K Number
K042077

Validate with FDA (Live)

Date Cleared
2004-10-27

(86 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BiTech bipolar scissors are designed for dissecting, cutting and bipolar coagulation of tissue during general surgical procedures.

The BiTech Bipolar Scissors has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.

Device Description

BiTech bipolar scissors enable cutting and dissection with simulataneous bipolar coagulation. They can also be used for precise pinpoint or zone coagulation of blood-vessels and tissue. The various working lengths allow the scissors to be used for a large number of applications in open surgery.

A petented construction priciple, which was developed by Dr. J. Manushakian*, ensures that the scissors are both easy to use and to repair. The electrical contacts for both scissor blades are conducted via one section of the handle, i. e. the connection cable only needs to be attached to one ring, which significantly reduces reference duriong handling. The scissor blades are flexibly joined by means of a standard steel scissor screw which does not need to be insulated and, as a result, is not subject to wear.

AI/ML Overview

The provided document describes a 510(k) notification for the BiTech Bipolar Scissors. A 510(k) submission primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting new clinical study data with specific acceptance criteria and performance metrics in the way a novel device might.

Therefore, the document does not contain the detailed information required for the requested output regarding acceptance criteria and the comprehensive study that proves the device meets those criteria. Specifically:

  • No specific acceptance criteria are listed for device performance in terms of accuracy, sensitivity, specificity, etc.
  • No study is described that details a test set, training set, ground truth establishment, or expert adjudication for performance evaluation.
  • There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study or a standalone algorithm-only study.

Instead, the submission relies on demonstrating compliance with recognized performance standards for safety and electrical equipment, and stating substantial equivalence to existing devices.

Here's what can be extracted from the document, tailored to the spirit of your request where possible, but acknowledging the limitations of this type of submission:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance / Compliance
Safety - GeneralDIN EN 60601-1Compliant
Safety - HF Surgical EquipmentDIN EN 60601-2-2Compliant
Safety - Electrosurgical DevicesANSI/AAMI HF18-2001Compliant
FunctionalitySubstantial Equivalence to Predicate Devices (Ethicon K981361, Enable K972558)Device functions as "dissecting, cutting and bipolar coagulation of tissue" similar to predicates.
SterilizationUser sterilization via validated steam process (DIN EN 554, 137°C/280°F, 3 bar, min. 15 min)Device is provided non-sterile and can be sterilized as per specified method.
Cleaning/MaintenanceUser cleaning via specified enzymatic and non-corrosive agents, soft brush, warm water, rinse, complete dryness.User instructions provided for cleaning and maintenance.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable. This submission relies on substantial equivalence and compliance with engineering standards, not on a clinical performance study with a test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No ground truth establishment by experts for a test set is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No test set described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI-assisted diagnostic device, nor is an MRMC study mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a surgical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable. For this type of device and submission, the "ground truth" for safety and efficacy is generally considered to be the established performance and safety record of the predicate devices and compliance with recognized industry standards.

8. The sample size for the training set

  • Not applicable. No training set is described.

9. How the ground truth for the training set was established

  • Not applicable. No training set is described.

Summary of the Study/Evidence:

The "study" in a 510(k) submission like this is primarily a bench-testing and documentation effort to demonstrate:

  • Compliance with recognized electrical and safety standards: DIN EN 60601-1, DIN EN 60601-2-2, and ANSI/AAMI HF18-2001. This would involve testing the device against the specific requirements and clauses within those standards.
  • Substantial equivalence to legally marketed predicate devices: Ethicon (K981361) and Enable (K972558). This involves comparing the technological characteristics, intended use, and performance claims of the BiTech Bipolar Scissors to these established predicates. The document implies that the "patented construction principle" and how electrical contacts are conducted are key differentiating features, but these are presented as improvements in usability and repair, not as fundamentally altering the safety or efficacy compared to predicates for the stated intended use.
  • Provision of validated user instructions: For cleaning, maintenance, and sterilization, ensuring the device can be safely and effectively prepared for use.

The FDA's letter states: "We have reviewed your Section 510(k) premarket notification of intent to market the device...and have determined the device is substantially equivalent...to legally marketed predicate devices..." This confirms the FDA's acceptance of the manufacturer's demonstration of substantial equivalence based on the provided documentation and compliance with the mentioned standards.

{0}------------------------------------------------

OCT 27 2004

Image /page/0/Picture/1 description: The image contains three text boxes. The first text box contains the logo "bissinger". The second text box contains the text "510(k) Notification" and "BiTech Bipolar Scissors". The third text box contains the text "Annex I, Page 2 of 4".

1. Device Name

Trade Name:BiTech Bipolar Scissors
Common Name:Bipolar Scissors

2. Classification

Device:Device, Electrosurgical, Cutting & Coagulation & Accessories
Device description:Electrosurgical cutting and coagulation device and accesso-ries.
Medical Specialty:Part 878, General & Plastic Surgery
Product Code:GEI
Regulation Number:878.4400
Device Class:2

3. Substantial Equivalence

BiTech Bipolar Scissors are substantially equivalent to other legally marketed Bipolar Scissors from different manufacturers, e.g.

4. Description of the Device

BiTech bipolar scissors enable cutting and dissection with simulataneous bipolar coagulation. They can also be used for precise pinpoint or zone coagulation of blood-vessels and tissue. The various working lengths allow the scissors to be used for a large number of applications in open surgery.

A petented construction priciple, which was developed by Dr. J. Manushakian*, ensures that the scissors are both easy to use and to repair. The electrical contacts for both scissor blades are conducted via one section of the handle, i. e. the connection cable only needs to be attached to one ring, which significantly reduces reference duriong handling. The scissor blades are flexibly joined by means of a standard steel scissor screw which does not need to be insulated and, as a result, is not subject to wear.

*Patented: US 6,355,035, EPO 99107816.3.

5. Intended Use

BiTech bipolar scissors are designed for dissecting, cutting and bipolar coagulation of tissue during general surgical procedures.

The BiTech Bipolar Scissors has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo with the word "bissinger" in lowercase letters. Above the word, there is a stylized eye graphic. The pupil of the eye is shaped like the letter "b". The logo is simple and uses a bold, sans-serif font.

510(k) Notification

BiTech Bipolar Scissors

Annex I, Page 3 of 4

6. Performance Standards

DIN EN 60601-1: Medical electrical equipment - Part 1: General requirements for safety (IEC 60601-1:1988 + A1:1991 + A2:1995); German version EN 60601-1:1990 + A1:1993 + A2:1995; Version: 01-Mar-1996;

DIN EN 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment (IEC 60601-2-2:1998); German version EN 60601-2-2:2000; Version: 01-Aug-2001:

ANSI/AAMI HF18-2001; Electrosuraical Devices: Version: 01-May-2001.

7. Sterilization by User

BiTech Bipolar Scissors are delivered in non-sterile conditions. The user may sterilize these devices by using a validated and applicable sterilization process.

Cleaning and Maintenance

Every surgical instrument should be disinfected and thoroughly cleaned after each use. Proper cleaning, inspection and maintenance will help ensure correct function of the surgical instrument. Clean, inspect and test each instrument carefully. Sterilise all instruments before surgery. A good cleaning and maintenance procedure will extend the useful life of the instrument.

Special attention must be paid to slots, stops, ends, hollow tubes and other highly inaccessible areas. Check insulation, cables and connectors for cuts, tears, abrasions, etc.

Do not use damaged instruments.

Cleaning and rinsing must take place immediately after each use for best effect. Failure to clean promptly may result in adherent particles or dried secretions that may resist cleaning and complicate or resist future sterilisation.

Instruments must be completely cleaned and rinsed of all foreign matter.

Use warm water and a commercially available instrument pre-soak or cleaning agent. Enzymatic cleaners (such as Enzol™) must be used to remove protein deposits. Follow the enzymatic cleaner's instructions; rinse thoroughly.

  • · Do not use corrosive cleaning agents (i.e. bleach). Cleaning solutions and rinses at or near a neutral pH (7.0) are best.
  • · Do not use abrasive cleaners.
  • · Only a soft bristle brush should be used.
  • · Immerse the entire device in detergent and clean while soaking, Rinse with sterile deionized water. Can be disinfected in the washing machine up to 203°F (95°C).
  • · Rinse thoroughly with distilled water.
  • · Prepare for storage and/or sterilisation.

Sterilisation

Only a validated steam-sterilization process according DIN EN 554 that uses a sterlization cycle of 137°C / 280°F, 3 bar, for min. 15 minutes has to be used.

(Note: Contact the manufacturer of your steam autoclave to confirm appropriate temperatures and sterilisation times.)

Caution: Autoclave temperatures should not exceed 280°F (137°C); handles, insulation or other non-metallic parts may be damaged

Do not sterilise with hot air.

Do not use 'Flash' autoclave procedures.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows three boxes with text inside. The first box contains the word "bissinger" and a logo. The second box contains the text "510(k) Notification" and "BiTech Bipolar Scissors". The third box contains the text "Annex I, Page 4 of 4".

8. Conclusion

Based on the available 510(k) summaries and 510(k) statements (21 CFR 807) and the information provided herein, we conclude that Bissinger BiTech Bipolar Scissors are substantially equivalent to the existing legally marketed devices under Federal Food, Drug and Cosmetic Act.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird with three curved lines representing its wings.

Public Health Service

OCT 2 7 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Guenter Bissinger Medizintechnik GmbH c/o Mr. Franz Menean Managing Director MEDAGENT GmbH & Co. KG 47. Griesweg Muehlheim, Baden-Wuerttemberg Germany 78570

Re: K042077

Trade/Device Name: BiTech Bipolar Scissors Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: July 26, 2004 Received: August 11, 2004

Dear Mr. Menean:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becament in ad the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass based in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, do noos mat have been that do not require approval of a premarket approval application (PMA). and Oosmeter For (1 roy that the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or cable for the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Franz Menean

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter with and in your finding of substantial equivalence of your device to a legally promated notification " results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you dome office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mornational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K042077

Device Name: BiTech Bipolar Scissors

Indications for Use:

BiTech bipolar scissors are designed for dissecting, cutting and bipolar coagulation of tissue during general surgical procedures.

The BiTech Bipolar Scissors has not been shown to be effective for tubal sterilization or the Bircon Elipola. Other its and should not be used for these procedures.

Prescription Use __ YES___ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use NO_ (21 SFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number Kc42077

Page 1 of 1

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.