K Number
K042168

Validate with FDA (Live)

Date Cleared
2005-01-14

(158 days)

Product Code
Regulation Number
888.3027
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Symphony™ VR Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

Device Description

Symphony™ VR Radiopaque Bone Cement is a PMMA bone cement made of the same chemical components as Advanced Biomaterial Systems, Inc. CONCERT® Radiopaque Bone Cement.

Symphony™ VR Radiopaque Bone Cement is provided as a two-component product. The polymer powder consists of a PMMA copolymer (polymethyl methacrylate and methyl methacrylate-styrene copolymer) with barium sulfate as the radiopacifier and benzoyl peroxide as the initiator. The liquid component consists of methyl methacrylate monomer, which includes hydroquinone as the stabilizer and N:N dimethyl-p-toluidine as the activator.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for a medical device called Symphony™ VR Radiopaque Bone Cement. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing a study proving the device meets specific performance acceptance criteria in the way a clinical trial for an AI algorithm would.

Therefore, many of the requested points regarding acceptance criteria, study design, ground truth, and human reader performance are not applicable to this type of regulatory submission for a bone cement.

Here's an analysis of the provided information in the context of the questions:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as pass/fail thresholds for clinical or diagnostic performance. Instead, the "acceptance criteria" are implied to be demonstrating substantial equivalence to legally marketed predicate devices in terms of:

    • Intended Use
    • Functional Characteristics (though not quantified with performance metrics)
    • Material Properties (chemical composition)
    • Biocompatibility
    • Clinical Application
  • Reported Device Performance:
    The document focuses on comparing the chemical composition of Symphony™ VR Radiopaque Bone Cement to its predicate devices. The key "performance" demonstrated here is the similarity in composition.

    Chemical CompositionSymphony™ VRKyphX® HV-RStryker Spineplex™Howmedica Osteonics Surgical Simplex® P
    Powder
    PolymethylMethacrylate / MethylMethacrylate-styrenecopolymer71.3% w/w68.0% w/w69.1% w/wNot provided for Simplex®
    Barium sulfate28.0% w/w30.0% w/w30.0% w/wNot provided for Simplex®
    Benzoyl peroxide0.7% w/w2.0% w/w0.9% w/wNot provided for Simplex®
    Liquid
    Methyl Methacrylate99.0% v/v99.1% v/v97.4% v/vNot provided for Simplex®
    N: N Dimethyl-p-toluidine1.0% v/v0.9% v/v2.6% v/vNot provided for Simplex®
    Hydroquinone100 ppm75 ppm75 ± 15 ppmNot provided for Simplex®

    Note: The document states "Symphony™ VR Radiopaque Bone Cement is a PMMA bone cement made of the same chemical components as Advanced Biomaterial Systems, Inc. CONCERT® Radiopaque Bone Cement." which is not listed in the comparison table, but implies a known baseline for the manufacturer.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable. This is a medical device (bone cement) submission, not an AI algorithm. There is no "test set" of patient data in the context of diagnostic performance. The "test" here is demonstrating chemical and functional similarity to predicates.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This is not an AI algorithm for diagnosis or image interpretation. "Ground truth" in the AI sense is not relevant.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not Applicable. The "ground truth" for this submission is the established properties of the predicate devices and the chemical synthesis and characterization of the new device.

8. The sample size for the training set

  • Not Applicable. This is not an AI algorithm, so there is no training set in the machine learning sense.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

K 042168 7.0 PREMARKET NOTIFICATION 510(K) SUMMARY

510(k) Number:TO BE ASSIGNED
Date Prepared:August 10, 2004
Applicant Information:
Applicant:Advanced Biomaterial Systems, Inc.100 Passaic AvenueChatham, NJ 07928
Contact:John P. CarrChief Operating Officer
Telephone:Facsimile:(973) 635-9040(973) 635-9878
Registration:To be assigned
Device Information:
Trade Name:Symphony™ VR Radiopaque Bone Cement
Common Name:PMMA Bone Cement (For Vertebroplasty)
Product Code:NDN
Classification Name:Filler, Bone Cement (For Vertebroplasty)
Regulation Class:Class II
Regulation Number:21 CFR §888.3027
Device Description:Symphony™ VR Radiopaque Bone Cement is a PMMA bonecement made of the same chemical components as AdvancedBiomaterial Systems, Inc. CONCERT® Radiopaque Bone Cement.Symphony™ VR Radiopaque Bone Cement is provided as a two-component product. The polymer powder consists of a PMMAcopolymer (polymethyl methacrylate and methyl methacrylate-styrene copolymer) with barium sulfate as the radiopacifier andbenzoyl peroxide as the initiator. The liquid component consists ofmethyl methacrylate monomer, which includes hydroquinone as thestabilizer and N:N dimethyl-p-toluidine as the activator.
Predicate Devices:K033801:Kyphon Inc. - KyphX® HV-R Bone Cement
K032945:Stryker Spineplex™ Radiopaque Bone Cement

of

{1}------------------------------------------------

Howmedica Osteonics Surgical Simplex® P N017004: Radiopaque Bone Cement

Symphony™ VR Radiopaque Bone Cement is indicated for the Intended Use: fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

The following table compares the chemical composition of Symphony™ VR Radiopaque Bone Cement compared to the predicate devices.

Chemical CompositionSymphony™ VRKyphX® HV-RStryker Spineplex™
Powder20 g (half-dose) Bottleof Sterile Powder20 g (half-dose) Packetof Sterile Powder20 g (half-dose) Packetof Sterile Powder
PolymethylMethacrylate / MethylMethacrylate-styrenecopolymer71.3% w/w68.0% w/w69.1% w/w
Barium sulfate28.0% w/w30.0% w/w30.0% w/w
Benzoyl peroxide0.7% w/w2.0% w/w0.9% w/w
Liquid8.2g (half-dose) Vial ofSterile Liquid9.0g (half-dose) Vial ofSterile Liquid9.4 (half-dose) Vial ofSterile Liquid
Methyl Methacrylate99.0% v/v99.1% v/v97.4% v/v
N: N Dimethyl-p-toluidine1.0% v/v0.9% v/v2.6% v/v
Hydroquinone100 ppm75 ppm75 ± 15 ppm

Summary:

Based on the device performance information provided in this premarket notification, SymphonyTM VR Radiopaque Bone Cement has been shown to be substantially equivalent to the currently marketed predicate devices. This device has the same intended use, functional characteristics, material properties, biocompatibility and clinical application as the predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white emblem or seal. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the outer edge. Inside the circle is a stylized symbol that resembles three abstract human figures or waves, stacked on top of each other.

Public Health Service

JAN 1 4 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. John P. Carr Chief Operating Officer Advanced Biomaterial Systems 100 Passaic Avenue Chatham, New Jersey 07928

Re: K042168

Trade/Device Name: Symphony™VR Radiopaque Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate Bone Cement Regulatory Class: II Product Code: NDN Dated: December 17, 2004 Received: December 20, 2004

Dear Mr. Carr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. John P. Carr

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark A Milhern

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510 (k) Number (if known): ___________________________________________________________________________________________________________________________________________________

Symphony™ VR Radiopaque Bone Cement Device Name:

Indications For Use:

Symphony™ VR Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device, Evaluation (ODE)

Mark N. Melkerson

Division of General, Restorative, and Neurological Devices

510(k) Number

§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.

(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”