K Number
K042273

Validate with FDA (Live)

Date Cleared
2004-08-31

(8 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Health Buddy® with Device Connectivity is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

Device Description

The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuff, patient weight scale and peak flow meter). The Health Buddy with Device Connectivity retrieves data from a specific medical device and stores it for later transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, and weight. In this submission, minor software changes are made in the Health Buddy ROM to allow it to connect to a Peak Flow Meter and monitor Peak Expiratory Flow (PEF) and FEV1 measurements. The Health Buddy has the ability to receive connections from optional, compatible devices, including those identified above, and others, such as digital thermometers, pulse oximeters and prothrombin/INR meters. The Health Buddy device has a data port that allows it to download readings from an identified attached device, activated only upon appropriate patient enrollment.

The Health Buddy with Device Connectivity is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to the Data Center via a toll-free number to send patient responses from the previous day and to retrieve the current day's dialogue.

The screen displays the information or questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient information is sent to the patient's health care provider.

AI/ML Overview

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document is a 510(k) premarket notification for a minor software modification, asserting substantial equivalence to previously cleared devices. It does not contain explicit performance acceptance criteria or a study with reported device performance metrics against such criteria.

The core of the submission is to demonstrate that the minor software changes to allow connection to a Peak Flow Meter do not alter the fundamental safety and effectiveness of the existing device. It relies on the concept of "substantial equivalence" rather than a de novo performance study.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable (N/A) – No specific test set to evaluate performance is described in this substantial equivalence submission.
  • Data Provenance: Not applicable (N/A) – No new clinical data or data from a test set is presented. The submission relies on establishing equivalence to previously cleared devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable (N/A) – No new ground truth establishment for a test set is described.
  • Qualifications of Experts: Not applicable (N/A)

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable (N/A) – No test set or ground truth adjudication is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Done: No. This type of study is not mentioned or implied in the provided document. The device is a data collection and transmission accessory, not an AI diagnostic tool intended to assist human readers in image interpretation or similar tasks.
  • Effect Size of Human Readers Improve with AI vs. Without AI Assistance: Not applicable (N/A)

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Standalone Study Done: No. The Health Buddy is described as a "communication tool" and an "accessory device" that collects and transmits historical data. It is not an algorithm that performs standalone diagnostic or interpretive functions.

7. Type of Ground Truth Used

  • Type of Ground Truth: Not applicable (N/A) – No ground truth is established or used in this submission. The device's function is to transmit existing physiological data, not to make a diagnosis or interpretation itself.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable (N/A) – The document does not describe a machine learning algorithm or a training set. The "software changes" mentioned are related to device connectivity for specific medical devices, not an AI/ML model.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth Was Established: Not applicable (N/A) – No training set or ground truth establishment for such a set is described.

Summary of Device's Approach to Meeting Requirements:

The document asserts that the modified Health Buddy® with Device Connectivity meets safety and effectiveness requirements by demonstrating substantial equivalence to predicate devices that have already received 510(k) clearance. The changes are described as "minor software changes" to allow connection to a Peak Flow Meter. The justification relies on the fact that the fundamental function of collecting and transmitting historical data for healthcare professionals (without providing automated treatment decisions) remains unchanged and aligns with the cleared functions of the predicate devices. Therefore, a new performance study with acceptance criteria and ground truth establishment was not deemed necessary for this specific submission.

{0}------------------------------------------------

Section 1.3

Summary of Safety and Effectiveness

REGULATORY AUTHORITY

Safe Medical Devices Act of 1990, 21 CFR 807.92

COMPANY NAME/CONTACT

Geoffrey J. Clapp Health Hero Network 2570 W. El Camino Real, Suite 111 Mountain View, CA 94040 (650) 559.1188 (phone) (650) 559.1050 (facsimile) clapp@healthhero.com

NAME OF DEVICE

Trade Name:Health Buddy® with Device Connectivity
Common Name:Data Management System; Accessory to Medical
Device
Classification Names:Refer to Table
RegulationNumberProductCodeClassification NameDeviceClass
862.1345CGAGlucose Test SystemII
870.1130DXNNoninvasive BloodPressureMeasurementSystemII
880.2700FRIPatient Weight ScaleI
868.1860BZHMeter, Peak Flow,SpirometryII
870.2910DRGPhysiological SignalTransmitters andReceiversII

{1}------------------------------------------------

PREDICATE DEVICES

  • Health Buddy® with Device Connectivity (K040086) .
  • Health Buddy® with Buddylink (K993128) ●
  • CareMatix Wellness System (#K040966) .
  • Viterion 100 TeleHealth Monitor (#K030419) .

DEVICE DESCRIPTION

The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuff, patient weight scale and peak flow meter). The Health Buddy with Device Connectivity retrieves data from a specific medical device and stores it for later transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, and weight. In this submission, minor software changes are made in the Health Buddy ROM to allow it to connect to a Peak Flow Meter and monitor Peak Expiratory Flow (PEF) and FEV1 measurements. The Health Buddy has the ability to receive connections from optional, compatible devices, including those identified above, and others, such as digital thermometers, pulse oximeters and prothrombin/INR meters. The Health Buddy device has a data port that allows it to download readings from an identified attached device, activated only upon appropriate patient enrollment.

The Health Buddy with Device Connectivity is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to the Data Center via a toll-free number to send patient responses from the previous day and to retrieve the current day's dialogue.

The screen displays the information or questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient information is sent to the patient's health care provider.

INDICATION FOR USE STATEMENT

Health Buddy® with Device Connectivity is indicated for use in non-clinical settings to collect and transmit historical data to healthcare professionals to help support effective management of their patients.

{2}------------------------------------------------

The Health Buddy® is an accessory device, intended to be a communication tool to enable healthcare providers to receive historical patient information. The product is used in conjunction with Health Hero Network's Online Service, a communication tool to enable health care providers to educate, motivate, and receive patient information. Health Buddy with Device Connectivity is not intended to provide automated treatment decisions, nor is it to be used as a substitute for a professional healthcare judgment. All patient medical diagnosis and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional.

SUBSTANTIAL EQUIVALENCE COMPARISON

This submission represents a modification to the software in the Health Buddy® with Device Connectivity, and is therefore substantially equivalent to the cleared Health Buddy device (#K040086 and #K993128). The device is substantially equivalent to the Viterion 100 TeleHealth Monitor (#K030419), as well as the CareMatix Wellness System (#K040966).

CONCLUSION

The Health Buddy with Device Connectivity is substantially equivalent to devices cleared under the Federal Food, Drug and Cosmetic Act. Therefore, safety and effectiveness are reasonably assured, justifying 510(k) clearance of this modified device.

005

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized caduceus-like symbol with three wavy lines representing the human services aspect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle, following its curvature.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 3 1 2004

Health Hero Network, Inc. c/o Mr. Geoffrey Clapp Chief Technical Officer & Chief Operating Officer 2570 W. El Camino Real, Suite 111 Mountain View, CA 96040

Re: K042273

K042273
Trade Name: Health Buddy® with Device Connectivity Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: August 20, 2004 Received: August 23, 2004

Dear Mr. Clapp:

We have reviewed your Section 510(k) premarket notification of intent to market the device in We have reviewed your Section 310(x) premained is substantially equivalent (for the indications referenced and nave decemblied the devices marketed predicate devices marketed in interstate for use stated in the encrosule, to regally manetice province Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 26, 1976, the enactified in accordance with the provisions of the Federal Food. Drug. devices that have been reclassified in accordance what a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a presisions of the Act and Cosmetic Act (Act) that to not require approval controls provisions of the Act. The
You may, therefore, market the device, subject to the general controls provisions of You may, merelore, market the devreet, seeject or use gentrements for annual registration, listing of the general controls provisions of the rise labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (SCC above) into existing major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional controller Lines and to 898. In addition, FDA may be found in the Code of Peachar Regarning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Mr. Geoffrey Clapp

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualles of a succession complies with other requirements of the Act
that FDA has made a determination that your device complies with over marke that FDA has made a decionmation administered by other Federal agencies. You must of any Federal statutes and regulations daminities on thot limited to: registration and listing (21 comply with an the Act 3 requirements, news 801); good manufacturing practice requirements as set any CFK Part 807), labeling (21 CFR Part 820); and 1f applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Declients on a described in your Section 5 10(k)
This letter will allow you to begin marketing your device as described in your section I ms letter will anow you to oogin maneting of substantial equivalence of your device to a legally premarket notification. The PDA midning of backership for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (1 CFR Part 801), please If you desire specific advice for your do 100 car. Also, please note the regulation entitled, contact the Office of Comphalled at (50 mg tification" (21CFR Part 807.97). You may obtain Misolallung by reference to premail.com the Act from the Division of Small other general information on your responsibilities and its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Nil R&, Ogden

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K04227=

Health Hero Network

Special 510(k): Device Modifications Health Buddy® with Device Connectivity

Section 1.2

Indications for Use Statement

510(k) Number:K042273
-------------------------

Device Name:

Health Buddy® with Device Connectivity

Indications for Use:

Health Buddy® with Device Connectivity is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

Prescription Use XX (Per 21 CFR 801.109)

.

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evalua

or

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K042273

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).