K Number
K050518

Validate with FDA (Live)

Date Cleared
2005-09-23

(206 days)

Product Code
Regulation Number
886.4670
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reprocessed Phacoemulsification Tips are intended to emulsify and excise cataract tissues in ophthalmic microsurgical procedures.
Reprocessed Phacoemulsification Tips are indicated for use to emulsify and excise cataract tissues in patients requiring eye surgery.

Device Description

Phacoemulsification Tips are used to emulsify and excise cataract tissue in ophthalmic microsurgical procedures. When connected to the ultrasonic handpiece of a phacoemulsification system and activated, the Phacoemulsification Tip vibrates at an ultrasonic frequency that emulsifies cataract tissue. The extracted tissue is then aspirated away through the hollow tip. Irrigation of the eye with a saline solution compensates for the loss of volume in the eye when the cataract tissue is removed.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Alliance Medical Corporation Reprocessed Phacoemulsification Tips, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
BiocompatibilityDemonstrated performance
Validation of ReprocessingDemonstrated performance
Sterilization ValidationDemonstrated performance
Function TestsDemonstrated performance
Packaging ValidationDemonstrated performance

The document states that "Performance testing demonstrates that Reprocessed Phacoemulsification Tips perform as originally intended." This indicates that the device met the acceptance criteria for each of these categories.

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for the bench and laboratory testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The provided text does not mention the use of experts to establish ground truth for the test set. The validation appears to be based on technical performance testing rather than expert-reviewed data.

4. Adjudication Method for the Test Set

The provided text does not mention any adjudication method for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The provided text does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The evaluation focuses on the reprocessed device's physical and functional equivalence to the predicate device, not on human reader performance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

The device described is a physical medical device (Phacoemulsification Tips), not an algorithm or AI system. Therefore, a standalone (algorithm only) performance study is not applicable and was not performed. The "standalone" performance here refers to the device's own functional performance, which was demonstrated through the listed performance tests.

7. Type of Ground Truth Used

The ground truth for the performance evaluation was based on technical specifications and functional output in comparison to the original, non-reprocessed predicate devices. This includes demonstrating:

  • Biocompatibility
  • Successful reprocessing
  • Sterility
  • Appropriate function
  • Packaging integrity

8. Sample Size for the Training Set

The device is a reprocessed physical medical device, not an AI or machine learning model that would require a "training set." Therefore, the concept of a training set sample size is not applicable.

9. How the Ground Truth for the Training Set Was Established

As explained above, there is no "training set" for this type of device. The performance evaluation relies on comparing the reprocessed devices to the established performance and specifications of the original predicate devices.

{0}------------------------------------------------

K050518

PART B: 510(k) SUMMARY

Submitter:Alliance Medical Corporation10232 South 51st StreetPhoenix, Arizona 85044
Contact:Elizabeth RenkenRegulatory Affairs Specialist(480) 763-5394 (o)(480) 763-5320 (f)erenken@alliance-medical.com
Date of preparation:February 25, 2005
Name of device:Trade/Proprietary Name: Reprocessed Phacoemulsification TipsClassification Name: Phacofragmentation System
Predicate DeviceK911808510(k) TitleAlcon® Series 20,000® Legacy®ManufacturerAlcon® Surgical, Inc.
Device description:Phacoemulsification Tips are used to emulsify and excisecataract tissue in ophthalmic microsurgical procedures. Whenconnected to the ultrasonic handpiece of a phacoemulsificationsystem and activated, the Phacoemulsification Tip vibrates at anultrasonic frequency that emulsifies cataract tissue. Theextracted tissue is then aspirated away through the hollow tip.Irrigation of the eye with a saline solution compensates for theloss of volume in the eye when the cataract tissue is removed.
Intended use:Reprocessed Phacoemulsification Tips are intended to emulsifyand excise cataract tissues in ophthalmic microsurgicalprocedures.
Indicationsstatement:Reprocessed Phacoemulsification Tips are indicated for use toemulsify and excise cataract tissues in patients requiring eyesurgery.
Technologicalcharacteristics:The design, materials, and intended use of ReprocessedPhacoemulsification Tips are identical to the predicate devices.The mechanism of action of Reprocessed PhacoemulsificationTips is identical to the predicate devices in that the samestandard mechanical design, materials, and sizes are utilized.There are no changes to the claims, intended use, clinicalapplications, patient population, performance specifications, ormethod of operation. In addition, Alliance Medical Corporation'sreprocessing of Phacoemulsification Tips includes removal ofadherent visible soil and decontamination. Each individualPhacoemulsification Tip is tested for appropriate function of itscomponents prior to packaging and labeling operations.

Alliance Medical Corporation Reprocessed Phacoemulsification Tips

{1}------------------------------------------------

Bench and laboratory testing was conducted to demonstrate Performance data: performance (safety and effectiveness) of Reprocessed Phacoemulsification Tips. This included the following tests:

  • Biocompatibility .
  • Validation of reprocessing .
  • Sterilization Validation .
  • . Function test(s)
  • . Packaging Validation

Performance testing demonstrates that Reprocessed Phacoemulsification Tips perform as originally intended.

Alliance Medical Corporation concludes that the modified Conclusion: devices (Reprocessed Phacoemulsification Tips) are safe, effective, and substantially equivalent to the predicate devices as described herein.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or abstract shapes, creating a sense of unity or interconnectedness.

SEP 2 3 2005

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Alliance Medical, Inc. c/o Ms. Elizabeth Renken Regulatory Specialist 10232 South 51st Street Phoenix. AZ 85044

Re: K050518

KU50518
Trade/Device Name: Alliance Reprocessed Phacofragmentation Needles
1778 Trade/Device Name: 178 Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: NKX Dated: May 4, 2005 Received: August 29, 2005

Dear Ms. Renken:

We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 310(tx) prematic is substantially equivalent (for the indications
referenced above and have determined the device is subscripted in interstate referenced above and nave detellinited the arreted predicate devices marketed in interstate for use stated in the enclosure) to tegally manced te of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment the engagions of the Federal Food. commerce prior to May 28, 1970, the encomence with the provisions of the Federal Food, Drug, devices that have been reciassified in accordance whatest approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act The and Cosmetic Act (Act) that do not require appear of the general controls provisions of the Act. The You may, therefore, market the device, subject to the gents for annual registration, listing of general controls provisions of the receined required.
devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see adver) mo entirely significal sifecting your device can
may be subject to such additional controls. Existing major regulations affective FD may be subject to such additional controlist Existing insults 898 In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 899 be found in the Code of Federal Regarates.
publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualic of a substitution on the requirements of the Act
that FDA has made a determination that your device accepains. You must that HDA has made a delermination inal your areas by other Federal agencies. You must
or any Federal statutes and regulations administered to: registration and listin or any Federal statures and regilances and unities and linited to: registration and listing (21)
comply with all the Act's requirements, including, but not securements as set comply with an the Act S requirements, monanagements practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good if emplicable, the electronic CFR Part 807); labeling (21 CFR 1 at 607); good manat 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000, 1050 forth in the quality systems (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Elizabeth Renken

This letter will allow you to begin marketing your device as described in your Section 510(k)
the same of the many of a first of a least of souvelenee of your device to a le This letter will allow you to begin marketing your article as assisteme of your device to a legally
premarket notification. The FDA finding of substantial end this permits v premarket notification. The FDA inding of substanted by and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs the same of the same and as a the same of a regulation entitled If you desire specific advice for your device on our acoming the regulation entitled,
contact the Office of Compliance at (301) 827-8910. Also, please note of the Mourmay ob contact the Office of Computation in (301) 027 07 (21CFR Part 807.97). You may obtain
"Misbranding by reference to premarket notification the Division of Small "Misbranding by relerence to prematics nontheaton" < Pos Act from the Division of Small
other general information on your responsibilities under the Act from the (800) 638-20 other general information on your responsionities and its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance at to votes himdustry/support/index.html.

Sincerely yours,

David M. Whipple

Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

ATTACHEMENT I

K050518

The Alcon Phacoemulsification Tip Models

to be

Reprocessed by Alliance Medical Corporation:

Reprocessed by Alliance Medical Corp.
MicroTip (0.9MM OD)8065740478
MicroTip (0.9MM OD)30RTS
MicroTip (0.9MM OD)45RTS
MicroTip (0.9MM OD)30KTS
MicroTip (0.9MM OD)45KTS
ABS MicroTip (0.9MM OD)8065790019
ABS MicroTip (0.9MM OD)8065790020
ABS MicroTip (0.9MM OD)8065790021
ABS MicroTip (0.9MM OD)8065790022
ABS MicroTip (0.9MM OD)8065790023
Flared ABS MicroTip (0.9MM OD)8065740836
Flared ABS MicroTip (0.9MM OD)8065740837
Flared ABS MicroTip (0.9MM OD)8065740838
Flared ABS MicroTip (0.9MM OD)8065740839
Flared ABS MicroTip (0.9MM OD)8065740840
TurboSONICS Standard U/S Tip (1.1mm OD)8065740476
TurboSONICS Standard U/S Tip (1.1mm OD)15RT
TurboSONICS Standard U/S Tip (1.1mm OD)30RT
TurboSONICS Standard U/S Tip (1.1mm OD)45RT
TurboSONICS Standard U/S Tip (1.1mm OD)30KT
TurboSONICS Standard U/S Tip (1.1mm OD)45KT
TurboSONICS Standard ABS Tip (1.1mm OD)8065740791
TurboSONICS Standard ABS Tip (1.1mm OD)8065740792
TurboSONICS Standard ABS Tip (1.1mm OD)8065740793
TurboSONICS Standard ABS Tip (1.1mm OD)8065740794
TurboSONICS Standard ABS Tip (1.1mm OD)8065740795

{5}------------------------------------------------

TurboSONICS Standard ABS Flared Tip (1.1mm OD)8065740805
TurboSONICS Standard ABS Flared Tip (1.1mm OD)8065740806
TurboSONICS Standard ABS Flared Tip (1.1mm OD)8065740807
TurboSONICS Standard ABS Flared Tip (1.1mm OD)8065740808
TurboSONICS Standard ABS Flared Tip (1.1mm OD)8065740809

and the comments of the comments of the comments of

and the comments of the comments of the comments of the comments of

...

:

{6}------------------------------------------------

Indications for Use Statement 2.

510(k) Number (if known):

Device Name: Reprocessed Phacoemulsification Tips

Indications for Use: Reprocessed Phacoemulsification Tips are indicated for use to emulsify and excise cataract tissues in patients requiring eye surgery.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CFR 801.109)

Jivision Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises

510(k) Number K050518

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

Alliance Medical Corporation Reprocessed Phacoemulsification Tips Traditional 510(k)

or

§ 886.4670 Phacofragmentation system.

(a)
Identification. A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.(b)
Classification. Class II.