K Number
K042642

Validate with FDA (Live)

Date Cleared
2004-11-03

(37 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aware™ Transport Monitor System (includes Aware POD and Aware Transport Monitor) is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional medical facility, such as a hospital, clinic, surgical center or doctor's office. It can be used in multiple areas such as operating room (OR), post anesthesia care unit (PACU), emergency department (ED), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care.

The Aware POD is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. Physiological parameter data acquired by the Aware POD includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, temperature, and respiration. This device acquires, processes and stores information for all aforementioned parameters and transmits this information to a transport or bedside central processing unit for viewing and alarm surveillance purposes.

The Aware Transport Monitor is intended for use as part of a transport monitoring system for intra-healthcare facility transport. When used with the Aware POD or the TRAM acquisition module, this device is intended to provide uninterrupted monitoring of physiologic parameter data for adult, pediatric, and neonatal patients during transport from one area of the healthcare facility to another. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, temperature, and respiration. Both the Aware POD and TRAM acquisition module acquire, process and store information for all aforementioned parameters.

Device Description

The Aware™ Transport Monitor System is a complete, high-acuity patient monitoring system composed of two main parts: the Aware POD and Aware Transport Monitor.

The Aware™ Patient Observation Device (POD) is an acquisition module that measures and processes a patient's physiological parameters. Physiological parameter data acquired by the Aware POD includes 12 Lead ECG, respiration, up to four invasive blood pressure channels (as options), non-invasive blood pressure, Masimo SpO2, dual temperature and cardiac output (optional).

The Aware™ Transport Monitor is designed to be lightweight, rugged, and provide continuous monitoring capability when coupled with a compatible acquisition module. The Aware Transport Monitor is designed to facilitate quick, simple connection of either the Aware POD or TRAM 2001 module (K011000). The Aware Transport Monitor provides a means to alert the clinician of limit violations via audible and visual alarms.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Aware™ Transport Monitor System

This document outlines the acceptance criteria and a summary of the study used to demonstrate that the Aware™ Transport Monitor System meets these criteria, based on the provided 510(k) summary.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly present a table of specific numerical acceptance criteria. Instead, it broadly states that the device "complies with the voluntary standards as detailed in Section 9 of this submission" and that "The results of these measurements demonstrated that the Aware Transport Monitor System is as safe, as effective, and performs as well as the predicate devices."

Based on the information available, the acceptance criteria are implicitly tied to a combination of:

  • Compliance with Voluntary Standards: Adherence to established industry and regulatory standards relevant to physiological patient monitors. (Specific standards are not detailed in the provided text.)
  • Equivalency to Predicate Devices: Performance characteristics (safety and effectiveness) should be comparable to the legally marketed predicate devices (K011000 TRAM 2001 Module).

Implicit Acceptance Criteria & Reported Performance:

Acceptance Criteria CategoryReported Device Performance
Safety and Effectiveness"The results of these measurements demonstrated that the Aware Transport Monitor System is as safe, as effective, and performs as well as the predicate devices."
Compliance with Standards"The Aware Transport Monitor System complies with the voluntary standards as detailed in Section 9 of this submission." (Specific standards not provided in the excerpt.)
Functional Equivalence"The Aware Transport Monitor System employs the same functional scientific technology as its predicate devices."
Performance TestingNot explicitly quantified, but performed as part of the overall testing.
Safety TestingNot explicitly quantified, but performed as part of the overall testing (e.g., electrical safety).
Environmental TestingNot explicitly quantified, but performed as part of the overall testing.

2. Sample Size Used for the Test Set and Data Provenance

The document indicates "Clinical Use Validation" and various levels of testing (unit, integration, final acceptance, performance, safety, environmental). However, it does not specify a distinct "test set" and a corresponding sample size for clinical validation nor does it provide details on the data provenance (e.g., country of origin, retrospective or prospective nature of clinical data). The testing described seems to be more focused on engineering verification and validation rather than a separate clinical performance study with a defined patient cohort.

3. Number of Experts Used to Establish Ground Truth and Qualifications

The provided text does not mention the use of experts to establish ground truth for a test set, nor does it specify their number or qualifications. The clinical validation is mentioned, but the methodology for assessing clinical outcomes or establishing "truth" for the device's physiological parameter measurements is not detailed beyond general testing.

4. Adjudication Method

The document does not describe any adjudication method used for a test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study is mentioned. The device is a physiological monitor, not an imaging or diagnostic AI-assisted device where such studies are typically performed to assess human reader improvement. There is no information about AI assistance or human-in-the-loop performance improvement.

6. Standalone (Algorithm Only) Performance Study

Given that the device is a physiological patient monitor, its primary function is data acquisition and display. The concept of "standalone (algorithm only)" performance without a human in the loop is not directly applicable in the same way it would be for an AI diagnostic algorithm. The device's performance is inherently tied to its ability to accurately measure and process physiological parameters, which is implicitly evaluated through the "performance testing" and "clinical use validation." However, a specific standalone performance study report with metrics akin to AI algorithms is not provided.

7. Type of Ground Truth Used

The type of "ground truth" used for validating the accuracy of physiological parameter measurements is not explicitly stated. For physiological monitors, ground truth typically involves comparison against established reference devices or invasive measurements (e.g., arterial line for blood pressure, direct temperature probes, or gold-standard ECG recording systems). The document generally refers to "measurements" and "clinical use validation" without specifying the reference methods.

8. Sample Size for the Training Set

The document does not refer to a "training set" as the device is not described as utilizing machine learning or AI that would require a distinct training phase with a dataset. The development process mentions "Requirements Reviews," "Design Reviews," and various levels of testing, which are standard for medical device engineering but do not imply an AI model training paradigm.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a training set for an AI or machine learning model, the method for establishing ground truth for such a set is not applicable and hence not provided.

{0}------------------------------------------------

Section 2 Summary and Certification

:

510(k) Summary of Safety and Effectiveness
Date:September 24, 2004
Submitter:GE Medical Systems Information Technologies8200 West Tower AvenueMilwaukee, WI 53223 USA
Contact Person:Andrew KluessendorfSr. Regulatory Affairs SpecialistGE Medical Systems Information TechnologiesPhone: (414) 362-3063Fax: (414) 362-2585
Device:Trade Name:Aware™ Transport Monitor System (includes Aware POD and AwareTransport Monitor)
Common/Usual Name:Physiological Patient Monitor (Multi-parameter Module)
Classification Names:Physiological Patient Monitor
Predicate Devices:K011000 TRAM 2001 Module
Device Description:The Aware™ Transport Monitor System is a complete, high-acuitypatient monitoring system composed of two main parts: the Aware PODand Aware Transport Monitor.The Aware™ Patient Observation Device (POD) is an acquisitionmodule that measures and processes a patient's physiologicalparameters. Physiological parameter data acquired by the Aware PODincludes 12 Lead ECG, respiration, up to four invasive blood pressurechannels (as options), non-invasive blood pressure, Masimo SpO2,dual temperature and cardiac output (optional).The Aware™ Transport Monitor is designed to be lightweight, rugged,and provide continuous monitoring capability when coupled with acompatible acquisition module. The Aware Transport Monitor isdesigned to facilitate quick, simple connection of either the Aware PODor TRAM 2001 module (K011000). The Aware Transport Monitorprovides a means to alert the clinician of limit violations via audible andvisual alarms.
Intended Use:The Aware™ Transport Monitor System (includes Aware POD andAware Transport Monitor) is intended for use under the directsupervision of a licensed healthcare practitioner, or by personneltrained in proper use of the equipment in a professional medical facility,such as a hospital, clinic, surgical center or doctor's office. It can beused in multiple areas such as operating room (OR), post anesthesiacare unit (PACU), emergency department (ED), critical care, surgicalintensive care, respiratory intensive care, coronary care, medicalintensive care, pediatric intensive care, or neonatal intensive care.

K042642
page 1 of 2

{1}------------------------------------------------

The Aware POD is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. Physiological parameter data acquired by the Aware POD includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, temperature, and respiration. This device acquires, processes and stores information for all aforementioned parameters and transmits this information to a transport or bedside central processing unit for viewing and alarm surveillance purposes.

The Aware Transport Monitor is intended for use as part of a transport monitoring system for intra-healthcare facility transport. When used with the Aware POD or the TRAM acquisition module, this device is intended to provide uninterrupted monitoring of physiologic parameter data for adult, pediatric, and neonatal patients during transport from one area of the healthcare facility to another. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, temperature, and respiration. Both the Aware POD and TRAM acquisition module acquire, process and store information for all aforementioned parameters.

Technology:

The Aware Transport Monitor System employs the same functional scientific technology as its predicate devices.

Test Summary:

The Aware Transport Monitor System complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures are applied to the development of the Aware Transport Monitor System:

  • Risk Analvsis
  • . Requirements Reviews
  • Design Reviews
  • . Testing on unit level (Module verification)
  • . Clinical Use Validation
  • . Integration testing (System verification)
  • . Final acceptance testing (Validation)
  • . Performance testing
  • . Safety testing
  • . Environmental testing

Conclusion:

The results of these measurements demonstrated that the Aware Transport Monitor System is as safe, as effective, and performs as well as the predicate devices.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 3 2004

GE Medical Systems Information Technologies c/o Mr. Andrew Kluessendorf Sr. Regulatory Affairs Specialist 8200 West Tower Avenue Milwaukee, WI 53223

Re: K042642

Trade Name: Aware "" Transport Monitor System (includes Aware POD and Aware Transport Monitor) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: II (two) Product Code: MHX Dated: September 24, 2004 Received: September 27, 2004

Dear Mr. Kluessendorf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for acted in to May 28, 1976, the enactment date of the Medical Device Amendments, or to conniner or prior to may 20, 1977) in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, iterere, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Andrew Kluessendorf

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

B/Simmons for

Dr. D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Aware™ Transport Monitor System (includes Aware POD and Aware Transport Monitor)

Indications for Use:

The Aware™ Transport Monitor System (includes Aware POD and Aware Transport Monitor) is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional medical facility, such as a hospital, clinic, surgical center or doctor's office. It can be used in multiple areas such as operating room (OR), post anesthesia care unit (PACU), emergency department (ED), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care.

The Aware POD is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. Physiological parameter data acquired by the Aware POD includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, temperature, and respiration. This device acquires, processes and stores information for all aforementioned parameters and transmits this information to a transport or bedside central processing unit for viewing and alarm surveillance purposes.

The Aware Transport Monitor is intended for use as part of a transport monitoring system for intra-healthcare facility transport. When used with the Aware POD or the TRAM acquisition module, this device is intended to provide uninterrupted monitoring of physiologic parameter data for adult, pediatric, and neonatal patients during transport from one area of the healthcare facility to another. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, temperature, and respiration. Both the Aware POD and TRAM acquisition module acquire, process and store information for all aforementioned parameters.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER

PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B. Summa

Page / of /

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.