K Number
K051206

Validate with FDA (Live)

Date Cleared
2005-11-02

(175 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Y2K2 Enteral In-Line Filter is intended for use to filter liquid feeding solutions prior to delivery into a patient's pre-inserted feeding tube, to minimize the risk of tube occusion by macro particulates.

This will be applicable to any size or length pediatric or adult feeding tube (e.g., nasogastric, nasoduodenal, nasojejunal, gastrostomy, jejunostomy, transgastric duodenal, or transgastic jejunal).

Device Description

The proposed and predicate devices are of essentially the same size, construction, and function. They are made of comparable (Re: Pall) or identical (Re: Cobe Cardiovascular) inert biomaterials. The Y2K2 Enteral In-Line Filter differs in having a larger pore size, to remove coarse (but not fine) particulates from enterally delivered fluids.

AI/ML Overview

Here's the analysis of the provided text regarding the Y2K2 Enteral In-Line Filter:

Acceptance Criteria and Device Performance Study

The provided 510(k) summary (K051206) for the Y2K2 Enteral In-Line Filter does not contain a detailed study report with specific acceptance criteria and performance data in the format typically seen for algorithm-based devices. This document is a premarket notification for a Class II medical device, primarily focused on establishing substantial equivalence to predicate devices.

The "study" in this context refers to the comparison made for substantial equivalence to predicate devices based on technical characteristics and function, rather than a clinical performance study with statistical endpoints.

1. Table of Acceptance Criteria and Reported Device Performance

Feature/CriterionAcceptance Criteria (Implied)Reported Device Performance
Primary FunctionFilter fluids delivered into a feeding tube to remove potentially obstructing particulates.The Y2K2 Enteral In-Line Filter "will filter fluids to be delivered into a feeding tube to remove potentially obstructing particulates." This is identical to its Indications for Use.
Pore SizeSufficient to remove "coarse (but not fine) particulates"The Y2K2 filter "differs in having a larger pore size, to remove coarse (but not fine) particulates from enterally delivered fluids." Specific pore size (e.g., microns) is not stated.
Material BiocompatibilityMade of inert biomaterials.Made of "comparable (Re: Pall) or identical (Re: Cobe Cardiovascular) inert biomaterials."
Size & ConstructionEssentially the same as predicate devices."The proposed and predicate devices are of essentially the same size, construction, and function."
Intended UseTo be used with various feeding tubes (nasogastric, gastrostomy, etc.) for pediatric or adult patients."This will be applicable to any size or length pediatric or adult feeding tube (e.g., nasogastric, nasoduodenal, nasojejunal, gastrostomy, jejunostomy, transgastric duodenal, or transgastic jejunal)."

Summary of the "Study" for Substantial Equivalence:

The document describes the device and states that its technical characteristics, particularly its size, construction, function, and materials, are "essentially the same" as, or "comparable" to, the predicate devices (Pall PharmAssure Capsule Filter K943127 and Cobe Cardiovascular Pre-bypass Filter K850139). The primary difference highlighted is the larger pore size of the Y2K2 filter, designed to remove "coarse" rather than fine particulates. This comparison forms the basis of the "study" for establishing substantial equivalence, indicating that because it is similar to already approved/cleared devices, it meets the necessary safety and effectiveness standards for its intended use.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable in the context of this 510(k) summary. There is no traditional "test set" of patient data in the document for the performance of an algorithm. The evaluation is based on a technical comparison to predicate devices and potentially bench testing (not detailed in this summary) to demonstrate filtration capability.
  • Data Provenance: Not applicable. The "data" here refers to the specifications and characteristics of the device itself and its comparison to predicate devices, not patient data from a particular country or retrospective/prospective study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. This is not an AI/algorithm-based device where experts would establish ground truth for a test set. The evaluation is primarily engineering and regulatory in nature.
  • Qualifications: If any expert input was involved, it would likely be from engineers, materials scientists, and regulatory affairs specialists who assessed the device specifications against regulatory requirements and predicate devices.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There is no test set in the sense of medical images or other data requiring expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, a Multi-Reader Multi-Case comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI devices where human readers' performance with and without AI assistance is evaluated. The Y2K2 is an in-line filter, not a diagnostic AI device.
  • Effect Size: Not applicable.

6. Standalone (Algorithm Only) Performance Study

  • Standalone Study: No, a standalone performance study for an algorithm was not done. The device itself is a physical filter, not a software algorithm. Its "performance" would be assessed through fluid dynamics, particulate retention, and material compatibility testing, not typically through an algorithm-only study.

7. Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" in this context is the established safety and performance profile of the predicate devices and the general understanding of filtration principles and biomaterial compatibility. The Y2K2 device is deemed to perform acceptably because it is "essentially the same" or "comparable" to devices already deemed safe and effective by the FDA. Compliance with manufacturing standards and ISO standards (though not explicitly detailed for filtering capability in this summary) would also contribute to this.

8. Sample Size for the Training Set

  • Sample Size: Not applicable. This is not an AI or machine learning device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment: Not applicable. As there is no training set, there is no ground truth to establish for it.

{0}------------------------------------------------

NOV - 1 2005

510(k) SUMMARY INFORMATION

APPLICANT NAME:Moss Medical Products, Inc.
ADDRESS:4049 NY 150West Sand Lake, New York 12196
PHONE:518-674-0904
FAX:914-684-1464
E-MAIL:mosstube@mossmed.com
ACTIVITY OF APPLICANTInitial Distributor
CONTACT PERSON:GERALD MOSS, Ph.D., M.D.
ADDRESS:1 Reynal RoadWhite Plains, New York 10605
PHONE:914-997-0392
FAX:914-684-1464
E-MAIL:gerald_moss@mossmed.com
NAME OF DEVICE
TRADE NAME:Y2K2 Enteral Tube Fluid Filter
COMMON NAME:In-Line Filter
CLASSIFICATION NAME:{21 CFR 876.5980}{Tubes, Gastrointestinal (and accessories)
PRODUCT CODE:{KNT}
MANUFACTURER:Filtertek, Inc.P.O. Box 31011411 Price RoadHebron, IL 60034-0310
PREDICATE DEVICES:Pall PharmAssure Capsule Filter (K943127

I will maintain records to support a claim (TL, 15 to 127)

Cobe Cardiovascular Pre-bypass Filter (K850139)

INDICATIONS FOR USE: The device will filter fluids to be delivered into a feeding tube to remove potentially obstructing particulates.

TECHNICAL CHARACTERISTICS: The proposed and predicate devices are of essentially the same size, construction, and function. They are made of comparable (Re: Pall) or identical (Re: Cobe Cardiovascular) inert biomaterials. The Y2K2 Enteral In-Line Filter differs in having a larger pore size, to remove coarse (but not fine) particulates from enterally delivered fluids.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows a logo with a stylized bird-like figure composed of three curved lines, suggesting movement or flight. The logo is encircled by text that reads 'Department of Health'. The text is arranged along the curve of the circle, with the bird figure positioned inside the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 1 2005

Moss Medical Products, Inc. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K051206

Trade/Device Name: Y2K2 Enteral In-Line Filter Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 76 KNT Dated: October 15, 2005 Received: October 17, 2005

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. hogdon.

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known):

K051206

Device Name:_Y2K2 Enteral In-Line Filter

Indications For Use:

The Y2K2 Enteral In-Line Filter is intended for use to filter liquid feeding solutions prior to delivery into a patient's pre-inserted feeding tube, to minimize the risk of tube occusion by macro particulates.

This will be applicable to any size or length pediatric or adult feeding tube (e.g., nasogastric, nasoduodenal, nasojejunal, gastrostomy, jejunostomy, transgastric duodenal, or transgastic jejunal).

Prescription Use
X
(Part 21 CFR 801 Subpart D)

ANDIOR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

5

(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.