Intended Use

The ISIS iQ UNO Handheld Electronic Breast Pump is intended to express and collect milk from the breasts of a lactating woman.

Device Description

The ISIS iQ UNO Handheld Electronic Breast Pump is a safe and effective powered breast pump used for expressing and collecting breast milk from the breasts of a lactating woman. The diaphragm within the pump is activated by a 6V DC electric motor controlled by a microprocessor. The device is comprised of a vacuum pump, 6V DC power supply, hand controller, Avent ISIS standard manual breast pump parts, and feeding bottles. The hand controller is designed to replicate the handle of the Avent standard manual ISIS Breast Pump, and is used to control the frequency and level of vacuum. A sealed AC/DC transformer or a battery pack powers the ISIS iQ UNO Breast Pump.

The ISIS iQ UNO Handheld Electronic Breast Pump operates in two modes. In the manual mode, the user controls the frequency and level of vacuum by varying the stroke/cycle of the control handle lever. After a comfortable manual pumping cycle has been established, the user engages the automatic pumping mode by depressing the control button on the hand controller. In the automatic mode, the pumping cycle is stored and controlled by the microprocessor. The user can switch between the automatic mode and manual mode at any time by repeatedly depressing the control button.

All milk-contacting and human tissue-contacting components are manufactured from materials that meet the appropriate regulations regarding food contact and/or biocompatibility.

AI/ML Overview

The provided text describes the ISIS iQ UNO Handheld Electronic Breast Pump and its 510(k) submission, not a study involving a device that meets acceptance criteria with reported performance metrics like sensitivity or specificity. This document is a regulatory submission demonstrating substantial equivalence to predicate devices, rather than a performance study.

Therefore, many of the requested sections (Table of acceptance criteria, sample sizes, number of experts, adjudication methods, MRMC study, standalone performance, ground truth types for test and training sets) cannot be fulfilled from the provided text, as they pertain to performance studies that were not conducted for this device.

Here's an analysis based on the information available:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

This information is not provided in the document. The submission focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and compliance with electrical, mechanical, environmental, food-contact, and biocompatibility standards, rather than defining and meeting specific performance acceptance criteria like accuracy, sensitivity, or specificity.

2. Sample Size Used for the Test Set and Data Provenance

No test set for performance was used as no clinical studies were performed (as stated in Section 8.9).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable as no clinical studies were performed and therefore no ground truth establishment was needed for device performance evaluation.

4. Adjudication Method for the Test Set

Not applicable as no clinical studies were performed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was done. The document explicitly states, "No clinical studies were performed."

6. Standalone (Algorithm Only) Performance Study

No standalone performance study was done. The device is a physical breast pump, not an algorithm. The document explicitly states, "No clinical studies were performed."

7. Type of Ground Truth Used

Not applicable as no clinical studies were performed.

8. Sample Size for the Training Set

Not applicable as no clinical studies were performed and this is a physical device, not an AI/ML algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable as no clinical studies were performed.


Summary of Non-Clinical Performance Data (as provided in Section 8.8):

While not framed as "acceptance criteria" for a performance study, the device met the following standards, which serve as its non-clinical acceptance criteria for regulatory clearance:

CategoryStandard Met
Electrical, Mechanical, and Environmental PerformanceIEC 60601-1: 1998 + AI: 1991, A2: 1995, Corrigendum: 1994; ANSI/UL 60601:2003; EN-60601-1-1: 1990 + Al: 1993, Al 1: 1993, A12: 1993, A12: 1995, A13: 1996, Corrigendum: 1994; CAN/CSA C22.2 No. 601.1 issued: 1994/01/05 Rev: 1997/01/01; EN60601-1-2: 2001; and FCC Part 15 for a class B device.
Material Biocompatibility and Food ContactAll milk-contacting and human tissue-contacting components meet appropriate regulations regarding food contact and/or biocompatibility.

The study that "proves the device meets the acceptance criteria" in this context is the non-clinical testing that demonstrated compliance with these listed electrical, mechanical, environmental, and material standards. This testing was conducted by the manufacturer, Avent America, Inc., and the results were submitted as part of their 510(k) premarket notification (K052047). The FDA reviewed these non-clinical performance data and the technological characteristics, concluding that the device is safe and effective and substantially equivalent to predicate devices, thus allowing it to be marketed. No specific details about the methodology or sample sizes for this non-clinical testing are provided in the summary.

{0}------------------------------------------------

K 65204

510(k) Summary 8

Submitter Name, Address, and Contact Information 8. I

Submitter:Avent America, Inc.475 Supreme DriveBensenville, IL 60106
Contact Person:Ms. Miriam SharifQuality Engineer(630) 694-7282

Summary Preparation Date 8.2

This 510(k) summary was prepared on July 28, 2005.

(866) 501-2982 Fax

8.3 Name of Device

The trade name of the device is the ISIS iQ UNO Handheld Electronic Breast Pump. The common name is Powered Breast Pump. The classification name is Powered Breast Pump.

Name of Predicate Device 8.4

Medela Pump In Style Advanced Breast Pump, by Medela Inc., K031614; and Natural Comfort/Easy Comfort Battery/Electric Breast Pump, by The First Years, K032566.

Description of Device 8.5

The ISIS iQ UNO Handheld Electronic Breast Pump is a safe and effective powered breast pump used for expressing and collecting breast milk from the breasts of a lactating woman. The diaphragm within the pump is activated by a 6V DC electric motor controlled by a microprocessor. The device is comprised of a vacuum pump, 6V DC power supply, hand controller, Avent ISIS standard manual breast pump parts, and feeding bottles. The hand controller is designed to replicate the handle of the Avent standard manual ISIS Breast Pump, and is used to control the frequency and level of vacuum. A sealed AC/DC transformer or a battery pack powers the ISIS iQ UNO Breast Pump.

The ISIS iQ UNO Handheld Electronic Breast Pump operates in two modes. In the manual mode, the user controls the frequency and level of vacuum by varying the stroke/cycle of the control handle lever. After a comfortable manual pumping cycle has been established, the user engages the automatic pumping mode by depressing the control button on the hand controller. In the automatic mode, the pumping cycle is stored and controlled by the microprocessor. The user can switch between the automatic mode and manual mode at any time by repeatedly depressing the control button.

All milk-contacting and human tissue-contacting components are manufactured from materials that meet the appropriate regulations regarding food contact and/or biocompatibility.

{1}------------------------------------------------

8.6 Intended Use of the Device

The ISIS iQ UNO Breast Pump is intended to express and collect milk from the breasts of a lactating woman.

Summary of Technological Characteristics 8.7

The technology of the ISIS iQ UNO Breast Pump differs slightly from the Medela predicate device. The ISIS iQ UNO provides total user control of the vacuum cycle and the ability to change between manual and automatic pumping. The Medela predicate device is programmed to deliver two pre-programmed pumping cycles in an automatic mode only, and allows the user to have limited control of pump operation. The Medela predicate device is programmed to change from a high frequency pumping cycle (let-down mode) to a lower frequency pumping cycle (expression mode) two minutes after power has been applied ("Natural Expression" technology). The user can change the pump cycle only from the let-down mode to the expression mode at any time during the first two minutes of operation. The user can vary the vacuum level during either operational mode.

The technology of the ISIS iQ UNO Breast Pump differs slightly from the First Years predicate. The First Years predicate features a fully automatic auto-nursing cycle consisting of a onesecond suction and release. A dial knob on the pump handle controls the suction level.

The ISIS iQ UNO also differs slightly from the Medela predicate in the technology employed to achieve let-down and milk flow. The Medela predicate device employs "Natural Expression" technology described above to stimulate let-down and millk flow. The ISIS iQ UNO employs a patented let-down massage cushion designed to massage the area around the nipple in addition to allowing the user to control the frequency of the pumping cycle to stimulate let-down. The First Years predicate is similar to the ISIS iQ UNO regarding how let-down is stimulated. Both products employ a soft-textured breast shield.

The ISIS iQ UNO and both predicates operate with a sealed AC/DC transformer or with batteries.

The technological differences described above do not raise new concerns regarding the safety and effectiveness.

Summary of Non-clinical Performance Data 8.8

The ISIS iQ UNO meets all applicable electrical, mechanical, and environmental performance requirements given in the IEC 60601-1: 1998 + AI: 1991, A2: 1995, Corrigendum: 1994; ANSI/UL 60601:2003; EN-60601-1-1: 1990 + Al: 1993, Al 1: 1993, A12: 1993, A12: 1995, A13: 1996, Corrigendum: 1994; CAN/CSA C22.2 No. 601.1 issued: 1994/01/05 Rev: 1997/01/01; EN60601-1-2: 2001; and FCC Part 15 for a class B device. Additionally, all milk-contacting and human tissue-contacting components are manufactured from materials that meet the appropriate regulations regarding food contact and/or biocompatibility.

8.9 Summary of Clinical Performance Data

No clinical studies were performed.

8.10 Conclusion

Based upon the information presented in this dossier, it is concluded that the ISIS iQ UNO is safe and effective for the intended use, and is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized eagle in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES". The eagle is depicted with three curved lines representing its body and wings, giving it a modern and abstract appearance. The logo is black and white and appears to be a scan or photocopy.

SEP 2 2 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Avent America, Inc. % Mr. Richard Morroney Vice President Clinivation, Inc. One Speen Street, Suite 160 FRAMINGHAM MA 01701 Re: K052047

Trade/Device Name: ISIS iQ UNO Handheld Electronic Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: HGX Dated: July 29, 2005 Received: July 29, 2005

Dear Mr. Morroney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition: FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

.
.
.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and in you to organ and equivalence of your device of your device to a legally premarket nothleation: "The PDF mailing sification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrree tor your are of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 807.77). Tou may ooually outer gallerial and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

K052047

Device Name: ISIS iQ UNO Handheld Electronic Breast Pump

Indications For Use:

The ISIS iQ UNO Handheld Electronic Breast Pump is intended to express and collect
s and collection in the times areas THE TOTO KE breasts of a lactating woman.

Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Hodgson

(Division Sign-Off) () Division of Reproductive, and Radiological Dev 510(k) Number

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).