrical Equipment-Part 2. Particular Requirements for Safety-Section 25. Specification for Electrocardiographs. (1999
IEC 60601-2-37Medical Electrical Equipment - Part 2-37; Particular requirements for the safetyof ultrasonic medical diagnostic and monitoring equipment (2001)
NEMA UD 2-2004Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
NEMA UD 3-2004Standard for Real-Time Display of Thermal and Mechanical Acoustic OutputIndices on Diagnostic Ultrasound Equipment, American Institute of Ultrasoundin Medicine

The following are other standards with which the SonoSite Hand-Carried Ultrasound System (C1 Series) complies, where applicable:

  • 行: -

i

Miscellaneous Design Standards

ﻳﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ

:

:

Reference No.Title
ASTM D5276-98Standard Test Methods for Drop Test of Loaded Containers by Free Fall(1998)
ASTM D999-96Standard Methods for Vibration Testing of Shipping Containers (1996)
CISPR 11Industrial, Scientific and Medical (ISM) Radio-Frequency Equipment -Electromagnetic Disturbance Characteristics - Limits and Methods ofMeasurement (2003)
JIS-T-601-1Japanese Standards for Medical Electrical Equipment
RTCA/DO160DRadio Technical Commission for Aeronautics, Environmental Conditionsand Test Procedures for Airborne Equipment, Section 21.0 Emission ofRadio Frequency Energy, Category B (1997)
UL 60601-1Underwriters Laboratories, Medical Electrical Equipment-Part 1:General Requirements for Safety (2003)
UL 94Underwriters Laboratories, Inc., Tests for Flammability of PlasticMaterials for Parts in Devices and Appliances, 5th Edition

{3}------------------------------------------------

Intended Use: 5)

The intended uses of the SonoSite ultrasound system referenced herein remain The Intention accords of cleared indications, except for the addition of "Small Organ dhenanged from proviously clource to the indications for use of the transducer C60/5-2 (Dreast, Infrold, Coclinion SonoSite Hand-Carried Ultrasound System (C1 Series) as a new intended use.

The intended uses of the SonoSite Hand-Carried Ultrasound System (C1 Series) as defined by FDA guidance documents, are:

Fetal - OB/GYNTrans-rectal
AbdominalTrans-vaginal
Intra-operative(Abdominal organs and vascular)Musculo-skel. (Conventional)
Intra-operative (Neuro.)Musculo-skel. (Superficial)
PediatricCardiac Adult
Small Organ(breast, thyroid, testicles, prostate)Cardiac Pediatric
Neonatal CephalicPeripheral vessel

Technologica! Characteristics: 6)

There are no technological or feature changes in this Submission to any of the legally marketed ultrasound systems, transducers, or accessories identified in Section 3 of this 1 1 2 3 3 Summary.

7) Testing:

The referenced SonoSite Hand-Carried Ultrasound System (C1 Series) has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, as well as thermal, electrical and mechanical safety, and has been found to conform to applicable medical device safety standards, as referenced in Section 4. Reports were previously included in the referenced predicate submissions. No additional clinical testing is required, as the added indication for use is not a novel indication as shown by the predicate devices in Section 3. The anatomical site is amenable to current transducer and post-processing ultrasound technology available with the SonoSite Hand-Carried Ultrasound System (C1 Series) and predicate devices. Additionally, the modes of operation that are indicated with this clinical application for the SonoSite Hand-Carried Ultrasound System (C1 Series) is consistent with those identified with the predicate devices.

Conclusion: 8)

SonoSite believes that the testing and analysis described in Section 7 demonstrates that the SonoSite Hand-Carried Ultrasound System (C1 Series) (K010374 and that the ochoone frank ounts and Organ (breast, thyroid, testicles, prostate)" indication on the transducer C60/5-2 5.0-2.0 MHz Curved Array, are substantially equivalent with respect to safety and effectiveness to the predicate device identified in Section 3.

510(K) Summary Of Safety And Effectiveness

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a caduceus, with three curved lines representing the human form and a wavy line at the bottom.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 1 2005

SonoSite, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K052109

Trade Name: SonoSite Hand-Carried Ultrasound System (C-1 Series) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX . Dated: August 2, 2005 Received: August 4, 2005

Dear Mr. Job:

:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the SonoSite Hand-Carried Ultrasound System (C-1 Series), as described in your premarket notification:

Transducer Model Number

C60/5-2. 5.0-2.0 MHz Curved Array

{5}------------------------------------------------

.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification: The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

{6}------------------------------------------------

Page 3 - Mr. Job

.

.

.

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

·

Sincerely yours,

David A. Lyon

or Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

. . . . . . :

.

.

.

11 - 1 - 1 - 1 - 1 -

. :

{7}------------------------------------------------

System:SonoSite Hand-Carried Ultrasound System (C1 Series)
Transducer:N/A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Spec.)Other(Spec.)
OphthalmicOphthalmic
FetalPPPB+M; B+PWDNote 1
AbdominalPPPPB+M; B+PWD;B+CWDNote 1
Intra-operative(Abdominal organs andvascular)PPPB+M; B+PWDNote 1
Fetal ImagingIntra-operative (Neuro.)PPPPB+M; B+PWD;B+CDNote 1, 2
LaparoscopicPPPB+M; B+PWDNote 1
& OtherPediatricPPPPB+M; B+PWD;B+CWDNote 1
Small Organ (breast,thyroid, testicles,prostate)PPPB+M; B+PWDNote 1
Neonatal CephalicPPPB+M; B+PWDNote 1
Adult Cephalic
Trans-rectalPPPB+M; B+PWDNote 1
Trans-vaginalPPPB+M; B+PWDNote 1
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel. (Convent.)PPPB+M; B+PWDNote 1
Musculo-skel. (Superfic.)PPPB+M; B+PWDNote 1
Intra-luminal
Other (spec.)
Cardiac AdultPPPPB+M; B+PWD;B+CWDNote 1
CardiacCardiac PediatricPPPPB+M; B+PWD;B+CWDNote 1
Trans-esophageal (card.)
Other (spec.)
PeripheralVesselPeripheral vesselPPPB+M; B+PWDNote 1
Other (spec.)

Table 4.3- 1 Diagnostic Ultrasound Indications for Use Form - System

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

: 上一篇:

Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging, and imaging for guidance of biopsy cleared in K014116. Imaging uidance for peripheral nerve block procedures and imaging to assist in the placement of needles and catheters in vascular or other analomical structures was previously cleared in K033367.

Note 2: The clinical application Intra-operative (Neuro.) was previously cleared in K043452

Prescription Use (Per 21 CFR 801.109)

Daniel A. Sessa

Bearroriuctive Abdon

{8}------------------------------------------------

System:SonoSite Hand-Carried Ultrasound System (C1 Series)
Transducer:C60/5-2 5.0-2.0 MHz Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track I only)Specific(Tracks I & III)BMPWDCWDColor DopplerCombined (Spec.)Other (Spec.)
OphthalmicOphthalmic
FetalPPPB+M; B+PWD;Note 1
AbdominalPPPB+M; B+PWD;Note 1
Intra-operative(Abdominal organs and vascular)PPPB+M; B+PWD;Note 1
Fetal Imaging& OtherIntra-operative (Neuro.)
LaparoscopicPediatricPPPB+M; B+PWD;Note 1
Small Organ (breast,thyroid, testicles,prostate)NNNB+M; B+PWD;Note 1
Neonatal CephalicAdult CephalicTrans-rectalTrans-vaginalTrans-urethralTrans-esoph. (non-Card.)Musculo-skel. (Convent.)Musculo-skel. (Superfic.)Intra-luminalOther (spec.)
CardiacCardiac AdultPPPB+M; B+PWD;Note 1
Cardiac PediatricPPPB+M; B+PWD;Note 1
Trans-esophageal (card.)Other (spec.)
PeripheralVesselPeripheral vesselOther (spec.)

Table 4.3- 2 Diagnostic Ultrasound Indications for Use Form - C60/5-2 Transducer

previously cleared by FDA; E= added under Appendix E

Additional Comments:

Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and Directional Color Power Dopler, 3-D Insue Farmonic maging, imaging for guidance of biopsy and institution of assist in the placement of needes and cather and of the reasonable minging, muging of gabanes of blogs and maging to assist in the placements of the proceed

Prescription Use (Per 21 CFR 801.109)

Daniel A. Graham

(Division Sign-Off) Division of Reproductive. and Radiological Devices 510(k) Number

3:42 :

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.