K Number
K052000

Validate with FDA (Live)

Date Cleared
2005-10-25

(92 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARCHITECT CA 19-9xR assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma on the ARCHITECT i System. The ARCHITECT CA 19-9xR assay is to be used as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods.

Patients known to be genotypically negative for the Lewis blood group antigen will be unable to produce the CA 19-9 antigen even in the presence of malignant tissue. Phenotyping for the presence of the Lewis antigen may be insufficient to detect true Lewis antigen negative individuals. Even patients who are genotypically positive for the Lewis antigen may produce varying levels of CA 19-9 based on gene dosage effect.

Device Description

The ARCHITECT CA 19-9xR assay is a two-step immunoassay for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex®.

In the first step, sample and 1116-NS-19-9 coated paramagnetic microparticles are combined. 1116-NS-19-9 reactive determinants present in the sample bind to the 1116-NS-19-9 coated microparticles. After washing, 1116-NS-19-9 acridinium-labeled conjugate is added to create a reaction mixture in the second step. Following another wash cycle, pretrigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). A direct relationship exists between the amount of 1116-NS-19-9 reactive determinants in the sample and the RLUs detected by the ARCHITECT i System optics.

AI/ML Overview

The ARCHITECT® CA 19-9™xR Assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma. It is intended to be used as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods.

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily focuses on demonstrating substantial equivalence to a predicate device and summarizing performance characteristics rather than explicit 'acceptance criteria' in the format of a clinical trial endpoint table. However, based on the reproducibility and comparison studies, we can infer performance targets and the achieved results.

Study/MetricAcceptance Criteria (Inferred)Reported Device Performance
ReproducibilityTotal precision %CV less than or equal to 10%Total precision %CV was determined to be less than or equal to 10%.
Comparison (Correlation with Predicate)High correlation coefficient (e.g., typically >0.90 for substantial equivalence in quantitative assays) and acceptable slope/intercept with the predicate device.Correlation coefficient of 0.96 with the Fujirebio Diagnostics, Inc. CA 19-9 RIA. Slope of 1.2 (99% CI: 1.08, 1.37). Y-axis intercept of -5.1 U/mL (99% CI: -7.4, -3.4).
Reference Ranges (Apparently Healthy)A high percentage of apparently healthy individuals should fall below a specified cutoff (e.g., 37 U/mL).94.4% of apparently healthy subjects (n=360) had values of 37 U/mL or less.
Association between Marker Change and Disease State Change (Concordance)Demonstrated reasonable concordance rates (Total, Positive, Negative) between CA 19-9 value changes and disease state changes.* Overall (Observation Pairs): Total Concordance (C) = 61.0%, Positive Concordance (C+) = 48.5%, Negative Concordance (C-) = 63.6% (based on 187 observation pairs).* Per-Patient Analysis: Total Concordance (C) = 68.92% (95% CI: 57.10% - 79.17%), Positive Concordance (C+) = 68.18% (95% CI: 45.13% - 86.14%), Negative Concordance (C-) = 69.23% (95% CI: 54.90% - 81.28%) (based on 74 patients).

2. Sample Size Used for the Test Set and Data Provenance:

  • Reproducibility: Not explicitly stated as a test set size, but samples were tested consisting of two panels of standards, one panel of serum with added determinants, and controls. Each sample was tested two separate times per day for 20 nonconsecutive days, using two lots of reagents, in replicates.
  • Comparison Study: 259 serum specimens. Data provenance is not specified, but typically for clinical studies supporting US regulatory submissions, samples would be sourced from a diverse US population or explicitly identified as international if applicable. The document does not state whether it was retrospective or prospective.
  • Reference Ranges (Apparently Healthy Population): 360 serum specimens from apparently healthy individuals. Data provenance not specified.
  • Patient Groups (Disease Distribution): 978 individual serum samples from various non-malignant and malignant disease groups. Data provenance not specified.
  • Pancreatic Cancer Serial Specimens (Concordance Study): 74 patients, yielding 261 evaluable observations (average 3.5 observations per patient) for the observation-pair analysis, and 74 for the per-patient analysis. Data provenance not specified. The average age was 61.8 years (range 17-85 years), 55% were men, 45% women. Staging was available for 70 of the 74 patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document describes an in vitro diagnostic (IVD) assay measuring a tumor marker, CA 19-9. The "ground truth" for such assays typically refers to the confirmed clinical diagnosis of pancreatic cancer and the progression/regression of the disease.

  • The document does not explicitly state the number or qualifications of experts used to establish the clinical ground truth (e.g., pancreatic cancer diagnosis, disease progression/no progression) for the patient samples used in the reference ranges, patient group distribution, or concordance studies.
  • However, for the "Association between Change in Marker Value and Change in Disease State" study, the ground truth for "Change in Disease State (W)" (Progression vs. No Progression) was established. This likely involved clinical assessment by medical professionals (e.g., oncologists, radiologists) based on various clinical methods like imaging, biopsy, and other clinical observations. The document itself is a 510(k) summary, which often omits granular details about clinical expert involvement if not directly related to the device's technical performance.

4. Adjudication Method for the Test Set:

  • The document does not describe an adjudication method for establishing the "Change in Disease State (W)" in the concordance study. Clinical diagnoses and disease state changes are typically determined by treating clinicians based on standard clinical practice, which may inherently involve a form of consensus or judgment, but a specific, formalized adjudication process involving multiple independent reviewers is not mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay that provides a quantitative measurement of a biomarker. MRMC studies are typically performed for imaging devices or algorithms where human readers interpret images or data, and the AI's impact on their performance is being evaluated. This document focuses on the analytical and clinical performance of the assay itself.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, this entire submission effectively describes a standalone performance evaluation of the ARCHITECT CA 19-9xR assay. As an IVD test, it generates results independently. The "performance" described (reproducibility, correlation, reference ranges, association with disease state) is the performance of the assay system itself, not in conjunction with human interpretation of its results in a structured reader study. Its intended use is "as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods," implying its results are interpreted by clinicians.

7. The Type of Ground Truth Used:

  • For the technical performance (reproducibility, comparison): The ground truth is the true concentration of CA 19-9 reactive determinants or reference method values.
  • For the clinical performance (reference ranges, patient groups, concordance study): The ground truth for disease classifications and changes in disease state would be clinical diagnosis and disease progression/regression determined by treating physicians based on a combination of clinical methods (e.g., pathology reports, imaging studies, clinical assessment, treatment response, outcomes data). The document refers to "Change in Disease State (W)," which is derived from these clinical assessments.

8. The Sample Size for the Training Set:

  • This submission describes a mature IVD product (ARCHITECT CA 19-9xR Assay) that is compared to an existing predicate device (Fujirebio Diagnostics, Inc. CA 19-9 RIA). IVD assays, especially for biomarkers, are typically developed using rigorous analytical validation rather than machine learning "training sets" in the conventional sense.
  • Therefore, the concept of a "training set" as understood in AI/ML contexts does not directly apply to this type of device and study. The assay itself is a chemical measurement system, not a learning algorithm that requires a separate training phase. The studies described are for validation and performance assessment.

9. How the Ground Truth for the Training Set Was Established:

  • As explained in point 8, there isn't a "training set" in the AI/ML sense for this device. The ground truth for developing and optimizing such an assay would involve extensive analytical validation against reference materials, known concentrations, and clinical samples with well-characterized disease states. This process is part of product development and analytical verification, not typically described as establishing ground truth for a "training set" in a 510(k) summary.

{0}------------------------------------------------

007 2 5 2005 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in This Sammary with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: __ OS2دهده (

Submitter Information

Address:Fujirebio Diagnostics, Inc.201 Great Valley ParkwayMalvern, PA 19355
Contact person:Diana L. Wolaniuk, (610) 240-3917
Summary preparation date:October 25, 2005
Name of Device
Trade/Proprietary Name:ARCHITECT® CA 19-9™XR Assay
Common/Usual Name:CA 19-9 Assay
Classification Name:System, Test, Carbohydrate Antigen (CA19-9), For MonitoringAnd Management Of Pancreatic Cancer
Predicate DeviceFujirebio Diagnostics, Inc. CA 19-9 RIA

Device Description

The ARCHITECT CA 19-9xR assay is a two-step immunoassay for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex®.

In the first step, sample and 1116-NS-19-9 coated paramagnetic microparticles are combined. 1116-NS-19-9 reactive determinants present in the sample bind to the 1116-NS-19-9 coated microparticles. After washing, 1116-NS-19-9 acridinium-labeled conjugate is added to create a reaction mixture in the second step. Following another wash cycle, pretrigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). A direct relationship exists between the amount of 1116-NS-19-9 reactive determinants in the sample and the RLUs detected by the ARCHITECT i System optics.

For additional information on system and assay technology, refer to the ARCHITECT System Operations Manual, Section 3.

{1}------------------------------------------------

Intended Use

Reagent Kit

The ARCHITECT CA 19-9xR assay is a chemiluminescent microparticle immunoassay The ARCHITECT OF 15 SXF doody to a Shorts -19-9 reactive determinants in human (CMIA) for the qualificatio dotorminative i System. The ARCHITECT CA 19-9xR assay is to be serum of plasma on the Arcorn EST pancreatic cancer patients in conjunction with other clinical methods.

Patients known to be genotypically negative for the Lewis blood group antigen will be unable Fallents known to be genotyploany nogation for seence of malignant tissue. Phenotyping for to produce the OA 10 0 anligen may be insufficient to detect true Lewis antigen negative the presence of the Eowlo antigon may of the Lewis antigen may produce varying levels of CA 19-9 based on gene dosage effect.

Calibrator Kit

The ARCHITECT CA 19-9xp Calibrators are for the calibration of the ARCHITECT i System The AROMITED OF OF TO OX Calleration of 1116-NS-19-9 reactive determinants in when used for the qualitian. Refer to the ARCHITECT CA 19-9xe reagent package insert for additional information.

Control Kit

The ARCHITECT CA 19-9xR Controls are for the estimation of test precision and the The ARCHITEOT OF 15 ox 30 ox Sentralia and the ARCHITECT i System (reagents, delection of Systemation and for the quantitative measurement of 111-NS-19-9 canbrutors, and instruments in human serum or plasma. Refer to the ARCHITECT CA 19-9xR reagent package insert for additional information.

Statement of Substantial Equivalence

The ARCHITECT CA 19-9xR assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of 1116-NS-19-9 reactive determinants in human (Onlin) for the quantitiatio assessment i System. The ARCHITECT CA 19-9xR assay is to be used as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods.

Patients known to be genotypically negative for the Lewis blood group antigen will be unable to produce the CA 19-9 antigen even in the presence of malignant tissue. Phenotyping for the presence of the Lewis antigen may be insufficient to detect true Lewis antigen negative individuals. Even patients who are genotypically positive for the Lewis antigen may produce varying levels of CA 19-9 based on gene dosage effect.

ARCHITECT CA 19-9 Assay kit is substantially equivalent to the Fujirebio Diagnostics, Inc. CA 19-9 RIA. Both of the devices are IVD products and are indicated for the quantitative determination of CA 19-9 assay values (1116-NS-19-9 reactive determinants) and used in conjunction with other clinical methods in the management of pancreatic cancer patients.

A comparison of the features of the ARCHITECT CA 19-9xR assay device and the Fujirebio Diagnostics, Inc. CA 19-9 RIA follows.

{2}------------------------------------------------

Abbott LaboratoriesFujirebio Diagnostics, Inc.
ARCHITECT CA 19-9" xR AssayCA 19-9™ RIA
(Proposed Device)(Predicate Device)K020566
Device TypeIn vitro diagnosticIn vitro diagnostic
Classification andClass II, NIGClass II, NIG
Product Code
Principle of OperationChemiluminscent MicroparticleRadioimmunoassay (RIA)
Immunoassay (CMIA)
Product UsageClinical and Hospitals laboratoriesClinical and Hospitals laboratories
Intended UseThe ARCHITECT® CA 19-9 xrThe Fujirebio Diagnostics CA 19-
assay is a chemiluminescent9 ™ RIA, an in vitro diagnostic test
microparticle immunoassayfor the quantitative measurement
(CMIA) for the quantitativeof the CA 19-9 tumor associated
determination of 1116-NS-19-9antigen, in human serum orplasma, is indicated for the serial
reactive determinants in humanmeasurement of CA 19-9 to aid in
serum or plasma on thethe management of patients
ARCHITECT i System. Thediagnosed with cancers of the
ARCHITECT CA 19-9xR assay isto be used as an aid in theexocrine pancreas. The test is
management of pancreatic canceruseful to aid in:
patients in conjunction with otherMonitoring of disease status in
clinical methods.those patients having confirmed
Patients known to be genotypicallypancreatic cancer who have levels
negative for the Lewis blood groupof serum or plasma CA 19-9 above
antigen will be unable to producethe cutoff, at the time of diagnosis.
the CA 19-9 antigen even in theCA 19-9 values must be
presence of malignant tissue.interpreted in conjunction with all
Phenotyping for the presence ofother clinical and laboratory data
the Lewis antigen may bebefore a medical decision is
insufficient to detect true Lewisdetermined.
antigen negative individuals.
Even patients who are
genotypically positive for the
Lewis antigen may produce
varving levels of CA 19-9 based
on gene dosage effect.Human Serum or Plasma (Citrate,
Type of SpecimenHuman serum or plasma (Sodium,Heparin, ACD-A or EDTA)
Lithium or EDTA)Routine Phlebotomy TechniquesRoutine Phlebotomy Techniques
Specimen Collection
MethodCapture Antibody1116-NS-19-9 mouse monoclonal1116-NS-19-9 mouse monoclonal
Conjugate Antibody1116-NS-19-9 (F(ab')2) mouse1116-NS-19-9 mouse monoclonal
monoclonal
Standards6 levels (0 - 1200 U/mL)6 levels (0 - 240 U/mL)
Controls3 levels (Low = 40 U/mL, Medium2 levels (Low = 40-50 U/mL, High
= 150 U/mL, High = 750 U/mL)= 80-90 U/mL)
Interpretation of ResultsCalibrator CurveStandard Curve

{3}------------------------------------------------

Summary of Performance characteristics

Reproducibility:

Precision was determined as described in the National Committee for Clinical Laboratory Frecision was determinde as aboshibe in amples were tested consisting of two panels of Standards (NOOLO) 1 rollood LP ) . D = panel of serum to which 1116-NS-19-9 reactive determinants were added (panel 3), and the ARCHITECT CA 19-9xR Controls, using two lots uctemninatio word datos (panel of, wo separate times per day, for 20 nonconsecutive days or reagents, in rophiedlor of work lot used a single calibration curve throughout the study. on two maturnemo. Each teagoed by calculating the standard deviation (SD) and percent coefficient of variation (%CV) values for each sample.

The total precision %CV of the ARCHITECT CA 19-9xp assay was determined to be less than or equal to 10%.

Comparison Study

A total of 259 serum specimens were tested using the ARCHITECT CA 19-9xx assay and the Fujirebio Diagnostics, Inc. CA 19-9 RIA. Passing-Bablok linear regression analysis was r affebrook on all specimens (2.0 - 9115.7 U/mL for the ARCHITECT CA 19-9xR assay and 1.17 - 10,782 U/mL for the Fujirebio Diagnostics, Inc. CA 19-9 RIA).

Passing-Bablok linear regression analysis comparing the ARCHITECT CA 19-9xR assay to the Fujirebio Diagnostics, Inc. CA 19-9 RIA yielded a correlation coefficient of 0.96, a slope of 1.2 (99% confidence interval of 1.08, 1.37), and Y-axis intercept of -5.1 U/mL (99% confidence interval of -7.4, -3.4).

Reference Ranges:

Apparently Healthy Population:

The distribution of CA 19-9xR assay values determined in 360 serum specimens from apparently healthy individuals is shown in the table below:

Distribution of ARCHITECT CA 19-9xR Values
Percent (%)
Number ofSubjects0 - 37.0U/mL37.1 - 100U/mL100.1 - 500U/mL500.1 - 1200U/mL> 1200 U/mL
ApparentlyHealthySubjects36094.4%5.6%0.0%0.0%0.0%

In this study, 94.4% of the specimens from apparently healthy subjects (n=360) had values of 37 U/mL or less.

{4}------------------------------------------------

Patient Groups:

The distribution of CA 19-9 assay values determined in 978 individual serum samples from The distinution of OA 15-5 assuy Value doceminoun in the table below:

Distribution of ARCHITECT CA 19-9XR Values
Number of Subjects0 - 37.0 U/mL37.1 - 100 U/mL100.1 - 500 U/mL500.1 - 1200 U/mL> 1200 U/mL
Nonmalignant Disease
Rectal Polyps3397.0%3.0%0.0%0.0%0.0%
Pancreatitis3100.0%0.0%0.0%0.0%0.0%
Gallbladder2195.2%0.0%0.0%0.0%4.8%
Diabetes3894.7%5.3%0.0%0.0%0.0%
Pulmonary40100.0%0.0%0.0%0.0%0.0%
Cirrhosis15392.8%4.6%0.7%0.7%1.3%
Hepatitis6892.6%7.4%0.0%0.0%0.0%
Renal3491.2%8.8%0.0%0.0%0.0%
OtherGastrointestinal5196.1%3.9%0.0%0.0%0.0%
Malignant Disease
Colorectal16981.1%7.7%5.3%1.2%4.7%
Pancreatic6643.9%6.1%12.1%10.6%27.3%
Gastric6966.7%11.6%10.1%2.9%8.7%
Hepatocellular3063.3%16.7%3.3%10.0%6.7%
Pulmonary7084.3%5.7%4.3%1.4%4.3%
Mammary10286.3%10.8%2.0%1.0%0.0%
Ovarian3187.1%6.5%3.2%3.2%0.0%

{5}------------------------------------------------

Pancreatic Cancer Serial Specimens

This analysis is based on 74 patients. There were a total of 261 evaluable observations. The average number of observations per patient is 3.5.

The average age of the subjects at time of diagnosis was 61.8 years (Exact 55% Cl: 59.5 The average age of the oubjock at the 85 years. Fifty-five percent (55% or 41/74) of the 74 years to o4.1 years) with a range of 17 try-five percent (45% or 33/74) were women. Staging patients were from the the romaining the 74 patients. The majority of the patients were stage was available from the churclor 10 of the 4.1% and 6.8% were stage 1 and 11 respectively.

Association between Change in Marker Value and Change in Disease State

A 2x2 table was constructed to show the association between a positive change in a patient's A 2x2 table was obnothed to the disease from one observation to the next. A positive OA 19-9 value and progression of the value that is at least 2.5 times greater than the total %CV of the test. For the test assay this value is 14.0%. The following Table (entitled "Distribution of W by V") presents the results for the 187 observation pairs in this study.

Three estimates of Concordance are given for the following Table.

Total Concordance:$C = (16+98) / 187 = 114/187 = 61.0%$
Positive Concordance:$C_{+} = 16/33 = 48.5%$
Negative Concordance:$C_{-} = 98/154 = 63.6%$
Distribution of W by V
Change in CA 19-9(V)Change in Disease State (W)
ProgressionNo ProgressionTotal
≥ 14.0%165672
< 14.0%1798115
Total33154187

Distribution of W by V

Per Patient Analysis

r of attent thatysie
The table below (entitled " Per Patient Distribution) demonstrates this distribution for the 74 patients in this study.

Per-Patient Distribution
Change in Disease State
Change in CA 19-9ProgressionNo ProgressionTotal
≥ 14.0%151631
< 14.0%73643
Total225274

Per-Patient Distribution

{6}------------------------------------------------

Estimates of per-patient concordances can be obtained. The for these Estimates or per-patient concordances ban bo otableding table (entilled' estimates can be delemined using the binomia distribution of the sonfidence Estimate of Fel-Fatlent Fostine, Negative and Fotal Centre vals about each estimate.

Estimates of Per-Patient Positive, Negative and Total Concordance with 95% Confidence Intervals

StatisticEstimateLower BoundUpper Bound
C68.92%57.10%79.17%
C+68.18%45.13%86.14%
C69.23%54.90%81.28%

{7}------------------------------------------------

Public Health Service

Image /page/7/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services (HHS). The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three strands intertwined, representing the three branches of government. The text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular pattern around the caduceus.

Fujirebio Diagnostics, Inc. c/o Ms. Diana L. Wolaniuk Clinical and Regulatroy Affairs Specialist 201 Great Valley Pkwy, Malvern. PA 19355-1307

OCT 25 2005

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Re: K052000

Trade/Device Name: ARCHITECT® CA 19-97Mxx Assay, ARCHITECT® CA 19-9TMxR Calibrator Kit and ARCHITECT® CA 19-9™xR Control Kit Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-associated Antigen Immunological Test System Regulatory Class: Class II Product Code: NIG, JIT, JJX Dated: July 21, 2005 Reccived: July 27, 2005

Dear Ms. Wolaniuk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce as onated as 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic nate (Act) that do not require approval of a premarket approval application (PMA). You may, the (free) market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be frag 10 addy of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{8}------------------------------------------------

Page 2 -

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

lobatz Beckerf

Robert L. Becker, Jr., MD, P2 Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K052000

Device Name: ARCHITECT® CA 19-9 1 xR

Indications For Use:

ARCHITECT CA 19-9xR Reagent Kit

ARCHITECT CA 19-9xe Reagent All
The ARCHITECT CA 19-9xe assay is a chemiuminescent micropaticle immunoassay (CMA)
The ARCHITECT CA 19-9xe assay is a cheminents in human serum The ARCHITECT CA 19-9x assy is a Cremination of 116-NS-19 reactive defeninents in human serum of for the quantitative determination of TTT-HRCHITECT CA 19-5x, assay is to be used as an
plasma on the ARCHITECT i System. The ARCHITECT CA 19-5x, assay is to be used as an plasma on the ARCHITECT 7 System. The Arconnellon with other clinical methods.
aid in the management of pancreatic cancer patients in councenting will be unable in aid in the management of pancreatic caliter patients in vol Janothern will be unable to
Patients known to be genotypically negative for the Lewis blood group and the Patients known to be genotypically negative of malional tissue. Phenotyping for the produce the CA 19-9 antigen may be insufficient to detect true lewis antigen negative presence of the Lewis antigen Thay be Insultient to dotson and antigen may produce
individuals. Even patients who are genolypically positive for the Lewis antigen may produc individuals: " Even patiento was on gene dosage effect.

ARCHITECT CA 19-9xR Calibrator Kit

ARCHITECT CA 19-9xe Calibrators are the calibration of the ARCHITECT i System
The ARCHITECT CA 19-9x2 Calibrators are for the calibration of 1146 NS 10 Broadiye deferminants when used for the quantitative determination of 1116-NS-199 readive deferminants in human when used for the quantitative detemmation of 1710-No 16-9 readers of the many of the many of information.

ARCHITECT CA 19-9xR Control Kit

ARCHITECT CA 19-9xe Control All
The ARCHITECT CA 19-9xe Control Street on the estimation of the detection and the detection The ARCHITECT CA 19-9x4 Controls are for the estimation of ortent (reagents, calibrators, and of systematic analytical devations of the Artical version of 1116-NS-19-9 reactive instrument), when used for the quantitative measurement of the ARCHITECT CA 19-9xR reagent package insert for additional information.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Maria Chan

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1

510(k) K0152000

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.