K Number
K050776

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-06-06

(70 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AUSS-6 Ultrasonic Surgical Aspiration System and Accessories are indicated for the fragmentation and aspiration of soft and hard (i.e. bone) tissues in various General and Specialty surgery applications. It is also indicated for use in the debridement of wounds (such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers), soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.

The Alliger AUSS-6 Ultrasonic Surgical System and Accessories are indicated for use in the fragmentation and aspiration of both soft and hard (e.g.: bone) tissue in the following surgical specialties:

  • Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecology External genitalia
    • condyloma
    • benign tumors (lipomas, fibromas, and leiomyomas)
    • malignant primary and metastatic tumors of all types and the following cystic lesions:
    • Bartholin's cysts
    • Vestibular adenitis
    • Inclusion cysts
    • Sebaceous cysts
      Abdominal area
      any abnormal growth, cystic or solid, benign or malignant, involving the ovary, fallopian tube, uterus, or the supporting structures of the uterus.
      Thoracic Surgery
      Limited pulmonary reception such as segmetectomies, nonanatomical subsegmentectomies and metastatectomies.
      Wound Care
      The Misonix Inc. AUSS-6 Utrasonic Surgical Aspirator is also indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.
Device Description

The AUSS-6 Ultrasonic Surgical System is comprised of a generator, which feeds a 22.5 kHz electrical signal to a piezoelectric crystals mounted in a hand-held handpiece; the crystals then vibrate at the same frequency. The titanium tip attached to the handpiece amplifies the vibration. An irrigation unit is provided to introduce irrigation solution to the operative site. An aspirator system removes fragmented material and waste liquids from the area. Accessories include probe tips, wrenches, sterile and non sterile tube sets and sterile Surgical Procedure bags.

AI/ML Overview

The Misonix Inc. AUSS-6 Ultrasonic Surgical Aspirator System and Accessories is a medical device for fragmentation and aspiration of soft and hard tissues in various surgical applications, and for wound debridement.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance/Testing
Safety StandardsCompliance with medical electrical equipment safety standards.The device was designed and tested to pass the following Voluntary Standards: - UL 2601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety - EN 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety
Electromagnetic Compatibility (EMC)Compliance with EMC requirements.The device was designed and tested to pass the following Voluntary Standards: - EN 60601-1-2:2001 Electromagnetic Compatibility - FCC Part 18 EMC Requirements
Non-Clinical PerformanceThe device should demonstrate appropriate functional characteristics for an ultrasonic surgical aspirator.Non-Clinical Tests Performed for Determination of Substantial Equivalence included: - Output Frequency Measurements - Output Power Measurements (No Load to Maximum Load) - Tip Displacement Measurements - Irrigation Flowrate Measurements - Life Tests - Vacuum Flowrate and Pressure Measurements - Input Power Measurements - EMI Tests - Dielectric Tests on Mains Circuits - Patient Current Leakage and Patient Sink Current Measurements - Power Line Ground Leakage Measurements - Dielectric Tests on Patient Circuits
Software ValidationIf applicable, software validation.This device does not contain software.
Sterilization ValidationDevice sterility and appropriate validation.Validation statements are contained in Exhibit J.
Clinical Equivalence (Wound Debridement)The device should be safe and efficacious for wound debridement, and substantially equivalent to predicate devices for this indication, and to sharps debridement. Other indications (soft and hard tissue ablation) are generally accepted as safe and efficacious based on predicate device history.For wound debridement, the claim states: "Based upon a review of the published literature Misonix Inc. can state that the use of an Ultrasonic Surgical Aspirator for Wound Debridement is safe and efficacious. We can also state that the AUSS-6 is substantially equivalent in this regard to the CUSA NS-100, the Misonix Inc. AUSS-5 in soft and hard tissue ablation. The AUSS-6 is also substantially equivalent to sharps debridment of wounds caused by various mechanisms such as burns, radiation and diabetes. Based upon the clinical experiences outlined herein, the Misonix Inc. AUSS-6 Ultrasonic Surgical System and Accessories pose no new issues of safety or efficacy when used for wound debridement." The results of clinical tests for Wound Debridement are included in Exhibit L.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not explicitly stated as numerical data. The submission refers to "several papers" and "clinical tests for Wound Debridment" outlined in Exhibit L. The nature or size of these clinical tests/studies is not detailed in the provided summary.
  • Data Provenance: The document states that Ultrasonic Surgical Aspirators have been employed for 30 years and are "well documented in the public domain." For wound debridement, "published literature" and "clinical tests for Wound Debridment" (Exhibit L) were used. The country of origin is not specified but given the context of FDA submission, it would likely be studies from North America or globally recognized literature. The studies appear to be a mix of retrospective (published literature review) and prospective (the "clinical tests" for AUSS-6).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

  • This information is not provided in the summary. The clinical data relies on "published literature" and "clinical tests," but details about expert involvement in establishing ground truth are absent.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • This information is not provided in the summary.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC or human-AI comparison study was done. This device is an ultrasonic surgical aspirator, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This question is not applicable. The device is a physical surgical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the general efficacy and safety of ultrasonic surgical aspirators in soft and hard tissue ablation, the "ground truth" is based on long-standing clinical use and published literature.
  • For the wound debridement indication, the "ground truth" is derived from published literature discussing the use of ultrasonic surgical systems for debriding wounds, and the results of "clinical tests" on the AUSS-6 device for wound debridement. This implies an assessment of clinical outcomes. Further specifics on how "safe and efficacious" was defined as ground truth (e.g., specific healing rates, reduction in infection, etc.) are not detailed in this summary, but would be expected in Exhibit L.

8. The sample size for the training set:

  • This question is not applicable. The device is a physical surgical instrument, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established:

  • This question is not applicable (as above).

{0}------------------------------------------------

JUN 6 - 2005

K050776

510(k) SUMMARY - Misonix Inc. AUSS-6 Ultrasonic Surgical Exhibit E Aspirator System and Accessories

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92.

!. Submitter's Identification

Submitter's Name:MISONIX INCORPORATED
Address:1938 New Highway, Farmingdale, NY 11735
Telephone Number:516-694-9555
Contact Person:Ronald R. Manna
Date Prepared:February 5, 2005

Name of Device 2.

:

Proprietary Name:Misonix Inc. AUSS-6 Ultrasonic Surgical AspiratorSystem and Accessories
Common/Usual Name:Ultrasonic Surgical SystemUltrasonic Surgical Aspirator
Classification Name:Instrument, Ultrasonic Surgical

3. Predicate Device Information

Predicate DevicesCUSA NS-100 Ultrasonic Surgical AspiratorSystem K801623Alliger Ultrasonic Surgical System AUSS-5K012028
4. Device DescriptionThe AUSS-6 Ultrasonic Surgical System iscomprised of a generator, which feeds a 22.5 kHzelectrical signal to a piezoelectric crystals mountedin a hand-held handpiece; the crystals then vibrateat the same frequency. The titanium tip attached tothe handpiece amplifies the vibration. An irrigationunit is provided to introduce irrigation solution tothe operative site. An aspirator system removesfragmented material and waste liquids from thearea. Accessories include probe tips, wrenches,sterile and non sterile tube sets and sterile SurgicalProcedure bags.

{1}------------------------------------------------

5.Intended Use:The AUSS-6 Ultrasonic Surgical AspirationSystem and Accessories are indicated for thefragmentation and aspiration of soft andhard (i.e. bone) tissues in various Generaland Specialty surgery applications. It is alsoindicated for use in the debridement ofwounds (such as, but not limited to, burnwounds, diabetic ulcers, bedsores andvaginal ulcers), soft tissue debridement andcleansing of the surgical site in applicationsin which, in the physician's judgment wouldrequire the use of an ultrasonic aspiratorwith sharp debridement.
6.Comparison to Predicate DeviceAUSS-6 Ultrasonic Surgical AspiratorSystem and Accessories are similar indesign, material and operating parameters tothe Misonix Inc. AUSS-5 UltrasonicSurgical Aspirator, the CUSA NS-100Ultrasonic Surgical Aspirator.

7. Safety and Performance Data

The Misonix Inc. AUSS-6 Ultrasonic Surgical Aspirator System and Accessories have been designed and tested to pass the following Voluntary Standards:

UL 2601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety EN 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety EN 60601-1-2:2001 Electromagnetic Compatibility FCC Part 18 EMC Requirements

7. Software ValidationThis device does not contain software.
8. Sterilization ValidationsValidation statements are contained in Exhibit J

9. Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

.

{2}------------------------------------------------

Output Frequency Measurements Output Power Measurements (No Load to Maximum Load) Tip Displacement Measurements Irrigation Flowrate Measurements Life Tests Vacuum Flowrate and Pressure Measurements Input Power Measurements EMI Tests Dielectric Tests on Mains Circuits Patient Current Leakage and Patient Sink Current Measurements Power Line Ground Leakage Measurements Dielectric Tests on Patient Circuits

9. Discussions of Clinical Tests Performed

The FDA has cleared all indications for use in the predicates since the early 1980's, with the exception of the express indication of Wound Debridement. Ultrasonic Surgical Aspirators have been employed for use in soft and hard tissue ablation for the past 30 years and are well documented in the public domain. Therefore, no clinicals were conducted in anticipation of this submission for those indications.

The use of an Ultrasonic Surgical System for debriding and cleansing of burns, ulcers or septic wounds has been investigated for decades. Several papers have discussed the use of the product in such applications and tabulated the results in Exhibit L. The AUSS-6 has also been the subject of clinical tests for Wound Debridment. The results of these tests are included in Exhibit L as well.

10. Conclusions

Based upon a review of the published literature Misonix Inc. can state that the use of an Ultrasonic Surgical Aspirator for Wound Debridement is safe and efficacious. We can also state that the AUSS-6 is substantially equivalent in this regard to the CUSA NS-100, the Misonix Inc. AUSS-5 in soft and hard tissue ablation. The AUSS-6 is also substantially equivalent to sharps debridment of wounds caused by various mechanisms such as burns, radiation and diabetes. Based upon the clinical experiences outlined herein. the Misonix Inc. AUSS-6 Ultrasonic Surgical System and Accessories pose no new issues of safety or efficacy when used for wound debridement.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The symbol is enclosed within a circle, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is written around the circumference of the circle.

JUN & - 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Ronald R. Manna Vice President Regulatory Affairs Misonix Incorporated 1938 New Highway Farmingdale, New York 11735

Re: K050776

Trade/Device Name: AUSS-6 Ultrasonic Surgical Aspirator Systems and Accessories Regulatory Class: Unclassified Product Code: LFL Dated: March 23, 2005 Received: March 28, 2005

Dear Mr. Manna:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the encrease to the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Orth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Ronald R. Manna

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Signature

Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K050776

Device Name: AUSS-6 Ultrasonic Surgical Aspirator System and Accessories

Indications for Use:

The Alliger AUSS-6 Ultrasonic Surgical System and Accessories are indicated for use in the fragmentation and aspiration of both soft and hard (e.g.: bone) tissue in the following surgical specialties:

  • Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecology External genitalia
    • condyloma ت
    • benign tumors (lipomas, fibromas, and leiomyomas)
    • ﺪ malignant primary and metastatic tumors of all types and the following cystic lesions:
    • Bartholin's cysts -
    • Vestibular adenitis -
    • Inclusion cysts -
    • Sebaceous cysts «

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

General, Restorative
Dental Devices

Page 1 of 2

{6}------------------------------------------------

510(k) Number (if known): K050776

Device Name: AUSS-6 Ultrasonic Surgical Aspirator System and Accessories

Indications for Use: (continued)

Abdominal area

any abnormal growth, cystic or solid, benign or malignant, involving the ovary, fallopian tube, uterus, or the supporting structures of the uterus.

Thoracic Surgery

Limited pulmonary reception such as segmetectomies, nonanatomical subsegmentectomies and metastatectomies.

Wound Care

The Misonix Inc. AUSS-6 Utrasonic Surgical Aspirator is also indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rüben

050776

Page 2of 2

N/A