The IDS 1,25-Dihydroxy Vitamin D kit is a complete assay system intended for purification of 1,25-dihydroxyvitamin D (1,25D) in human serum or plasma by immunoextraction followed by quantitation by enzymeimmunoassay. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of 1,25D deficiency associated with renal disease in adult populations.
Device Story
The IDS 1,25-Dihydroxy Vitamin D EIA is an in vitro diagnostic assay system used in clinical laboratories. It processes human serum or plasma samples to measure 1,25-dihydroxyvitamin D levels. The device utilizes an immunoextraction step to purify the analyte, followed by an enzyme immunoassay (EIA) for quantitation. The output is a quantitative measurement of 1,25D, which clinicians use alongside other clinical and laboratory data to assess 1,25D deficiency in patients with renal disease. The device is intended for professional use by laboratory personnel.
Clinical Evidence
No clinical data provided in the document; the submission relies on bench-level performance characteristics of the assay system.
Technological Characteristics
The device is an enzyme immunoassay (EIA) system. It utilizes immunoextraction for sample purification followed by EIA for quantitation. It is designed for in vitro diagnostic use in human serum or plasma.
Indications for Use
Indicated for the assessment of 1,25-dihydroxyvitamin D (1,25D) deficiency associated with renal disease in adult populations using human serum or plasma.
Regulatory Classification
Identification
A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.
Special Controls
*Classification.* Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.
Related Devices
K042519 — GAMMA-B 1,25-DIHYDROXY VITAMIN D · Immunodiagnostic Systems , Ltd. · Dec 1, 2004
K123253 — IDS ISYS 1,25 DIHYDROXY VITAMIN D, IDS-ISYS, 1,25 DIHYDROXY VITAMIN D CONTROL SET, AND IDS ISYS 1,25 DIHYDROXY VITAMIN D · Immunodiagnostic Systems , Ltd. · Jun 18, 2013
K021163 — OCTEIA 25-HYDROXY VITAMIN D · Immunodiagnostic Systems , Ltd. · May 10, 2002
K172992 — Diazyme EZ Vitamin D Assay · Diazyme Laboratories · Jan 11, 2018
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k051110
B. Purpose for Submission:
New device
C. Analyte:
1,25-Dihydroxy Vitamin D
D. Type of Test:
Enzymeimmunoassay
E. Applicant:
Immunodiagnostic Systems, Ltd.
F. Proprietary and Established Names:
IDS 1,25-Dihydroxy Vitamin D EIA
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1825, Vitamin D test system
21 CFR § 862.1660, Quality control material
2. Classification:
Class II
3. Product Code:
MRG, Vitamin D test system
JJX, Single analyte controls
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indications for Use
2. Indication(s) for use:
The IDS 1,25-Dihydroxy Vitamin D kit is a complete assay system intended
for purification of 1,25-dihydroxyvitamin D (1,25D) in human serum or
{1}
Page 2 of 7
plasma by immunoextraction followed by quantitation by enzymeimmunoassay. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of 1,25D deficiency associated with renal disease in adult populations.
3. Special condition for use statement(s):
For in vitro diagnostic use only. Results are to be used in conjunction with other clinical and laboratory data.
4. Special instrument Requirements:
Instrumentation common to most high complexity laboratories: centrifuge, sample mixer, vortex, photometric microplate reader and data analysis equipment
I. Device Description:
The IDS 1,5-Dihydroxy Vitamin D EIA kit is a complete assay system of the purification of 1,25D in patient samples by immunoextraction followed by quantitation by EIA. Patient samples are dilapidated and 1,25D extracted from potential cross-reactants by incubation for 90 minutes with a highly specific solid phase monoclonal anti-1,25D. The immunoextraction gel is then washed and purified 1,25D eluted directly into glass assay tubes. Reconstituted eluates and calibrators are incubated overnight with a highly specific sheep anti-1,25D. Then a portion of this is incubated for 90 minutes with shaking in microplate wells which are coated with a specific anti-sheep antibody. 1,25D linked to biotin is then added and the plate shaken for a further 60 minutes before aspiration and washing. Enzyme (horse peroxidase) labelled avidin is added and binds selectively to complexed biotin and, following a further wash step, color is developed using a chromogenic substrate (TMB). The absorbance of the stopped reaction mixtures are read in a microtiter plate reader, colour intensity developed being inversely proportional to the concentration of 1, 25D.
Controls included in the device are prepared from human source materials. They have tested negative by FDA recommended assays for the presence of antibodies to HIV (I and II), to HCV, and to Hepatitis B surface antigen.
J. Substantial Equivalence Information:
1. Predicate device name(s):
GAMMA-B 1,25-Dihydroxy Vitamin D RIA
2. Predicate K number(s):
k042519
{2}
Page 3 of 7
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Quantitative determination of 1,25D for assessment of deficiency associated with renal disease | Quantitative determination of 1,25-D for assessment of deficiency associated with renal disease |
| Test Principle | Immunoassay – Competitive binding technique | Immunoassay – Competitive binding technique |
| Antibody Format | Liquid Phase: Polyclonal anti-1,25-Dihydroxyvitamin D [Sheep] in an aqueous buffer | Liquid Phase: Polyclonal anti-1,25-Dihydroxyvitamin D [Sheep] in an aqueous buffer |
| Antigen in calibrators | 1,25- DihydroxyvitaminD3 | 1,25-Dihydroxyvitamin D3 |
| Sample type | Serum or plasma (EDTA or Heparin) | Serum or plasma (EDTA or Heparin) |
| Sample Volume | 100 μL | 100 μL |
| Differences | | |
| Item | Device | Predicate |
| Detection System | Colorimetric | Isotope |
| Test Method | Enzymeimmunoassay (EIA) | Radioimmunoassay (RIA) |
K. Standard/Guidance Document Referenced (if applicable):
CLSI C28-A2 – How to Define and Determine Reference Intervals in the Clinical Laboratory
CLSI EP9-A – Method Comparison and Bias Estimation Using Patient Samples
CLSI EP7-A – Interference Testing in Clinical Chemistry
CLSI EP5-A – Evaluation of Precision Performance of Clinical Chemistry Devices
CLSI EP6-A – Evaluation of the Linearity of Quantitative Analytical Methods
L. Test Principle:
Enzymeimmunoassay
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision was evaluated in accordance with CLSI EP-5A
Evaluation of Precision Performance of Clinical Chemistry Devices. Three human serum controls were assayed over 17 assay days spanning more than 49 operating days. The assays were performed by multiple operators using multiple reagents lots.
{3}
Page 4 of 7
| Control | n | Mean (pmol/L) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% |
| 1 | 28 | 19.0 | 2.0 | 10.7 | 3.8 | 19.7 |
| 2 | 28 | 53.2 | 5.6 | 10.5 | 9.1 | 17.1 |
| 3 | 28 | 152 | 14.1 | 9.3 | 26.7 | 17.6 |
Recovery was assayed by adding 1,25D (D₃) to samples prior to extraction and assay. Results are summarized below (1,25D concentrations are in pmol/L).
| Sample Concentration | 1,25D (D₃) added | Measured pmol/L | Recovery pmol/L | Recovery % |
| --- | --- | --- | --- | --- |
| 62.7 | 46.5 | 106.4 | 43.6 | 98 % |
| 62.7 | 93.0 | 140.7 | 78.0 | 84 % |
| 46.2 | 54.4 | 100.6 | 54.4 | 100% |
| 46.2 | 108.8 | 161.3 | 115.1 | 106 % |
| Mean | | | | 96% |
b. Linearity/assay reportable range:
Linearity studies were designed using CLSI EP6A, Evaluation of the Linearity of Quantitative Analytical Methods. Samples containing varying concentrations of 1,25D prior were assayed in duplicate. The resulting mean concentrations were compared to predicated concentrations. Samples were prepared by diluting a patient sample with assay buffer prior to extraction and assay. The reportable range was determined to be <6 - 333 pmol/L.
c. Traceability (controls, calibrators, or method):
Calibrators and controls are identical to those in the predicate device Gamma-B 1,25-Dihydroxy Vitamin D kit. (K042519). Calibrators are assigned values against primary reference calibrators in the IDS 1,25-Dihydroxy Vitamin D EIA. The mean potency for each calibrator level is calculated from the results of at least 9 independent assays. The assays are carried out on different days by different operators using a variety of reagent batches and expiries.
Acceptable ranges for the controls are assigned by assaying the controls with the IDS 1,25-Dihydroxy Vitamin D EIA. A minimum of 20 values are obtained by performing the assay on different days, by different operators using a variety of reagent batches. The value claimed is the mean of the measurements and the acceptable range is given as the mean ± 24%.
Expiration date of 12 months for the calibrators and 18 months for the controls when stored at 4 °C is based on real time studies that
{4}
Page 5 of 7
show no or minimal deterioration of the calibrators up to 17 months of storage and up to 25 months for the controls. Stability of calibrators and controls after they have been reconstituted and stored at -20 °C, including several freeze thaw cycles, has been established past the recommended 8 week expiry.
d. Detection limit:
The analytical sensitivity of this device was determined to be 11 pmol/L is defined as the concentration corresponding to the mean counts from 10 measurements of the zero calibrator minus 2 standard deviations.
e. Analytical specificity:
Specificity of the antibodies was tested against 2 common metabolites of Vitamin D as well as against the intended analytes. The specificity of the kit was assessed with the following analytes at 50% binding of the zero calibrator.
| Molecule | Cross-reactivity |
| --- | --- |
| 1,25-Dihydroxyvitamin (D_{3}) | 100.0 % |
| 1,25-Dihydroxyvitamin D_{2} | 39 % |
| 24,25-Dihydroxyvitamin D_{3} | 0.058 % |
| 25-Hydroxyvitamin D_{3} | 0.009 % |
Interference testing was conducted in accordance with CLSI EP7-A. Common endogenous substances – haemoglobin, bilirubin, lipid and urea were tested. No interference was found at the levels tested.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The IDS 1,25-Dihydroxy Vitamin D EIA kit was compared against a recognized radioimmunoassay for the quantitative determination of 1,25-dihydroxyvitamin D following CLSI EP9-A2, “Method Comparison ans Bias Estimation Using Patient Samples”. A population of 152 samples, selected to represent a wide range of 1,25-dihydroxyvitamin D [10-402 pmol/L], were assayed by each method. Passing & Bablok regression analysis was performed on the comparative data with the following result:
IDS = 0.94X + 7.2 (95% confidence intervals of the slope and intercept were 0.89 to 1.01, and 2.1 to 12.7 respectively);
{5}
Page 6 of 7
Correlation coefficient (r) = 0.95.
b. Matrix comparison:
A study comparing serum to plasma (EDTA & Heparin) for IDS 1,25-Dihydrox Vitamin D EIA was performed to establish matrix equivalency. No statistical difference in the means was observed between different sample types.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable)
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The 95% reference interval for Normal Adults was determined from 120 apparently healthy adults of US origin, was calculated by a nonparametric method following CLSI guideline C28-A2, "How to Define and Determine Reference Intervals in the Clinical Laboratory".
Normal Adults 39 – 193 pmol/L (n=120)
End-stage of renal disease <6 – 22 pmol/L (n = 24)
These values are suggested guidelines. It is recommended that each laboratory establish the normal range for the area in which it is located.
N. Proposed Labeling:
The labeling is sufficient and satisfies the requirement of 21 CFR part 809.10
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{6}
Page 7 of 7
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.