K Number
K050447

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-03-08

(14 days)

Product Code
Regulation Number
870.4350
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Performa Adult Hollow Fiber Membrane Oxygenator is intended for use in adult surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods of up to 6 hours.

Device Description

The Performa Adult Hollow Fiber Oxygenator is a cardiopulmonary bypass blood oxygenator with an integral heat exchanger.

AI/ML Overview

This document describes the Performa Adult Hollow Fiber Membrane Oxygenator, a medical device, and its evaluation for substantial equivalence to a predicate device. The information provided outlines the non-clinical and in vitro testing conducted. It is important to note that this document is a 510(k) summary, which typically focuses on demonstrating equivalence to an already approved device rather than a comprehensive de novo submission requiring extensive proof of safety and effectiveness from scratch.

Here's an analysis of the provided text in the context of acceptance criteria and a study:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of acceptance criteria with numerical targets. Instead, it states that the tests "met established specifications." The "established specifications" likely refer to the requirements outlined in ISO 7199 and the FDA Guidance for Cardiopulmonary Bypass Oxygenators.

Acceptance Criteria CategoryReported Device Performance
BiocompatibilityMet established specifications (ISO 10993-1:1997 requirements, FDA May 1, 1995 Memorandum)
SterilityMet established specifications
PyrogenicityMet established specifications (nonpyrogenic fluid path)
EO ResidualsMet established specifications
Package IntegrityMet established specifications
Gas Transfer CharacteristicsMet established specifications (ISO 7199 requirements)
Pressure DropMet established specifications (ISO 7199 requirements)
Plasma LeakageMet established specifications (ISO 7199 requirements)
Heat Exchanger PerformanceMet established specifications (ISO 7199 requirements)
Hemolysis/Cell DepletionMet established specifications (ISO 7199 requirements)
Operating Blood VolumeMet established specifications (ISO 7199 requirements)
Mechanical IntegrityMet established specifications (ISO 7199 requirements)
Aging PerformanceTests performed on devices accelerated aged to an equivalent of three years real-time aging, and results met established specifications.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not explicitly state the sample size (number of devices) used for the non-clinical and in vitro tests. It refers to "Applicable tests" and "The Performa Oxygenator aged to 3 years was tested," implying that a certain number of devices underwent these evaluations. Without specific numbers, it's impossible to quantify the sample size.
  • Data Provenance: The tests seem to have been conducted by Dideco S.r.l., which is based in Mirandola (MO), Italy. The data is in vitro and non-clinical, meaning it's from laboratory testing rather than human subjects. It is prospective testing, as the device was manufactured and then subjected to these tests to assess its performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable to the provided document. The study described is an in vitro and non-clinical engineering and performance evaluation of a device. It does not involve human subjects, expert review of images, or clinical outcomes that would require "ground truth" established by medical experts in the way that, for example, an AI diagnostic device would. The "ground truth" here is the established scientific and engineering principles and the specifications outlined in ISO and FDA guidance documents.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials or studies where there are subjective interpretations of data (e.g., medical images, patient symptoms) that require consensus among multiple experts. The tests described are objective, measurable physical and chemical properties of the device, assessed against predefined quantitative specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. An MRMC study is relevant for diagnostic or screening devices where human readers (e.g., radiologists) interpret cases (e.g., medical images) with and without the assistance of an AI algorithm. The Performa Adult Hollow Fiber Membrane Oxygenator is a therapeutic device (for cardiopulmonary bypass), not a diagnostic tool, and its evaluation does not involve human readers interpreting cases or AI assistance in that context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable. "Standalone performance" in this context typically refers to the performance of an AI algorithm independent of human intervention. The Performa device is a physical medical device, not an AI algorithm. Its performance is evaluated mechanically, chemically, and biologically through in vitro tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

As mentioned in point 3, the concept of "ground truth" for this type of device evaluation differs from clinical studies involving diagnostic accuracy. The "ground truth" or reference standards used are:

  • Established specifications/standards: Primarily the requirements from ISO 10993-1:1997 (for biocompatibility) and ISO 7199 (for in vitro performance characteristics such as gas transfer, pressure drop, plasma leakage, heat exchange, hemolysis, operating blood volume, and mechanical integrity).
  • FDA Guidance: The FDA May 1, 1995 Memorandum (for biocompatibility) and the Guidance for Cardiopulmonary Bypass Oxygenators 510(k) submissions -- Final Guidance for industry and FDA Staff November 13, 2000 provided the framework and specific criteria for the in vitro testing.

These standards serve as the objective "ground truth" against which the device's performance is measured.

8. The sample size for the training set

This section is not applicable. A "training set" refers to data used to train a machine learning or AI model. The Performa oxygenator is a physical medical device, not an AI algorithm, and therefore does not have a "training set."

9. How the ground truth for the training set was established

This section is not applicable. As there is no training set for an AI algorithm, the concept of establishing ground truth for it does not apply here.

{0}------------------------------------------------

Performa Adult Hollow Fiber Membrane Oxygenator Dideco S.r.l.

MAR 8 - 2005

510(k) SUMMARY

SUBMITTER:Dideco S.r.l.86, Via Statale 12 Nord41037 Mirandola (MO) Italy
CONTACT PERSON:Luigi VecchiPhone: 011 39 0535 29811Fax: 011 39 0535 25229
DATE PREPARED:February 21, 2005
DEVICE TRADE NAME:Performa Adult Hollow Fiber Membrane Oxygenator
COMMON NAME:Hollow Fiber Oxygenator
CLASSIFICATION NAME:Cardiopulmonary Bypass Oxygenator.
PREDICATE DEVICE:Apex Ph.I.S.I.O. Adult Hollow Fiber Oxygenator

DEVICE DESCRIPTION:

The Performa Adult Hollow Fiber Oxygenator is a cardiopulmonary bypass blood oxygenator with an integral heat exchanger.

INDICATION FOR USE:

The Performa Adult Hollow Fiber Oxygenator is intended for use in adult surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods of up to 6 hours.

TECHNOLOGICAL CHARACTERISTICS:

The Performa adult hollow fiber oxygenator is identical in design to the Apex Ph.1.S.I.O. adult hollow The Forformal adult heller modifications made to the device consist of an overall reduction in the size of the device, updating of the instructions for use reflecting this modification and changes in format, ergonomic and aesthetic variations that do not change the function of the device. The fundamental scientific technology is unchanged from the predicate device.

The oxygenator is ethylene oxide sterilized and has a nonpyrogenic fluid path. It is for single use only.

NONCLINICAL TEST RESULTS:

Applicable tests were carried out in accordance with the requirements of ISO 10993-1:1997 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of raw materials. Tests were performed on devices accelerated aged to an equivalent of three years real time aging. Sterility, pyrogenicity, EO residuals and package integrity testing were also conducted. The results of this testing met established specifications.

IN VITRO TEST RESULTS:

In vitro testing was carried out in accordance with the requirements of ISO 7199 and the Guidance for Cardiopulmonary Bypass Oxygenators 510(k) submissions -- Final Guidance for industry and FDA Staff November 13, 2000 to provide the data necessary to demonstrate compliance of the predicate device with safety and effectiveness requirements. The Performa Oxygenator aged to 3 years was tested for gas transfer characteristics, pressure drop and plasma leakage, heat exchanger performance, hemolysis/ cell depletion, operating blood volume and mechanical integrity. The results The modifications being made to the Performa of these tests met established specifications.

{1}------------------------------------------------

Oxygenator do not affect the performance of the device; therefore, the functional and biocompatibility Oxygenator us not anoot the Ph.I.S.I.O. apply only to the Performa Oxygenator.

CONCLUSION:

The Performa membrane oxygenator is substantially equivalent to the Apex P.h.l.S.l.O predicate The Penomia Thembrane Oxygenator 15 Gabocandary - Sce is biocompatible according to its intended device. Biocompanility studies demonstrated the effectiveness of production techniques to assure that the oxygenator is sterile and non-pyrogenic.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a stylized human figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 8 - 2005

Dideco S.r.I. c/o Mr. Barry Sall Senior Consultant Parexel International Corporation 195 West Street Waltham, MA 02451-1163

Re: K050447

Performa Adult Hollow Fiber Membrane Oxygenator Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II (two) Product Code: DTZ Dated: February 21, 2005 Received: February 22, 2005

Dear Mr. Sall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classificd (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Barry Sall

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issuance of a substanted activation of the requirements of the Act
that FDA has made a determination that your device on plier Federal agencies. that FDA has made a determination that your device compines with one oncess. You must
or any Federal statutes and regulations administered by other Federal on and listi or any Federal statutes and regulations administration application and listing (21
comply with all the Act's requirements, including, but not limited to: reguirements as set comply with all the Act 's requirements, including, but not minted novelse requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice, the ele CFR Part 807); labeling (21 CFR Pat 801); good manufacturing practice liqueded, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicab forth in the quality systems (QS) regulation (2) CFR 1 at 620), and 21 (2FR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections 31-342 of the 100), 21 0100
This letter will allow you to begin marketing your device of your Section 510(k)
This letter will a This letter will allow you to begin marketing your device of your device of your device to a legally
premarket notification. The FDA inning of substantial equir device to a premarket notification. The FDA finding of substanial equivalence of your device and thus, permits your device to proceed to the market.

lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r and the If you desire specific advice for your device on our adocting included to the specificals on the contact the Office of Compliance at (501) 394-40-1 could office of Compliance at
promotion and advertising of your device, please contact the Office of Compliance promotion and advertising of your device, please contact, "Misbranding by reference to
(301) 594-4639. Also, please note the regulation entitled, "Misbranding of (301) 594-4639. Also, please note the regulation entitler, "WilsOthers. Policy on your
premarket notification" (21CFR Part 807.97) you may obtain. Other general information premarket notification" (21CFR Pat 80 /.97) you may bivision of Small Manufacturers,
responsibilities under the Act may be obt its toll from the Division of Small or (301) 44 responsibilities under the Act may be obtained from the Roll of Sames of Sames of Sames of Same of Constitutions of American Property of Constitutions of Constitutions of Con International and Consumer Assistance at the results of the main.html

Sincerely yours,

Vitina R. Vichner

(V)_Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): Kのら0447

Device Name: Performa Adult Hollow Fiber Membrane Oxygenator

Indications For Use:

The Performa Adult Hollow Fiber Membrane oxygenator is intended to be used in adult surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods up to 6 hours.

Prescription Use __ X (Part 21 CFR 801 Subpart D)


AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Divire D. Vordone1

(Division Sign-Off) Jivision of Cardiovascular Devices

510(k) Number_k (25447

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”