The MedX HOME System Console System is an infrared lamp system, The Medx HOME System Console of Stations of Station of Call heating as per 21 CFR 890.5500. as per 21 CFR 090155001 Title temporary increase in local blood when neat is indicated "relief of minor muscle and joint aches, pains Circulation, temporary relier of muscles; for muscle spasms, and minor pain and stiffness associated with arthritis".
MedX HOME console powers one to two SLD Accessories. It uses the MedX 600 accessory, called the MedX 601 SLD accessories for the HOME unit and a smaller console, based on the MedX 1100 Console. The MedX HOME console powers one to two SLD accessories. The MedX HOME Console includes semiconductors and assembly, electronics, controls, liquid crystal display, front panel and labels. The front panel contains the membrane switch with command keys to enter data and select options. The port on the right side of the console is dual purpose. First, with the use of a programming key, the clinician can access software to sets the specific treatment parameters for individual patients. Without this key plugged in the software is not visible. The second purpose of the port is for the patient to pluq in the SLD accessories. This only provides access to a limited area of the software, allowing the patient to select one of potentially 5 different treatment numbers associated with a pre-programmed treatment. The left side of the console panel has the ON/OFF switch and the plug for the electrical outlet. The liquid crystal display provides 4 lines of 15 characters each line for prompting, options, and selections. The main label is located on the rear panel of the console. The MedX HOME uses the MedX 600 SLD accessory, called the MedX 601 for this purpose, to generate therapeutic heating, to comply with the heating category, ILY requirements. The SLD accessory use 633 nm and 870 nm superluminous diodes to produce the heat required for topical heating, resulting in the temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, and minor pain and stiffness associated with arthritis. The SLD device generates 500 mW of power to produce therapeutic heat. The accessories are powered by the console and can be set to pulse from 1 – 2000 pulses per second. The console automatically turns itself off after the set treatment time has been delivered.
Here's a breakdown of the acceptance criteria and study information for the MedX HOME device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Stated Requirements) | Reported Device Performance |
|---|---|
| Safety: | |
| Designed in accordance with UL 2601 | Meets UL 2601 requirements (Canada) |
| Meets international and US medical electrical equipment standards for safety | Meets requirements of international and US medical electrical equipment standards for safety |
| Internal testing for mechanical, electrical, controls, thermal safety, environmental conditions, electromagnetic compatibility, temperature control, irradiation distribution patterns | Found to be safe in all tested areas for intended use |
| Performance/Effectiveness: | |
| Warm surface temperature to therapeutic heating for topical heating | Demonstrated ability to warm the surface temperature to therapeutic heating |
| Generate therapeutic heating for: | |
| - Temporary increase in local blood circulation | Clinical practice and academic research demonstrate IR lamp can provide this |
| - Temporary relief of minor muscle and joint aches, pains and stiffness | Clinical practice and academic research demonstrate IR lamp can provide this |
| - Relaxation of muscles | Clinical practice and academic research demonstrate IR lamp can provide this |
| - For muscle spasms | Clinical practice and academic research demonstrate IR lamp can provide this |
| - Minor pain and stiffness associated with arthritis | Clinical practice and academic research demonstrate IR lamp can provide this |
| Deliver 500 mW of power | SLD device generates 500 mW of power |
| Able to pulse from 1 – 2000 pulses per second | Accessories can be set to pulse from 1 – 2000 pulses per second |
| Automatic shut-off after set treatment time | Console automatically turns itself off after the set treatment time has been delivered |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable. The submission states, "No clinical research was conducted for this specific 510(k) submission."
- Data Provenance: Not applicable for clinical testing. For non-clinical testing, it was "Internal testing." The device also notes Health Canada Medical Device Branch Therapeutic Devices Directorate approval (Canada).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Not applicable, as no clinical research was conducted for this specific submission to establish ground truth for a test set. The claims for therapeutic heating effects are based on "Clinical practice and academic research" for infrared lamps generally, not specific to this device's clinical trial.
4. Adjudication Method for the Test Set
- Not applicable, as no clinical research was conducted for this specific submission.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The submission explicitly states "No clinical research was conducted for this specific 510(k) submission."
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- This device is a therapeutic heating device, not an AI/algorithm-based diagnostic or assistive system. Therefore, the concept of "standalone (algorithm only)" performance is not relevant. The device operates as intended by delivering physical energy (heat).
7. Type of Ground Truth Used
- For the device's safety and functional performance, the ground truth appears to be based on engineering standards compliance (UL 2601, international and US medical electrical equipment standards) and internal testing results demonstrating the device's physical output (e.g., warming surface temperature, power output).
- For the therapeutic claims (e.g., temporary relief of muscle aches), the ground truth is implicitly based on existing clinical consensus and academic research regarding the known effects of therapeutic heating from infrared lamps, as applied to the predicate device and the general class of devices.
8. Sample Size for the Training Set
- Not applicable. The MedX HOME is a physical medical device that delivers therapeutic heat, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as the device doesn't use a training set in the context of machine learning.
{0}------------------------------------------------
DISCUSSION OF SAFETY AND EFFECTIVENESS 4.
510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA A.
| Model No. /Names: | MedX HOME (Console and 1 - 2 SLDAccessories - MedX 601) |
|---|---|
| Classification: | Lamp Infrared, Heating Category ILYPhysical Medicine Device, 21 CFR 890.5500(Class II) |
| Predicate Device: | MedX 1000 Series Console & LED Accessories(uses the MedX 600 LED Accessory) |
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
Description of the Device
માટતાં:
MedX HOME console powers one to two SLD Accessories. It uses the MedX 600 accessory, called the MedX 601 SLD accessories for the HOME unit and a smaller console, based on the MedX 1100 Console.
The MedX HOME console powers one to two SLD accessories. The MedX HOME Console includes semiconductors and assembly, electronics, controls, liquid crystal display, front panel and labels. The front panel contains the membrane switch with command keys to enter data and select options.
The port on the right side of the console is dual purpose. First, with the use of a programming key, the clinician can access software to sets the specific treatment parameters for individual patients. Without this key plugged in the software is not visible.
MedX HOME 510 (k) Submission
{1}------------------------------------------------
The second purpose of the port is for the patient to pluq in the SLD accessories. This only provides access to a limited area of the software, allowing the patient to select one of potentially 5 different treatment numbers associated with a pre-programmed treatment.
The left side of the console panel has the ON/OFF switch and the plug for the electrical outlet.
The liquid crystal display provides 4 lines of 15 characters each line for prompting, options, and selections.
The main label is located on the rear panel of the console.
The MedX HOME uses the MedX 600 SLD accessory, called the MedX 601 for this purpose, to generate therapeutic heating, to comply with the heating category, ILY requirements.
The SLD accessory use 633 nm and 870 nm superluminous diodes to produce the heat required for topical heating, resulting in the temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, and minor pain and stiffness associated with arthritis.
The SLD device generates 500 mW of power to produce therapeutic heat. The accessories are powered by the console and can be set to pulse from 1 – 2000 pulses per second. The console automatically turns itself off after the set treatment time has been delivered.
No national body has established safety or performance standards for infrared lamps in the US or Canada. The device is designed in accordance with UL 2601, and meets the requirements of this standard in Canada with field inspections completed by Entella on all equipment prior to shipping.
The products have been:
- Health Canada Medical Device Branch Therapeutic . Devices Directorate approved (Canada)
- . ISO 9001 and ISO 13485
- . QSR compliant
MedX HOME 510 (k) Submission
10
KOS OOZZ
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white graphic that appears to be a map or diagram. It contains a series of dots and lines arranged in a somewhat scattered pattern. There is a distinct arrow-like shape pointing upwards on the right side of the image. The overall impression is that of a simplified representation of a network or spatial arrangement.
11 KOSOOZZ
Technology Characteristics Summary
The technological characteristics are based on the predicate device -K020017.
MedX HOME has been tested in the areas of mechanical, electrical, controls, and thermal safety, environmental conditions and electromagnetic compatibility, temperature control and irradiation distribution patterns. The MedX HOME has been found to be safe in all of the above areas for the intended use referenced in this submission.
Discussion of Non-Clinical and Clinical Data
Internal testing has demonstrated the ability of MedX HOME to warm the surface temperature to therapeutic heating. No clinical research was conducted for this specific 510(k) submission.
Conclusions Demonstrating Safety, Effectiveness and Performance
The testing carried out for MedX HOME indicates that they meet design and performance functional requirements. Clinical practice and academic research demonstrate that therapeutic heating from an infrared lamp can be used successfully to provide topical heating for the temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles, for muscle spasms, minor pain and stiffness associated with arthritis. The proposed device meets the requirements of international and US medical electrical equipment standards for safety, and key performance and safety requirements.
MedX HOME 510 (k) Submission
{3}------------------------------------------------
DEVICE SPECIFICATIONS B.
The MedX HOME Console
The MedX HOME Console powers the SLD accessories that produce the therapeutic heating. The console supports the electronics, controls and labels. See Appendix 1(a) for the MedX HOME Main Board, 1 (b) the MedX HOME Daughter Board and 1 (c) the Block Diagram of the MedX HOME system.
The front panel contains the membrane switch with four buttons for Select/Start, menu, pause/clear and < and > keys. The liquid display window has 4 lines with 15 characters per line. The main label is located on the console rear panel. The ON/OFF switch is on the left side of the console with the electrical plug input. The right side of the console has a dual purpose port. When the clinician plugs in the set-up key, software is accessible to select specific treatment parameters, including treatment time and pulsing for up to 5 different treatments (A, B, C, D, and E) and number of days the unit will function prior to requiring re-setting by the clinician. When the SLD accessory is plugged into the port and the console is turned ON, only the pre-set treatments (A through E) appear on the LED display. (Figure 1 - MedX HOME Console Front View and SLD Accessories (Front and Back Views).
The MedX HOME SLD Accessory
The MedX HOME powers and controls the MedX 601 SLD accessory, consisting of 9 - 633 nm visible red SLD and 40 - 870 nm infrared SLDs. The MedX HOME SLD accessory treatment area or radiant surface is 19.4 cm². The enclosure material is made of ABS plastic, with the radiant surface covered with a polycarbonate window. (Figure 1 - The MedX 601 Accessory Radiant Surface and Back Views are included.
llegi:
Image /page/3/Picture/7 description: The image shows the text "MedX HOME" and "510 (k) Submission". The text is black and is on a white background. The text is likely part of a document or presentation.
K 0500ZZ
4/
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized design featuring three horizontal lines that curve downwards, resembling a stylized human form.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 5 2005
Ms. Anita Saltmarche RN, BScN, MHSc VP of Clinical and Scientific Affairs MedX Health, Incorporated 3350 Ridgeway Drive Unit 3 Mississauga Ontario, Canada L5L 5Z9
Re: K050022
Trade/Device Name: MedX Home Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: April 15, 2005 Received: April 20, 2005
Dear Ms.Saltmarche:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaused in ad the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use cared in to May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered processified in accordance with the provisions of the Federal Food, Drug, de nees mat have been rocation of require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, diererere, mains of the Act include requirements for annual registration, listing of general obliavil profitering practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your devibe to such additional controls. Existing major regulations affecting your device n may of sayon to basis of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i rease of advisod that i Dr renation that your device complies with other requirements of the Act that I DA has made a associations administered by other Federal agencies. You must of any I odetar statuate is requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 – Ms. Anita Saltmarche RN, BScN, MHSc
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and will be finding of substantial equivalence of your device to a legally premarket notification - 12 a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at rioliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Stupt Rurch
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510 (k) Number:
Device Names: MedX HOME
INDICATION FOR USE
Indication for Use
The MedX HOME System Console System is an infrared lamp system, The Medx HOME System Console of Stations of Station of Call heating as per 21 CFR 890.5500. as per 21 CFR 090155001 Title temporary increase in local blood when neat is indicated "relief of minor muscle and joint aches, pains Circulation, temporary relier of muscles; for muscle spasms, and minor pain and stiffness associated with arthritis".
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × OR
Over-the-counter Use -(Optional Format 1-2-96)
(Division Sign-off)
Division of General Restorative Devices
510(k) Number:
Stypt Rhodes
(Division S Division of General, Restorative, and Neurological Devices
510(k) Number K050022