K Number
K042815

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-11-05

(24 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Quest Diagnostics Tumor Marker Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

Device Description

This product is prepared from human serum with added chemicals, constituents of human and animal origin. This product is provided in lyophilized form for added stability.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (New Device)Reported Device Performance (New Device)
Open Vial Stability14 days at 2°C to 8°C14 days at 2°C to 8°C
Post-Reconstitution & Freezing Stability30 days at -10°C to -20°C30 days at -10°C to -20°C
Shelf LifeThree years and three months when stored at 2°C to 8°CThree years and three months when stored at 2°C to 8°C
Intended UseTo monitor the precision of laboratory testing procedures for listed analytes.Intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
FormLyophilizedLyophilized
MatrixHuman SerumHuman Serum
PreservativesDoes not contain preservativesDoes not contain preservatives
Storage (Unopened)2°C to 8°C until expiration date2°C to 8°C until expiration date

Note: The document primarily focuses on establishing "substantial equivalence" to a predicate device, rather than explicit numerical performance acceptance criteria common in diagnostic accuracy studies. The core acceptance criteria for this type of device are related to its stability and intended use aligning with the predicate.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not explicitly state the sample size used for the test set in the stability studies. It only mentions that "Stability studies have been performed to determine the open vial stability and shelf life for the Quest Diagnostics Tumor Marker Control."

The document also does not specify the data provenance (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable to this type of device and study. The device is a quality control material, and its "ground truth" relates to its inherent stability and consistency, demonstrated through laboratory testing rather than expert-derived diagnoses.

4. Adjudication Method

This information is not applicable to this type of device and study. Adjudication methods are used in studies involving human interpretation or diagnosis to resolve discrepancies, which is not the nature of a quality control material evaluation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation and aims to assess the impact of AI assistance on human reader performance. This document concerns a quality control material.

6. If a Standalone Study Was Done

Yes, a standalone study (i.e., algorithm only without human-in-the-loop performance) was effectively done in the form of stability studies. These studies evaluated the intrinsic properties of the "Quest Diagnostics Tumor Marker Control" itself (its stability under various conditions) without human interpretation being the primary endpoint. The device doesn't have an "algorithm" in the traditional sense, but its performance as a quality control is assessed standalone.

7. The Type of Ground Truth Used

The "ground truth" for this device is established through:

  • Laboratory Testing/Reference Methods: The stability and concentration of the analytes within the control material would be measured using established laboratory methods over time and under various storage conditions. The expected values, as determined by these methods at the time of manufacture and over the claimed stability period, serve as the ground truth against which the device's performance is assessed.
  • Comparison to Predicate Device: The ultimate "ground truth" for regulatory purposes in this 510(k) submission is demonstrating substantial equivalence to the "Lyphochek Tumor Marker Control." This implies that the new device's stability and characteristics should fall within acceptable parameters relative to the predicate, as determined by industry standards and established laboratory practices for quality control materials.

8. The Sample Size for the Training Set

The provided document does not mention a training set. This is because the device is a quality control material, not an AI/ML algorithm that requires training data. The "study" here involves analytical performance (stability) rather than a learning process.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for this type of device, this question is not applicable.

{0}------------------------------------------------

K042815

Page 1 of 2

Page 1 of 2

Summary of Safety and Effectiveness Quest Diagnostics Tumor Marker Control

NOV - 5 2004

1.0 Submitter

Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1557 Fax:

Contact Person

Maria Zeballos Regulatory Affairs Specialist (949) 598-1367 Telephone:

Date of Summary Preparation

November 2, 2004

2.0 Device Identification

Product Trade Name:Quest Diagnostics Tumor Marker Control
Common Name:Multi-analyte Controls, (Assayed and unassayed)
Classifications:Class I
Product Code:JJY
Regulation Number:21 CFR 864.1660

3.0 Device to Which Substantial Equivalence is Claimed

Lyphochek Tumor Marker Control Bio-Rad Laboratories Irvine, California

510 (k) Number: K011579

4.0 Description of Device

This product is prepared from human serum with added chemicals, constituents of human and animal origin. This product is provided in Iyophilized form for added stability.

5.0 Intended Use

Quest Diagnostics Tumor Marker Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

{1}------------------------------------------------

Comparison of the new device with the Predicate Device 6.0

Quest Diagnostics Tumor Marker Control claims substantial equivalence to the Lyphochek Tumor Marker Control currently in commercial distribution (K011579).

Table 1. Similarities and Differences between new and predicate device.
CharacteristicsQuest DiagnosticsTumor Marker Control(New Device)Bio-Rad LaboratoriesLyphochek Tumor Marker Control(Predicate Device K011579)
Similarities
Intended UseQuest Diagnostics Tumor Marker Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.Lyphochek Tumor Marker Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
FormLyophilizedLyophilized
MatrixHuman SerumHuman Serum
PreservativesDoes not Contains preservativesDoes not Contains preservatives
Storage (Unopened)2°C to 8°CUntil expiration date2°C to 8°CUntil expiration date
Reconstituted Vial Claim14 days at 2°C to 8°CAll analytes will be stable for 14 days at 2 to 8°C, with the following exceptions: Ferritin and CA 27-29 will be stable for 6 days. ACTH, Free PSA, PSA, and Calcitonin should be assayed immediately following reconstitution
After Reconstituting and FreezingAll analytes 30 days at -10 to -20°CAll analytes 30 days at -10 to -20°C
Differences
AnalytesContains the following analytes:CA 15-3CA 125CA 19-9CA 27.29 [Footnote #2 listed in chart]Does not contain the following analytes:ACTHAFP (Alpha Fetoprotein)AldosteroneBeta-2-MicroglobulinCA 50CA 72-4CalcitoninCASACEA (Carcinoembryonic Antigen)CYFRA 21-1FerritinhCG (Human Chorionic Gonadotropin)hCG - Beta SubunitNSE (Neuron Specific Enolase)PAP (Prostatic Acid Phosphatase)ProlactinContains the following analytes:ACTHAFP (Alpha Fetoprotein)AldosteroneBeta-2-MicroglobulinCA 15-3CA 19-9CA 27-29CA 50CA 72-4CA 125CalcitoninCASACEA (Carcinoembryonic Antigen)CYFRA 21-1FerritinhCG (Human Chorionic Gonadotropin)hCG - Beta SubunitNSE (Neuron Specific Enolase)PAP (Prostatic Acid Phosphatase)ProlactinPSA (Prostate Specific Antigen)

7.0 Statement of Supporting Data

Stability studies have been performed to determine the open vial stability and shelf life for the Quest Diagnostics Tumor Marker Control. Product claims and a summary of the protocols used to establish claims are as follows:

  • Open vial Stability: 14 days at 2 to 8°C. .
  • After reconstituting and freezing: 30 days at -10 to -20°C.
  • . Shelf Life: Three years and three months when stored at 2 to 8 °C

All supporting data is retained on file at Bio-Rad Laboratories.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (USA). The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized representation of an eagle, with three curved lines forming its body and wings. The eagle is facing to the right.

Public Health Service

NOV - 5 2004

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Elizabeth Platt Regulatory Affairs Manager/Quality Assurance Bio-Rad Laboratories. QSD 9500 Jeronimo Road Irvine, CA 92618-2017

Re: K042815

Trade/Device Name: Quest Diagnostics Tumor Marker Control Levels 1, 2 and 3 Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: October 5, 2004 Received: October 12, 2004

Dear Ms. Platt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Robert L. Becker Jr.

Robert L. Becker, Jr., M.D., Ph.D., Ph.D Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

KO42815 510(k) Number (if known): Quest Diagnostics Tumor Marker Control Device Name: Quest Diagnostics Turnor Marker Control is intended for use as a quality control Indications For Use: serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page 1 of

Mono M Clan

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K042815

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.