MODULAR ARTHRODESIS NAIL

K042409 · Biomet, Inc. · HSB · Dec 2, 2004 · Orthopedic

Device Facts

Record IDK042409
Device NameMODULAR ARTHRODESIS NAIL
ApplicantBiomet, Inc.
Product CodeHSB · Orthopedic
Decision DateDec 2, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Intended Use

The primary indication for these devices is the replacement of segmental defects of long bone, including midshaft replacement, and arthrodesis of the knee. Indications include: 1. Treatment of patients who are not candidates for total knee arthroplasty. 2. Irretrievably failed total knee arthroplasty. 3. Limb salvage in oncology surgery. 4. Trauma. 5. Any other condition where there is little soft tissue or bony tissue available for support, and arthrodesis is the treatment of choice. These devices are intended for cemented use only.

Device Story

Modular Arthrodesis Nail provides rigid fixation for knee arthrodesis and long bone segmental defect replacement. Device consists of femoral and tibial adapter segments connected by a central locking collar. Adapter segments available in two resection lengths (1 cm or 3 cm) and two body styles (standard or elliptical). Used in surgical settings by orthopedic surgeons. Device requires cement for fixation. Provides structural support for patients with failed arthroplasty, trauma, or oncological defects.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench-top mechanical testing.

Technological Characteristics

Multi-component intramedullary fixation rod. Material: Ti-6Al-4V titanium alloy (ASTM F-136). Design: Modular femoral and tibial adapter segments with central locking collar. Resection lengths: 1 cm or 3 cm. Body styles: standard or elliptical. Fixation: Cemented use only.

Indications for Use

Indicated for patients requiring knee arthrodesis or long bone segmental defect replacement (including midshaft) due to failed total knee arthroplasty, oncology-related limb salvage, trauma, or conditions with insufficient soft/bony tissue support. Contraindicated for patients who are candidates for total knee arthroplasty. Intended for cemented use only.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 042409 :12 DEC - 2 2004 Image /page/0/Picture/2 description: The image shows the word "BIOMET" in a stylized, blocky font. The letters are all capitalized and outlined in black, with the interior of the letters being white. The "O" in the middle is a perfect circle, and the "E" and "T" have a unique, geometric design. #### 510(k) Summary | Applicant/Sponsor: | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587 | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Gary Baker<br>Biomet Manufacturing Corp.<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587<br>Phone: (574) 267-6639 ext. 1568<br>FAX: (574) 372-1683 | | Proprietary Name: | Modular Arthrodesis Nail | | Common Name: | Titanium Intramedullary Rod | | Classification Name: | Rod, Fixation, Intramedullary and Accessories (21 CFR §888.302) | ## Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Wichita® Fusion Nail - Howmedica Inc. (K993603) Titanium Intramedullary Nails - Biomet Inc. (K982953) T2 Arthrodesis Nail System - Howmedica Inc. (K020384) Orthopedic Salvage System - Biomet Inc. (K002757) ### Device Description: The Modular Arthrodesis Nail is a multi-component device designed to provide arthrodesis fixation of the knee. The device has a central locking collar that connects fernoral and tibial adapter segments. The adapter segments are available in two resection lengths (1 cm or 3 cm), each with two body styles (standard or elliptical). All components of the Modular Arthrodesis Nail are manufactured from Ti-6Al-4V titanium alloy conforming to ASTM F-136. ### Intended Use: The primary indication for these devices is the replacement of segmental defects of long bone, including midshaft replacement, and arthrodesis of the knce. Indications include: - 1. Treatment of patients who are not candidates for total knee arthroplasty. - Irretrievably failed total knee arthroplasty 2. - 3. Limb salvage in oncology surgery, MAILING ADDRESS PO. Box 587 Warsaw, IN 46581-0587 SHIPPING ADDRESS 56 E. Bell Drive W:nisaw, IN 46582 대 OFFICE 574.267.6639 FAX 574.267.81.37 82 E MAIL biomet@bionet.com {1}------------------------------------------------ - 4. Trauma - 5. Any other condition where there is little soft tissue or bony tissue available for support, and arthrodesis is the treatment of choice. These devices are intended for cemented use only. Summary of Technologies: The Modular Arthrodesis Nail has the same intended use (arthrodesis), the same functional characteristics (rigid fixation) as the predicate devices, and is made of the same titanium alloy material as the predicate Titanium Intramedullary Nails and Orthopedic Salvage System (OSS) components used with the Modular Arthrodesis Nail. Non-Clinical Testing: Mechanical testing indicated that the Modular Arthrodesis Nails are substantially equivalent to the predicate devices for the uses indicated. Clinical Testing: Clinical testing was not required for these components to support substantial equivalence. All trademarks are property of Biomet, Inc. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 2 2004 Mr. Gary Baker Regulatory Specialist Biomet Manufacturing Corp. 56 East Bell DR. P.O. Box 587 Warsaw, IN 46581 Re: K042409 Trade/Device Name: Modular Arthrodesis Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: September 2, 2004 Received: September 3, 2004 Dear Mr. Baker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 - Mr. Gary Baker This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Page 1 of 1 # Statement of Indications For Use 510(k) Number (IF KNOWN): Koy2489 Device Name: Modular Arthrodesis Nail Indications for Use: The primary indication for these devices is the replacement of segmental defects of long bone, including midshaft replacement, and arthrodesis of the knee. Indications include: - Treatment of patients who are not candidates for total knee arthroplasty l . - Irretrievably failed total knee arthroplasty 2. - Limb salvage in oncology surgery 3. - Trauma 4. - Any other condition where there is little soft tissue or bony tissue available for 5. support, and arthrodesis is the treatment of choice. These devices are intended for cemented use only. Prescription Use X (Per 21 CFR 801 Subpart D) OR Over-the-Counter Use (Per 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evanuation (ODE) lellin Wht (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number *C042405*
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%