K Number
K042522

Validate with FDA (Live)

Date Cleared
2005-07-06

(293 days)

Product Code
Regulation Number
862.1113
Age Range
2 - 120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Jaundice Meter (JM-103) is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue as a shown to correlate with serum bilirubin in newborn infants. The device is intended for use in hospitals, clinics or doctor's offices under a physician's direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. The device is intended to be used in conjunction with other clinical signs and laboratory measurements. Newborn infants whose JM-103 Jaundice Meter test results are indicative of hyperbilliubinemia should be evaluated by their physician(s) for appropriate patient management. JM-103 test results should be confirmed by other methods, such as serum bilirubin, prior to treatment determinations. The JM 103 is a prescription Medical Device. The JM 103 is not intended for home use. The JM 103 may only be used at the sternum measurement site for Physician's office applications.

Device Description

The JM-103 Jaundice Meter is designed to provide a transcutaneous measurement of bilirubin displayed in Mg/dl or umol/L. This measurement is intended as a screening tool to determine when a serum bilirubin measurement should be taken, or, sequential bilirubin measurements over time to provide indication of change. This device is not intended for determinations of whether treatment is indicated. The determination of treatment must be based on a serum bilirubin measurement.

AI/ML Overview

The provided text does not contain detailed information about acceptance criteria or a study proving the device meets those criteria. The document is a 510(k) summary for the Draeger Air Shields Infant Care Inc. Minolta Draeger Air Shields JM 103 Jaundice Meter. It primarily focuses on demonstrating substantial equivalence to a predicate device (K021622 JM 103 Jaundice Meter) and provides an "Indications for Use Statement."

Here's an analysis based on the absence of the requested information:

1. A table of acceptance criteria and the reported device performance

  • Information not provided. The document does not specify any quantitative acceptance criteria or report specific performance metrics from a study conducted for this 510(k) submission. It states "No modification to hardware, software, or procedures for use were implemented for this submission on the JM 103 Device previously cleared by the FDA as a medical device," implying that performance was established during the predicate device's clearance.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Information not provided. No details are given about a test set sample size or its provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Information not provided. The document does not mention the use of experts or ground truth establishment for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Information not provided. There is no mention of adjudication methods.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. The device is a Jaundice Meter, a non-AI diagnostic tool for transcutaneous bilirubin measurement, not an AI-assisted imaging device or system. No MRMC study would be relevant in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. The JM 103 is a standalone device in that it provides a measurement directly. However, it's explicitly stated that it's "not intended as a standalone for diagnosis" and its results "should be confirmed by other methods, such as serum bilirubin, prior to treatment determinations." This indicates it's a screening tool that operates independently but requires human medical interpretation and follow-up. Since it's not an AI algorithm, "algorithm only" performance is not a relevant concept here.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Information not provided for this submission. The document states the device "has been shown to correlate with serum bilirubin in newborn infants." This implies that the ground truth for its initial development and validation (likely for the predicate device) would have been serum bilirubin measurements, which are established clinical standards for diagnosing hyperbilirubinemia.

8. The sample size for the training set

  • Information not provided. The device is not an AI/machine learning device, so the concept of a "training set" in that context does not apply.

9. How the ground truth for the training set was established

  • Information not provided. As above, the concept of a training set is not applicable.

Summary based on the provided text:

The submission for the JM 103 Jaundice Meter is a 510(k) for a device that had "No modification to hardware, software, or procedures for use" from a previously cleared device. Therefore, the core performance characteristics and "acceptance criteria" are presumed to have been met by the predicate device (K021622 JM 103 Jaundice Meter). This 510(k) is about demonstrating substantial equivalence rather than presenting new performance data from clinical studies. The device's function is to provide a transcutaneous bilirubin measurement that correlates with serum bilirubin, acting as a screening tool, not a definitive diagnostic or treatment determination device.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness Section I

K 04252 2

Applicant:

Draeger Air Shields Infant Care Inc. 330 Jacksonville Road Hatboro, Pa. 19040 Registration No: 2510954

JUL 6 - 2005

Contact Person:

Thomas McIntosh Ph 215-682-8634 Fax 215-682-8689

Device trade/proprietary name:

Minolta Draeger Air Shields JM 103

Device common / usual/ classification name:

Jaundice Meter

Classification:

Clinical Chemistry Test Systems 21 CFR 862.113 Bilirubin in the Neonatal Test System, MQM, Class I

Performance Standards:

None applicable

Predicate Device:

K021622 JM 103 Jaundice Meter

Device Description

The JM-103 Jaundice Meter is designed to provide a transcutaneous measurement of bilirubin displayed in Mg/dl or umol/L. This measurement is intended as a screening tool to determine when a serum bilirubin measurement should be taken, or, sequential bilirubin measurements over time to provide indication of change. This device is not intended for determinations of whether treatment is indicated. The determination of treatment must be based on a serum bilirubin measurement.

Description of Modifications:

No modification to hardware, software, or procedures for use were implemented for this submission on the JM 103 Device previously cleared by the FDA as a medical device.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines representing its body and wings. The eagle faces right and is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL 6 - 2005

Mr. Thomas McIntosh Regulatory Affairs Draeger Air Shields Infant Care, Inc. 330 Jacksonville Road Hatboro, PA 19040

Re: K042522

Trade/Device Name: JM 103 Jaundice Meter Regulation Number: 21 CFR § 862.1113 Regulation Name: Bilirubin (total and unbound) in the neonate test system Regulatory Class: I Product Code: MQM Dated: March 24, 2005 Received: April 11, 2005

Dear Mr. McIntosh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Carol C. Benem

Carol C. Benson, M.A. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known): K042522

Device Name: JM 103 Jaundice Meter

A.1 Indication for Use Statement

The Jaundice Meter (JM-103) is a non-invasive transcutaneous bilirubinometer. It The Jaunulce Meter (SM-105) is a non-firsten o newborn infants. The unit provides a measures yellowness of ouboutano sus as a shown to correlate with serum bilirubin in newborn infants.

The device is intended for use in hospitals, clinics or doctor's offices under a The device is Internation for assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of inflants. The deviloc is not internation with other clinical signs and laboratory measurements.

Newborn infants whose JM-103 Jaundice Meter test results are indicative of Newbom inflants whose on: 100 Saated by their physician(s) for appropriate patient hyperbilliubinemia should be evaluation by their projects should be confirmed by other mathods, such as serum bilirubin, prior to treatment determinations.

The JM 103 is a prescription Medical Device The JM 103 is not intended for home use. The JM 103 is not intended for nome assi
The JM 103 may only be used at the sternum measurement site for Physician's office applications.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 5 of 40

Keal Sumpf
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)K_____________________________________________________________________________________________________________________________________________________________________

§ 862.1113 Bilirubin (total and unbound) in the neonate test system.

(a)
Identification. A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).(b)
Classification. Class I.