K050946

Validate with FDA (Live)

Date Cleared
2005-05-27

(42 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

This premarket notification is for the addition of the antimicrobial agent meropenem at concentrations of 0.0313-2 µg/mL to Streptococcus ID/AST or AST only Phoenix panels. Meropenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Streptococcus pneumoniae (excluding penicillin-resistant strains) Viridans group streptococci

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing Streptococcus species the system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST-S Broth used for performing AST tests only. .
  • BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth . determination.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and containing driver organisme or Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Indicator for the deterative of the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35℃. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The device is an Antimicrobial Susceptibility Test (AST) system. The acceptance criteria are based on Essential Agreement (EA) and Category Agreement (CA) with a CLSI reference broth microdilution method.

Acceptance Criteria (Required Performance)Reported Device Performance (Meropenem 0.0313-2 µg/mL for Streptococcal Organisms)
Essential Agreement (EA) %: Not explicitly stated as a minimum percentage but implied to be high for "substantially equivalent performance."97.2% (n=1935)
Category Agreement (CA) %: Not explicitly stated as a minimum percentage but implied to be high for "substantially equivalent performance."95.3% (n=1935)
Intra-site Reproducibility: Greater than 90%Greater than 90% (for this antimicrobial agent)
Inter-site Reproducibility: Greater than 95%Greater than 95% (for this antimicrobial agent)
Overall Reproducibility (Accuracy section): 95% or greater within ± 1 dilution95% or greater within ± 1 dilution

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set (Clinical Studies): The table for Essential Agreement (EA) and Category Agreement (CA) shows n=1935 for the "Antimicrobial" (Meropenem) for streptococcal organisms. This represents the number of isolates tested.
  • Data Provenance: "Clinical, stock and challenge isolates were tested across multiple geographically diverse sites." This indicates a mix of retrospective (stock and challenge) and prospective (clinical) data from various geographical locations.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not specify the number or qualifications of experts used to establish the ground truth. The ground truth method, CLSI reference broth microdilution, is a standardized laboratory procedure, not typically relying on expert consultation for each case's outcome.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for disagreements. The comparison is made against a standardized "CLSI reference broth microdilution method." Discrepancies would likely be investigated for technical or procedural errors, rather than expert adjudication in the traditional sense of medical image interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This study focuses on the performance of an automated microbiology system (BD Phoenix) against a reference standard (CLSI broth microdilution), not on human reader performance with or without AI assistance. The device is a standalone diagnostic tool, not an AI-assisted diagnostic aid for human readers in the context of these types of studies.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone performance study was done. The entire study describes the performance of the "BD Phoenix Automate Microbiology System" (the device/algorithm) in isolation, compared to the CLSI reference method. The results (EA, CA) are for the device's output without direct human interpretation influencing the final MIC or category.

7. The Type of Ground Truth Used

The type of ground truth used is the CLSI reference broth microdilution method. This is a laboratory-based, standardized, and highly accepted method for determining antimicrobial susceptibility.

8. The Sample Size for the Training Set

The document does not provide information on the training set size. This is common for this type of 510(k) submission, as the focus is on the validation performance of the final device against a reference method. The device's underlying "algorithm" (how it interprets readings) would have been developed and "trained" prior to this validation.

9. How the Ground Truth for the Training Set Was Established

The document does not specify how the ground truth for the training set was established, nor does it describe the training process. This information is typically proprietary to the manufacturer and not required for 510(k) submissions focusing on validation against established standards. However, it can be inferred that similar CLSI reference methods or other established laboratory techniques would have been used during the development and calibration of the system.

{0}------------------------------------------------

Koso946

510(k) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: 303-390-3240Fax: 410-316-4499
CONTACT NAME:Linda Dillon,R&D Manager, Pasco Laboratories
DATE PREPARED:April 12, 2005
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System –Meropenem 0.0313-2 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321May 23, 2002), Ofloxacin (K020323, April 14, 2002), andLevofloxacin (K020322, March 27, 2002).
INTENDED USE:The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing Streptococcus species the system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST-S Broth used for performing AST tests only. .
  • BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth . determination.

{1}------------------------------------------------

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and containing driver organisme or Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Indicator for the deterative of the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35℃. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Streptococcus ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of streptoccal isolates. Each site tested the isolates in triplicate on three different days using one lot of Streptococcus Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the streptococcal isolates tested.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across Clinical, stock and chances were mance of the Phoenix antimicrobial susceptibility test with the Streptococcus Phoenix Panel format containing this antimicrobial agent. Phoenix System results the or eprococcas i noums i and romand to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and The portomation of the I novement of the more results for all isolates tested. Essential Agreement Category HEFORMER (OT) To enform Automated Microbiology System agrees exactly or within ± (DA) occurs within to the reference result. Category Agreement (CA) occurs when the BD ono two 1016 unanced Microbiology System agrees with the reference method with respect to the I hoomize - racentated retire (susceptible, intermediate, resistant or nonsusceptible).

Table 1 summarizes the performance for the isolates tested in this study.

Performance of BD Phoenix System for Streptococcal Organisms by Drug Table 1:

Antimicrobialproduction of the controlled on the comments of the comments of the comments of the comments of the consisted on the consisted on the consistence of the consisted on the consConcentrationEA (n)was were would the state of the comments of the first and and the bearing the first and12 . 1 . 1EA (%)THE REAL PROPERTY CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CARATION CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CARATING CONTRACT CARATING CARCA (n)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CA (%)The Real Property of Children a more a more a
The First of Children of Children American Comments of Children Comments ofA - For For Act Accomments of Concession - For Portuge MeropenemThe program and the commend of the comments of the comments of the country of the country of the count0.0313-2 ug/mL19351 07 21193505 3- - -

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

{3}------------------------------------------------

SUMMARY INFORMATION FOR MEROPENEM Available Range 0.0313-2 µg/mL

Performance

Accuracy

and the comments of the comments of the comments of the many of the many of the many of the many of the many of the many of the many of the many of the many of the many of thAntimicrobialConcentrationand and the many of the country of the country of the county of theTHE CHILING OF CHILIFORNIA CONSULTION COLLEGIONAL CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONEA (n)The first and the country of the county of the county of the county of the county of the county of the county of the county of the county of the county of the county of the cA BRICH THE FOR WALLEA (%)Status and Children and Children and Children the States of the Sunday of------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CA (n)CA (%)
Comments of the Research and Children and ChildrenAMeropenem0.0313-2 ug/mL*1035ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ. 07 ?ﮐﺴﻪ » ﺍ102511/2/05 ?ل ه ه له له

Reproducibility

Testing performed at multiple clinical sites demonstrated 95% reproducibility or greater within ± 1 dilution.

Breakpoints - CLSI

OrganismSIR
Streptococcus pneumoniae$\leq$ 0.250.5$\geq$ 1
Streptococcus spp. Other thanStreptococcus pneumoniae *$\leq$ 0.5----

*Per CLSI Performance Standards for Antimicrobial Susceptibility Testing: Fifteenth Informational Supplement, M100-S15, January 2005, in Table 2H for Streptococcus species other than Streptococcus pneumoniae, under Comment 5, the following statement appears describing meropenem breakpoints for Streptococcus species other than Streptococcus pneumoniae: "For some organism/antimicrobial agent combinations, the absence of resistant strains precludes defining any results categories other than 'susceptible'. For strains yielding results suggestive of a 'nonsusceptible' category, organism identification and antimicrobial susceptibility test results should be confirmed."

Recommended Quality Control Organisms

Quality Control StrainMIC RangeSource
Streptococcus pneumoniae ATCC 496190.06-0.25CLSI

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

MAY 2 7 2005

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Linda K. Dillon R&D Manager, Pasco Laboratories BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

K050946 Re: Trade/Device Name: BD Phoenix™ Automated Microbiology System Meropenem (0.0313-2 µg/mL) - Streptococcus ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: April 12, 2005 Received: April 15, 2005

Dear Ms. Dillon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

Sales, a Hogg

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

510(k) Number: KC50946

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent meropenem (0.0313-2 ug/mL) - Streptococcus ID/AST or AST only Phoenix Panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus , Enterococcus, and Streptococcus.

This premarket notification is for the addition of the antimicrobial agent meropenem at concentrations of 0.0313-2 µg/mL to Streptococcus ID/AST or AST only Phoenix panels. Meropenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Streptococcus pneumoniae (excluding penicillin-resistant strains) Viridans group streptococci

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Signature

Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety

BD Diagnostic SystemsBecton, Dickinson and Company510(k) Page 9 K050946
-------------------------------------------------------------------------------

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”