K Number
K050829

Validate with FDA (Live)

Date Cleared
2005-05-06

(35 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nurse's Assistant® 1.7 is intended to be used to route video from multiple sources to multiple destinations, to be used to turn on and off power to specific 110 volt receptacles within the OR, and to be used to adjust specific operating parameters of video displays, analog or digital recording, communication and playback equipment, endoscopic cameras, endoscopic light sources, surgical lamps, observation cameras, surgical cameras and operating room lights.

The Nurse's Assistant® 1.7 is indicated for use in an operating room for video assisted surgery. This includes minimally invasive procedures in all surgical specialties, as an adjunct display of interventional techniques and traditional open procedures.

The Nurse's Assistant® 1.7 is indicated for use in general, cardiovascular, ENT, gastroenterology, urology, plastic, obstetrics, gynecology, and orthopedic surgery, and general thoracoscopy, general cardiothoracic surgery, general laparoscopy, arthroscopy, laparoscopy, nasopharyngoscopy, ear endoscopy and sinusoscopy.

Device Description

The Nurse's Assistant® 1.7 O.R. Control System is a computerized control system that provides operating room (OR) staff with a simple touch panel interface from which to activate, adjust and monitor certain settings of equipment located in the OR.

AI/ML Overview

The provided text is a 510(k) premarket notification for the Nurse's Assistant® 1.7 O.R. Control System. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than conducting new clinical trials to prove safety and effectiveness from scratch. Therefore, the information typically requested in your query (e.g., sample sizes, ground truth establishment, MRMC studies, specific acceptance criteria with performance metrics) is not present in this document, as it is not required for a 510(k) submission that demonstrates equivalence to existing predicate devices.

Instead, the document focuses on comparing the new device's intended use, indications for use, and technological characteristics to those of predicate devices. The "study" in this context is the comparison itself, which concludes that the Nurse's Assistant® 1.7 is substantially equivalent.

Here's an organized breakdown based on the available information, with explanations for what is not present:


Acceptance Criteria and Study for Nurse's Assistant® 1.7 O.R. Control System

Device: Nurse's Assistant® 1.7 O.R. Control System

Approach: This 510(k) submission leverages the concept of substantial equivalence to predicate devices rather than proving de novo safety and effectiveness through clinical trials with specific performance metrics. Therefore, the "acceptance criteria" and "study" are framed around demonstrating equivalence to already approved devices.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
Intended Use: The device must have an intended use substantially equivalent to the predicate devices."The Nurse's Assistant® 1.7 O.R. Control System has similar indications for use, intended use and technological characteristics as the predicate devices."
Indications for Use: The device must have indications for use substantially equivalent to the predicate devices."The Nurse's Assistant® 1.7 O.R. Control System has similar indications for use, intended use and technological characteristics as the predicate devices."
Technological Characteristics: The device must possess technological characteristics substantially equivalent to the predicate devices."The technological characteristics of the Nurse's Assistant® 1.7 are equivalent to the predicate devices listed above."
Safety and Effectiveness: The device must demonstrate substantial equivalence in terms of safety and effectiveness compared to predicate devices."The Nurse's Assistant® 1.7 is substantially equivalent to the predicate devices in terms of safety, effectiveness, and performance."
Performance: Test results for the new device must support substantial equivalence to predicate devices."Tests performed on the Nurse's Assistant® 1.7 demonstrate substantial equivalence to the predicate devices listed above."

Explanation: For a 510(k) submission, the "acceptance criteria" are primarily established by the regulatory requirement to demonstrate substantial equivalence to legally marketed predicate devices. Specific quantitative performance metrics (e.g., sensitivity, specificity, accuracy, precision figures) and numerical acceptance thresholds for these metrics are typically not provided in a 510(k) summary when equivalence is established through comparison of design and intended use, rather than de novo performance testing. The "reported performance" here refers to the conclusion of equivalence based on the comparison and any verification/validation testing performed.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable/Not mentioned. The document does not describe specific patient or image test sets in the typical sense of a clinical performance study. The "tests performed" refer to verification and validation activities to confirm the device operates as designed and is equivalent to predicates.
  • Data Provenance: Not applicable/Not mentioned. The document does not refer to clinical data (e.g., from specific countries, retrospective/prospective studies) for performance evaluation.

3. Number of Experts and Qualifications for Ground Truth

  • Not applicable/Not mentioned. The determination of substantial equivalence does not involve establishing ground truth from experts for a diagnostic output.

4. Adjudication Method for the Test Set

  • Not applicable/Not mentioned. No clinical test set is described.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No. The document does not describe an MRMC study.
  • Effect Size: Not applicable. An MRMC study was not conducted, so no effect size for human reader improvement with AI assistance is reported. This device is an OR control system, not a diagnostic AI tool.

6. Standalone Performance Study (Algorithm Only)

  • Standalone Performance Study: No. This device is an O.R. Control System; it's an integrated hardware/software system to manage other medical equipment, not a standalone algorithm in the typical AI sense that would have an "algorithm only" performance study. The "tests performed" refer to engineering and functional verification to ensure the system operates correctly.

7. Type of Ground Truth Used

  • Not applicable/Not mentioned. The concept of "ground truth" (e.g., pathology, outcomes data) is not relevant to this 510(k) submission, which focuses on functional equivalence of an OR control system, not diagnostic accuracy.

8. Sample Size for the Training Set

  • Not applicable/Not mentioned. This device is a control system, not a machine learning model that would typically have a "training set" of data in the context of AI.

9. How Ground Truth for the Training Set Was Established

  • Not applicable/Not mentioned. As there is no described training set, there is no ground truth establishment for it.

Conclusion from the Document:

The Nurse's Assistant® 1.7 O.R. Control System gained clearance based on demonstrating substantial equivalence to three predicate devices: Val Med Nurse's Assistant O.R. Control System (K010754), Olympus Integrated Endosurgery System EndoALPHA (K981993), and Karl Storz OR1 (KSEA SCB-ACC) (K023704). The "study" proving it meets "acceptance criteria" is the comprehensive comparison of its intended use, indications for use, and technological characteristics to these legally marketed devices, concluding that it is as safe and effective.

{0}------------------------------------------------

Nurse's Assistant® 1.7 O.R. Control System 510(k) #: _

March 28, 2005

K050829

Page 1 of 2

SUMMARY OF SAFETY AND EFFECTIVENESS

In accordance with the requirements of the Safe Medical Device Act of 1990 and In accordance with the roquil offects Systems is hereby submitting the 510(k) 21 CFR 607.92, Oonlined Intogrator of Stems of Stems of Statent® 1.7 O.R. Control System.

Submitter: A.

ConMed Integrated Systems Registration Number: 3031835

Company Contact: B.

C. Jeff Lipps Director of Regulatory Affairs and Quality Assurance 1815 NW 169th Place, Suite 4020 Beaverton, OR 97006 503-614-1106 Phone 503-614-1109 Fax

Device Name: C.

D.

Trade Name:Nurse's Assistant® 1.7 O.R. Control System
Common Name:Surgical Control Center
Classification Names:1. 876.1500 – Endoscope and accessories2. 878.4580 – Surgical lamp3. 884.1720 – Gynecological laparoscope andaccessories
Proposed Class/Device:Class II
Product Codes:GCJ, KOG, FET, FSY, HET, FTA
Predicate/Legally Marketed Devices:
    1. Val Med Nurse's Assistant O.R. Control System K010754
      1. Olympus Integrated Endosurgery System EndoALPHA K981993
      1. Karl Storz OR1 (KSEA SCB-ACC) K023704

{1}------------------------------------------------

K050829

Page 2 of 2

Device Description: E.

The Nurse's Assistant® 1.7 O.R. Control System is a computerized control system that provides operating room (OR) staff with a simple touch panel interface from which to activate, adjust and monitor certain settings of equipment located in the OR.

Intended Use:

The Nurse's Assistant® 1.7 is intended to be used to route video from multiple sources to multiple destinations, to be used to turn on and off power to specific 110 volt receptacles within the OR, and to be used to adjust specific operating parameters of video displays, analog or digital recording, communication and playback equipment, endoscopic cameras, endoscopic light sources, surgical lamps, ebservation cameras, surgical cameras and operating room lights.

Indications for Use:

The Nurse's Assistant® 1.7 is indicated for use in an operating room for video assisted surgery. This includes minimally invasive procedures in all surgical specialties, as an adjunct display of interventional techniques and traditional open procedures.

The Nurse's Assistant® 1.7 is indicated for use in general, cardiovascular, ENT, gastroenterology, urology, plastic, obstetrics, gynecology, and orthopedic surgery, and general thoracoscopy, general cardiothoracic surgery, general laparoscopy, arthroscopy, laparoscopy, nasopharyngoscopy, ear endoscopy and sinusoscopy.

Substantial Equivalence / Device Technological Characteristics and г. Comparison to Predicate Device(s):

The Nurse's Assistant® 1.7 O.R. Control System has similar indications for use, intended use and technological characteristics as the predicate devices; Val Med Nurse's Assistant O.R. Control System - K010754, Olympus Integrated Endosurgery System EndoALPHA - K981993 and Karl Storz OR1 (KSEA SCB-ACC) - K023704.

The technological characteristics of the Nurse's Assistant® 1.7 are equivalent to the predicate devices listed above.

Tests performed on the Nurse's Assistant® 1.7 demonstrate substantial equivalence to the predicate devices listed above.

Conclusion:

The Nurse's Assistant® 1.7 is substantially equivalent to the predicate devices in terms of safety, effectiveness, and performance.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features the department's name encircling a symbol. The symbol consists of three stylized human figures, representing the department's focus on health and well-being. The figures are connected and appear to be moving forward.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY - 6 2005

Mr. C. Jeff Lipps Director, Regulatory Affairs and Quality Assurance ConMed Integrated Systems Incorporated 1815 NW 169th Place, Suite 4020 Beaverton, Oregon 97006

Re: K050829

Trade/Device Name: Nurse's Assistant® 1.7 O.R. Control System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: April 30, 2005 Received: May 2, 2005

Dear Mr. Lipps:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use battler in the May 28, 1976, the enactment date of the Medical Device Amendments, or to conniner of the to May 2011-12-11 accordance with the provisions of the Federal Food, Drug, de nees that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, merestere, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may be subject to bash of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA can oo found in the announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be actives and i Drimination that your device complies with other requirements of the Act that I Dri has made a cond regulations administered by other Federal agencies. You must or any I cach stututed and regirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (21 CFR Part 820); and if applicable, the electronic form in the qualisy by rovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Mr. C. Jeff Lipps

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter with anow you to begin manisting of substantial equivalence of your device to a legally prematics notification. "The PDF inturlig gour device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you ucate specific advise in 90% 276-0115 . Also, please note the regulation entitled, Colliact the Office of Compunation in (21CFR Part 807.97). You may obtain Misolanding by relevelee to promance no promance in the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Colisa.http://www.fda.gov/cdrh/industry/support/index.html.

Sineerely yours,

E. K.

Miriam C. Provost, Ph.D. 8 Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K050829

INTENDED USE / INDICATIONS FOR USE

510(K) # (if known): K050829

Device Name: Nurse's Assistant® 1.7 O.R. Control System

Intended Use /Indications for Use:

The Nurse's Assistant® 1.7 is intended to be used to route video from multiple The Narse's Assistant - 1.7 % intentions, to be used to turn on and off power to specific 110 volt receptacles within the OR, and to be used to adjust specific operating voil receptables within the US , analog or digital recording, communication and parameters of viaoo displayer, andoscopic light sources, surgical playback equipmont, chaosespurgical cameras and operating room lights.

The Nurse's Assistant® 1.7 is indicated for use in an operating room for video The Nurse 3 Aoolotatis includes minimally invasive procedures in all surgical assisted surgery. This molders wif interventional techniques and traditional open procedures.

The Nurse's Assistant® 1.7 is indicated for use in general, cardiovascular, ENT, The Nurse 3 / Solotant - 1. Tastic, obstetrics, gynecology, and orthopedic surgery, gastroenterology, drology, placeral cardiothoracic surgery, general laparoscopy, arthroscopy, laparoscopy, genefanyngoscopy, ear endoscopy and sinusoscopy.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
----------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

  • Jon Davison

406087

Prescription Use
(Part 21 CFR 801.109)

OR

Over-the-Counter Use (Optional Format 1-2-96)

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.