K Number
K042564

Validate with FDA (Live)

Date Cleared
2005-09-30

(374 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device use is for the treatment of hemorrhoids by applying it directly to the swollen hemorrhoidal tissues. By applying the device to the tissue, the inflammation is reduced. The base section is used for the treatment of external hemorrhoids and the neck/probe section is used for the treatment of internal hemorrhoids. The lubricant is used to ease the introduction of the probe into the rectum.

Device Description

Hemor-Rite Cryotherapy is an anatomically designed device for applying cold therapy (Cryotherapy) directly to the swollen hemorrhoidal veins of external hemorrhoids and the internal hemorrhoids within the rectal cavity. The direct application of cold provides prompt relief of itching, pain, and swelling. In addition, Hemor~Rite Cryotherapy is beneficial for the treatment of anal and perianal fissures due to the vasoconstriction and analgesia properties of the device. Hemor-Rite Cryotherapy has been designed taking the following factors into consideration: human anatomy and the medical concepts for treating the ailment.

Hemor-Rite Cryotherapy is an anatomically designed medical device ergonamicly shaped for application of cold therapy (Cryotherapy) directly to swollen hemorroidal veins of external hemorrhoids and internal hemorthoids within the rectal canal. The direct application of cold provides prompt relief of itching, pain, and swelling. The device consists of a sealed applicator containing a coolant formula. The HemorRite device is placed into the user's freezer until frozen. After the Hemor-Rite Cryotherapy device is frozen, the user inserts the device to te affected hemorrhoidal area. The probe is maintained in the area for 6-8 minutes, until the probe reaches body temperature, or up to a maximum of 10 minutes. The HemorRite Cryotherapy device is a reusable device.

Hemor~Rite is comprised of two pieces of plastic; the crown and the base.

AI/ML Overview

The provided text describes the Hemor~Rite Cryotherapy device and a study to demonstrate its substantial equivalence to a predicate device (ANUICE). However, the document does not contain a comprehensive algorithm performance study with specific acceptance criteria, test set details, ground truth establishment for AI/ML devices, as would be expected for a submission for an AI-powered medical device. Instead, the study focuses on hardware and biological compatibility assessments.

Here's a breakdown of the information that can be extracted, acknowledging the limitations regarding AI/ML-specific criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Material/Device IntegrityAccelerated Aging and IntegrityResults conclude Hemor~Rite is substantially equivalent to ANUICE.
Mechanical StrengthInvestigation of Compression StrengthResults conclude Hemor~Rite is substantially equivalent to ANUICE.
BiocompatibilityCytotoxicity (ISO 10993)Results conclude Hemor~Rite is substantially equivalent to ANUICE.
BiocompatibilityISO Intracutaneous (ISO 10993)Results conclude Hemor~Rite is substantially equivalent to ANUICE.
BiocompatibilityMaximization Sensitization (ISO 10993)Results conclude Hemor~Rite is substantially equivalent to ANUICE.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the context of an AI/ML algorithm. The studies listed (Accelerated Aging, Compression Strength, Cytotoxicity, Intracutaneous, Maximization Sensitization) are typical for medical device material and physical property testing. The sample sizes for these tests are not provided, nor is the data provenance (e.g., country of origin, retrospective/prospective). These are laboratory-based tests on the device materials themselves.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not applicable to the type of studies described. The "ground truth" for these tests would be established by standardized testing protocols and analytical methods, not by expert consensus on clinical cases.

4. Adjudication Method for the Test Set

This is not applicable as the studies are not clinical trials or expert review processes.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size of Human Improvement with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study was not conducted as this is not an AI/ML powered device. The device is a physical cryotherapy device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was done

No, a standalone algorithm performance study was not done for the same reason as above.

7. The Type of Ground Truth Used

The "ground truth" for the tests performed relates to the adherence of the device's materials and design to established safety and performance standards (e.g., ISO 10993 for biocompatibility), and its physical characteristics compared to the predicate device. It is based on laboratory testing and standardized analytical methods, not clinical outcomes or expert consensus on patient data.

8. The Sample Size for the Training Set

This is not applicable as there is no AI/ML algorithm being trained.

9. How the Ground Truth for the Training Set was Established

This is not applicable as there is no AI/ML algorithm being trained.

Summary of the Study that Proves the Device Meets Acceptance Criteria:

The study proving the device meets the acceptance criteria is a benchtop and biocompatibility testing program. The listed tests include:

  • Accelerated Aging and Integrity: Likely to assess the device's durability and material stability over time under simulated aging conditions.
  • Investigation of Compression Strength: To evaluate the device's mechanical robustness and ability to withstand forces during use.
  • Cytotoxicity (ISO 10993): To determine if the device's materials release substances that are harmful to cells.
  • ISO Intracutaneous (ISO 10993): To assess local skin irritation effects from the device materials.
  • Maximization Sensitization (ISO 10993): To determine the potential for the device materials to cause allergic sensitization.

The conclusion of these tests was that "Hemor~Rite is substantially equivalent to ANUICE and is safe, as effective, and performs as well as or better than the predicate device." This implies that the device successfully passed these tests and demonstrated comparable or superior characteristics to the predicate device, thereby fulfilling the safety and performance requirements for marketing. The basis for substantial equivalence for this device is primarily its technological characteristics and the results of these non-clinical performance assessments.

{0}------------------------------------------------

EXHIBIT C

Image /page/0/Picture/1 description: The image shows the word "HEMORRITE" in large, bold letters. Below the word is the word "Cryotherapy" in a smaller font. Above the word "HEMORRITE" is the text "SEP 5-6 2003".

K042564
pg 1 of 5

510 (k) Summary

OWNER:

FAMA Holdings International Corporation 4613 N University Drive Suite 252 Coral Springs, Florida 33067 Telephone: (866) 436-6748 Fax: (305) 402-2253

CONTACT PERSON:

Christine M. Humphrey, Esq. (786)245-0440

DATE PREPARED:

September 27, 2005

1. Device Name

Proprietary Name: Hemor~Rite Cryotherapy Common/Usual Name: Hemorrhoid Device Device Classification: Device, Thermal, Hemorrhoids (LKX)

2. Predicate Device:

Applicant: Cryotherapy Pain Relief Products, Inc. Proprietary Name: ANUICE 510(k): K981428

3. Device Description and Indications For Use:

DESCRIPTION:

Hemor-Rite Cryotherapy is an anatomically designed device for applying cold therapy (Cryotherapy) directly to the swollen hemorrhoidal veins of external hemorrhoids and the internal hemorrhoids within the rectal cavity. The direct application of cold provides prompt relief of itching, pain, and swelling. In addition, Hemor~Rite Cryotherapy is beneficial for the treatment of anal and perianal fissures due to the vasoconstriction and analgesia properties of the device. Hemor-Rite Cryotherapy has been designed taking the following factors into consideration: human anatomy and the medical concepts for treating the ailment.

{1}------------------------------------------------

TREATMENT PERIOD:

Use the Hemor-Rite Cryotherapy device for a minimum of four the first week of treatment. After the first week, use the Hemor-Rite Cryotherapy device at least twice a day until the symptoms disappear. Product reusable for a maximum of six months after first application. Note: Please allow a minimum of two hours between each treatment.

INSTRUCTIONS FOR USE TO TREAT INTERNAL HEMORRHOIDS:

  • Insert the Hemor~Rite Cryotherapy device in its case and place in the freezer for a minimum of two (2) 1. hours.
    1. After the minimum time has elapsed, remove the HemorRite Cryotherapy device from its case. Holding it by the base, apply several drops of lubricant to the tip and allow the drops to drip down the neck of the HemorRite Cryotherapy Device.
    1. Lie on one side, with one leg extended and the other leg bent towards the abdomen (See Image 1). Take the Hemor-Rite Cryotherapy device by the base, holding it with the wider end pointing towards the front (towards the genitals) and the narrower end toward the back (See Image 3). Insert the lubricated device slowly and smoothly into the anal canal until it is completely inserted.
  • Use the Hemor~Rite Cryotherapy device for 6 to 8 minutes, or until it reaches body temperature, for a 4. maximum of 10 minutes.
  • న్. Remove the Hemor~Rite Cryotherapy device slowly and smoothly. Wash it with soap and warm water or any external antiseptic. Remove excess moisture with a paper towel and place the dry device in its case.
  • Return the Hemor-Rite Cryotherapy device to the freezer and follow the above procedure for each use. 6.

WARNING: This device should be discarded if the user detects any cracks or breaks in the device. Should Hemor-Rite Cryotherapy device crack or break and the coolant comes in contact with the skin rinse with water. If the coolant comes into contact with eyes flush with plenty of water.

WARNINGS:

  • . The Hemor~Rite Cryotherapy device is for personal anal use only and should be kept in its sanitary case when not in use.
  • . Do not introduce the Hemor~Rite Cryotherapy device in to the rectal canal without lubrication.
  • . The cooling material contained in the Hemor~Rite Cryotherapy device may be toxic.
  • Only the neck of the Hemor~Rite Cryotherapy device should be inserted into the rectal canal. .

{2}------------------------------------------------

K042564
pg 3 of 5

  • For advanced hemorrhoidal conditions, grades 3 and 4, the Hemor~Rite Cryotherapy device should be . used for external hemorrhoidal treatment only. Once the swelling and the condition of the hemorrhoids have improved, you may then proceed to the internal hemorrhoidal treatment. Third degree (3) hemorrhoids prolapse with defecation but recede only by manual reduction. Fourth degree (4) hemorrhoids are permanently prolapsed and cannot be reduced into the anal canal.
  • Do not use if you have an elevated reaction to extreme cold or Cryotherapy. .
  • Keep out of reach of children. ●
  • Hemor~Rite Cryotherapy is a drug free treatment. No prescription needed. ●
  • Hemor~Rite Cryotherapy device should not be taken apart. .
  • Hemor~Rite Cryotherapy device should not be used in pregnant women or children without prior . approval of a physician.
  • Hemor~Rite Cryotherapy device is flammable once ignited, may burn rapidly. ●
  • Stop Hemor~Rite Cryotherapy treatment and consult a physician if symptoms continue past seven (7) . days or bleeding continues.

INSTRUCTIONS FOR USE TO TREAT EXTERNAL HEMORRHOIDS:

  • Insert the Hemor~Rite Cryotherapy device in its case and place in the freezer for a minimum of two (2) l . hours.
  • After the minimum time has elapsed, remove the Hemor~Rite Cryotherapy device from its case. 2.
  • Lie on one side, with one leg extended and the other leg bent towards the abdomen (See Image 2). Take 3. the Hemor~Rite Cryotherapy device by the neck or base (See Image 3), holding it with the wider end pointing towards the front (towards the genitals) and the narrower end toward the back. (See Image 3). Place and hold the device in this position over the anus (See Image 2).
  • Use the Hemor~Rite Cryotherapy device for 6 to 8 minutes, or until it reaches body temperature, for a 4. maximum of 10 minutes.
  • Remove the Hemor~Rite Cryotherapy device slowly and smoothly. Wash it with soap and warm water న్. or any external antiseptic. Remove excess moisture with a paper towel and place the dry device in its case.
  • Return the Hemor~Rite Cryotherapy device to the freezer and follow the above procedure for each use. 6.

TREATMENT PERIOD:

Use the Hemor-Rite Cryotherapy device for a minimum of four the first week of treatment. After the first week, use the Hemor-Rite Cryotherapy device at least twice a day until the symptoms disappear. Product reusable for a maximum of six months after first application. Note: Please allow a minimum of two hours between each treatment.

  • The Hemor~Rite Cryotherapy device is for personal anal use only and should be kept in its sanitary case . when not in use.

{3}------------------------------------------------

K042564
pg 4 of 5

  • Do not introduce the Hemor~Rite Cryotherapy device in to the rectal canal without lubrication. .
  • The cooling material contained in the Hemor-Rite Cryotherapy device may be toxic. .
  • Only the neck of the Hemor~Rite Cryotherapy device should be inserted into the rectal canal. .
  • For advanced hemorrhoidal conditions, grades 3 and 4, the Hemor~Rite Cryotherapy device should be . used for external hemorrhoidal treatment only. Once the swelling and the condition of the hemorrhoids have improved, you may then proceed to the internal hemorrhoidal treatment. Third degree (3) hemorrhoids prolapse with defecation but recede only by manual reduction. Fourth degree (4) hemorrhoids are permanently prolapsed and cannot be reduced into the anal canal.
  • Do not use if you have an elevated reaction to extreme cold or Cryotherapy. ●
  • . Keep out of reach of children.
  • Hemor~Rite Cryotherapy is a drug free treatment. No prescription needed. .
  • Hemor~Rite Cryotherapy device should not be taken apart. ●
  • Hemor~Rite Cryotherapy device should not be used in pregnant women or children without prior . approval of a physician.
  • Hemor~Rite Cryotherapy device is flammable once ignited, may burn rapidly. .
  • Stop Hemor-Rite Cryotherapy treatment and consult a physician if symptoms continue past seven (7) ● days or bleeding continues.

Hemor-Rite Cryotherapy is an anatomically designed medical device ergonamicly shaped for application of cold therapy (Cryotherapy) directly to swollen hemorroidal veins of external hemorrhoids and internal hemorthoids within the rectal canal. The direct application of cold provides prompt relief of itching, pain, and swelling. The device consists of a sealed applicator containing a coolant formula. The HemorRite device is placed into the user's freezer until frozen. After the Hemor-Rite Cryotherapy device is frozen, the user inserts the device to te affected hemorrhoidal area. The probe is maintained in the area for 6-8 minutes, until the probe reaches body temperature, or up to a maximum of 10 minutes. The HemorRite Cryotherapy device is a reusable device.

4. Predicate Device (ANUICE) Technological Characteristics

Anuice is comprised of three pieces of plastic; two of them together and one of them as well by expansion of a pipe inside another.

DESIGN: The pipe of insertion is totally straight, without curves on the end. The base is taller in height than in width and once it is utilized, it should be stored in the appropriate position. The union of both components (tube and body) is done in the zone where the disposition is inserted.

SIZE: The inserted tube is 2 %'' in height and 2 7/8" in length. The diameter width is (maximum) 15/16".

5. HEMOR~RITE Technological Characteristics

Hemor~Rite is comprised of two pieces of plastic; the crown and the base.

{4}------------------------------------------------

042564
pg 5 of 5

DESIGN: The crown of insertion is slanted with curves on the end. The base is taller in height than in width. The union of both components (crown and body) occurs at the base.

SIZE: The crown is 1 51/64" in height and 1 9/32" in length. The diameter width is (maximum) 1 9/32". The base is 9/32" height and 2 7/16" in length. The diameter width is (maximum) 1 9/32".

Clinical Performance 6.

Substantial equivalence was based on an assessment of clinical performance data, including:

  • Accelerated Aging and Integrity a.
  • Investigation of Compression Strength b.
  • Cytotoxicity c.
  • d. ISO Intracutaneous
  • e. Maximization Sensitization

The results of the testing conclude that Hemor~Rite is substantially equivalent to ANUICE and is safe, as effective, and performs as well as or better than the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the upper portion of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 3 0 2005

FAMA Holdings International Corp. c/o Christine M. Humphrev. Esq. C. Humphrey & Associates, P.A. 801 Brickell Avenue Suite Nine Hundred MIAMI FL 33131

Re: K042564

Trade/Device Name: HEMOR~RITE CRYOTHERAPY Regulation Number: None Regulatory Class: Unclassified Product Code: LKX Dated: September 26, 2005 Received: September 30, 2005

Dear Ms. Humphrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{6}------------------------------------------------

Page 2 - Ms. Christine Humphrey

(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. In addition, we have determined that your device kit contains a glycerine lubricant which is subject to regulation as a drug.

Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:

Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

David W. Torgeson

for Na
Dir

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "HEMOR~RITE" in a bold, outlined font. Below the word, in a smaller font, is the word "Cryotherapy". There are two horizontal lines above and below the word "Cryotherapy".

INDICATIONS FOR USE

510(k) Number (if known): K042564

Device Name: HEMOR~RITE CRYOTHERAPY

Indications for Use:

The device use is for the treatment of hemorrhoids by applying it directly to the swollen hemorrhoidal tissues. By applying the device to the tissue, the inflammation is reduced. The base section is used for the treatment of external hemorrhoids and the neck/probe section is used for the treatment of internal hemorrhoids. The lubricant is used to ease the introduction of the probe into the rectum.

Prescription Use - AND/OR (Part 21 CFR 801 Subpart D)

ND/OR

Over-The-Counter Use
(21 CFR 801 Subpart C) ✓

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David B. Henson

Division Sign-Off Division of Reproductive, Abdominal. and Radiological Devices 510(k) Number

N/A