Intended Use

The Flexi-Stat SpO2 Sensor is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in adult, pediatric, and neonatal patients.

Device Description

The Flexi-Stat(tm) SpO2 Sensor is an electro-optical sensor that functions without skin penetration, electrical contact, or heat transfer. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The optical components are housed in adhesive film. The sensor cable is terminated in a DB-9 style connector, with an adapter cable for Datexcompatible models.

AI/ML Overview

The provided text describes a 510(k) summary for the Flexi-Stat SpO2 Sensor. However, it does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a detailed study description. The document focuses on demonstrating substantial equivalence to a predicate device rather than providing a comprehensive clinical study report with specific acceptance criteria.

Based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document broadly states that the device was "found to be equivalent to predicate device accuracy claims." It does not provide specific numerical acceptance criteria (e.g., A_rms values, bias, precision) or the detailed performance metrics of the Flexi-Stat against those criteria. The predicate device's accuracy claims are not specified in this document.

Acceptance CriteriaReported Device Performance
Specific SpO2 accuracy (e.g., A_rms < X%)"Equivalent to predicate device accuracy claims"
Specific pulse rate accuracy (e.g., ± Y bpm)"Equivalent to predicate device accuracy claims"
Biocompatibility requirementsConforms to established industry standards
Electrical safety requirementsConforms to established industry standards
EMC testing requirementsConforms to established industry standards

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified. The document mentions "clinical hypoxia studies" but does not provide the number of subjects or data points used in these studies.
  • Data Provenance: The studies were conducted in an "independent research lab." The country of origin of the data is not specified, but the submitter is Elekon Industries, USA, Inc., and the FDA is a US regulatory body, suggesting the studies likely occurred in the US. The studies are prospective in nature, involving "clinical hypoxia studies" where the device was compared to arterial blood samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not applicable in this context. The ground truth for oxygen saturation (SpO2) measurements in this type of device comparison is typically established by a laboratory co-oximeter analyzing arterial blood samples, which is a gold standard measurement, not by expert interpretation of images or other subjective data.
  • Qualifications of Experts: N/A.

4. Adjudication method for the test set:

Not applicable. The ground truth is established by a laboratory co-oximeter, which provides an objective measurement, not a subjective interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a SpO2 sensor, which provides a direct physiological measurement, not an AI-assisted diagnostic tool that would involve human readers interpreting images or data.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the performance testing described for the Flexi-Stat SpO2 Sensor is a standalone performance assessment. The device itself (the sensor) provides the SpO2 and pulse rate measurements, which were then compared against a gold standard (arterial blood samples analyzed on a laboratory co-oximeter). There is no "human-in-the-loop" component for the measurement itself, beyond applying the sensor and operating the connected oximeter.

7. The type of ground truth used:

The ground truth used was arterial blood samples analyzed on a laboratory co-oximeter. This is considered a highly accurate and objective gold standard for measuring functional arterial oxygen saturation.

8. The sample size for the training set:

Not applicable. This device is a hardware sensor, and the performance testing described is for its accuracy in measuring SpO2 and pulse rate. The concept of a "training set" is typically relevant for machine learning or AI-based devices that learn from data. This device's operation is based on electro-optical principles, not an algorithm that requires a training set in the AI sense.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set in the context of this device.

{0}------------------------------------------------

DEC 2 9 2004

K042705

Appendix C Page 1 of 2

510(k) Summary

Submitter Information:

Elekon Industries, USA, Inc. 3848 Del Amo Blvd. Torrance CA 90503 USA

Contact:

Tom Dietiker, President Tel: 310-370-8022 Fax:310-861-5619

Date Prepared:

September 13, 2004

Product Name:

Common Name: SpO2 Sensor (accessory to pulse oximeter) Trade Name(s): Flexi-Stat SpO2 Sensors

Predicate Device:

The proposed devices are identical to the devices cleared under K041522, with the exception of added pediatric claims and a change to the wrap material from nonadhesive to adhesive.

Description:

The Flexi-Stat(tm) SpO2 Sensor is an electro-optical sensor that functions without skin penetration, electrical contact, or heat transfer. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The optical components are housed in adhesive film. The sensor cable is terminated in a DB-9 style connector, with an adapter cable for Datexcompatible models.

Intended Use:

The Flexi-Stat SpO2 Sensor is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring.

{1}------------------------------------------------

Comparison to Predicate Device:

The Flexi-Stat SpO2 Sensor uses the same theory and principle of operation as the predicate device. Design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.

Performance Data & Conclusions:

Performance testing was conducted during clinical hypoxia studies conducted in an independent research lab. The Flexi-Stat was compared to arterial blood samples analyzed on a laboratory co-oximeter and found to be equivalent to predicate device accuracy claims.

Biocompatibility, electrical safety, and EMC testing was also performed to demonstrate conformance with established industry standards.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 9 2004

Elekon Industries USA, Incorporated C/O Ms. Krista Oakes Amica Solutions 2300 McDermott Road, #200-207 Plano, Texas 75025

Re: K042705

Trade/Device Name: Flexi-Stat SpO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: November 29, 2004 Received: December 6, 2004

Dear Ms. Oakes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Oakes

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Cluts

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) # (if known):

Device Name: Flexi-Stat SpO2 Sensor

Indications for Use:

The Flexi-Stat SpO2 Sensor is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in adult, pediatric, and neonatal patients.

Prescription Use x (21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Oive

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control. Dental I 510(k) Number

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).