K Number
K052108

Validate with FDA (Live)

Date Cleared
2005-12-07

(126 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BGP-100 blood glucose and blood pressure measurement system consists of a meter with wrist cuff and test strips. The system is intended for use in the quantitative measurement of glucose in whole blood taken from the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes mellitus, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates.

Also the system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillometric method of measurement.

Device Description

BGP-100 Sein Blood Glucose & Pressure Monitor system BGP-100 combines the function of a blood glucose monitoring system and a blood pressure meter in one unit. Supplies with the meter are test strips, lancets, lancing device, storage case and control solution. The blood glucose monitoring system is turned on by strip insertion; Once the user confirms that the numerical code on the display, and strip bottle all match, a test strip is inserted into the meter and glucose testing can proceed. The user then supplies finger-tip blood or control solution to the strip and the meter starts the measurement, which is completed in 5 seconds. Sein Electronics Co., Ltd has provided instructions and illustrations explaining, that the blood drop will be pulled into the strip sample entry by capillary action until meter beeps and the result will be show on liquid crystal display ("LCD") after 5 seconds. Results are automatically stored in the meter's memory for tracking purposes. The capacity of memory is 150 blood glucose values and average value of past 14 days or 30days. Control Solution is sold separately from the kit. BGP-100 adopts the wrist type cuff for blood pressure meter part. The cuff and control unit are combined into a single wrist-mounted assembly. The user interface panel has power switch, mode switch, memory switch, for blood pressure meter part and LCD for displaying the systolic blood pressure, diastolic blood pressure, pulse rate, date and time. This device has the memory function that permits memory and display of the 60 most recent measurement results. User should wrap the pressure cuff around the wrist with the palm facing up and pushes the "Power" button. The cuff will be full with appropriate air pressure automatically. Pressurization is automatically governed. If the initial inflation pressure (180mmHg) is inadequate for measurement, i.e. lower that the patient's systolic pressure, the pump will automatically repressurize to a preset level (30 mmHg) above the initial level. Symbols in the LCD indicate pressurization status at all times. The air of cuff is automatically deflated during blood pressure measurement. No special training beyond basic ability to follow instruction is required. Since the products are designed for home use, detailed instructions on avoidance of practices that adversely affect the accuracy of measurements are included in the instruction manual.

AI/ML Overview

The provided text is a 510(k) summary for the "Blood Glucose and Blood Pressure Monitor System, Model BGP-100." While it describes the device and its intended use, it does not include detailed information on acceptance criteria for device performance or a study demonstrating that these criteria have been met. It primarily focuses on demonstrating substantial equivalence to predicate devices, but the performance data itself is absent from this document.

Therefore, I cannot fulfill your request for:

  • A table of acceptance criteria and reported device performance.
  • Sample sizes used for the test set and data provenance.
  • Number of experts used to establish ground truth and their qualifications.
  • Adjudication method.
  • MRMC comparative effectiveness study results.
  • Standalone performance details.
  • Type of ground truth used.
  • Sample size for the training set.
  • How ground truth for the training set was established.

This information would typically be found in a more detailed performance study report within the full 510(k) submission, not necessarily in the publicly available summary. The summary states "Performance" as a technological characteristic in a comparison table but leaves it blank, suggesting the details are in the main submission.

{0}------------------------------------------------

DEC 7 2005

K052/08

510(k) Summary

As Required by 21 section 807.92 ( c )

  • Sein Electronics Co., Ltd 1 - Submitter Name: 133-3. Pyungchon-dong, Anyang-city, Kyunggi-do, Korea 2- Address: (82) 31-421-0389 3- Phone: (82) 31-467-2107 4- Fax: 5- Contact Person: Tae voung Choi August 1, 2005 6- Date summary prepared: Mansour Consulting LLC 7 - Official Correspondent: 1308 Morningside Park Dr. Alpharetta, GA 30022 USA 8- Address: 770-777-4146 9- Phone: 678-623-3765 10- Fax: Jay Mansour, President 11- Contact Person: 12- Device Trade or Proprietary Name: BLOOD GLUCOSE AND BLOOD PRESSURE MONITOR SYSTEM, MODEL BGP-100 Blood Glucose Monitor, Blood glucose test strips, quality 13- Device Common or usual name: control material, and Blood Pressure Monitor 14- Device Classification Name: System, test, blood glucose, over the counter Glucose oxidase, glucose Single (specified) analyte controls System, measurement, blood-pressure, non-invasive 15- Substantial Equivalency is claimed against the following device: Blood Pressure Monitor, Model SE-311, manufactured by Sein Electronics Co., Ltd. 510k #K042138 Clever Chek TD-3213, 510k# K042795
    One Touch Ultra Blood Glucose monitoring system, 510k #K002134

16- Description of the Device:

BGP-100 Sein Blood Glucose & Pressure Monitor system BGP-100 combines the function of a blood glucose monitoring system and a blood pressure meter in one unit. Supplies with the meter are test strips, lancets, lancing device, storage case and control solution. The blood glucose monitoring system is turned on by strip insertion; Once the user confirms that the numerical code on the display, and strip bottle all match, a test strip is inserted into the meter and glucose testing can proceed. The user then supplies finger-tip blood or control solution to the strip and the meter starts the measurement, which is completed in 5 seconds. Sein Electronics Co., Ltd has provided instructions and illustrations explaining, that the blood drop will be pulled into the strip sample entry by capillary action until meter beeps and the result will be show on liquid crystal display ("LCD") after 5 seconds. Results are automatically stored in the meter's memory for tracking purposes. The capacity of memory is 150 blood glucose values and average value of past 14 days or 30days. Control Solution is sold separately from the kit. BGP-100 adopts the wrist type cuff for blood pressure meter part. The cuff and control unit are combined into a single wrist-mounted assembly. The user interface panel has power switch, mode switch, memory switch, for blood pressure meter part and LCD for displaying the systolic blood pressure, diastolic blood pressure, pulse rate, date and time. This device has the memory function that permits memory and display of the 60 most recent measurement results. User should wrap the pressure cuff around the wrist with the palm facing up and pushes the "Power" button. The cuff will be full with appropriate air pressure automatically. Pressurization is automatically governed. If the initial inflation pressure (180mmHg) is inadequate for measurement, i.e. lower that the patient's systolic pressure, the pump will automatically repressurize to a preset level (30 mmHg) above the initial level. Symbols in the LCD indicate pressurization status at all times. The air of cuff is automatically deflated during blood pressure measurement. No special training beyond basic ability to follow instruction is required. Since the products are designed for home use, detailed instructions on avoidance of practices that adversely affect the accuracy of measurements are included in the instruction manual.

Image /page/0/Picture/8 description: The image shows the handwritten text "PAGE 13". The text is written in a cursive style. The number 13 is written to the right of the word "PAGE".

{1}------------------------------------------------

17- Intended use of the device: (refer to FDA form attached)

The BGP-100 blood glucose and blood pressure measurement system consists of a meter with rric DOF Too brood groot of the system is intended for use in the quantitative measurement of whice our and toot othpost the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes mellitus, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates. Also the system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillometric method of measurement.

18- Safety and Effectiveness of the device:

This device is safe and effective as the predicate device cited above. This is better expressed in the tabulated comparison (Paragraph 14 below)

14- Summary comparing technological characteristics with predicate device:

Please find below a tabulated comparison supporting that this device is substantially equivalent to other medical devices in commercial distribution. Also, Equivalency overview chart path is attached. Indeed, this device is SUBSTANTIALLY EQUIVALENT to the predicate device. Refer to the explanations/details within the main submission.

FDA file reference number510k # K042138510k # K042795 and 002134
TECHNOLOGICALCHARACTERISTICSComparison result
Indications for useIdentical
Target populationIdentical
Design
Materials
Performance
Sterility
Biocompatibility
Mechanical safety
Chemical safety
Anatomical sites
Human factorsIDENTICALSIMILAR
Energy used and/ordelivered
Compatibility withenvironment and otherdevices
Where used
Standards met
Electrical safety
Thermal safety
Radiation safety

Refer to the submission for more details.

Page 14

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

2005 DEC 7

Sein Electronics Co., Ltd. c/o Mr. Jay Mansour, MSQA, BE, LA, RAC President Mansour Consulting LLC 1308 Morningside Park Drive Alpharetta, GA 30022

Re: K052108

Trade/Device Name: Blood Glucose and Blood Pressure Monitor System, Model BGP-100 Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA, JJX, DXN Dated: October 26, 2005 Received: October 31, 2005

Dear Mr. Mansour,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow your to begin inding of substantial equivalence of your device to a legally promative nevice results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I vision of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Guti

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K052108

Device Name: Blood Glucose and Blood Pressure Monitor System, Model BGP-100

Indications For Use:

The BGP-100 blood glucose and blood pressure measurement system consists of a meter with wrist cuff and test strips. The system is intended for use in the quantitative measurement of glucose in whole blood taken from the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes mellitus, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates.

Also the system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillometric method of measurement.

Prescription Use (Part 21 CFR 801 Subpart D) ANDIOR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C. Benson
Division Sign-Off

office of in vitro Diagnostic Device Evaluation and Safety

Page 1 of __ 1________________________________________________________________________________________________________________________________________________________________

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.