K Number
K041786

Validate with FDA (Live)

Device Name
CALCAMEA PLATE
Manufacturer
Date Cleared
2004-09-08

(69 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CALCANEA™ -PLATE is indicated for use in fixation of fractures or osteotomies of the calcaneus.

Device Description

The CALCANEA 100 PLATE is a Titanium low profile anatomical plate dedicated for fixation of calcaneal fractures. The CALCANEATM PLATE enables a reconstruction of height and width of the calcaneus and a possible remodeling to the lateral calcaneus. Its anatomical design provides optimal bone coverage while its low profile provides a minimal irritation of soft tissues. The CALCANEA 110 PLATE is available in three different sizes (1-2-3) for optimal anatomical fit. Numerous holes enable a versatile screw fixation. The fixation is provided by Titanium screws available in two different design : locking screws and variable angle screws. The locking screws have threaded heads adapted to the threaded holes of the plate whereas the other design provides a variable angle for a versatile fixation. The range of screws include a diameter of 3.5 mm and a length from 20 mm to 45 mm (by 5 mm). The fixation screws are self tapping with a smooth tip to protect soft tissues. The screw head design enables a low profile screw/plate/bone interface. Both plates and screws present a color code for size identification. The lot number and the reference are marked for easier product traceability.

AI/ML Overview

The Calcanẽa™️-Plate is a titanium low-profile anatomical plate for calcaneal fracture fixation.

Here's an analysis of the provided information regarding its acceptance criteria and supporting study:

1. Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mechanical CharacteristicsSatisfactory. The various tests performed on the CALCANEA™ Plate were satisfactory and allowed marketing of this device.
Torsion StrengthAt least as good as competitors. The device has mechanical characteristics at least as good as those of competitors, including the Synthes Calcaneus Locking Plate (K991407), which was defined as a predicate.
Bending StrengthAt least as good as competitors. The device has mechanical characteristics at least as good as those of competitors, including the Synthes Calcaneus Locking Plate (K991407), which was defined as a predicate.
Comparison with Predicate DevicesEquivalent. The device was found to have mechanical characteristics at least as good as the Synthes Calcaneus Locking Plate (K991407) and is considered similar in design and function to other predicate devices like SYNTHES Locking Calcaneal Plates (K991407 - K020401) and DEPUY ACE Calcaneal Peri-Articular Plate (K993465 - K981775).
Intended UseSame as predicate devices. The CALCANEA™-PLATE, the Synthes Locking calcaneal plates, and the DePuy ACE calcaneal plate have the same intended use and are indicated for fixation of fractures and osteotomies of the calcaneus.
Material (Titanium Alloy TiAI6V4)Meets specification (matches DePuy ACE). Both the CALCANEA™-PLATE and DePuy ACE calcaneal plate are made of Titanium alloy TiAI6V4, while the Synthes locking calcaneal plates are manufactured in Stainless steel 316 L.
Fixation Screws CompatibilityCompatible with 3.0mm, 3.5mm, or 4.0mm screws (matches predicate devices). All systems (CALCANEA™-PLATE and predicate devices) are fixed with 3.0mm, 3.5mm or 4.0mm screws.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify a sample size for a test set in the context of human data or clinical trials. The "tests performed" refer to mechanical characteristic determinations of the device itself. Therefore, concepts like country of origin or retrospective/prospective data are not applicable here. This appears to be a pre-market notification (510(k)) based on bench testing and comparison to predicate devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to the type of study described. The ground truth for mechanical characteristics would be established by industry standards, engineering specifications, and validated testing methodologies, not by clinical experts.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods are typically used in clinical studies involving interpretation of medical data, where there's potential for inter-observer variability. The study focuses on objective mechanical properties.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, And the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This information is not applicable. This submission is for a medical device (bone plate) and not for an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to AI assistance would not be relevant.

6. If a Standalone (i.e., algorithm-only without human-in-the-loop performance) Was Done

This information is not applicable. This is a physical medical device, not an algorithm, so "standalone algorithm-only performance" is not a relevant concept.

7. The Type of Ground Truth Used

The ground truth used for the study of the Calcanẽa™️-Plate was primarily mechanical and material specifications, and comparative performance against legally marketed predicate devices. This means:

  • Mechanical Standards: Implied adherence to engineering standards for torsion strength and bending strength.
  • Predicate Device Performance: The mechanical characteristics of existing, approved devices (SYNTHES Locking Calcaneal Plates and DEPUY ACE Calcaneal Peri-Articular Plate) served as a benchmark for "at least as good as" performance.
  • Material Composition: Verification of the use of Titanium alloy TiAI6V4.
  • Design and Function Equivalency: Demonstrated similarity in intended use, design (anatomical, low profile, screw fixation), and screw compatibility.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" in the context of mechanical testing for a physical device. This term is relevant for machine learning algorithms.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Newdeal. The logo consists of the word "newdeal" in a sans-serif font, with a circle to the left of the word. The circle has four dots around it, two of which are solid black and two of which are white with black speckles. The "®" symbol is to the right of the word "deal".

SEP = 8 2004

Image /page/0/Picture/2 description: The image shows the text "K041786 page 1 of 3" in a handwritten style. The text is written in black ink on a white background. The text is slightly slanted to the right.

10, PLACE D'HELVETIE · 69006 LYON · FRANCE t +33 (0)4 37 47 51 51 ● t +33 (0)4 37 47 51 52 newdeal@newdeal.infowww.newdeal.info

SUMMARY OF SAFETY AND EFFECTIVENESS 3.

  • SPONSOR IDENTIFICATION: A. NewDeal SA 10, place d'Helvétie 69006 LYON FRANCE
    Tél. : +33 4 37 47 51 51 Fax : +33 4 37 47 51 52

ESTABLISHMENT REGISTRATION NUMBER: 9615741 B.

  • OFFICIAL CONTACT PERSON C. Norman F. Estrin, Ph. D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac . MD 20854
    Tel. : (301) 279 -2899 Fax : (301) 294-0126 estrin@yourFDAconsultant.com

CALCANEA™-PLATE PROPRIETARY (TRADE) NAME: D.

  • Plate for calcaneal fracture COMMON NAME: E.

CLASSIFICATION NAME AND REFERENCE F.

Plate, Fixation, Bone (21 CFR, Section 888. 3030)

  • PROPOSED REGULATORY CLASS: Class II G.
  • DEVICE PRODUCT CODE: HRS H.
  • 87 OR Orthopedic PANEL CODE: I.

0 0007

{1}------------------------------------------------

K041786
page 2 of 3

J. DESCRIPTION OF DEVICE:

The CALCANEA 100 PLATE is a Titanium low profile anatomical plate dedicated for fixation of calcaneal fractures.

The CALCANEATM PLATE enables a reconstruction of height and width of the calcaneus and a possible remodeling to the lateral calcaneus.

Its anatomical design provides optimal bone coverage while its low profile provides a minimal irritation of soft tissues.

The CALCANEA 110 PLATE is available in three different sizes (1-2-3) for optimal anatomical fit. Numerous holes enable a versatile screw fixation.

The fixation is provided by Titanium screws available in two different design : locking screws and variable angle screws.

The locking screws have threaded heads adapted to the threaded holes of the plate whereas the other design provides a variable angle for a versatile fixation.

The range of screws include a diameter of 3.5 mm and a length from 20 mm to 45 mm (by 5 mm). The fixation screws are self tapping with a smooth tip to protect soft tissues. The screw head design enables a low profile screw/plate/bone interface.

Both plates and screws present a color code for size identification. The lot number and the reference are marked for easier product traceability.

  • The CALCANEA™-PLATE is indicated for use in K. INDICATIONS FOR USE: fixation of fractures or osteotomies of the calcaneus.
  • The CALCANEATM-PLATE is similar in design and L. PREDICATE DEVICE: function to the predicates SYNTHES Locking Calcaneal Plates (K991407 - K020401) and the DEPUY ACE Calcaneal Peri-Articular Plate (K993465 - K981775). Based on the information provided in this premarket notification, Newdeal considers the subject devices to be equivalent to the existing DePuy Ace and Synthes calcaneal plates.

M. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:

CALCANEATM-PLATE, the Synthes The Locking calcaneal plates and the DePuy ACE calcaneal plate have same intended of use and are indicated for fixation of

11 11008

{2}------------------------------------------------

kost 1786 fractures and osteotomies of the calcaneus. Page 303

fractures and osteotomies of the calcaneus.

of CALCANEA™-PLATE and DePuy ACE Both calcaneal plate in Titanium alloy TIAI6V4 whereas the Synthes locking calcaneal plates are manufactured in Stainless steel 316 L.

-All those systems are fixed with 3.0mm, 3.5mm or 4.0mm screws.

N. SUMMARY Determination of the mechanical characteristics (torsion OF STUDIES: strength, bending strength, comparison with a predicate) of the CALCANEATM -PLATE was conducted.

The various tests performed on the CALCANEA™ Plate were satisfactory and allow marketing this device, which has mechanical characteristics at least as good as the ones of the competitors, among which the Synthes Calcaneus Locking Plate (K991407) defined as a predicate.

June 28, 2004 DATE OF PREPARATION OF THIS SUMMARY:

  1. 1.000

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is the department's logo, which is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP = 8 2004

NewDeal S.A. C/o Norma F. Estrin, PhD, RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854

Re: K041786 Trade/Device Name: CALCANEA™ - PLATE Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: June 28, 2004 Received: July 6, 2004

Dear Dr. Estrin:

We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to conninered process in they 20, 2017) in accordance with the provisions of the Federal Food, Drug, de vices that have been i that do not require approval of a premarket approval application (PMA). and Cosmetic rear (110) that 80 hevice, subject to the general controls provisions of the Act. The r ou may, dicrore, mains of the Act include requirements for annual registration, listing of general ocultive profitive practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device is elassinod (tional controls. Existing major regulations affecting your device can may or subject to basil at Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I Dr Has Intacted and regulations administered by other Federal agencies. You must or uny I out all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality by router (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Norma F. Estrin, PhD, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter will and in your of substantial equivalence of your device of your device to a legally premated predicated device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at noliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
Mark N Millman

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number (if known):

K04/784

CALCANEA™ -PLATE Device Name: Indications for Use:

The CALCANEA™ -PLATE is indicated for use in fixation of fractures or osteotomies of the calcaneus.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-the-Counter Use

(Optional Format 1-2-96)

Mark A. Millerson
Division Sign Off

Division of General, Restorative, and Neurological Devices

.............................................................................................................................................................................. 510(k) Number_

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.