RNA1C HEMOGLOBIN A1C CONTROL FOR BAYER DCA2000 AND DCA2000+ ANALYZER

K050100 · Bionostics, Inc. · JJX · Feb 25, 2005 · Clinical Chemistry

Device Facts

Record IDK050100
Device NameRNA1C HEMOGLOBIN A1C CONTROL FOR BAYER DCA2000 AND DCA2000+ ANALYZER
ApplicantBionostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateFeb 25, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers is intended to be used to monitor and evaluate the analytical performance of the Bayer DCA 2000 and DCA 2000+ Analyzers.

Device Story

RNA1c Hemoglobin A1c Control is a two-level, non-biological aqueous liquid solution containing a synthetic peptide identical to the HbA1c epitope and dye to simulate total hemoglobin. It serves as a quality control material for Bayer DCA 2000 and DCA 2000+ Analyzers. The device is used by laboratory personnel to perform periodic quality control checks. The user applies the solution to the analyzer's test cartridge via a dropper tip. The analyzer measures the control using its standard immunoassay and colorimetry methods; the resulting HbA1c% value is compared against expected ranges to verify the precision and analytical performance of the system. This process helps ensure the accuracy of patient HbA1c testing, supporting clinical decision-making in diabetes management.

Clinical Evidence

No clinical data. Bench testing only, including closed bottle stability, stability after opening, correlation to predicate device, and test precision/range.

Technological Characteristics

Two-level aqueous liquid control solution. Contains synthetic peptide (HbA1c epitope) and dye in non-biological aqueous base. Packaged in plastic bottle with dropper tip. Storage at 2-8°C. Liquid state stability: 12 months at 2-8°C.

Indications for Use

Indicated for use as an objective quality control material to monitor and evaluate the analytical performance and precision of Bayer DCA 2000 and DCA 2000+ Analyzers measuring HbA1c. Intended for in vitro diagnostic use in laboratory settings.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 5 2005 Image /page/0/Picture/1 description: The image shows the logo for Bionostics Ferraris Quality Solutions. The logo features the word "BIONOSTICS" in bold, sans-serif font on the left side of the image. To the right of the word is a black, semi-circular shape. Below the semi-circle and to the right is the text "FERRARIS Quality Solutions" in a smaller font. K050100 # 510(k) Summary1 | (a) (1) Submitter's name, address | |-----------------------------------| | Bionostics, Inc. | | 7 Jackson Road | | Devens, MA 01434 | Contact Person Kathleen Storro Director, QA & Regulatory Affairs (978) 772-7070 x 220 Date of preparation of this summary: 17 January 2005 - Device trade or proprietary name: (2) RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers ## Device common or usual name or classification name: Glycosylated Hemoglobin Control | CLASSIFICATION | | | | |------------------------------------------------|----------------|-------|-----------| | PRODUCT NOMENCLATURE | NUMBER | CLASS | PANEL | | Control, Single Analyte (Assayed or Unassayed) | 862.1660 (JJX) | I | Chemistry | ## I. Substantial Equivalence RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers is substantially equivalent in function, safety and efficacy to Bayer DCA 2000 Hemoglobin A1c Normal and Abnormal Control Kit. [K021484] | Comparison of Technological Characteristics with Predicate Device | |-------------------------------------------------------------------| |-------------------------------------------------------------------| | Characteristic | New Device | Predicate Device | |-------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | Name: | RNA1c Hemoglobin A1c Control for<br>Bayer DCA 2000 and DCA2000<br>Analyzers | Bayer DCA 2000+ HbA1c Control | | 510(k), Date: | | K021484, 06/14/2002 | | Description: | Red, aqueous solution with non-<br>biological synthetic peptide | Lyophilized Hemoglobin Control | | Intended Use: | As a quality control solution for use<br>to verify the performance of the<br>Bayer DCA 2000+ Analyzer for the<br>measurement of HbA1c | As a quality control lysate to monitor<br>the precision of laboratory<br>procedures for measurement of<br>HbA1c | | Number of levels: | 2 | 2 | | Analytes: | Glycated hemoglobin (HbA1c%) | Glycated hemoglobin (HbA1c%) | | Container: | plastic bottle | 3.5 mL clear borosilicate glass vial<br>with plastic screw cap and phenolic<br>moisture barrier liner. | <sup>1</sup> This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. {1}------------------------------------------------ | Characteristic | New Device | Predicate Device | |-------------------------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Serology Testing of<br>Human Source<br>Material | No serology testing is necessary.<br>Non-biological solution with<br>synthetic peptide | Non-reactive for:<br>• Hepatitis B Surface Antigen<br>• Antibody to Hepatitis C<br>• Antibody to HIV-1 & HIV-2<br>• Syphilis (TPHA & RPR) | | Storage: | 2° - 8°C | 2° - 8°C | | Stability of Lyophilized<br>Product: | N/A – liquid solution | 3 years at 2° - 8°C | | Stability in Liquid State<br>(reconstituted) | 12 months at 2° - 8°C | 13 weeks at 2° - 8°C | | Color: | Red | Dark red (after reconstitution) | #### II. Description of the new device Glycosylated Hemoglobin (HbA1c) is a human hemoglobin A (HbA0) with glucose irreversibly bound to the N-terminal valine of the beta-chain. %HbA1c is the percentage of glycolsylated hemoglobin molecules as a percentage of the total hemoglobin concentration. %HbA1c is quantified on medical devices by turbidimetric inhibition immunoassay (TINIA), ion-exchange HPLC, boronate affinity HPLC or gel electrophoresis. The Bayer DCA 2000+ [K940971] determines HbA1c concentration by immunoassay and total hemoglobin concentration by colorimetry. Using the values obtained for each of these two analytes, the percentage of the total hemoglobin that is glycated is calculated and reported as HbA1c%. RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers is a two-level, aqueous liquid control solution consisting of a synthetic peptide identical to the HbA1c epitope in a non-biological aqueous solution with dye to provide the appropriate total hemoglobin value. The concentration of dye and peptide are optimized for use on the Bayer DC A 2000 and DCA 2000+ Analyzers to provide measurement values for HbA1c equivalent to the predicate device, Bayer DCA 2000 Hemoglobin A1c Normal and Abnormal Control Kit. RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers provides a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program. The product is packaged in a plastic bottle with dropper tip for application of the solution to the test cartridge. {2}------------------------------------------------ RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers is a non-hazardous aqueous solution containing no biological materials and requires no reconstitution prior to use. - (5) Intended use of the device RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers is intended to be used to monitor and evaluate the analytical performance of the Bayer DCA 2000 and DCA 2000+ Analyzers. #### (6) Technological characteristics of the device. RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers is a two-level, aqueous liquid control solution consisting of a synthetic peptide identical to the HbA1c epitope in a non-biological aqueous solution with dye to provide the appropriate total hemoglobin value. The concentration of dye and peptide are optimized for use on the Bayer DCA 2000 and DCA 2000+ Analyzer to provide measurement values for HbA1c equivalent to the predicate device, Bayer DCA 2000 Hemoglobin A1c Normal and Abnormal Control Kit. ### (b) (1) Summary of non-clinical tests submitted with the premarket notification for the device. Tests were conducted to verify specific performance requirements: - a) Closed bottle stability - b) Stability after opening - c) Correlation to predicate device - d) Test precision and range ## Summary of clinical tests submitted with the premarket notification for the (b) (2) device. N/A #### (b) (3) Conclusions drawn from the clinical and non-clinical trials. Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, composed of three curved lines. FEB 2 5 2005 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Kathleen Storro Director, QA and Regulatory Affairs Bionostics, Inc. 7 Jackson Road Devens, MA 01432 Re: k050100 Trade/Device Name: RNA1c Hemoglobin A1c Control for Bayer DCA2000 and DCA2000+ Analyzers Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assay and unassayed) Regulatory Class: Class I Product Code: JJX Dated: January 17, 2005 Received: January 18, 2005 Dear Ms. Storro: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Jean M.Cooper MS, DUM Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of *In Vitro* Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K050100 RNA1c Control for Bayer DCA2000 and DCA2000+ Analyzers Device Name: Indications For Use: RNA1c Control for Bayer DCA2000 and DCA2000+ Analyzers is intended to be used to monitor and evaluate the analytical performance of the Bayer DCA 2000 and DCA 2000+ Analyzers that measure HbA1c. The use of quality control materials is indicated as an objective assessment of the precision of methods and techniques in use and is an integral part of good laboratory practice. The two levels of controls allow performance monitoring within the clinically important range. For In Vitro Diagnostic Use Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Carol C. Benson Division Sign-Off Office of a firo Diognostic Devine Bunbation and Sately 510K K050100 Page 1 of
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%