K Number
K043039

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-01-18

(75 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synthes External Fixation Devices - MR Safe are intended for use in the construction of an external fixation frame for treatment of various fracture types that require external fixation.

LARGE
Provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for:

  • Stabilization of soft tissues and fractures .
  • Poyltrauma/multiple orthopedic trauma .
  • Vertically stable pelvic fractures, or as a treatment adjunct for vertically unstable pelvic . fractures
  • Arthrodeses and osteotomies with soft tissue problems; failures of total joints .
  • Neutralization of fractures stabilized with limited internal fixation .
  • Non-unions/septic non-unions .
  • Intra-operative reductions/stabilization tool to assist with indirect reduction .
  • Unilateral rectilinear bone segment transport or leg lengthening .

MEDIUM
Indicated for construction of an external fixation frame for the treatment of pediatric and adult fractures.

Device Description

Synthes External Fixation Systems, MR Safe are a system of components that form a construct intended to treat stable and unstable fractures. All frame elements are made from non-magnetic materials and are intended for use in the MR environment. This line extension consists of Synthes Pin Clamps with Outrigger Posts and additional Curved Carbon Fiber Rods. These components are also intended for use in the MR environment.

AI/ML Overview

Although the provided document discusses a 510(k) premarket notification for a medical device (Synthes External Fixation Component Line Extension - MR Safe), it does not contain the information requested to describe the acceptance criteria and the study that proves the device meets those criteria.

A 510(k) submission primarily demonstrates "substantial equivalence" to a legally marketed predicate device, rather than providing detailed studies with specific acceptance criteria and performance data as might be found in a De Novo or PMA submission.

Therefore, for each of your requested points, the answer is that the information is not available in the provided text.

Here's a breakdown of why based on the document:

  1. A table of acceptance criteria and the reported device performance: The document only describes the device and its intended use, and states that substantial equivalence has been demonstrated. It does not provide any specific performance metrics or acceptance criteria a study would measure against.

  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): No study is described, so no information on sample size or data provenance is present.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Since no study is detailed, there's no mention of experts or ground truth establishment.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: No study means no adjudication method is discussed.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is an external fixation component, not an AI diagnostic tool, so an MRMC study related to AI assistance would not be applicable or expected. The document does not mention any such study.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: As this is a physical medical device and not an algorithm, a standalone algorithm performance study is not relevant and not mentioned.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Without a study, there's no mention of ground truth.

  8. The sample size for the training set: This is not an AI/machine learning device, so the concept of a "training set" is not applicable, and none is mentioned.

  9. How the ground truth for the training set was established: See point 8.

In summary, the provided 510(k) summary and FDA clearance letter confirm the substantial equivalence of the Synthes External Fixation Component Line Extension - MR Safe to predicate devices, but they do not detail performance studies with acceptance criteria or ground truth establishment.

{0}------------------------------------------------

11043039

Image /page/0/Picture/1 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized design. A horizontal line is present below the word, adding a sense of grounding to the overall composition. The registered trademark symbol is located to the right of the word.

3.0510(k) SummaryPage 1 of 1
Sponsor:Synthes (USA)1690 Russell RoadPaoli, PA 19301(610) 647-9700
Device Name:Synthes (USA) External Fixation Component Line Extension - MR Safe
Classification:Class II, §888.3030 - Single/multiple component bone fixationappliances and accessories
Predicate Device:Synthes Reprocessed External Fixation DevicesHoffmann II External Fixation System (Stryker)EBI XFIX Vision Fixation System
Device Description:Synthes External Fixation Systems, MR Safe are a system ofcomponents that form a construct intended to treat stable andunstable fractures. All frame elements are made from non-magnetic materials and are intended for use in the MRenvironment. This line extension consists of Synthes Pin Clampswith Outrigger Posts and additional Curved Carbon Fiber Rods.These components are also intended for use in the MRenvironment.
Intended Use:Synthes External Fixation Devices - MR Safe are intended for usein the construction of an external fixation frame for treatment ofvarious fracture types that require external fixation
SubstantialEquivalence:Documentation is provided which demonstrates that the Synthes(USA) External Fixation Component Line Extension - MR Safe issubstantially equivalent to other legally marketed devices.The term "substantial equivalence" as used in this 510(k)notification is limited to the definition of substantial equivalencefound in the Federal Food, Drug and Cosmetic Act, as amendedand as applied under 21CFR 807, Subpart E under which a devicecan be marketed without premarket approval or reclassification. Adetermination of substantial equivalency under this notification isnot intended to have any bearing whatsoever on the resolution ofpatent infringement suits or any other patent matters. Nostatements related to, or in support of substantial equivalenceherein shall be construed as an admission against interest under theUS Patent Laws or their application by the courts.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

JAN 1 8 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Kathy Anderson Synthes (USA) 1690 Russell Road Paoli, Pennsylvania 19301

Re: K043039

Trade/Device Name: Synthes (USA) External Fixation Component Line Extension – MR Safe Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: November 2, 2004 Received: November 4, 2004

Dear Ms. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass batted in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may) ators, orovisions of the Act include requirements for annual registration, listing of general volurols proving practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vathar sthates act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Kathy Anderson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

O. Mark A. Mullerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. There is a line underneath the word "SYNTHES". The registered trademark symbol is to the right of the word.

Mark N. Mlbess

and Neurological Devices

Indications for Use

2.0

510(k) Number K043039

510(k) Number (if known):

Synthes (USA) External Fixation Component Line Extension - MR Safe Device Name:

Indications for Use:

Synthes External Fixation Devices - MR Safe are intended for use in the construction of an external fixation frame for treatment of various fracture types that require external fixation.

LARGE

Provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for:

  • Stabilization of soft tissues and fractures .
  • Poyltrauma/multiple orthopedic trauma .
  • Vertically stable pelvic fractures, or as a treatment adjunct for vertically unstable pelvic . fractures
  • Arthrodeses and osteotomies with soft tissue problems; failures of total joints .
  • Neutralization of fractures stabilized with limited internal fixation .
  • Non-unions/septic non-unions .
  • Intra-operative reductions/stabilization tool to assist with indirect reduction .
  • Unilateral rectilinear bone segment transport or leg lengthening .

MEDIUM

Indicated for construction of an external fixation frame for the treatment of pediatric and adult fractures.

X Prescription Use (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.