(29 days)
The intended use of the UCR Spinal System is as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur.
The intended use of the UCR Spinal System, when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System, is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies,
- severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusions by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion (indication for use only when used as a Spondylolisthesis Spinal Fixation Device System),
- spondylolisthesis,
- trauma (i.e., fracture or dislocation),
- spinal stenosis,
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
- spinal tumor.
- pseudoarthrosis, and/or
- failed previous fusion.
The intended use and indications of the UCR Spinal System, when used as a Spinal Interlaminal Fixation Orthosis or Hook Spinal System, are limited to T1-L5 and are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies,
- spondylolisthesis
- trauma (i.e., fracture or dislocation),
- spinal stenosis.
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- tumor,
- pseudarthosis, and/or
- failed previous fusion.
The UCR Spinal System includes titanium alloy polyaxial and polyaxial reduction screw assemblies. Both screw types will be offered in numerous widths and lengths. The screw assembly is comprised of a screw body, washer, and housing. Both screws are designed to be compatible with and work in conjunction with the components in the current UCR Spinal System. The product is supplied "NON-STERILE" and must be sterilized prior to use.
The UCR Spinal System will also include titanium alloy cross links, crossbars, rods, and caps. The cross links will come in one size, the crossbars in five sizes, the rods in two diameters, and the caps in two sizes. The cross link assemblies are comprised of a nut and hook while the crossbar assemblies are made up of two bars, two locking screws, a coupler, and a pin. All of these devices are designed to be compatible with and work in conjunction with the components in the current UCR Spinal System. The product is supplied "NON-STERILE" and must be sterilized prior to use.
The UCR Spinal System will offer a wide variety of instruments that range from in-situ rod benders to modular taps. These various instruments will be made primarily from various grades of stainless steel with a few components made from aluminum, titanium alloy, Radel, and silicone. These items are supplied "NON-STERILE" and must be sterilized prior to use.
The provided document is a 510(k) summary for the UCR Spinal System, which is a medical device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting a de novo study to establish new performance criteria. Therefore, the information you've requested regarding acceptance criteria, study details, expert involvement, and ground truth is generally not applicable in the context of this 510(k) submission.
Here's why and what information is available:
1. A table of acceptance criteria and the reported device performance
- Acceptance Criteria: For a 510(k) submission like this, the "acceptance criteria" revolve around demonstrating that the new device shares the same fundamental scientific technology, has the same intended use, and is as safe and effective as a legally marketed predicate device. Specific quantitative performance metrics (like sensitivity, specificity, accuracy for an AI device) and their acceptance thresholds are typically not established or reported in this format. The primary acceptance criterion here is "Substantially Equivalent."
- Reported Device Performance: The document states that the new components of the UCR Spinal System are "Substantially Equivalent" to the current UCR Spinal System (predicate device) in terms of Intended Use, Indications for Use, Design, Screw Sizes, Cross Links, Crossbars, Rods, Caps, Material, Sterilization, and Method of Sterilization. For "Mechanical Strength," it states "See test results," implying that testing was done to ensure the mechanical properties are similar to the predicate, thus meeting the substantial equivalence requirement.
Table of Acceptance Criteria and Reported Device Performance (based on provided text):
| Feature | Acceptance Criteria (for Substantial Equivalence) | Reported Device Performance (New UCR Spinal System vs. Current UCR Spinal System) |
|---|---|---|
| Intended Use | Intended use should be similar to the predicate device. | Similar |
| Indications for Use | Indications for use should be similar to the predicate device. | Similar |
| Design | Design principles and components should be similar to the predicate device. | Similar |
| Screw Sizes | Dimensional specifications should be within the range or similar to those of the predicate device. | Similar (details in prints) |
| Cross Links | Dimensional specifications should be within the range or similar to those of the predicate device. | Similar (details in prints) |
| Crossbars | Dimensional specifications should be within the range or similar to those of the predicate device. | Similar (details in prints) |
| Rods | Dimensional specifications should be within the range or similar to those of the predicate device. | Similar (details in prints) |
| Caps | Dimensional specifications should be within the range or similar to those of the predicate device. | Similar (details in prints) |
| Material | Material should be the same or a comparable, biocompatible material to the predicate device. | Titanium alloy (Similar) |
| Sterile | Sterility status should be similar to the predicate device. | Non-sterile (Similar) |
| Method of Sterilization | Method of sterilization should be suitable and similar to the predicate device. | High-temperature steam (Similar) |
| Mechanical Strength | Mechanical strength should be comparable to or meet the safety requirements of the predicate device. | See test results (implies equivalence was demonstrated) |
The remaining points are largely not applicable because this is a 510(k) for a physical medical device (spinal system components), not an AI/software device or a clinical study establishing new performance benchmarks.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. This document describes a submission for a physical implantable device (UCR Spinal System components). The "test set" in this context would refer to mechanical testing of the physical components, not a dataset of patient images or clinical data for an AI algorithm. The document states "See test results" for mechanical strength, indicating such testing was performed, but specific sample sizes and data provenance for these engineering tests are not detailed in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. Ground truth, in the sense of expert consensus on clinical diagnoses or image interpretations, is not relevant for a 510(k) submission for a physical spinal implant. "Ground truth" for this device would relate to engineering specifications and material properties, validated by standard mechanical testing protocols (e.g., ASTM standards), not medical experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., in radiological reads for AI studies). This is not relevant for mechanical testing of a medical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. MRMC studies are specifically for evaluating diagnostic accuracy with and without AI assistance in situations involving medical image interpretation. This submission is for a physical orthopedic implant, not an AI diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This device is a physical implant, not a standalone AI algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Not Applicable. As mentioned, "ground truth" for this device would be derived from physical and mechanical performance specifications and testing results, confirming that the new components meet established engineering standards and are substantially equivalent to the predicate device. It would not typically involve medical "ground truth" like pathology or outcomes data in the context of a 510(k) for device components.
8. The sample size for the training set
- Not Applicable. There is no "training set" in the context of a physical medical device submission like this. Training sets are relevant for machine learning algorithms.
9. How the ground truth for the training set was established
- Not Applicable. There is no training set or associated ground truth for this type of device submission.
{0}------------------------------------------------
UCR Spinal System
510(K) SUMMARY
Pursuant to 510(i) of the Federal Food, Drug, and Cosmetic Act, as amended, and in accordance with 21 CFR § 807.92.
| Submitter Information: | SeaSpine, Inc.Contact: Diana Smith2302 La Mirada DriveVista, CA 92081-7862Phone: 760-727-8399 Fax: 760-727-8809 |
|---|---|
| Company Registration Number: | 2032593 |
| Submission Correspondent: | SeaSpine, Inc.Contact: Diana Smith, Manager ofRegulatory Affairs and Quality Assurance2302 La Mirada DriveVista, CA 92081-7862Phone: 760-727-8399 Fax: 760-727-8809 |
| Date Summary Prepared: | November 19, 2004 |
| Classification Name: | Pedicle Screw Spinal SystemMNH (Class II) - 888.3070(b)(1)MNI (Class II) - 888.3070(b)(1)NKB (Class III) – 888.3070(b)(2)Spinal Interlaminal Fixation OrthosisKWP (Class II) – 888.3050 |
| Common/Usual Name: | Polyaxial Screws, Polyaxial ReductionScrews, Cross Links, Crossbars, Rods, Caps,Components, and Instruments |
| Device Trade Name: | UCR Spinal System |
The primary device used for comparison in this summary is SeaSpine Inc.'s existing UCR Spinal System (K993503, K021623, and K032739/S1).
- Intended Use: (The statements of intended use are identical.)
The intended use of the UCR Spinal System, polyaxial screws, polyaxial reduction screws, cross links, crossbars, rods, caps, set screws, and components, is substantially equivalent to the intended use of the prodicate device. The intended use of the UCR Spinal System is as a temporary or permanent posterior, non-cervical implant to
। 8
{1}------------------------------------------------
UCR Spinal System
correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur.
The intended use of the UCR Spinal System, when used as a Pedicic Screw Spinal System or Spondylolisthesis Spinal Fixation Device System, is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows:
- . degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies,
- severe spondylolisthesis (Grades 3 and 4) of the US-S1 vertebra in skeletally . mature patients receiving fusions by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion (indication for use only when used as a Spondylolisthesis Spinal Fixation Device System),
- . spondylolisthesis,
- . trauma (i.e., fracture or dislocation),
- . spinal stenosis,
- . deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
- . spinal tumor.
- pseudoarthrosis, and/or ◆
- . failed previous fusion.
The intended use and indications of the UCR Spinal System, when used as a Spinal Interlaminal Fixation Orthosis or Hook Spinal System, are limited to TI-I.S. and are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin . with degeneration of the disc confirmed by patient history and radiographic studies,
- spondylolisthesis .
- trauma (i.e., fracture or dislocation), .
- . spinal stenosis.
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), �
- . tumor,
- pseudarthosis, and/or
- failed previous fusion. �
2. Description:
The UCR Spinal System includes titanium alloy polyaxial and polyaxial reduction screw assemblics. Both screw types will be offered in numerous widths and lengths. The screw assembly is comprised of a screw body, washer, and housines. Both! screws are designed to be compatible with and work in conjunction with the components w
19
Page 2 of 4
{2}------------------------------------------------
UCR Spinal System
the current UCR Spinal System. The product is supplied "NON-STERILE" and must be sterilized prior to use.
The UCR Spinal System will also include titanium alloy cross links, crossbars, rods, and caps. The cross links will come in one size, the crossbars in five sizes, the rods in two diameters, and the caps in two sizes. The cross link assemblies are comprised of a nut and hook while the crossbar assemblies are made up of two bars, two locking screws, a coupler, and a pin. All of these devices are designed to be compatible with and work in conjunction with the components in the current UCR Spinal System. The product is supplied "NON-STERILE" and must be sterilized prior to use.
The UCR Spinal System will offer a wide variety of instruments that range from insitu rod benders to modular taps. These various instruments will be made primarily from various grades of stainless steel with a fcw components made from aluminum, titanium alloy, Radel, and silicone. These items are supplied "NON-STERILE" and must be sterilized prior to use.
3. Technological Characteristics:
The polyaxial screws, polyaxial reduction screws, cross links, crossbars, rods, caps, and components in this submission have been designed as an addition to the current UCR Spinal System. The devices being added to the UCR Spinal System have substantially equivalent technological characteristics to the predicate devices and are designed to be compatible with and work in conjunction with the current components in the UCR Spinal System. Refer to Table 1 in the following section, entitled Comparison Analysis, for a summation of technological characteristics such as design, dimensional specifications, and material.
4. Comparison Analysis:
The overall designs of the UCR Spinal System polyaxial screws, polyaxial reduction screws, cross links, crossbars, rods, caps, and components are substantially equivalent to the predicate devices. See Table 1 on the following page for a companison of the UCR Spinal System polyaxial screws, polyaxial reduction screws, cross links, crossbars, rods, caps, and components to the predicate devices.
Page 3 of 4
20
{3}------------------------------------------------
Ko43232
UCR Spinal System
| Feature | UCR Spinal System(new components) | UCR Spinal System(current) | SubstantiallyEquivalent |
|---|---|---|---|
| Intended Use | The intended use of the UCR Spinal Systemis as a temporary or permanent posterior,non-cervical implant to correct spinaldisorders and provide stabilization of thespine to permit the biological process ofspinal fusions to occur. | Similar | Yes |
| Indications for Use | • Degenerative disc disease (DDD) asdefined by back pain of discogenic originwith degeneration of the disc confirmed bypatient history and radiographic studies,• spondylolisthesis,• trauma (i.e., fracture or dislocation),• spinal stenosis,• deformities or curvatures (i.e., scoliosis,kyphosis, and/or lordosis).• tumor,• pseudarthosis, and/or• failed previous fusion. | Similar | Yes |
| Design | Polyaxial screws, polyaxial reductionscrews, cross links, crossbars, rods, caps,and components | Similar | Yes |
| Screw Sizes | See prints | Similar | Yes |
| Cross Links | See prints | Similar | Yes |
| Crossbars | See prints | Similar | Yes |
| Rods | See prints | Similar | Yes |
| Caps | See prints | Similar | Yes |
| Material | Titanium alloy | Similar | Yes |
| Sterile | Non-sterile | Similar | Yes |
| Method ofSterilization | High-temperature steam | Similar | Yes |
| Mechanical Strength | See test results | Similar | Yes |
Table 1: Summary of Design Comparison
21
Page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized symbol that resembles an abstract bird or human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 1 2004
Ms. Diana Smith Manager of Regulatory Affairs and Quality Assurance SeaSpine, Inc. 2302 La Miranda Drive Vista, California 92081
Re: K043232
Trade/Device Name: UCR Spinal System Regulation Number: 21 CFR 888.3050, 21 CFR 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Pedicle screw spinal system Regulatory Class: III Product Code: NKB, KWP, MNH, MNI Dated: November 17, 2004 Received: November 22, 2004
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Diana Smith
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Lila witte
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K043232
Device Name: UCR Spinal System
Indications for Use:
The intended use of the UCR Spinal System, polyaxial screws, polyaxial reduction screws, cross links, crossbars, rods, caps, set screws, and components, is substantially equivalent to the intended use of the predicate device. The intended use of the UCR Spinal System is as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur.
The intended use of the UCR Spinal System, when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System, is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin . with degeneration of the disc confirmed by patient history and radiographic studies,
- severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally . mature patients receiving fusions by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion (indication for use only when used as a Spondylolisthesis Spinal Fixation Device System),
- spondylolisthesis, .
- trauma (i.e., fracture or dislocation), ●
- spinal stenosis, .
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restoration and Neurological Devices
| 510(k) Number | K043232 |
|---|---|
| --------------- | --------- |
Page 1 of 2
Page 1 of 225
{7}------------------------------------------------
Indications for Use Statement continued
The intended use and indications of the UCR Spinal System, when used as a Spinal Interlaminal Fixation Orthosis or Hook Spinal System, are limited to T L J and are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin . with degeneration of the disc confirmed by patient history and radiographic studies,
- spondylolisthesis .
- trauma (i.e., fracture or dislocation), .
- . spinal stenosis.
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), .
- . tumor.
- pseudarthosis, and/or .
- failed previous fusion.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Page 2 of 2
N/A