K Number
K051958

Validate with FDA (Live)

Date Cleared
2006-07-21

(367 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MP RapidOXY Test Strip is an immunochromatographic one-step in-vitro test designed for qualitative determination of oxycodone in human urine specimens above a cut-off level of 100 ng/ml. The MP RapidOXY Test Strip may be used in a point-of-care (POC) setting and will provide preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) has been established as the preferred confirmatory method by the Substance Abuse Mental Health Services Administration (SAMHSA). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

The MP RapidBUP Test Strip is an immunochromatography based one step in vitro test. It is designed for qualitative determination of the major metabolite of buprenorphine, buprenorphine-3-beta-d-glucuronide, in human urine specimens at cut-off level of 10 ng/ml. The MP RapidBUP Test Strip may be used in a point-of-care (POC) setting and will provide preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) has been established as the preferred confirmatory method by the Substance Abuse Mental Health Services Administration (SAMHSA). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

Device Description

The MP RapidOXY Test Strip and the MP RapidBUP Test Strip are based on the principle of specific immunochemical reaction between antibody and antigen to analyze particular compounds in human urine specimens. The assays rely on the competition for binding antibody between drug conjugate and free drug which may be present in urine. When drug is present in the urine specimen, it competes with drug conjugate for the limited amount of antibodydve conjugate. When the amount of drug is equal or more than the cut-off (i.e., 100 ng/ml oxycodone or 10 ng/ml buprenorphine-3-beta-d-glucuronide, or B-3-beta-d-G), it will prevent the binding of drug conjugate to the antibody. Therefore, a positive urine specimen will not show a colored band on the test line zone, indicating a positive result, while the presence of a colored band indicates a negative result. A control line is present in the test window to work as procedural control in both assays. This colored band should always appear on the control line zone if the test device is stored in good condition and the test is performed appropriately.

AI/ML Overview

The provided text describes two drug-of-abuse test kits: the MP RapidOXY Test Strip (for oxycodone) and the MP RapidBUP Test Strip (for buprenorphine). However, the document does not contain explicit "acceptance criteria" in the typical sense of performance metrics (e.g., sensitivity, specificity thresholds) that a study aims to meet. Instead, the document outlines the "intended use" and "cut-off levels" which serve as the functional definitions for the device's performance. The "study" referenced within the document is implicitly the submission of "Performance Data and the Accuracy Data" to the FDA, but the details of these studies (e.g., specific methodologies, results, sample sizes, ground truth establishment) are not provided in the given text.

Therefore, I will populate the table and answer the questions based on the information available in the provided text, and explicitly state when information is not present.

Acceptance Criteria and Device Performance

Acceptance Criteria (Implicit from Intended Use)Reported Device Performance (as per Intended Use)Study Proving Performance (Implicit in Submission)
MP RapidOXY Test Strip
Qualitative determination of oxycodone in human urineQualitative determination of oxycodone in human urinePerformance and Accuracy Data (details not provided)
Above a cut-off level of 100 ng/mlDesigned for detection above 100 ng/mlPerformance and Accuracy Data (details not provided)
For point-of-care (POC) useIntended for point-of-care (POC) usePerformance and Accuracy Data (details not provided)
Provide preliminary analytical test resultProvides preliminary analytical test resultPerformance and Accuracy Data (details not provided)
Follow-up with GC/MS for confirmationRecommended follow-up with GC/MS for confirmatory resultPerformance and Accuracy Data (details not provided)
MP RapidBUP Test Strip
Qualitative determination of buprenorphine's metabolite (buprenorphine-3-β-d-glucuronide) in human urineQualitative determination of buprenorphine's metabolite (buprenorphine-3-β-d-glucuronide) in human urinePerformance and Accuracy Data (details not provided)
Above a cut-off level of 10 ng/mlDesigned for detection above 10 ng/mlPerformance and Accuracy Data (details not provided)
For point-of-care (POC) useIntended for point-of-care (POC) usePerformance and Accuracy Data (details not provided)
Provide preliminary analytical test resultProvides preliminary analytical test resultPerformance and Accuracy Data (details not provided)
Follow-up with GC/MS for confirmationRecommended follow-up with GC/MS for confirmatory resultPerformance and Accuracy Data (details not provided)

Detailed Study Information (Based on Provided Text):

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in the provided text. The document mentions "Performance Data and the Accuracy Data" but does not detail the sample sizes for these studies.
    • Data Provenance: Not specified in the provided text.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not specified in the provided text.
    • Qualifications of Experts: Not specified in the provided text.
    • The document states that Gas Chromatography/Mass Spectrometry (GC/MS) "has been established as the preferred confirmatory method by the Substance Abuse Mental Health Services Administration (SAMHSA)" and is recommended for follow-up. This implies GC/MS results would likely be the basis for ground truth, which is an analytical method rather than expert consensus.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not specified in the provided text.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not specified. This is a point-of-care immunoassay "Test Strip" for qualitative determination, not a system involving "human readers" or "AI assistance" in the typical sense of diagnostic imaging or complex analytical interpretation. The device provides a visual, qualitative result.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, implicitly. The devices (test strips) are designed to provide a "preliminary analytical test result" qualitatively and visually. This implies a standalone performance without human interpretation or adjustment being part of the primary result generation, although a human reads the visual result. The device itself is an "algorithm" in the sense of a chemical reaction with a defined output.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The document explicitly states that Gas Chromatography/Mass Spectrometry (GC/MS) is the "preferred confirmatory method." This indicates that analytical confirmation (GC/MS) would be the ground truth for the studies supporting the device's accuracy.
  7. The sample size for the training set

    • Not specified in the provided text. As these are immunoassays (chemical tests), the concept of a "training set" in the context of machine learning or AI is not directly applicable in the same way. Performance data would typically involve testing against known positive and negative samples, and samples spiked with controlled concentrations of the target analyte.
  8. How the ground truth for the training set was established

    • Not specified in the provided text. Assuming "training set" refers to samples used to optimize the test's performance characteristics (e.g., cut-off, cross-reactivity), the ground truth for such samples would likely also be established through analytical methods like GC/MS or by preparing samples with known concentrations of the drug/metabolite.

{0}------------------------------------------------

PART 1 – Administrative Information

JUL 2 1 2006

510(k) SUMMARY II.

Product Information A.

Rapid Diagnostics, a division of MP Biomedicals, LLC, has developed two additional drug-of-abuse tests kits for the detection of oxycodone and buprenorphine in human urine samples. These two test strips complement the current panel for DOA analysis, the MICROMEDIC® Drugs of Abuse Panel Test (K033566). The MICROMEDIC® Drugs of Abuse Panel Test is an immunochromatographic one-step in vitro diagnostic test designed for the qualitative determination of up to nine (9) drug substances in human urine specimens to include amphetamines, barbiturates, benzodiazepine, cocaine, methadone, methamphetamine, opiates, phencyclidine (PCP) and cannabinoid (THC).

As will be demonstrated in the following three sections (Parts I, II, and III), both the MP RapidOXY Test Strip and the MP RapidBUP Test Strip are similar in format and intended use as the parent MICROMEDIC® DOA Panel Test. There are to be added to the existing test panel once cleared.

Both the MP RapidOXY Test Strip and the MP RapidBUP Test Strip are based on the principle of specific immunochemical reaction between antibodies and antigens to analyze particular compounds in human urine specimen. The tests rely on the competition for antibody binding between drug conjugate and free drug which may be present in the urine specimen being tested.

{1}------------------------------------------------

PART 1 - Administrative Information

II. 510(k) SUMMARY

B. Intended Use:

Oxycodone 1.

The MP RapidOXY Test Strip is an immunochromatography based one step in vitro test. It is designed for qualitative determination of oxycodone in human urine specimens above a cut-off level of 100 ng/ml. Each oxycodone test kit is used to obtain a visual, qualitative result and is submitted for point-of-care (POC) use. It is not intended for over the counter sale to lay persons.

The assay will provide a preliminary analytical test result, with recommended follow-up using gas chromatography/mass spectrometry (GC/MS) for a confirmatory result.

2. Buprenorphine

The MP RapidBUP Test Strip is also an immunochromatic, one-step in vitro test. It is designed for qualitative determination of buprenorphine's metabolite, buprenorphine-3-beta-d-qlucuronide, in human urine specimens above a cut-off level of 10 nq/ml.

As with the oxycodone test kit, the buprenorphine kit is used to obtain a visual, qualitative result and is submitted for point-of-care (POC) use. It is not intended for over the counter sale to lay persons. In addition, the assay will provide only a preliminary analytical test result, with recommended follow-up using GC/MS.

{2}------------------------------------------------

PART 1 - Administrative Information

II. 510(k) SUMMARY

  • Name of Manufacturer: Rapid Diagnostics, Inc. 1429 Rollins Road, C. Burlingame, California, USA, a division of MP Biomedicals, LLC, hereafter RAPID DIAGNOSTICS.

D. Principle

The MP RapidOXY Test Strip and the MP RapidBUP Test Strip are based on the principle of specific immunochemical reaction between antibody and antigen to analyze particular compounds in human urine specimens. The assays rely on the competition for binding antibody between drug conjugate and free drug which may be present in urine. When drug is present in the urine specimen, it competes with drug conjugate for the limited amount of antibodydve conjugate.

When the amount of drug is equal or more than the cut-off (i.e., 100 ng/ml oxycodone or 10 ng/ml buprenorphine-3-β-d-glucuronide, or B-3-β-d-G), it will prevent the binding of drug conjugate to the antibody. Therefore, a positive urine specimen will not show a colored band on the test line zone, indicating a positive result, while the presence of a colored band indicates a negative result.

A control line is present in the test window to work as procedural control in both assays. This colored band should always appear on the control line zone if the test device is stored in good condition and the test is performed appropriately.

{3}------------------------------------------------

PART 1 -- Administrative Information

II. 510(k) SUMMARY

E. Submission Summary

In this Premarket Notification [510(k)], RAPID DIAGNOSTICS, Inc. will present information for clearance of the following two devices, the MICROMEDIC® MP RapidOXY Test Strip and the RapidBUP Test Strip, for use in a POC setting. The submission will consist of three parts, Part I, II and III, to include Administrative Information, the Performance Data and the Accuracy Data, respectively.

The submission consists of the required administrative and technical information, along with a detailed presentation of the clinical and laboratory data used to substantiate the POC claim.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized symbol in the center. The text around the perimeter is small and difficult to read, but it appears to be the name of an organization or agency. The symbol in the center consists of three curved lines that resemble a stylized human figure or a bird in flight. The logo is black and white and has a simple, clean design.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MP Biomedicals, LLC. c/o Ms. Robin J. Hellen Hellen Professional Services 9418 Lasaine Aveune Northridge, CA 91325

JUL 2 1 2006

Re: K051958 Trade/Device Name: RapidBUP Test Strip RapidOXY Test Strip Regulation Number: 21 CFR§ 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG Dated: June 15, 2006 Received: June 26, 2006

Dear Ms. Hellen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that I'DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Garcia

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

and contract and consisted in the plant of the property that a first and the

STATEMENT FOR INDICATIONS FOR USE (BUPRENORPHINE)

510(k) Number (if known):K051958
Device Name:RapidBUP Test Strip

Indications for Use:

The MP RapidBUP Test Strip is an immunochromatography based one step in vitro test. It
is designed for qualitative determination of the major metabolite of buprenorphine,
buprenorphine-3-β-d-glucuronide, in human urine specimens at cut-off level of 10 ng/ml.

The MP RapidBUP Test Strip may be used in a point-of-care (POC) setting and will provide
preliminary analytical test result. A more specific alternative chemical method must be
used in order to obtain a confirmed analytical result. Gas chromatography/mass
spectrometry (GC/MS) has been established as the preferred confirmatory method by the
Substance Abuse Mental Health Services Administration (SAMHSA). Clinical onsideration
and professional judgment should be applied to any drug of abuse test result, particularly
when preliminary positive results are indicated.

Concurrence of the CDRH, Office of In-Vitro Diagnostic Device (OIVD)

Prescription Use:XOROver the Counter Use:
-------------------------------------------------

Division Sign-Off

Diagnostic Safety

K051958

{7}------------------------------------------------

STATEMENT FOR INDICATIONS FOR USE (OXYCODONE)

510(k) Number (if known):K051958
Device Name:RapidOXY Test Strip

Indications for Use:

The MP RapidOXY Test Strip is an immunochromatographic one-step in-vitro test designed for qualitative determination of oxycodone in human urine specimens above a cut-off level of 100 ng/ml.

The MP RapidOXY Test Strip may be used in a point-of-care (POC) setting and will provide preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) has been established as the preferred confirmatory method by the Substance Abuse Mental Health Services Administration (SAMHSA). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

Concurrence of the CDRH, Office of In-Vitro Device (OIVD)

Prescription Use: XOROver the Counter Use: ______
-------------------------------------------------------

Division Sign-Off

Office of In Vitro Diagnostic
Device Evaluation and Safety

K051958

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).