K Number
K042357

Validate with FDA (Live)

Device Name
IDI-MRSA ASSAY
Date Cleared
2004-10-22

(52 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IDI-MRSA™ assay is a qualitative in vitro diagnostic test for the direct detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in healthcare settings. The test performed on the Smart Cycler® instrument with a nasal swab specimen from patients at risk for colonization, utilizes polymerase chain reaction (PCR) for the amplification of MRSA DNA and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.

IDI-MRSA is not intended to diagnose MRSA infections nor to guide or monitor treatment for Concomitant cultures are necessary only to recover organisms for MRSA infections. epidemiological typing or for further susceptibility testing.

Device Description

A nasal specimen is collected and transported to the laboratory using the Copan Venturi Transystem®. For testing, the swab is placed in sample preparation buffer. The specimen is concentrated and lysed. An aliquot of the lysate is added to PCR reagents which contain the MRSA-specific primers used to amplify the genetic target [a sequence near the insertion site of Staphylococcal Cassette Chromosome mec (SCCmec)], if present. The assay also includes an internal control (IC) used to detect PCR inhibitory specimens and to confirm the integrity of assay reagents in negative specimens. Amplified targets are detected with hybridization probes labeled with quenched fluorophores (molecular beacons). The amplification, the detection of fluorescence and the interpretation of signals are done automatically by the Smart Cycler® instrument. The whole processed takes about 60 to 75 minutes, depending on the number of specimens processed. This is in addition to specimen preparation which takes about 15 minutes. If appropriate, culture media can be inoculated during specimen preparation or up to 24 hours after its preparation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Infectio Diagnostic Inc. IDI-MRSA™ assay, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission describes a Special 510(k) for modifications to an already cleared device. Therefore, the "acceptance criteria" are primarily established against the performance of the original, cleared formulation of the IDI-MRSA™ assay. The reported device performance indicates equivalence to the original.

ParameterAcceptance Criteria (inferred from "Equivalent to original formulation" and "Meets claims")Reported Device Performance
Validation of cut-offEquivalent to original formulationEquivalent to original formulation
Verification of amplified productsNucleic acids equivalent to their theoretical sizes (as per original)Estimated number of nucleic acids equivalent to their theoretical sizes
Limit of detectionEquivalent to original formulationEquivalent to original formulation
Analytical sensitivity towards different MREJ polymorphic groupsEquivalent to original formulationEquivalent to original formulation
Validation of analytical specificityEquivalent to original formulationEquivalent to original formulation
Potentially interfering substancesEquivalent to original formulationEquivalent to original formulation
Performance with clinical specimens (Sensitivity)Meets claims (of original formulation)Sensitivity meets claims
Performance with clinical specimens (Specificity)Meets claims (of original formulation)Specificity meets claims
Performance with clinical specimens (Rate of unresolved specimens)Equivalent to original formulationRate of unresolved specimens equivalent to original formulation
Performance with clinical specimens (Results agreement with original)>96% agreement with original formulationResults agreement with original formulation >96%
ReproducibilityMeets claims (of original formulation)Meets claims

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size used for the test set of clinical specimens. It only mentions "Performance with clinical specimens" and provides results for sensitivity, specificity, unresolved specimens, and agreement with the original formulation.

The data provenance is not explicitly stated regarding country of origin or whether it was retrospective or prospective. However, given it's a Special 510(k) for modifications, it's highly probable the clinical specimen testing involved comparing the modified assay's performance against the original assay on a set of existing (retrospective) and/or newly collected (prospective) clinical samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For an in vitro diagnostic test like this, the "ground truth" for clinical specimens would typically be established by established microbiological culture methods, often interpreted by clinical microbiologists or infectious disease specialists. However, the document does not detail this process or the number/qualifications of experts involved.

4. Adjudication Method for the Test Set

This information is not provided in the document. Given the nature of a PCR-based assay, adjudication for "ground truth" in the context of clinical specimens usually involves standard microbiological culture results, not expert consensus on the device's output itself.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this device. The IDI-MRSA™ assay is a standalone in vitro diagnostic test that provides a qualitative result (detection of MRSA DNA). There is no "human reader" interpreting the output in the same way as, for example, a radiologist interprets an imaging scan. Therefore, there's no concept of human readers improving with or without "AI assistance" in this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was implicitly done. The IDI-MRSA™ assay is a fully automated system from detection of fluorescence to interpretation of signals by the Smart Cycler® instrument. The comparison study evaluated the modified assay's direct output (detection of MRSA DNA) against the original formulation. The assay itself performs the "algorithm" for positive/negative determination.

7. The Type of Ground Truth Used

For the "Performance with clinical specimens," the ground truth is implicitly microbiological culture results, which is considered the gold standard for detecting Staphylococcus aureus and determining methicillin resistance. The document states: "Concomitant cultures are necessary only to recover organisms for epidemiological typing or for further susceptibility testing," which suggests culture is the comparator, though not explicitly called the "ground truth" for the performance metrics.

8. The Sample Size for the Training Set

The document does not specify a separate "training set" or its sample size. For an IVD like this, analytical validation and optimization might involve various bacterial strains and spiked samples (which could be considered part of a development/training phase), but a distinct "training set" in the machine learning sense is not described. The studies primarily focus on validating the modified assay against the original cleared assay's performance characteristics.

9. How the Ground Truth for the Training Set was Established

As no explicit "training set" is described, the method for establishing its ground truth is not provided. However, for analytical parameters like Limit of Detection, Analytical Sensitivity, and Specificity, the ground truth would be established by carefully characterized bacterial isolates, known concentrations of target DNA, and well-identified non-target organisms.

{0}------------------------------------------------

K042357

OCT 2 2 2004

510(k) Summary

Infectio Diagnostic Inc. IDI-MRSA™ assay Special 510(k)

August 28, 2004

Submitted by:Infectio Diagnostic Inc.2050, boul. René-Lévesque O, 4e étageSainte-Foy, QuébecCanadaG1V 2K8
Contact:Patricia Dionne, PhD.
Name of Device:
Trade Name:Common Name:IDI-MRSATM AssayTest kit for the detection of methicillin-resistantStaphylococcus aureus
Product Code:NQX
Classification NameSystem, Nucleic Acid Amplification Test, DNA, MethicillinResistant Staphylococcus aureus, Direct Specimen
Predicate Device:Infectio Diagnostic Inc. IDI-MRSATM Assay

Device Description:

Intended Use:

IDI-MRSA™ assay is a qualitative in vitro diagnostic test for the direct detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in healthcare settings. The test performed on the Smart Cycler® instrument with a nasal swab specimen from patients at risk for colonization, utilizes polymerase chain reaction (PCR) for the amplification of MRSA DNA and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.

IDI-MRSA is not intended to diagnose MRSA infections nor to guide or monitor treatment for Concomitant cultures are necessary only to recover organisms for MRSA infections. epidemiological typing or for further susceptibility testing.

Test Description:

A nasal specimen is collected and transported to the laboratory using the Copan Venturi Transystem®. For testing, the swab is placed in sample preparation buffer. The specimen is concentrated and lysed. An aliquot of the Ivsate is added to PCR reagents which contain the MRSA-specific primers used to amplify the genetic target [a sequence near the insertion site of Staphylococcal Cassette Chromosome mec (SCCmec)], if present. The assay also includes an internal control (IC) used to detect PCR inhibitory specimens and to confirm the integrity of assay reagents in negative specimens. Amplified targets are detected with hybridization probes

{1}------------------------------------------------

labeled with quenched fluorophores (molecular beacons). The amplification, the detection of fluorescence and the interpretation of signals are done automatically by the Smart Cycler® nuorescence and the interpretainer of eighting 60 to 75 minutes, depending on the number of Institunent. The whole processed takes as a for epidemiological typing or for further Specifichs procosour. This in appropriate culture media can be inoculated during specimen preparation or up to 24 hours after its preparation.

Substantial Equivalence:

This Special 510(k) is submitted for the following modifications to the Infectio Diagnostic Inc. IDI-MRSA™ assay:

ModificationImpact of modification
Development of a 200-tests format(cleared device is a 50-tests format)Change in the workflow of the preparation of PCRreagents for specimens and controls. Positive andNegative Controls prepared by user.
Substitution with chemically modified DNApolymerase complexModification of formulation of existing materials and ofPCR protocol to optimize reaction conditions forsubstituted DNA polymerase.
Upgrade of software versionChange not pertinent to IDI assay

Risk analysis performed on the changes did not raise new issues of safety and effectiveness. The parameters listed below were evaluated in comparison studies of the modified assay versus the original formulation. The modified assay met product claims for all parameters

ParameterResult
Validation of cut-offEquivalent to original formulation
Verification of amplified productsEstimated number of nucleic acids equivalent to theirtheoretical sizes
Limit of detectionEquivalent to original formulation
Analytical sensitivity towards differentMREJ polymorphic groupsEquivalent to original formulation
Validation of analytical specificityEquivalent to original formulation
Potentially interfering substancesEquivalent to original formulation
Performance with clinical specimensSensitivity and specificity meet claims;Rate of unresolved specimens equivalent to originalformulation;Results agreement with original formulation >96%.
ReproducibilityMeets claims

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized lines that resemble a human figure. The lines are curved and connected, creating a sense of movement and unity.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 2 2 2004

Infectio Diagnostic (I.D.I.), Inc. c/o Sienna Partners, LLC Ms. Judi Smith Principal P.O. Box 103 Baldwin, MD 21013

K042357 Re:

Trade/Device Name: IDI-MRSATM Assay Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: Class II Product Code: NQX Dated: October 15, 2004 Received: October 16, 2004

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encroours) to regists of the Medical Device Amendments, or to conniner of Har to Har to Har 2011 accordance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). alla Cosmetic Hot (110) that to neview subject to the general controls provisions of the Act. The r ou may, mereleve, mailer of the Act include requirements for annual registration, listing of general controls provisions of tax tice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is clusion (600 a00 ro) als. Existing major regulations affecting your device can may be subject to save adon adon Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Tease oc advised that I Dr o ration that your device complies with other requirements of the Act that I DA has made a assod regulations administered by other Federal agencies. You must or any I odetail the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice CI K Part 8077, labornish in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) I ins letter will anow you to ogen mains of substantial equivalence of your device to a legally premitset notification. The PDA mianing of East Minister and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, II you desire specific information advertising of your device, please contact the Office of of questions on the promotion and Safety at (301) 594-3084. Also, please note the In Viro Diagliostic De Hood Diarabing by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I bu may oodain other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Saartys

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

1 - Indications for Use Statement

510(k) Number (if known): K042357

Device Name:___IDI-MRSA™

Indications For Use:

IDI-MRSA™ assay is a qualitative in vitro diagnostic test for the direct detection of nasali colonization by methicillin-resistant Staphylococus aureus (MRSA) to aid in the prevention colonization by memicilinf resistant Staphyrosooooon and the Smart aria Control of MROA in noctoris in noch specimen from patients at risk for colonization, Cycler@ instrument with a nasal onals of the amplification of MRSA DNA and utilizes "polymerase" onain" (roustler)" (rom) - for the detection of the amplified DNA.

IDI-MRSA™ assay is not intended to diagnose MRSA infections nor to guide or monitor IDI-MIKOA - assocy is net internet to concomitant cultures are necessary only to recover troalmons for epidemiological typing or for further susceptibility testing.

Prescription Use _ ٢

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie L. Poole

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of

510(k) K042357

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).