K Number
K051595

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-07-26

(405 days)

Product Code
Regulation Number
880.6860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TSO3 Model 125L is an Ozone Sterilizer intended for use in the sterilization processing of reusable medical devices in health care facilities. The TSO3 Ozone Sterilizer, Movol 1252. is designed for sterilization of both metal and non-metal medical devices at low temperatires. The sterilization cycle operates at very low pressure and low temperature, consequently it is suitable for processing medical devices sensitive to heat and moisture.

The TSO3 Ozone Sterilizer, Model 125L, is designed to sterilize instruments and devices with diffusion-restricted spaces, such as the hinged portion of forceps and sciators.

Device Description

TSO3 Ozone Sterilizer, model 125L is intended to sterilize medical devices that has been previously cleaned.

The sterilization chamber has a capacity of 125 liters (4.3 cu. ft.).

It requires USP grade oxygen, water and electricity. Model 125L could be installed as a free standing unit or recessed behind the wall. No exhaust gas ventilation duct is required as long as the room is adequately ventilated. By-products are oxygen and low humidity water vapor.

Model 125L is equipped with a unique factory-programmed control system.

Non-woven wrapping material or pouches and anodized aluminum containers are used as packaging for medical devices to be sterilized.

OZO-TEST® self-contained Biological Indicators (B. stearothermophilus) are recommended for use in evaluating cycle performance.

TSO3 Chemical Indicators are available for this process.

No aeration is needed following the sterilization cycle. The sterilized items are cool to the touch and can be removed immediately and used, or stored for future use.

AI/ML Overview

Here's an analysis of the provided text regarding the TSO3 Ozone Sterilizer, Model 125L, focusing on acceptance criteria and the supporting study:

Device Performance and Acceptance Criteria

The TSO3 Ozone Sterilizer, Model 125L, was tested for its ability to sterilize medical devices with extended lumens. The acceptance criterion for sterilization is the achievement of a Sterility Assurance Level (SAL) of 10^-6, which means the probability of a single viable microorganism remaining on an item after sterilization is 1 in a million.

The study demonstrated the device's ability to sterilize a range of stainless steel lumens.

Table of Acceptance Criteria and Reported Device Performance

Internal Diameter (mm)Internal Diameter (French) (Approximate Correspondence)Length (mm)Acceptance Criterion (SAL)Device Performance (Achieved SAL)
0.92.748510^-6Achieved 10^-6 SAL (implied by "successfully sterilize")
1350010^-6Achieved 10^-6 SAL (implied by "successfully sterilize")
2657510^-6Achieved 10^-6 SAL (implied by "successfully sterilize")
3965010^-6Achieved 10^-6 SAL (implied by "successfully sterilize")
41270010^-6Achieved 10^-6 SAL (implied by "successfully sterilize")

Note: The document states, "Testing on lumens were conducted employing half cycle to demonstrate achievement of a sterility assurance level (SAL) of 10^-6." This "half cycle" method is a common and stringent approach to demonstrate an SAL, as it proves that even under reduced sterilization conditions (half of a full cycle), the target SAL can still be met, implying that a full cycle would provide an even greater margin of safety.


Study Details

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the numerical sample size (number of lumens) used for the test set. It lists five distinct lumen configurations (combinations of internal diameter and length) that were tested. For each configuration, multiple tests would have been performed using biological indicators to establish the SAL.
  • Data Provenance: The study was conducted by TSO3 Inc. in Quebec, Canada. The data is likely prospective as it describes specific validation testing performed to support the extended lumen claims.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not mention the use of human experts or an expert panel to establish ground truth for the test set in the context of sterilization efficacy. Sterilization efficacy studies typically rely on quantitative laboratory methods (e.g., biological indicators) rather than subjective expert assessment.

4. Adjudication method for the test set:

  • Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for subjective diagnostic assessments where discrepancies among human readers need resolution. For sterilization efficacy, the "ground truth" is determined by the inactivation of biological indicators, which is an objective measure.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a sterilizer, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study comparing human reader performance with or without AI assistance is irrelevant to this device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • The device itself is a "standalone" sterilizer. Its performance is evaluated intrinsically through its ability to inactivate microorganisms, not as an algorithm assisting a human. The "effectiveness" section explicitly describes validation testing of the process (sterilizer) using an "overkill" approach and biological indicators, which is a standalone assessment of the device's function.

7. The type of ground truth used:

  • The ground truth for sterilization efficacy was established using biological indicators (B. stearothermophilus). These indicators contain a known population of highly resistant bacterial spores. The complete inactivation of these spores after sterilization, especially in a "half cycle" challenge, directly demonstrates the achievement of the required Sterility Assurance Level (SAL). This is an objective, laboratory-based method.

8. The sample size for the training set:

  • Not applicable. This device is a sterilizer, not an AI model that requires a training set. The "validation testing" described refers to the testing of the physical device's performance, not the training of an algorithm.

9. How the ground truth for the training set was established:

  • Not applicable, as there is no training set for an AI/algorithm in this context.

{0}------------------------------------------------

K051595

Image /page/0/Picture/1 description: The image shows a chemical structure diagram. The diagram includes the letters 'TS' on the left side. To the right of the letters, there is a chemical symbol with a circle and lines, indicating a chemical bond or group. The overall image appears to represent a chemical compound or molecule, possibly an intermediate or reactant in a chemical reaction.

EXTENDED LUMEN CLAIMS FO TSO3 OZONE STERILIZER, MODEL 125L

510(k) Summary

JUL 26 2006

Applicant's Name and Address

TSO3 Inc. 2505, Dalton Avenue Ste-Foy, Quebec, Canada G1P 3S5

Contact Person, Telephone, FAX

Marc Chaunet, Quality Assurance and Regulatory Affairs Tel : (418) 651-0003 FAX : (418) 653-5726 E-mail : mchaunet@tso3.com

U.S. Agent

Charles O. Hancock Inc. Contact person: Charles O. Hancock, RAC Tel : (585) 223-1850 FAX : (585) 223-6855 E-mail : chancock@frontiernet.net

Submission Date June 14, 2005

Trade Name

TSO3 Ozone Sterilizer, model 125L

Common Name

TSO3 125L Ozone Sterilizer

Classification Name

Sterilizer, Chemical Class II (as per 21CFR, part 880.6860 equivalent device)

Legally Marketed Equivalent Device Name(s) TSO3 Ozone Sterilizer, model 125L (K020875)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a chemical structure diagram. The diagram contains the letters 'TS' on the left side. To the right of the letters, there is a chemical structure with a central 'S' atom bonded to three 'O' atoms. The structure appears to represent a tosyl group, a common moiety in organic chemistry.

EXTENDED LUMEN CLAIMS FOR TSO3 OZONE STERILIZER, MODEL 125L

Description of Device

TSO3 Ozone Sterilizer, model 125L is intended to sterilize medical devices that has been previously cleaned.

The sterilization chamber has a capacity of 125 liters (4.3 cu. ft.).

It requires USP grade oxygen, water and electricity. Model 125L could be installed as a free standing unit or recessed behind the wall. No exhaust gas ventilation duct is required as long as the room is adequately ventilated. By-products are oxygen and low humidity water vapor.

Model 125L is equipped with a unique factory-programmed control system.

Non-woven wrapping material or pouches and anodized aluminum containers are used as packaging for medical devices to be sterilized.

OZO-TEST® self-contained Biological Indicators (B. stearothermophilus) are recommended for use in evaluating cycle performance.

TSO3 Chemical Indicators are available for this process.

No aeration is needed following the sterilization cycle. The sterilized items are cool to the touch and can be removed immediately and used, or stored for future use.

Effectiveness

Model 125L validation testing for extended lumen claims was performed using the « overkill » approach to demonstrate the effectiveness of the process.

Safety

Safety of the TSO3 Ozone Sterilizer, model 125L was demonstrated into the original 510(k) submission (K020875). The content of the actual submittal does not compromise the safety of the device.

{2}------------------------------------------------

TSO3 Ozone Sterilizer, Model 1251, bas the ability to sterilize successfully medical devices having a single stainless steel humen which falls into the following length and internal diameters *:

Internal diameter (mm)Internal diameter (French)(Approximative correspondence)Length (mm)
0.92.7485
13500
26575
39650
412700

*Note: Testing on lumens were conducted employing half cycle to demonstrate achievement of a sterility assurance level (SAL) of 10°.

The types of packaging compatible with the TSO; Model 1231 Ozone Sterilizer are the TSO3 sterifization pouch and rigid anodized aluminum containers using disposable cellulose filiter paper.

Page 2 of 2

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged around the circle's perimeter.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 6 2006

TSO3, Incorporated C/O Mr. Charles O. Hancock Charles O. Hancock Associates, Incorporated 33 Black Watch Trail Fairport, New York 14450

Re: K051595

Trade/Device Name: TSO3 Ozone Sterilizer, Model 125L Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: II Product Code: FLF Dated: July 10, 2006 Received: July 10, 2006

Dear Mr. Hancock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Hancock

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K051595

Device Name: TSO3 Ozone Sterilizer, Model 125L

Indications For Use:

The TSO3 Model 125L is an Ozone Sterilizer intended for use in the sterilization processing of reusable medical devices in health care facilities. The TSO3 Ozone Sterilizer, Movol 1252. is designed for sterilization of both metal and non-metal medical devices at low temperatires. The sterilization cycle operates at very low pressure and low temperature, consequently it is suitable for processing medical devices sensitive to heat and moisture.

The TSO3 Ozone Sterilizer, Model 125L, is designed to sterilize instruments and devices with diffusion-restricted spaces, such as the hinged portion of forceps and sciators.

Prescription Use (Part 21 CFR 801 Subpan D)

AND/OR

X Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED !

Concurrence of CDRH, Office of Device Evaluation (ODE)

Stil. Mighy, R
...on Sign-Off)
an of Anesthesiology, General Hospital,
Control, Dental Devices

nn of Anesthestology
Control. Dental Devices

Page 1 of 2

ANNEX A19 - 0001

§ 880.6860 Ethylene oxide gas sterilizer.

(a)
Identification. An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.(b)
Classification. Class II (performance standards).