K Number
K050668

Validate with FDA (Live)

Date Cleared
2006-02-03

(325 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GRT LITE Model PRO-8A is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and is clinically indicated for:

  • -- Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.
  • -- Adjunctive use in providing temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS).
Device Description

The Model PRO-8A is a hand-held, non-invasive, pain therapy system which utilizes four nonheating light emitting diodes (LED) consisting of two visible LED's and two infrared LED's in one system. It combines the clinically accepted therapeutic treatment of numerous predicate light therapy systems currently in commercial distribution and 510(k) approved. The system consists of a basic hand-held, battery operated, control unit with the four LED's emitting light through a special red acrylic lens which does not absorb any light transmission. The visible LEDs operate at a measured wavelength of 628nm (±5%) and the infrared LEDs operate at a measured wavelength of 850nm (±5%).

AI/ML Overview

The provided document (K050668) describes a 510(k) premarket notification for the GRT LITE Model PRO-8A Light Therapy System. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring a detailed study proving the device meets specific acceptance criteria in the same way a de novo or PMA submission might.

Therefore, the information typically requested about acceptance criteria and a study proving device performance (as in a clinical trial with specific endpoints and statistical analysis) is not applicable in this context.

Instead, the document focuses on demonstrating that the GRT LITE Model PRO-8A has the same intended use and similar technological characteristics as several already approved predicate devices.

Here's a breakdown of the available information regarding the device's characteristics and how its equivalence is established:

1. Table of Acceptance Criteria and Reported Device Performance:

Since this is a substantial equivalence submission, there isn't a table of "acceptance criteria" and "reported device performance" in the traditional sense of a clinical trial. Instead, the "performance" is assessed against the characteristics of the predicate devices.

CharacteristicPredicate Devices (Implied)GRT LITE Model PRO-8A Reported Performance
Intended Use- Adjunctive use for temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin. - Adjunctive use for temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS).- Adjunctive use for temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin. - Adjunctive use for temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS).
Wavelength (Visible LEDs)Clinically accepted therapeutic wavelengths in predicate devices628nm (±5%)
Wavelength (Infrared LEDs)Clinically accepted therapeutic wavelengths in predicate devices850nm (±5%)
Power OutputConforms to specifications of predicate devicesConforms to specifications of predicate devices (implicitly, as it provides the same treatment benefits)
Frequency (Pulsed/Continuous)Conforms to specifications of predicate devicesOffers six different operating modes with specific operating frequencies (pulsed or continuous)
Type of DevicePulsed therapeutic light therapy systems, non-heating infrared lampsHand-held, non-invasive, pain therapy system utilizing four non-heating LEDs (two visible, two infrared)
Safety and Standards ComplianceComplies with relevant FDA regulations and electrical safety standardsComplies with: - 21CFR - Subchapter J - Part 1010 - FCC Standard - 47CFR Part 15B - All Electrical Components Utilized Are UL® Approved

2. Sample Size for the Test Set and Data Provenance:

  • Not Applicable. This submission does not involve a clinical study with a "test set" of patients in the typical sense. Equivalence is established based on technological characteristics and intended use compared to previously approved devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not Applicable. No ground truth for a clinical test set was established as part of this 510(k) submission. The FDA reviewers are the "experts" assessing the information provided.

4. Adjudication Method for the Test Set:

  • Not Applicable. See point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a light therapy system, not an AI-based diagnostic tool for human readers. Therefore, an MRMC study comparing human reader performance with or without AI assistance is irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This device is a physical light therapy system, not an algorithm. Performance is assessed based on its physical characteristics and operational modes.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. The "ground truth" in this context is the FDA's regulatory acceptance, based on previous approvals of similar devices (the predicates). The submission aims to demonstrate that the GRT LITE Model PRO-8A conforms to the performance specifications and clinical parameters already approved for these predicate devices.

8. The sample size for the training set:

  • Not Applicable. No training set for an algorithm is involved.

9. How the ground truth for the training set was established:

  • Not Applicable. See point 8.

Summary of the "Study" (Equivalence Demonstration):

The "study" in this 510(k) submission is a comparison and verification process to demonstrate substantial equivalence to predicate devices:

  • Methodology: The manufacturer compared the GRT LITE Model PRO-8A's intended use and technical characteristics (wavelength, power output, frequency) to those of four identified predicate light therapy systems (Tuco Erchonia PL3000, Excalibur System, Microlight 830 Laser, Acculaser Pro LLLT Device).
  • Testing: Functional performance, electrical safety, and component specification verification were conducted. This included an "eight-stage manufacturing testing and verification GRT Solutions, Inc. procedure protocol" tracked by serial number. This manufacturing testing verified correct diode performance, operating mode parameters, power output, and total operating time.
  • Conclusion: The manufacturer concluded that the GRT LITE Model PRO-8A has the same intended use and similar functional and performance characteristics as the predicate devices, meeting or exceeding required design, testing, and labeling standards, and raising no new regulatory, safety, and/or clinical efficacy issues. The FDA concurred with this assessment, finding the device substantially equivalent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows handwritten text, seemingly a combination of letters and numbers. The text appears to read "k 050668", with the 'k' being a lowercase letter and the rest being numerical digits. The handwriting is somewhat cursive, giving the text a slightly stylized appearance.

510(k) Summ

FEB 3 2006

(as required per 21CFR; §807.92)

GRT LITE Model PRO-8A Light Therapy System

I. ApplicantGRT Solutions, Inc.425 Gardner StLos Angeles, CA 90036Phone: 323-653-2818Fax: 323-653-2298
II. Contact Name:Dr. George Gonzalez; Presidentmydoctorgeorge@hotmail.com
III. Device Name
Proprietary Name GRT LITE Model PRO-8A
Common/Usual Name(s) Light Therapy SystemTherapeutic Light System
Classification Name Infrared Lamp; (21CFR; §890.5500
Regulatory Class Class II
Product Code NHN
Establishment Registration Number 9070001

IV. Predicate Device/Substantial Equivalency

The GRT LITE Model PRO-8A is substantially equivalent to other pulsed therapeutic light therapy systems currently in commercial distribution. The Model PRO-8A has the same intended use and similar technological characteristics to predicate devices. It combines the clinically accepted therapeutic uses of several previously FDA 510(k) approved light therapy systems currently in commercial distribution into one compact system.

The technological equivalence to the predicate devices is substantiated by the wavelength and power output generated by the Model PRO-8A. The Model PRO-8A will provide the same treatment benefits and regimens for clinical presentations already approved by the Food and Drug Administration for the predicate devices.

{1}------------------------------------------------

The predicate devices the Model PRO-8A establishes equivalence to include:

Predicate Device510(k) #Manufacturer
Tuco Erchonia PL3000 .........................K012580Tuco Innovations
Excalibur System .........................K041530Stargate International, Inc.
Microlight 830 Laser .........................K010175Microlight Corporation of America
Acculaser Pro LLLT Device .........................K020657Acculaser, Inc.

V. Intended Use of the Device

The GRT LITE Model PRO-8A is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and is clinically indicated for:

  • -- Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.
  • -- Adjunctive use in providing temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS).

As with the predicate devices, pain therapy treatment can be prescribed for pain associated with the clinical presentations specified above by having the beams pulsed or continuous with time considerations. The GRT LITE Model PRO-8A's variables conform to the performance specifications of the clinical parameters used by the predicate devices in wavelength, frequency as a function of time, and power output.

VI. Description of the Device

The Model PRO-8A is a hand-held, non-invasive, pain therapy system which utilizes four nonheating light emitting diodes (LED) consisting of two visible LED's and two infrared LED's in one system. It combines the clinically accepted therapeutic treatment of numerous predicate light therapy systems currently in commercial distribution and 510(k) approved.

The system consists of a basic hand-held, battery operated, control unit with the four LED's emitting light through a special red acrylic lens which does not absorb any light transmission. The visible LEDs operate at a measured wavelength of 628nm (±5%) and the infrared LEDs operate at a measured wavelength of 850nm (±5%). The Model PRO-8A complies with all performance, labeling, and manufacturing standards set forth in 21CFR.

VII. Summary of Technical Characteristics of the Device To Referenced Predicate Devices

The GRT LITE Model PRO-8A and the aforementioned predicate devices emit visible and invisible photonic energy to human tissue. The comparing of the technologies is dependent on the laws of physics in that the variables are frequency, wavelength, power output, and time.

{2}------------------------------------------------

COSO668

The performance parameters and intended use of the GRT LITE Model PRO-8A are identical to all predicate devices and conforms with all FDA approved application protocols for the devices.

VIII. System Testing

The testing of the Model PRO-8A includes functional performance, electrical safety, and component specification verification. This includes an eight-stage manufacturing testing and verification GRT Solutions, Inc. procedure protocol that is tracked by run component and system serial number.

The operation of the Model PRO-8A is controlled using proprietary PROM technology. Once the CPU is programmed, the "hard software" cannot be changed and/or altered. This hard software affords the operator six different selection options for the Model PRO-8A operating modes. Each of the six operating modes have specific operating frequencies (pulsed or continuous), and visible and infrared light combinations of LED operation. One visible LED is always on, regardless of mode selection, indicating power to the system is on. Every system is checked and tracked, by serial number, for correct diode performance, all six mode operation parameters as they relate to the specific diode, power output, and total operating time.

The Model PRO-8A Light Therapy System is manufactured, performs, is labeled, and is tested to comply with the following standards:

  • . 21CFR - Subchapter J - Part 1010
  • . FCC Standard - 47CFR Part 15B
  • . All Electrical Components Utilized Are UL® Approved

IX. Conclusions

In accordance with testing and comparison to the predicate devices, and pursuant to 21CFR; §890.500, the GRT LITE Model PRO-8A Light Therapy System has the same intended use, with similar functional and performance characteristics.

The device meets or exceeds the design, testing, and labeling standards required by law. The GRT LITE Model PRO-8A Light Therapy System is manufactured and performs as intended and does not raise any new regulatory, safety, and/or clinical efficacy issues.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the branches of government. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

2006 FEB 3

Dr. George Gonzalez President GRT Solutions, Inc. 425 Gardner Street Los Angeles, California 90036

Re: K050668

Trade/Device Name: GRT LITE Model PRO-8A Regulation Number: 21 CFR 890.5500 Regulation Name: Lamp, non-heating for adjunctive use in pain therapy Regulatory Class: II Product Code: NHN Dated: December 19, 2005 Received: December 20, 2005

Dear Dr. Gonzalez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Dr. Gonzalez

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbarar Bnechum

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number: K050668

Device Name: GRT LITE Model PRO-8A Light Therapy System

Indications For Use:

The GRT Model PRO-8A is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and its indications for use:

  • Adjunctive use in providing temporary relief of minor chronic neck • and shoulder pain of musculoskeletal origin.
  • Adjunctive use in providing temporary relief of hand and wrist pain � associated with Carpal Tunnel Syndrome (CTS).

Prescription Use: ANDIOR X (Part 21 CFR 801 Subpart D)

Over The Counter Use:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Offic

Хаиала Висшић

and New

GRT LITE Model PRO-8A 510(k) Application of the program

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.