(189 days)
To provide interference fixation of soft tissue graft and bone-tendon-bone patellar graft during the Anterior Cruciate Ligament repair through arthroscopy or arthrotomy.
The TriABSorb™ composite screw is bio-absorbable sterile single use tapered and has a smooth threaded bone screw, which provides interference fixation of soft tissue nas a smooth througou bone patellar grafts during Anterior Cruciate Ligament repair through arthroscopy or arthrotomy. The device is cannulated, tapered with a smooth threaded design, which provides The device to ourmalatour tapers and bone-tendon-bone patellar grafts during Interience inxation of Sont trough arthrough arthroscopy or arthroscopy or arthrotomy procedures. The implants are made of similar biopolymer: PLLA (poly L-lactide acid), Tri-Calcium phosphate (TCP), Hydroxyapatite (HA). The screw comes in multiple lengths (20mm to 45mm), multiple diameters (6mm to 12mm) in each length and left and right thread design, resulting in a set of screws adapted to the morphology of the graft and the patient.
The provided text is a 510(k) summary for a medical device (TriABSorb™ Composite Screw) and the FDA's clearance letter. This type of document focuses on establishing substantial equivalence to previously cleared predicate devices, rather than presenting a detailed study proving the device meets specific performance acceptance criteria through clinical trials or extensive standalone testing.
Therefore, much of the requested information (acceptance criteria, specific study results, sample sizes, expert involvement, adjudication, MRMC studies, effect sizes) is not present in the provided text.
Here is what can be inferred or stated based on the given document:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly stated as quantifiable acceptance criteria with thresholds in this document. The primary "acceptance criteria" here are demonstrating substantial equivalence in material, design, and intended use to predicate devices.
- Reported Device Performance: No specific performance metrics (e.g., tensile strength, degradation rate, clinical success rates) are reported in a quantifiable manner as part of a formal study in this document. The document asserts that minor design modifications "do not affect the use, safety and effectiveness of the device" and that the composite screw has "similar characteristics" to the polymers used in predicate devices.
| Acceptance Criteria (Implied) | Reported Device Performance (from text) |
|---|---|
| Similar materials to predicate devices | Implants made of similar biopolymer: PLLA (poly L-lactide acid), Tri-Calcium phosphate (TCP), Hydroxyapatite (HA) as predicate devices. |
| Similar overall design to predicate devices | The TriABSorb™ composite screw and the predicate devices have the same overall design. Small differences do not affect use, safety, and effectiveness. |
| Similar intended use to predicate devices | "The ABS TriABSorb™ composite screw and predicate device have overall same intended use..." |
| Safety and Effectiveness not adversely affected by differences | Design differences "do not affect the use, safety and effectiveness." |
2. Sample sized used for the test set and the data provenance:
- Sample Size for Test Set: Not applicable/not provided. No specific test set of patients or specimens is described as part of a performance study in this document. The substantial equivalence relies on comparisons to existing predicate designs and materials.
- Data Provenance: Not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/not provided. No "ground truth" establishment by external experts for a test set is described. The FDA itself acts as the expert body reviewing the substantial equivalence claim.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/not provided. No adjudication method for a test set is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a surgical screw, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a surgical screw.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable/not provided in the context of a performance study. The "ground truth" for FDA clearance in this context is the established safety and effectiveness of the predicate devices, to which the new device is compared for substantial equivalence.
8. The sample size for the training set:
- Not applicable/not provided. There is no mention of a "training set" as this is not a machine learning or AI-based device.
9. How the ground truth for the training set was established:
- Not applicable/not provided.
In summary: The provided document is a 510(k) premarket notification, which establishes substantial equivalence rather than presenting clinical study results against predefined acceptance criteria for novel device performance. The "study" referenced is the compilation of information and comparisons demonstrating that the TriABSorb™ Composite Screw is as safe and effective as its predicate devices, based on material properties, design, and intended use.
{0}------------------------------------------------
page 1 of 2
AUG 3 1 2005
510(k) SUMMARY FOR ABS COMPOSITE TriABSorb™ Composite SCREW
ADVANCED BIOMATERIALS 510k #: K050468 Company: 265 Route de La Baronne, St. Jeannet 06640 France Adresse: +33-4-92-12-04-74 Phone: +33-4-92-12-48-70 Fax:
Mr Patrick Janin Contact:
Date of submission: February 18th 2005
Name of device:
TriABSorb™ Composite Screw
Common or usual name:
Composite Interference Screw
Classification:
Smooth or threaded bone fixation fastener: 888.3040
Predicate device:
| K013572 | MITEK BIOCRYL INTERFERENCE SCREWS (MITEK) |
|---|---|
| K032894 | OSTEO ACL SCREW MODEL (LINVATEC) |
| K002070 | BIOLOK SCREW (BIOCOMPOSITES) |
| K003641 | BIOLOK SCREW, BIOSTEON SCREW (BIOCOMPOSITES) |
Device description:
The TriABSorb™ composite screw is bio-absorbable sterile single use tapered and has a smooth threaded bone screw, which provides interference fixation of soft tissue nas a smooth througou bone patellar grafts during Anterior Cruciate Ligament repair through arthroscopy or arthrotomy.
Intended use:
The device is cannulated, tapered with a smooth threaded design, which provides The device to ourmalatour tapers and bone-tendon-bone patellar grafts during Interience inxation of Sont trough arthrough arthroscopy or arthroscopy or arthrotomy procedures.
The ABS TriABSorb™ composite screw and predicate device have overall same intended use and design despite some minor modifications which do not affect the use, safety and effectiveness of the device.
{1}------------------------------------------------
page 2 of 2
The implants are made of similar biopolymer: PLLA (poly L-lactide acid), Tri-Calcium phosphate (TCP), Hydroxyapatite (HA).
The screw comes in multiple lengths (20mm to 45mm), multiple diameters (6mm to 12mm) in each length and left and right thread design, resulting in a set of screws adapted to the morphology of the graft and the patient.
Substantial equivalence:
The polymers for the ABS TriABSorb™ composite screw has similar characteristics than the polymers use for the predicate devices.
The TriABSorb™ composite screw and the predicate devices have the same overall design. In addition, the small differences in design do not affect the use, safety and effectiveness, between the device and the predicate devices.
The principal difference is the mixture of the polymers and the insertion technique between the ABS screw and the predicate devices (the screwdriver driver hole).
The ABS TriABSorb™ composite screw is equivalent in material, design and intended use as the predicate device.
Based on these similarities and equivalences our TriABSorb™ composite screw and the predicate devices are substantially equivalent for the interference fixation of soft tissue and Bone Tendon Bone graft in the ACL reconstruction through arthroscopy or arthrotomy.
Indications for use:
To provide interference fixation of soft tissue graft and bone-tendon-bone patellar graft during the Anterior Cruciate Ligament repair through arthroscopy or arthrotomy.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 3 1 2005
Mr. Patrick Janin ADVANCED BIOMATERIALS 265 Route de la Baronne St-Jeannet, 06640 France
Re: K050468
RoJo 106
Trade/Device Name: TriABSorb™ Composite Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: June 14, 2005 Received: July 1, 2005
Dear Mr. Janin:
We have reviewed your Section 510(k) premarket notification of intent to market the device w o have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreases) 76, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetier for ( 10) has the device, subject to the general controls provisions of the Act. The r ou mayy are s provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or subject to been as begulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -- Mr. Patrick Janin
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin maketing of substantial equivalence of your device to a legally premarket nothleation: "The PDA miams of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire specific advice for your ac no no can and please note the regulation entitled, contact the Office of Complaned at (210) 276 of 807.97). You may obtain Misoraliums by releveloe to premainters in the Act from the Division of Small other general information on your responsibility of the mumber (800) 638-2041 of the hard Manufacturers, International and Consamer Fiscibitition of and only industry/support/index.html
Sincerely yours,
signature
Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510 K Number: K050468
Device name: TriABSorb™ Composite Screw
INDICATIONS FOR USE:
To provide interference fixation of soft tissue graft and bone-tendon-bone patellar graft during the Anterior Cruciate Ligament repair through arthroscopy or arthrotomy.
| Prescription Use(Part 21 CFR 801 Subpart D) | X |
|---|---|
| AND/OR | |
| Over-The-Counter Use(21 CFR 807 Subpart C) | No |
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
| 510(k) Number | K050468 |
|---|---|
| --------------- | --------- |
..
:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
N/A