(40 days)
For use as a quality control serum to monitor the precision of an individual laboratory's automated and manual-testing procedures.
Quest Diagnostics Serum Chemistry Control is prepared from human serum to which purified biochemical materials (tissue extracts of human and animal origin), chemicals. preservatives, and stabilizers have been added.
Here's an analysis of the provided text regarding the Quest Diagnostics Serum Chemistry Control, focusing on acceptance criteria and study details:
This device is a Quality Control (QC) material, not a diagnostic device that produces patient results or relies on AI. Therefore, many of the requested fields (e.g., number of experts, adjudication methods, MRMC studies, standalone performance, training set) are not applicable to this type of product.
1. Table of Acceptance Criteria and Reported Device Performance
| Criteria (Claimed Performance) | Reported Device Performance |
|---|---|
| Intended Use | To monitor the precision of an individual laboratory's automated and manual-testing procedures. |
| Open Vial Stability (Unopened Frozen) | 10 days when stored tightly capped at 2-8°C. |
| Shelf Life (Unopened Frozen) | Two years when stored at -10 to -20 °C. |
| Form | Liquid |
| Matrix | Human serum based |
| Other Ingredients | Stabilizers and preservatives |
| Analytes | ALT, Albumin, Alkaline Phosphatase, Amylase, AST, Direct Bilirubin, Total Bilirubin, Blood Urea Nitrogen, Calcium, Chloride, Cholesterol, Cholesterol, HDL, CO2, Creatine Kinase (CK), Creatinine, Gamma-Glutamyltransferase, Glucose, Iron, Lactate Dehydrogenase (LDH), Lipase, Magnesium, Phosphorous, Potassium, Sodium, T3 Uptake, T4 Total, Total Protein, Triglycerides, Iron-Binding Capacity, Unsaturated (UIBC), Uric Acid. |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated with a specific number of samples. The determination of stability involves testing material over time, involving multiple aliquots and measurements.
- Data Provenance: The studies were performed by Bio-Rad Laboratories. The country of origin of the data is not specified but is presumably the USA, where Bio-Rad Laboratories is located. The studies are prospective in nature, as they involve testing the product over its claimed stability periods (e.g., real-time shelf life studies).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This is a quality control material where "ground truth" typically refers to the expected values and acceptable ranges for analytes, which are established through extensive analytical testing and statistical analysis, not by expert interpretation in the same way a diagnostic imaging device would be. The "ground truth" for the analytes within the control is determined by the manufacturer's reference methods and value assignment processes.
4. Adjudication method for the test set
- Not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation or subjective assessments, generally for diagnostic accuracy. This device's performance is determined through objective laboratory measurements and statistical analysis against pre-defined stability criteria.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a quality control material, not a diagnostic device that would involve human readers or AI assistance in interpreting patient data.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This device is a physical reagent used to monitor the performance of analytical instruments. It does not employ an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for this type of device is the expected concentration/activity of each analyte within the control material, along with established stability profiles. These are determined through:
- Reference analytical methods: Using highly accurate and precise analytical techniques to assign target values to the analytes in the control.
- Statistical analysis: To define the acceptable range of variation (precision) for each analyte over its claimed shelf life and open vial stability.
- Real-time stability studies: Monitoring the material's performance over its stated shelf life and open vial periods to confirm that analyte values remain within established acceptance limits.
8. The sample size for the training set
- Not applicable. This device does not use machine learning or AI, and therefore does not have a "training set."
9. How the ground truth for the training set was established
- Not applicable. As there is no training set, this question is not relevant.
{0}------------------------------------------------
K03387
Page 1 of 2
Summary of Safety and Effectiveness Quest Diagnostics Serum Chemistry Control
1.0 Submitter
Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1555 Fax:
Contact Person
Maria Zeballos Regulatory Affairs Specialist Telephone: (949) 598-1367
Date of Summary Preparation
October 15, 2003
2.0 Device Identification
| Product Trade Name: | Quest Diagnostics Serum Chemistry Control |
|---|---|
| Common Name: | Multi-Analyte Controls, (Assayed and Unassayed) |
Classifications: Class I Product Code: JJY Requlation Number: CFR 862.1660
3.0 Device to Which Substantial Equivalence is Claimed
Bio-Rad Laboratories Liquid Assayed Multiqual Control Irvine, California Docket Number: K011867
4.0 Description of Device
Quest Diagnostics Serum Chemistry Control is prepared from human serum to which purified biochemical materials (tissue extracts of human and animal origin), chemicals. preservatives, and stabilizers have been added.
5.0 Statement of Intended Use
Quest Diagnostics Serum Chemistry Control is intended for use as a quality control serum to monitor the precision of an individual laboratory's automated and manualtesting procedures.
{1}------------------------------------------------
6.0 Comparison of the new device with the Predicate Device
Quest Diagnostics Serum Chemistry Control claims substantial equivalence to the Liquid Assayed Multiqual Control currently in commercial distribution (K011867).
| Quest DiagnosticsSerum Chemistry Control(New Device) | Bio-RadLiquid Assayed Multiqual Control(Predicate Device K011867) | |
|---|---|---|
| Characteristics | ||
| Similarities | ||
| Intended Use | Quest Diagnostics Serum Chemistry Control isintended for use as a quality control serum tomonitor the precision of an individual laboratory'sautomated and manual-testing procedures. | Liquid Assayed Multiqual Control is intended for use asan assayed quality control serum to monitor theprecision of laboratory testing procedures for analyteslisted in this package insert. |
| Form | Liquid | Liquid |
| Matrix | Human serum based | Human serum based |
| Other ingredients | Stabilizers and preservatives | Stabilizers and preservatives |
| Differences | ||
| Storage(Unopened Frozen) | -10 °C to -20°CUntil expiration date | -20°C or colderUntil expiration date |
| Open Vial Claim | 10 days at 2-8° C | 14 days 2-8° C |
| Storage(Unopened Thawed) | No claim | 30 days at 2-8° C |
| Analytes | Contains the following analytes that areequivalent to the predicate device:ALT; Albumin; Alkaline Phosphatase; Amylase;AST; Direct Bilirubin; Total Bilirubin; Blood UreaNitrogen; Calcium; Chloride; Cholesterol;Cholesterol, HDL; CO2; Creatine Kinase (CK);Creatinine; Gamma-Glutamyltransferase;Glucose; Iron; Lactate Dehydrogenase (LDH);Lipase; Magnesium; Phosphorous; Potassium;Sodium; T3 Uptake; T4 Total; Total Protein;Triglycerides; Iron-Binding Capacity,Unsaturated (UIBC); Uric Acid. | Contains the following additional analytes not claimedin the new product:Acetaminophen; Acid Phosphatase; Amikacin;Amylase, Pancreatic; Bilirubin, Neonatal; Calcium,ionized; Carbamazepine; Cholesterol, HDL;Cholesterol, LDL; CK-MB Isoenzyme; Cortisol; Digoxin;Ethyl Alcohol; Gentamicin; a-1-Antitripsin; HBDH;ApoA; ApoB; C3 Complement; C4 Complement;Ceruloplasmin; Cholinesterase; Copper; Ferritin;Globulin; Lithium; Osmolality; Phenobarbital;Phenytoin; Phospholipids; PAP; Salicylate; T3 Free; T3Uptake/T4 Uptake; T4 Free; Theophylline; TSH;Tobramycin; Valproic Acid; Haptoglobin; IgA ; IgG; IgM:TIBC; LAP Arylamidase; Prealbumin; ProteinElectrophoresis, Transferrin; Vitamin B12 Zinc. |
| Table 1. Similarities and Differences between new and predicate device. | |||||
|---|---|---|---|---|---|
7.0 STATEMENT OF SUPPORTING DATA
Stability studies have been performed to determine the open vial stability and shelf life for the Quest Diagnostics Serum Chemistry Control. Product claims are as follows:
- 7.1 Open vial: 10 days when stored tightly capped at 2-8°C.
- 7.2 Shelf Life: Two years when stored at -10 to -20 °C.
- 7.3 Real time studies will be ongoing to support the shelf life of this product.
All supporting data is retained on file at Bio-Rad Laboratories.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the top half of the circle. In the center of the logo is a stylized image of an eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC - 2 2003
Ms. Elizabeth Platt Regulatory Affairs Manager/Quality Assurance Bio-Rad Laboratories, QSD 9500 Jeronimo Road Irvine, CA 92618-2017
Re: K033387
Trade/Device Name: Quest Diagnostics Serum Chemistry Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: October 15, 2003 Received: October 24, 2003
Dear Ms. Platt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practicc requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510 (k) Number (if known): K033387
Device Name: Quest Diagnostics Serum Chemistry Control
Indications for Use:
ﺎ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ પ્રમુદ્
For use as a quality control serum to monitor the precision of an individual laboratory's automated and manual-testing procedures.
(PLEASE DO NOT WRITE BELOW THE LINE-CONINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription use X or Over-the Counter use
Carol C Benson for Jean Cooper, DVM
Division Sign-Off
Office of In Vitro Diagnostic Device
Evaluation and Safety
510(k) K033387
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.