K Number
K051235

Validate with FDA (Live)

Device Name
ON TRAK TESTCARD
Manufacturer
Date Cleared
2005-06-08

(26 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OnTrak TesTcard menus are various combinations of the different drugs hat are listed below. The TesTcard products are in vitro diagnostic tests for the qualitative detection of drug or drug metabolite in urine. The TesTcard devices simultaneously test for the presence of multiple drugs or drug metabolites. The TesTcard profile (cutoff) consists of the following: Amphetamines (d,l-amphetamine): 1000 ng/mL, Barbiturates (secobarbital): 200 ng/mL, Benzodiazepines (oxazepam): 100 ng/mL, Cocaine metabolite (benzoylecgonine): 300 ng/mL, Methamphetamine (d-methamphetamine): 500 ng/mL, Morphine (morphine): 300 ng/mL, PCP (phencyclidine): 25 ng/mL, Tricyclic Antidepressants (TCA)(Imipramine): 1000 ng/mL, THC (11-nor-delta9-THC-9-carboxylic acid): 50 ng/mL. AND/OR TesTcard products provide only preliminary analytical test result. A more specific alternate test method must be used in order to obtain a confirmed analytical result. Prescription Use V (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

Device Description

The OnTrak TesTcard assays contained in this submission are in vitro diagnostic tests intended for the qualitative detection of drug or drug metabolite in urine. The TesTcard devices simultaneously test for the presence of multiple drugs or drug metabolites. The TesTcard profile (cutoff) consists of amphetamines (d,l-amphetamine 1000 ng/mL), barbiturates (secobarbital 200 ng/mL), benzodiazepines (oxazepam 100 (benzoylecgonine cocaine metabolite ng/mL). methamphetamine (d-methamphetamine 500 ng/mL), morphine (morphine 300 ng/mL), PCP (phencyclidine 25 ng/mL), tricyclic antidepressants (TCA) (imipramine 1000 ng/mL) and THC (11-nor-A-THC-9-carboxylic acid 50 ng/mL). The TesTcard assays are based on the principle of microparticle capture inhibition. The test relies on the competition between drug, which may be present in the urine being tested, and drug conjugate immobilized on a membrane in the test chamber. When the TesTcard contacts the urine sample, the sample is absorbed into the TesTcard sample pad. The absorbed sample travels through the reagent strips contained in the device by capillary action. In the reagent strip, the sample rehydrates and mobilizes antibody-coated blue microparticles. The microparticle-urine suspension continues to migrate through the reagent strip and comes in contact with the immobilized drug conjugates. In the absence of drugs in the urine, the antibody coated microparticles bind to the drug conjugates and blue bands are formed in the result areas. When drugs are present in the specimen, they bind to the respective antibody-coated microparticles. If sufficient drug is present, the microparticles are inhibited from binding the appropriate drug conjugate and no blue band is formed in the result area below the drug name. A positive specimen causes the membrane to remain white. An additional antibody/antigen reaction occurs at the "VALID" area. The "TEST VALID" blue band forms when antibodies imbedded in the reagent membrane bind to the antigen on the blue microparticles. The presence of the "TEST VALID" band indicates that the test has completed, the reagents in the "TEST VALID" area are valid, and the results are ready to interpret.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the OnTrak TesTcard, a device for qualitative detection of drugs or drug metabolites in urine. However, it does not contain any information about a study proving the device meets specific acceptance criteria.

The document outlines the device's intended use, technological characteristics, and claims substantial equivalence to a predicate device, but it lacks detailed performance data, sample sizes for testing, ground truth establishment, or expert involvement.

Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance or answer most of the other questions regarding a study, as this information is not present in the provided text.

Based on the information provided, here's what can be extracted:

1. A table of acceptance criteria and the reported device performance

  • Not provided in the document. The document lists the cutoff concentrations for various drugs, which are essential for defining "positive" or "negative" results, but it does not present acceptance criteria (e.g., sensitivity, specificity thresholds) or actual performance data against these criteria.
Acceptance CriteriaReported Device Performance
Information not provided in the documentInformation not provided in the document

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not provided in the document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not provided in the document.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not provided in the document.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable/Not provided. This device is an in-vitro diagnostic test, not an AI-assisted diagnostic tool that would involve human "readers" in the context of radiology or similar fields.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable/Not provided. This is an IVD device, not an algorithm, so "standalone" performance in the AI sense is not relevant. The device operates as a self-contained test.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not explicitly stated within the provided text. For qualitative drug detection tests like this, the ground truth for a study would typically be established by a reference method (e.g., Gas Chromatography/Mass Spectrometry - GC/MS) performed on the urine samples. The document only mentions that "A more specific alternate test method must be used in order to obtain a confirmed analytical result" which implies that such reference methods are for confirmation but doesn't detail their use in validation.

8. The sample size for the training set

  • Not provided in the document. (This type of device does not typically have a "training set" in the machine learning sense.)

9. How the ground truth for the training set was established

  • Not provided in the document. (See above, not applicable in the ML sense.)

{0}------------------------------------------------

JUN 8 - 2005

510(k) Summary

K051235
page 1 of 2

According to the requirements of 21 CFR 807.92, the following Introduction According to the requirements detail to understand the basis for a determination of substantial equivalence.

  1. Submitter name, address, contact

Varian, Inc. 25200 Commercentre Drive Lake Forest, CA 92630 (949) 770-9381

Contact: Lorna Gamboa

Date Prepared: May 11, 2005

Proprietary Name: OnTrak TesTcard 2) Device Name

Panel: Toxicology

Panel: Toxicology
Classification Name:ProductCodeRegulationNumber
Enzyme Immunoassay, AmphetamineDKZ862.3100
Enzyme Immunoassay, BarbiturateDIS862.3150
Enzyme Immunoassay, BenzodiazepinesJXM862.3170
Enzyme Immunoassay, Cocaine and CocaineMetaboliteDIO862.3250
Fluorometry, MorphineDJJ862.3640
Enzyme Immunoassay, PhencyclidineLCMUnclassified
Enzyme Immunoassay, CannabinoidsLDJ862.3870
Gas Chromatography, MethamphetamineLAF862.3610
Radioimmunoassay, Tricyclic AntidepressantDrugsLFG862.3910
3) PredicateDeviceWe claim substantial equivalence to this currently marketed device:On Trak TesTcard 9, K012396, 11/05/2001On Trak TesTcard 9, K050321, 04/18/2005
4) DeviceDescriptionThe OnTrak TesTcard assays contained in this submission are in vitrodiagnostic tests intended for the qualitative detection of drug or drugmetabolite in urine. The TesTcard devices simultaneously test for thepresence of multiple drugs or drug metabolites. The TesTcard profile(cutoff) consists of amphetamines (d,l-amphetamine 1000 ng/mL),barbiturates (secobarbital 200 ng/mL), benzodiazepines (oxazepam 100

Image /page/0/Picture/11 description: The image contains the word "CONFIDENTIAL" in a large, sans-serif font. The letters are outlined, giving them a hollow appearance. The word is horizontally oriented and fills most of the frame.

{1}------------------------------------------------

K051235
page 2 of 2

(benzoylecgonine cocaine metabolite ng/mL). methamphetamine (d-methamphetamine 500 ng/mL), morphine (morphine 300 ng/mL), PCP (phencyclidine 25 ng/mL), tricyclic antidepressants (TCA) (imipramine 1000 ng/mL) and THC (11-nor-A-THC-9-carboxylic acid 50 ng/mL).

The TesTcard assays are based on the principle of microparticle capture inhibition. The test relies on the competition between drug, which may be present in the urine being tested, and drug conjugate immobilized on a membrane in the test chamber.

When the TesTcard contacts the urine sample, the sample is absorbed into the TesTcard sample pad. The absorbed sample travels through the reagent strips contained in the device by capillary action. In the reagent strip, the sample rehydrates and mobilizes antibody-coated blue microparticles. The microparticle-urine suspension continues to migrate through the reagent strip and comes in contact with the immobilized drug conjugates. In the absence of drugs in the urine, the antibody coated microparticles bind to the drug conjugates and blue bands are formed in the result areas.

When drugs are present in the specimen, they bind to the respective antibody-coated microparticles. If sufficient drug is present, the microparticles are inhibited from binding the appropriate drug conjugate and no blue band is formed in the result area below the drug name. A positive specimen causes the membrane to remain white.

An additional antibody/antigen reaction occurs at the "VALID" area. The "TEST VALID" blue band forms when antibodies imbedded in the reagent membrane bind to the antigen on the blue microparticles. The presence of the "TEST VALID" band indicates that the test has completed, the reagents in the "TEST VALID" area are valid, and the results are ready to interpret.

All drug test strips contained in the TesTcard products have been 5) Technological previously reviewed by FDA under the 510(k) numbers indicated in Characteristics Section 3 of this summary.

Like the predicate device, TesTcard devices utilize microparticle capture inhibition.

6) Substantial Equivalence

The TesTcard devices have the same intended use and incorporate the same fundamental scientific technology as the predicate device.

TestcardPredicate
Intended UseQualitative detection of drugs in urineSame
Scientific TechnologyMicroparticle capture inhibitionSame
Sample MatrixUrineSame

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, composed of three curved lines.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 8 - 2005

Ms. Lorna Gamboa RA/QA Manager Varian, Inc. 25200 Commercentre Drive Lake Forest, CA 92630

Re: K051235

Trade/Device Name: OnTrak TesTcard™ Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Regulatory Class: Class II
Product Code: DKZ, DIS, JXM, DIO, LAF, DJJ, LCM, LFG, LDJ Dated: May 11, 2005 Received: May 13, 2005

Dear Ms. Gamboa:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 310(x) premaince is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) is regars analysis and the Medical Device American Comments, or to commence program to May 20, 1978, the eccordance with the provisions of the Federal Food, Drug, devices that have been reclassified in accessfired in a premarket approval application (PMA). alle Costlictic Act (71ct) that do not required to the general controls provisions of the Act. The 1 ou may, merceroro, manov as act include requirements for annual registration, listing of general connols provisions of ractice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (bee acover) in s. Existing major regulations affecting your device It inay be subject to Such additional comiser - Marts 800 to 895. In addition, FDA can be louing in Title 21, Coursements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Frease of advised mat I Dri 3 issualled or wouldevice complies with other requirements of the Act that I Dri has made a associations administered by other Federal agencies. You must or any Federal statutes and regulations adminities or of not limited to: registration and listing (21 comply with an the Free STequirements 801 and 809); and good manufacturing practice and requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 5 10(k) I his letter will allow you to begin naketing your actrice of your device of your device to a legally
premarket notification. The FDA finding of substantial equivalence of yo premarket notification. "The I DA miding of our device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, If you destic specific information and advertising of your device, please contact the Office of In or questions on the promotion and Safety at (240) 276-0484. Also, please note the Vitro Diagnostic Dovec Dranation and Bases to premarket notification" (21CFR Part $07.97). regulation entitled, "Misoranant of Frience on your responsibilities under the Act from the You may ootan other general information on Joan Copyrer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Carol C. Benem

Carol C. Benson, M.A. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number:K051233
-------------------------

Device Name: __OnTrak TesTcard™

Indications for Use:

The OnTrak TesTcard menus are various combinations of the different drugs hat are listed The On Frax TesTcard menus are various collegnostic tests for the qualitative detection of drug below. The Tes Icard products are in viro diagnoste the resence of multiple drugs or drug of the fictabolite in article (cutoff) consists of the following:

Cutoff Concentrations
Amphetamines (d,I-amphetamine):1000 ng/mL
Barbiturates (secobarbital):200 ng/mL
Benzodiazepines (oxazepam):100 ng/mL
Cocaine metabolite (benzoylecgonine):300 ng/mL
Methamphetamine (d-methamphetamine):500 ng/mL
Morphine (morphine):300 ng/mL
PCP (phencyclidine):25 ng/mL
Tricyclic Antidepressants (TCA)(Imipramine):1000 ng/mL
THC (11-nor-Δ9-THC-9-carboxylic acid):50 ng/mL

AND/OR

TesTcard products provide only preliminary analytical test result. A more specific alternate rest card products provide only porder to obtain a confirmed analytical result.

Prescription Use V (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Shutt Clam

Page 1 of 1

Division Sign-Off

Office of In Vitro Diagno Device Evaluation and

510(k) K051235

CONFIDENTIAL

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).