K Number
K050903

Validate with FDA (Live)

Device Name
XAWEX G 100
Date Cleared
2005-06-23

(73 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Xavex G100 is a dental ceramic that can be used by dental technicians to fabricate all-ceramic restorations. It is a partially sintered yttria stabilized zirconium oxide powder that has to be machined with the CAD/CAM technique to its required shape and then sintered to full density. Xavex G100 is recommended for manufacturing substructures of single tooth and bridgework with up to two pontics, which can be used in the anterior as well as in the posterior tooth region.

Device Description

Xavex G100 is a dental ceramic that is composed of partially sintered yttria (yttrium oxide) stabilized zirconium oxide. Xavex G100 is designed for manufacturing substructures of all-ceramic dental appliances for the sole use of particular patients. It has to be machined with the CAD/CAM technique and then sintered to its full density. Xavex G100 is designed for use as single tooth or bridgeworks with up to two pontics in the anterior as well as the posterior teeth region. Xavex G100 is biocompatible and insoluble in water. In addition, the strength of densely sintered Zirconium oxide is extremely high and makes delicate and filigree framework shaping possible. This, together with the white color of the zirconium oxide offers the basis for aesthetically pleasing, safe and effective dental restorations. Xavex G100 substructures can be veneered with suitable dental porcelains with the layering technique. Xavex G100 meets all applicable requirements of the standard ISO 6872: 1997 "Dental ceramic". The partially sintered Xavex G100 blanks will be offered as disks with a diameter of 100 mm and 5 different thicknesses: Xavex G100-25 with a thickness of 25 mm Xavex G100-20 with a thickness of 20 mm Xavex G100-18 with a thickness of 18 mm Xavex G100-14 with a thickness of 14 mm Xavex G100-10 with a thickness of 10 mm

AI/ML Overview

I am sorry, but the provided text from the "Premarket Notification 510 (k) Xavex G100" does not contain information about acceptance criteria, device performance, or any studies conducted to prove the device meets specific criteria. The document describes the device (Xavex G100), its intended use, and the regulatory approval process (510(k) summary and FDA letter of substantial equivalence). It states that the device "meets all applicable requirements of the standard ISO 6872: 1997 'Dental ceramic'," but it does not detail those requirements or present a study proving compliance.

Therefore, I cannot fulfill your request for the specific information regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment based on the provided text.

{0}------------------------------------------------

Premarket Notification 510 (k) Xavex G100

Image /page/0/Picture/1 description: The image shows a date, JUN 23 2005. The text is in a bold, sans-serif font. The date is likely extracted from a document or photograph, indicating the date of creation or relevance of the content.

Image /page/0/Picture/2 description: The image shows the word "WIELAND" in large, bold, black letters. Below the word "WIELAND" is the phrase "Dental + Technik" in a smaller, lighter font. The text is horizontally aligned and centered.

Wieland Dental + Technik Gmbel & Co. KG Schwerninger Straße 13 3-751-49 Pressure Размасы 16.2014 1. 154,9008. 2000: 6年度 2017年8月 sept. 2011 as 1998

6050903

5. 510 (k) Summary

Wieland Dental + Technik GmbH & Co. KG Submitter of 510(k): Schwenninger Str. 13 D-75179 Pforzheim Germany Phone: +49-7231-3705-0

Contact person:Dr. Gerhard Polzer
Phone:+49-7231-3705-219
Fax:+49-7231-357959
e-mail:gerhard.polzer@wieland-dental.de

2005-04-07 Date of Summary:

Trade name:

Xavex G100

Classification name: Product code: C.D.R section: Classification:

Porcelain powder for clinical use ElH 872.6660 Class II

Legally marketed Cercon Base equivalent device:

K013230 510(k) number:

{1}------------------------------------------------

Premarket Notification 510 (k) Xavex G100

Image /page/1/Picture/1 description: The image shows the logo for WIELAND Dental + Technik. The word "WIELAND" is in large, bold, black letters. Below that, in smaller letters, are the words "Dental + Technik."

Wieland Sental - "Ashink GIMBH & Co Schwomenner Microped S : 75, 为何: 31mm Doctforlo in 0-75170 95 Child

510 (k) Summary

Device description

Xavex G100 is a dental ceramic that is composed of partially sintered yttria (yttrium oxide) stabilized zirconium oxide.

Xavex G100 is designed for manufacturing substructures of all-ceramic dental appliances for the sole use of particular patients. It has to be machined with the CAD/CAM technique and then sintered to its full density.

Xavex G100 is designed for use as single tooth or bridgeworks with up to two pontics in the anterior as well as the posterior teeth region.

Xavex G100 is biocompatible and insoluble in water. In addition, the strength of densely sintered Zirconium oxide is extremely high and makes delicate and filigree framework shaping possible. This, together with the white color of the zirconium oxide offers the basis for aesthetically pleasing, safe and effective dental restorations.

Xavex G100 substructures can be veneered with suitable dental porcelains with the layering technique.

Xavex G100 meets all applicable requirements of the standard ISO 6872: 1997 "Dental ceramic".

The partially sintered Xavex G100 blanks will be offered as disks with a diameter of 100 mm and 5 different thicknesses:

Xavex G100-25 with a thickness of 25 mm Xavex G100-20 with a thickness of 20 mm Xavex G100-18 with a thickness of 18 mm Xavex G100-14 with a thickness of 14 mm Xavex G100-10 with a thickness of 10 mm

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 3 2005

Dr. Gerhard Polzer Director Regulatory Affairs Wieland Dental +Technik GmbH & Company KG Schwenninger Strasse 13 D-75179 Pforzheim GERMANY

Re: K050903

Trade/Device Name: Xavex G100 Regulation Number: 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: April 8, 2005 Received: April 11, 2005

Dear Mr. Polzer:

We have reviewed your Section 510(k) premarket notification of intent to market the device wo have ro row and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of Amendments, or to are and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{3}------------------------------------------------

Page 2 -- Mr. Polzer

Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that 1 DT is issualites or that your device complies with other requirements meall that IDA has made a deceminaresulations administered by other Federal agencies. of the Act of ally receitar statutes and reguirements, including, but not limited to: registration You must comply with an the Hec b required (21 CFR Part 801); good manufacturing practice and listing (21 CFR Part 607), laceling (21 CFR Part 820), and if 1 CFR Part 820); and if requirents as set form in the quality by occasions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow Jour of Substantial equivalence of your device to a premarket notification. The PDF results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), II you desire specific advice for your de at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may other general miormational and Consumer Assistance at its toll-free Drisber (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Cars

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Xavex G100 Device Name:

Indications For Use:

Xavex G100 is a dental ceramic that can be used by dental technicians to fabricate all-ceramic restorations. It is a partially sintered yttria stabilized zirconium oxide powder that has to be machined with the CAD/CAM technique to its required shape and then sintered to full density.

Xavex G100 is recommended for manufacturing substructures of single tooth and bridgework with up to two pontics, which can be used in the anterior as well as in the posterior tooth region.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rurr
(Division Sign-Off)

ﻠﺴﻌ Page 1 of

person of Anesthesiology, General Hospital, information of Anesthesiology, Gene

510(k) Number: J

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.