K Number
K052074

Validate with FDA (Live)

Date Cleared
2005-09-01

(31 days)

Product Code
Regulation Number
886.5918
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.

JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.

JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.

JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.

Device Description

Cleaning, Disinfecting and Storing Solutions; and Wetting, Lubricating and Rewetting Solutions for Rigid Gas Permeable Contact Lenses.

AI/ML Overview
  1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria with quantified performance metrics to be considered "met" in the traditional sense for a medical device. Instead, the approach taken for these contact lens solutions is a demonstration of substantial equivalence to existing legally marketed predicate devices. The "performance" is implicitly deemed acceptable if the new products exhibit comparable chemical, toxicological, and microbiological characteristics to the predicates.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    ChemistryCompatibility with RGP lenses; effective daily cleanerJSZ-Stat-Tree/JSZ-Wet-Tree and JSZ-Stat-Or/JSZ-Wet-Or systems were compatible with all groups of RGP lenses. Both JSZ-Stat-Tree and JSZ-Stat-Or were shown to be effective daily cleaners.
    ToxicologyComparable to or passing established safety tests for predicates- Cytotoxicity: NA (cleaning solutions), Passed (wetting solutions) - Acute Oral Toxicity: Passed - Ocular Irritation (21d rabbit): Passed - Ocular Irritation (72hr rabbit): Passed - Anesthetic Effect: NA (cleaning solutions), Passed (wetting solutions) - Guinea Pig Maximization: Passed - Corneal Epithelial Wound Healing: Passed - Corneal Penetration: Passed
    MicrobiologyAcceptable disinfection efficacy, neutralizer efficacy, preservative efficacy, and sterility/stability- Disinfection Efficacy: Passed (JSZ-Stat-Tree, JSZ-Stat-Or), NA (wetting solutions) - Neutralizer Efficacy: Passed (JSZ-Stat-Tree, JSZ-Stat-Or), NA (wetting solutions) - Preservative Efficacy: Passed (all products) - Sterility/Stability: Passed (all products)
    Clinical StudiesNot required if formulations are identical to previously approved devicesNo clinical data included; explicitly stated "Clinical data is not required for solutions using the same active ingredients at the same concentrations as currently marketed products."
    Substantial EquivalenceIdentical to or substantially equivalent to predicate devices with similar formulations and indicationsSolutions are identical to DeStat-3, DeStat-4, Stay-Wet 3, and Stay-Wet-4 solutions; substantially equivalent to Optimum by Lobob and Claris Cleaning and Soaking Solution.
  2. Sample size used for the test set and the data provenance:
    The document does not specify discrete "test sets" in the context of a typical clinical study. The toxicology and microbiology tests were likely conducted on laboratory samples of the solutions themselves, not on human subjects or clinical data sets. The provenance of the data is from laboratory testing of the JSZ products, and comparisons to legally marketed predicate devices in the US market. The study is not a clinical study; it's a pre-market notification (510(k)) relying on substantial equivalence to existing products, with supportive lab data.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable (N/A). This submission relies on laboratory testing and comparison to predicate devices, not on expert consensus for a "ground truth" in the diagnostic sense. The regulatory body (FDA) reviews the submitted data.

  4. Adjudication method for the test set:
    N/A. There was no clinical test set requiring adjudication in the context of this 510(k) submission.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    N/A. This is a submission for contact lens solutions, not an AI-powered diagnostic device. No MRMC study was performed.

  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    N/A. This is not an algorithm or AI device. The "performance" refers to the solutions' chemical, toxicological, and microbiological properties.

  7. The type of ground truth used:
    The "ground truth" for demonstrating safety and effectiveness relies on established scientific methods for toxicology (e.g., rabbit ocular irritation, acute oral toxicity) and microbiology (e.g., disinfection efficacy, preservative efficacy), as well as the prior regulatory acceptance of chemically identical or substantially equivalent predicate devices. There is no single "ground truth" in the context of pathology or outcomes data as would be found in a diagnostic study.

  8. The sample size for the training set:
    N/A. There is no "training set" in the context of an AI/ML algorithm or a traditional clinical trial here. The submission is based on laboratory tests of the formulated solutions.

  9. How the ground truth for the training set was established:
    N/A. As there is no training set for an AI/ML model, there is no ground truth established for such a set.

{0}------------------------------------------------

SEP 0 1 2005

K052074

JSZ-Stat-Tree, JSZ-Stat-Or, JSZ-Wet-Tree, JSZ-Wet Or 510(k)

SUMMARY, FREEDOM OF INFORMATION

JSZ-STAT-TREE, AND JSZ-STAT-OR, CLEANING, DISINFECTING AND STORING SOLUTIONS;

AND JSZ-WET-TREE AND JSZ-WET-OR, WETTING, LUBRICATING AND REWETTING SOLUTIONS

FOR RIGID GAS PERMEABLE CONTACT LENSES

1. Common/Usual NamesJSZ-Stat-Tree, and JSZ-Stat-Or, Cleaning, Disinfecting and Storing Solutions;and JSZ-Wet-Tree and JSZ-Wet-Or, Wetting, Lubricating and Rewetting Solutions
2. Trade/ProprietaryJSZ-Stat-Tree, and JSZ-Stat-Or, Cleaning, Disinfecting and Storing Solutions;and JSZ-Wet-Tree and JSZ-Wet-Or, Wetting, Lubricating and Rewetting Solutions
3. FDA ClassificationClass II(Performance Standards)21 CFR 886.5918Rigid Gas Permeable (RGP) contact lens solutions
4. Performance standardsNone established under section 514
  1. Substantial equivalence

These products are substantially equivalent to several currently marketed products, such as DeStat-3 and Destat 4 Cleaning, Disinfecting and Storage Solutions, Stay-Wet 3 and Stay-Wet 4 Wetting and Rewtting Drops. Optimum by Lobob Cleaning, Disinfecting and Storage Solution, Wetting and Rewetting Drops, and Claris Cleaning and Soaking Solution.

6. Indications for Use

JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.

JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.

JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.

JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.

{1}------------------------------------------------

SUMMARY

Contained in this submission are comparisons of the products to the predicate devices, information on the chemistry and manufacturing, toxicology and microbiology. Because the predicate devices, with identical formulations, have been previously approved for market, and have been marketed over the last decade, no clinical data is included in this submission.

  • I. Chemistry
    The JSZ-Stat-Tree/JSZ-Wet-Tree and the JSZ-Stat-Or/JSZ-Wet-Or systems were compatible with all groups of RGP lenses. Both ISZ-Stat-Tree and JSZ-Stat-Or were shown to be effective daily cleaners.

Toxicology II.

The toxicological testing of the the JSZ-Stat-Tree/JSZ-Wet-Tree, and JSZ-Stat-Or/JSZ-Wet-Or systems for the cleaning, disinfecting and storing, and wetting, lubricating and rewetting of rigid gas permeable lenses is summarized below.

TestDeStat-3/DeStat-4Stay-Wet-3/Stay-Wet 4
CytotoxicityNAapassed
Acute Oral Toxicitypassedpassed
Ocular Irritation (21d rabbit)passedb
Ocular Irritation (72hr rabbit)passedb
Anesthetic EffectNAapassed
Guinea Pig Maximizationpassedc
Corneal Epithelial WoundHealingpassedc
Corneal Penetrationpassedc

a NA - test not applicable to cleaning solutions

b Test used JSZ-Stat-Tree/JSZ-Wet-Tree, or JSZ-Stat-Or/JSZ-Wet-Or

° Tests on benzyl alcohol

III. Microbiology

The microbiological testing of the the JSZ-Stat-Tree/JSZ-Wet-Tree, and JSZ-Stat-Or! JSZ-Wet-Or systems for the cleaning, disinfecting and storing, and wetting, lubricating and rewetting of rigid gas permeable lenses is summarized below.

। (

{2}------------------------------------------------

JSZ-Stat-Tree, JSZ-Stat-Or, JSZ-Wet-Tree, JSZ-Wet Or 510(k)

TestJSZ-Stat-TreeJSZ-Stat-OrJSZ-Wet-TreeJSZ-Wet-Or
Disinfection Efficacy(microbial, viral)passedpassedNAaNA
Neutralizer EfficacypassedpassedNANA
Preservative Efficacypassedpassedpassedpassed
Sterility/Stabilitybpassedpassedpassedpassed

ª NA - Not applicable to rewetting solutions

6 Protocol for real-time stability of JSZ products included

IV. Clinical Studies

Clinical data is not required for solutions using the same active ingredients at the same concentrations as currently marketed products.

V. Substantial Equivalence

The solutions are identical to the DeStat-3, DeStat-4, Stay-Wet 3, and Stay-Wet-4 solutions previously approved, and substantially equivalent to the Optimum by Lobob Cleaning, Disinfecting and Storage Solution, Optimum by Lobob Wetting and Rewetting Drops, and Claris by Menicon Cleaning and Soaking Solution.

VI. Manufacturing

Manufacturing details are included in the manufacturing section.

VII. Shelf Life

The shelf life is assumed to be the same as in the predicate filings and a protocol for on-going real time shelf-life is included.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 0 1 2005

John Szabocsik, Ph.D. President SZABOCSIK AND ASSOCIATES 203 North Wabash Avenue Chicago, Illinois 60601

Re: K052074

Trade/Device Name: JSZ-Stat-Tree, and JSZ-Stat-Or, Cleaning, Disinfecting and Storing Solutions; and JSZ-Wet-Tree and JSZ-Wet-Or, Wetting, Lubricating and Rewetting Solutions

Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable (RGP) contact lens solutions Regulatory Class: Class II Product Code: MRC Dated: July 29, 2005 Received: August 1, 2005

Dear Dr. Szabocsik :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - John Szabocsik, Ph.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

David M. Whipple

David M. Whipple Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

JSZ-Stat-Tree, JSZ-Stat-Or, JSZ-Wet-Tree, JSZ-Wet Or 510(k)

510(k) NUMBER (IF KNOWN) K052074

DEVICE NAME

JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution

JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops.

INDICATIONS FOR USE

JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution is indicated for use in the JSZ-Stat-Tree Cleaning, Disameeting, Sessings of Stages
cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.

JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution is indicated for use in the JSZNal-Of Cleaning, Districting, Stellige Beer ago Bearther and silicone acrylate (RGP) contact lenses.

JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops are indicated for use to wet JSZ-Wel-Tree Wetting. Dubrieading acrylate rigid gas permeable (RGP) contact lenses Indord/sincolle act yiate and barrest and lubricate the lens while on the eye.

JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops are indicated for use to wet JSZ-Wel-Or Welting: Edwrieauing, songlate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Amilham

Concurrence of CDRH, Office of Device Evaluation (ODE)

Over-The-Counter-Use U (Per 21 CFR 801.109) OR Prescription Use _ (Optional Format 1-2-96)

stmlest

ision of Ophthalm

10(k) Number K052074

§ 886.5918 Rigid gas permeable contact lens care products.

(a)
Identification. A rigid gas permeable contact lens care product is a device intended for use in the cleaning, conditioning, rinsing, lubricating/rewetting, or storing of a rigid gas permeable contact lens. This includes all solutions and tablets used together with rigid gas permeable contact lenses.(b)
Classification. Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”