GAMMA-B 1,25-DIHYDROXY VITAMIN D

K042519 · Immunodiagnostic Systems , Ltd. · MRG · Dec 1, 2004 · Clinical Chemistry

Device Facts

Record IDK042519
Device NameGAMMA-B 1,25-DIHYDROXY VITAMIN D
ApplicantImmunodiagnostic Systems , Ltd.
Product CodeMRG · Clinical Chemistry
Decision DateDec 1, 2004
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 862.1825
Device ClassClass 2

Indications for Use

The IDS Gamma-B 1,25-Dihydroxy Vitamin D kit is a complete assay system intended for the purification of 1,25-dihydroxyvitamin D (1,25D) in human serum or plasma by immunoextraction followed by quantitation by 1251 radioimmunoassay. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in,the assessment of 1,25D deficiency associated with renal disease in adult populations.

Device Story

The IDS Gamma-B 1,25-Dihydroxy Vitamin D kit is an in vitro diagnostic radioimmunoassay (RIA) system used in high-complexity clinical laboratories. The process involves immunoextraction of 1,25D from patient serum or plasma samples using a solid-phase monoclonal anti-1,25D gel, followed by elution. The purified 1,25D is then incubated with sheep anti-1,25D and 125I-labeled 1,25D. Bound and free fractions are separated via Ac-Cel incubation, centrifugation, and decantation. A gamma counter measures the radioactivity, which is inversely proportional to the 1,25D concentration. Clinicians use these quantitative results alongside other clinical and laboratory data to assess 1,25D deficiency in patients with renal disease.

Clinical Evidence

No clinical trials were performed. Evidence consists of bench testing, including precision (n=20 per control level), linearity, analytical sensitivity (11 pmol/L), and specificity. Method comparison against the predicate device (n=112) yielded a Passing & Bablok regression of y = 0.96x - 6.3 (r=0.82). Reference intervals were established using 143 healthy adults and 44 patients with end-stage renal disease.

Technological Characteristics

Competitive radioimmunoassay (RIA) using 125I isotope. Materials include solid-phase monoclonal anti-1,25D gel and sheep anti-1,25D. Requires centrifuge, sample mixer, vortex, and gamma counter. Standards referenced: NCCLS C28-A2, EP9-A, EP7-P, EP5-A, and EP6-P2. Storage at 4°C.

Indications for Use

Indicated for the quantitative determination of 1,25-dihydroxyvitamin D in human serum or plasma to assist clinicians in assessing 1,25D deficiency associated with renal disease in adult populations.

Regulatory Classification

Identification

A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.

Special Controls

*Classification.* Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and (2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k042519 B. Purpose for Submission: New device C. Analyte: 1,25-Dihydroxy Vitamin D D. Type of Test: Radioimmunoassay E. Applicant: Immunodiagnostic Systems, Ltd. F. Proprietary and Established Names: Gamma-B 1,25-Dihydroxy Vitamin D RIA G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1825, Vitamin D test system 2. Classification: Class II 3. Product Code: MRG, Vitamin D test system 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): See Indications for Use 2. Indication(s) for use: The IDS Gamma-B 1,25-Dihydroxy Vitamin D kit is a complete assay system intended for purification of 1,25-dihydroxyvitamin D (1,25D) in human serum or plasma by immunoextraction followed by quantitation by $^{125}\mathrm{I}$ radioimmunoassay. {1} Page 2 of 6 Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of 1,25D deficiency associated with renal disease. 3. Special condition for use statement(s): For in vitro diagnostic use only. Results are to be used in conjunction with other clinical and laboratory data. 4. Special instrument Requirements: Instrumentation common to most high complexity laboratories: centrifuge, sample mixer, vortex, gamma counter I. Device Description: The IDS Gamma-B 1,5-Dihydroxy Vitamin D kit is a complete assay system of the purification of 1,25D in patient samples by immunoextraction followed by quantitation by $^{125}\mathrm{I}$ RIA. Patient samples are dilapidated and 1,25D extracted from potential cross-reactants by incubation for 3 hours with a highly specific solid phase monoclonal anti-1,25D. The immunoextraction gel is then washed and purified 1,25D eluted directly into glass assay tubes. Reconstituted eluates and calibrators are incubated overnight with a highly specific sheep anti-1,25D. $^{125}\mathrm{I}-1,25\mathrm{D}$ is added and incubation continued for 2 hours. Separation of bound from free is achieved by a short incubation with Ac-Cel® followed by centrifugation, decantation and counting. Bound radioactivity is inversely proportional to the concentration of 1,25D. Controls included in the device are prepared from human source materials. They have tested negative by FDA recommended assays for the presence of antibodies to HIV (I and II), to HCV, and to Hepatitis B surface antigen. J. Substantial Equivalence Information: 1. Predicate device name(s): DiaSorin 1,25-Dihydroxyvitamin D $^{125}$I RIA 2. Predicate K number(s): k014030 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Quantitative determination of 1,25D for assessment of deficiency associated with renal disease | Quantitative determination of 1,25-D for assessment of deficiency associated with renal disease | | Test Method | Competitive RIA | Competitive RIA | | Detection | 125I isotope | 125I isotope | | Antibody | Polyclonal sheep anti-1,25D | Polyclonal rabbit anti- | {2} Page 3 of 6 | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | (D_{2}/D_{3}) | 1,25D (D_{2}/D_{3}) | | Antigen in calibrators | 1,25D (D_{3}) | 1,25D (D_{3}) | | Sample type | Serum or plasma | Serum or plasma | | Sample Volume | 500 μL | 500 μL | | Calibration Range | 15 – 500 pmol/L (6 – 208 pg/mL) | 12 – 480 pmol/L (5 – 200 pg/mL) | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Assay duration | > 24 hrs | > 2 hrs | | Analytical Sensitivity | 11 pmol/L (4.6 pg/mL) | 5 pmol/L (2 pg/mL) | | Reference Range | 48 – 150 pmol/L (20.2 – 63.0 pg/mL) | 60 – 157 pmol/L (25.1 – 66.1 pg/mL) | ## K. Standard/Guidance Document Referenced (if applicable): NCCLS C28-A2 – How to Define and Determine Reference Intervals in the Clinical Laboratory NCCLS EP9-A – Method Comparison and Bias Estimation Using Patient Samples NCCLS EP7-P – Interference Testing in Clinical Chemistry NCCLS EP5-A – Evaluation of Precision Performance of Clinical Chemistry Devices NCCLS EP6-P2 – Evaluation of the Linearity of Quantitative Analytical Methods ## L. Test Principle: Radioimmunoassay ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision was evaluated in accordance with NCCLS EP-5A Evaluation of Precision Performance of Clinical Chemistry Devices. Three human serum controls were assayed over 20 assay days spanning more than 90 operating days. The assays were performed by multiple operators using multiple reagents lots. | Control | n | Mean (pmol/L) | Within-run | | Total | | | --- | --- | --- | --- | --- | --- | --- | | | | | SD | CV% | SD | CV% | | A | 20 | 23 | 2.2 | 9.7 | 3.0 | 13.0 | | B | 20 | 61 | 6.5 | 10.7 | 6.1 | 10.1 | | C | 20 | 122 | 6.7 | 5.5 | 12.0 | 9.8 | {3} Page 4 of 6 Recovery was assayed by adding 1,25D (D₃) to samples prior to extraction and assay. Results are summarized below (1,25D concentrations are in pmol/L). | Sample Concentration | 1,25D (D₃) added | Observed Concentration | Recovery (of added analyte) | % Recovery | | --- | --- | --- | --- | --- | | 9.6 | 38.6 | 47.4 | 37.8 | 98 % | | 14.8 | 38.6 | 55.8 | 41.1 | 106 % | | 84.4 | 38.6 | 128.4 | 44.0 | 114 % | | 2.6 | 38.6 | 35.5 | 32.9 | 85 % | | 8.6 | 38.6 | 40.8 | 32.2 | 83 % | | 59.5 | 38.6 | 103.1 | 43.6 | 113 % | | 62.8 | 38.6 | 104.4 | 41.6 | 107 % | | 91.7 | 38.6 | 129.3 | 37.6 | 97 % | | Mean | | | | 101 % | b. Linearity/assay reportable range: Linearity studies were designed using NCCLS EP6-P Evaluation of the Linearity of Quantitative Analytical Methods. Samples containing varying concentrations of 1,25D prior were assayed in duplicate. The resulting mean concentrations were compared to predicated concentrations. Samples were prepared by diluting a patient sample with assay buffer prior to extraction and assay. The reportable range was determined to be 16 – 253 pmol/L. c. Traceability (controls, calibrators, or method): Calibrators are assigned values against primary reference calibrators in the Gamma-B 1,25D assay. The mean potency for each calibrator level is calculated from the results of at least 9 independent assays. The assays are carried out on different days by different operators using a variety of reagent batches and expiries. Acceptable ranges for the controls are assigned by assaying the controls with the Gamma-B 1,25-Dihydroxy Vitamin D assay. A minimum of 20 values are obtained by performing the assay on different days, by different operators using a variety of reagent batches. The value claimed is the mean of the measurements and the acceptable range is given as the mean ± 24%. Expiration date of 12 months for the calibrators and 18 months for the controls when stored at 4 °C is based on real time studies that show no or minimal deterioration of the calibrators up to 17 months of storage and up to 25 months for the controls. {4} Page 5 of 6 Stability of calibrators and controls after they have been reconstituted and stored at -20 °C, including several freeze thaw cycles, has been established past the recommended 8 week expiry. d. Detection limit: The analytical sensitivity of this device was determined to be 11 pmol/L is defined as the concentration corresponding to the mean counts from 10 measurements of the zero calibrator minus 2 standard deviations. e. Analytical specificity: Specificity of the antibodies was tested against 2 common metabolites of Vitamin D as well as against the intended analytes. Specificity was determined by comparing the level at which 50% inhibition of $^{125}\mathrm{I}$ binding was reached with that of the intended analyte. The antibody was raised to 1,25D ($\mathrm{D}_3$). Results are summarized below (presented as a percentage of inhibition by 1,25D ($\mathrm{D}_3$)). | Molecule | Cross-reactivity | | --- | --- | | 1,25-Dihydroxyvitamin (D_{3}) | 100.0 % | | 1,25-Dihydroxyvitamin D_{2} | 91.0 % | | 24,25-Dihydroxyvitamin D_{3} | < 0.0123 % | | 25-Hydroxyvitamin D_{3} | 0.0010 % | Interference testing was conducted in accordance with NCCLS EP7-P. Common endogenous substances – haemoglobin, bilirubin, lipid and urea were tested. No interference was found at the levels tested. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Accuracy was evaluated based on NCCLS Protocol EP9-A2. The study was conducted by comparing to the marketed device. A comparative study of Gamma-B 1,25-Dihydrox Vitamin D kit and the marketed device on 112 patient samples were performed. Passing &amp; Bablok regression analysis was performed with the following result: IDS = 0.96X – 6.3 (95% confidence intervals of the slope are 0.85 and 1.08 and of the intercept are -15.3 and 1.7) r = 0.82; n = 112. {5} Page 6 of 6 b. Matrix comparison: A study comparing serum to plasma (EDTA &amp; Heparin) for Gamma-B 1,25-Dihydrox Vitamin D RIA was performed to establish matrix equivalency. No statistical difference in the means was observed between different sample types. 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable) Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The 95% reference interval for Normal Adults was determined from 143 apparently healthy adults (13 samples were from the US) and was calculated by a nonparametric method following NCCLS guideline C28-A2, "How to Define and Determine Reference Intervals in the Clinical Laboratory". Normal Adults 48 – 150 pmol/L (n=143) End-stage of renal disease &lt;11 – 32 pmol/L (n = 44) These values are suggested guidelines. It is recommended that each laboratory establish the normal range for the area in which it is located. N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...