K Number
K041907

Validate with FDA (Live)

Device Name
TRIO MONITOR
Manufacturer
Date Cleared
2004-07-23

(8 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trio™ monitor is intended for use in healthcare settings under the direct supervision of a licensed healthcare practitioner. The intended use of the monitor is to monitor physiologic parameter data on adult and pediatric patients. Physiologic data includes: electrocardiogram, invasive blood pressure, non-invasive blood pressure (NIBP), pulse oximetry, heart rate (derived from ECG, SpO2 or NIBP), respiration and temperature as summarized in the operating instructions manual. The information can be displayed, stored, trended and printed.

The monitor is not intended for home use. The monitor is not intended to be an apnea monitor. It was not designed or validated for use as an apnea monitor

Device Description

The Trio monitor is a three or four trace patient monitor that can be mounted on a rolling stand, wall mount bracket, bed rail, or operated as a tabletop device. The patient parameters that can be monitored on the Trio are: ECG (3-lead or 5- lead selectable), Temperature, SpO2, Invasive Blood Pressure, Respiration and Non-Invasive Blood Pressure. Digital displays are provided for Heart Rate, Pulse Rate, Pulse Oximetry, Non-Invasive Blood Pressure, Respiration Rate, Invasive Blood Pressure (optional) and Temperature. Waveform displays are provided for ECG, Pleth, Invasive Blood Pressure (optional) and Respiration. The optional built-in thermal recorder provides hard copies of all digital data and waveforms, as well as Tabular and Graphic Trend information.

AI/ML Overview

The provided text for K041907 focuses on establishing substantial equivalence for the Trio Monitor, particularly regarding its expanded capability to derive Heart Rate from Non-Invasive Blood Pressure (NIBP). While it mentions testing, it does not offer the detailed acceptance criteria and study results in the structured format requested. The document primarily references previous submissions and states that the device is "as safe, as effective and performs as well, as the predicate devices."

Therefore, I cannot fulfill all sections of your request based solely on the provided text. I will, however, extract what information is present and indicate where details are missing.


Acceptance Criteria and Study for K041907 (Trio Monitor - NIBP Derived Heart Rate)

The provided submission does not explicitly state detailed, quantitative acceptance criteria or present a full study report with specific performance metrics for the NIBP-derived heart rate functionality. Instead, it relies on demonstrating substantial equivalence to predicate devices and adherence to voluntary standards.

The document indicates that testing and documentation related to the ability of the Trio monitor to derive Heart Rate from Non-Invasive Blood Pressure were included in sections five, seven, and eight of the original 510(k) submission K032338, as well as in the current submission. However, these sections are not provided in the input text.

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Not explicitly stated in provided text)Reported Device Performance (Not explicitly stated in provided text)
Accuracy of NIBP-derived Heart Rate vs. Reference Heart RatePresumed to meet AAMI standards or predicate performanceStated as "performs as well as predicate devices"
Bias (Mean Difference)Not providedNot provided
Precision (Standard Deviation)Not providedNot provided
Agreement Limits (e.g., Bland-Altman)Not providedNot provided

Note: The submission states that "The results of these measurements demonstrated that the Trio monitor is as safe, as effective and performs as well, as the predicate devices." This is a general statement of equivalency rather than specific numerical performance data or explicit acceptance criteria.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified in the provided text.
  • Data Provenance: Not specified in the provided text. This would typically include information about the patient population, site of data collection (e.g., country of origin), and whether the data was collected retrospectively or prospectively.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not specified.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • MRMC Study: No, this type of study is not relevant for a physiological monitoring device that produces objective measurements like heart rate. MRMC studies are typically used to evaluate diagnostic imaging algorithms where human readers interpret images.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done

  • Standalone Study: Yes, the evaluation of the NIBP-derived Heart Rate would inherently be a standalone performance assessment of the algorithm against a reference standard (e.g., ECG-derived Heart Rate). However, specific details of this study (e.g., its design, results) are not provided in the text. The statement "The ability to derive Heart Rate from the Non-Invasive Blood Pressure parameter is the subject of this 510(k) submission" and "Testing and documentation, related to the ability of the Trio monitor to derive Heart Rate from Non-Invasive Blood Pressure, is included in this 510(k) submission" implies such benchmarking was performed algorithmically.

7. The Type of Ground Truth Used

  • Type of Ground Truth: While not explicitly stated, for a physiological parameter like heart rate derived from NIBP, the ground truth would most likely be a well-established and highly accurate reference method, such as Heart Rate derived from Electrocardiogram (ECG). The document mentions that in the previous submission, heart rate was derived from ECG and SpO2, suggesting ECG as a primary reference.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not specified in the provided text. The document refers to "development process" and "testing and documentation," but does not detail the training set used for the NIBP-derived heart rate algorithm if machine learning was involved in its development. However, NIBP-derived heart rate typically relies on signal processing algorithms rather than complex machine learning models requiring large "training sets" in the modern AI sense.

9. How the Ground Truth for the Training Set was Established

  • Ground Truth for Training Set: Not specified. If the algorithm is based on signal processing principles rather than machine learning, a "ground truth for a training set" in the typical sense would not be applicable. The algorithm would be built upon known physiological relationships and signal characteristics where fundamental reference measurements (like those from ECG) are implicitly the gold standard.

In summary, the provided document K041907 serves as a 510(k) summary for a modification, referencing a previous submission (K032338). It states that testing was performed to demonstrate the safety and effectiveness of the NIBP-derived heart rate, concluding that it performs "as well as predicate devices." However, it lacks the detailed quantitative data, sample sizes, and methodological specifics typically found in full study reports when evaluating novel algorithms or device performance against explicit acceptance criteria.

{0}------------------------------------------------

K041907
page 1 of 3

JUL 2 3 2004

510(k) Summary of Safety and Effectiveness

Date:June 10, 2004
Submitter:Patient Monitoring DivisionDatascope Corp.
Contact Person:Susan E. Mandy
Director, Clinical & Regulatory Affairs
Patient Monitoring Division
Datascope Corp.
Telephone: (201) 995-8025
Fax: (201) 995-8605

Device trade name: Trio Monitor

Common/usual name: Physiological Patient Monitor

Classification name:

21 CFR 870.1100Blood pressure alarm
21 CFR 870.1130Noninvasive blood pressure measurement systems
21 CFR 870.1110Blood pressure computer
21 CFR 870.2300Cardiac monitor (including cardiotachometer and rate alarm)
21 CFR 870.2340Electrocardiograph
21 CFR 870.2700Oximeter
21 CFR 880.2910Clinical electronic thermometers
  • Device Description: Mindray Bio-Medical Co. Ltd., previously known as Caymans Mindray Medical Electronic Company of Shenzen, China, manufactures the Trio monitor for Datascope Corp.

{1}------------------------------------------------

Background:

The FDA cleared the Trio monitor to market under 510(k) K032338 on February 11, 2004. Datascope Corp. has modified the Trio monitor to include Heart Rate derived from Non-Invasive Blood Pressure. Datascope has also made some minor modifications to the Trio monitor as a result of feedback received from both domestic and international customers, and from an October 2002 focus group, referenced on page 228 of K032338, submitted to the FDA on July 28, 2003. These modifications include the replacement of the "Freeze" key with a "Normal Screen" key, minor user interface display formatting modifications, e.g., changes to fonts, colors, screen layout, as well as user preferred default selections.

Purpose:

The ability to derive Heart Rate from the Non-Invasive Blood Pressure parameter is the subject of this 510(k) submission. In the previous Trio submission, Heart Rate was derived from ECG and SpO2. The description of the Trio monitor remains the same as in the original 510(k) K032338, submitted to the FDA, July 28, 2003 and is provided below. Datascope Corp. incorporates by reference Trio monitor 510(k) submission K0323338.

The Trio monitor is a three or four trace patient monitor that can be mounted on a rolling stand, wall mount bracket, bed rail, or operated as a tabletop device. The patient parameters that can be monitored on the Trio are: ECG (3-lead or 5- lead selectable), Temperature, SpO2, Invasive Blood Pressure, Respiration and Non-Invasive Blood Pressure. Digital displays are provided for Heart Rate, Pulse Rate, Pulse Oximetry, Non-Invasive Blood Pressure, Respiration Rate, Invasive Blood Pressure (optional) and Temperature. Waveform displays are provided for ECG, Pleth, Invasive Blood Pressure (optional) and Respiration. The optional built-in thermal recorder provides hard copies of all digital data and waveforms, as well as Tabular and Graphic Trend information.

Intended Use:

The Trio monitor is intended for use in healthcare settings under the direct supervision of a licensed healthcare practitioner. The intended use of the monitor is to monitor physiologic parameter data on adult and pediatric patients. Physiologic data includes: electrocardiogram, invasive blood pressure, non-invasive blood pressure (NIBP), pulse oximetry, heart rate (derived from ECG, SpO2 or NIBP), respiration, and temperature as summarized in the operating instructions manual. The information can be displayed, stored, trended and printed.

{2}------------------------------------------------

The Trio monitor employs the same functional technology as the predicate Technology: devices.

The Trio monitor complies with the voluntary standards identified in Test Summary: section six of this submission. During the development process of the Trio monitor described in K032338, the following activities were completed:

  • Requirements specification review .
  • . Hardware and software testing
  • Code design and code reviews .
  • Environmental testing .
  • . Safety testing
  • Performance testing .
  • Risk Analysis .
  • Hardware and Software validation .

Testing and documentation, related to the ability of the Trio monitor to derive Heart Rate from Non-Invasive Blood Pressure, is included in this 510(k) submission, sections five, seven, and eight.

The results of these measurements demonstrated that the Trio monitor is as Conclusion: safe, as effective and performs as well, as the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 3 2004

Datascope Corp. c/o Mr. Bahram Barzideh Underwriters Laboratories, Inc. 1285 Walt Whitman Rd. Melville, NY 11747

Re: K041907

Trade Name: Trio Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: 21 OF CF 81 8126 including cardiotachometer and rate alarm) Regulatory Class: II (two) Product Code: MWI -Dated: July 13, 2004 Received: July 15, 2004

Dear Mr. Barzideh:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications felerenced above and nave usements of a legally marketed predicate devices marketed in interstate for use stated in the encrobate) to regionent date of the Medical Device Amendments, or to commerce phor to may 20, 1978, are excerdance with the provisions of the Federal Food, Drug, devices mat nave been receasined trequire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, mereloro, manes of the Act include requirements for annual registration, listing of general controls provisions of wactice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassified (600 a00 rols. Existing major regulations affecting your device can Inay be subject to sach adational cegulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Mr Bahram Barzideh

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that 1271 3 lossanted or our device complies with other requirements of the Act that TDA has made a determinations administered by other Federal agencies. You must of ally Federal statutes and regulations and limited to: registration and listing (21 comply with an the Act stequirements, morealing, and manufacturing practice requirements as set CFK Part 807), falcing (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control providens (esting your device as described in your Section 510(k) I ms letter will anow you to oegin maning of substantial equivalence of your device to a legally premarket nothleadon. The PDA mailing of baction for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for Jour act (301) 594-4648. Also, please note the regulation entitled, Colliact the Office of Compilance market notification" (21CFR Part 807.97). You may obtain Misoranumly by reference to prenessibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Neil Ry Ogl for
Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K041907

Indication for Use

510(k) Number (if known): K041907

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

INDICATION FOR USE STATEMENT

The Trio™ monitor is intended for use in healthcare settings under the direct supervision of a licensed healthcare practitioner. The intended use of the monitor is to monitor physiologic parameter data on adult and pediatric patients. Physiologic data includes: electrocardiogram, invasive blood pressure, non-invasive blood pressure (NIBP), pulse oximetry, heart rate (derived from ECG, SpO2 or NIBP), respiration and temperature as summarized in the operating instructions manual. The information can be displayed, stored, trended and printed.

The monitor is not intended for home use. The monitor is not intended to be an apnea monitor. It was not designed or validated for use as an apnea monitor

AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil L. Ogden

ivision of Cardiovascular Devices

510(k) Number K041907

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).