K Number
K041778

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-08-31

(61 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device measures systolic and diastolic blood pressures and pulse rate. It uses the oscillometric method. It is indicated for use by adults (those 18 years of age and older) at home or in an outpatient setting.

Device Description

The Reizen Talking Blood Pressure Monitor SF860T measures systolic and diastolic blood pressures and pulse rate. It uses the oscillometric method. It is indicated for use by adults (those 18 years of age and older) at home or in another outpatient setting. It has the same technological characteristics as the predicate device.

AI/ML Overview

The Reizen Talking Blood Pressure Monitor, Model SF860T, met its acceptance criteria by demonstrating conformance to established international and national standards for blood pressure measurement devices.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria StandardReported Device Performance
ISO 9001/ISO 13485 Quality System StandardCertified as meeting the standards
European Union Medical Devices Directive (93/42/EEC)Certified as complying with applicable requirements
ANSI/AAMI SP10 StandardPassed testing to show conformance

2. Sample Size and Data Provenance:
The provided 510(k) summary does not specify the exact sample size used for testing against the ANSI/AAMI SP10 standard, nor does it detail the data provenance (e.g., country of origin, retrospective/prospective). The other certifications (ISO, EU MDD) relate to quality management systems and general compliance, not specific device performance on a human data set for accuracy.

3. Number of Experts and Qualifications:
The summary does not mention the use of experts to establish ground truth for testing against the ANSI/AAMI SP10 standard. Conformance to this standard typically involves technical testing and, if clinical validation is performed, it would involve comparison to reference methods (e.g., auscultatory measurements by trained observers), but the role and number of "experts" in the context of clinical ground truth establishment are not detailed here.

4. Adjudication Method:
The adjudication method for any clinical testing against the ANSI/AAMI SP10 standard is not specified in the provided document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
A multi-reader multi-case (MRMC) comparative effectiveness study was not performed or reported in this 510(k) summary. This type of study is typically associated with AI-powered diagnostic devices, where the effect of AI assistance on human reader performance is evaluated. The Reizen Talking Blood Pressure Monitor is a stand-alone measurement device, not an AI-assisted diagnostic tool for human readers.

6. Standalone Performance:
Yes, a standalone performance assessment was conducted, as indicated by the device passing testing to show conformance with the ANSI/AAMI SP10 Standard. This standard specifically addresses the performance requirements for automated sphygmomanometers, implying that the algorithm's (oscillometric method) accuracy was tested independently.

7. Type of Ground Truth Used:
For the ANSI/AAMI SP10 standard, the "ground truth" for blood pressure measurements is typically established through a reference measurement method. Historically, this has involved auscultation by trained observers using a mercury sphygmomanometer (or a validated aneroid manometer) in accordance with standardized protocols. The device's readings would then be compared against these reference measurements.

8. Sample Size for the Training Set:
The provided document does not mention a training set sample size. This is a conventional medical device whose core technology (oscillometric method) is well-established, rather than a machine learning or AI device that requires a distinct "training set" for model development.

9. How Ground Truth for Training Set was Established:
As there is no mention of a training set for a machine learning model, the concept of establishing ground truth for a training set is not applicable in the context of this 510(k) summary. The oscillometric algorithm itself is developed based on established physiological principles and extensive validation over time across the medical device industry.

{0}------------------------------------------------

AUG 3 1 2004

510(k) Summary

Sponsor Information:

Maxi-Aids 42 Executive Boulevard Farmingdale, NY 11735 Elliot Zaretsky Phone (631) 752-0521 Facsimile (631) 752-0689

Date Prepared:

June 25, 2004

Device Name:

Proprietary Name: Reizen Talking Blood Pressure Monitor, Model SF860T

Common/Usual Name: Blood pressure monitor

Classification Name: Noninvasive blood pressure measurement system (per 21 C.F.R. § 870.1130)

Device Description, Intended Use, and Technological Characteristics:

The Reizen Talking Blood Pressure Monitor SF860T measures systolic and diastolic blood pressures and pulse rate. It uses the oscillometric method. It is indicated for use by adults (those 18 years of age and older) at home or in another outpatient setting. It has the same technological characteristics as the predicate device.

Predicate Device:

A&D Medical LifeSource Talking Digital Blood Pressure Monitor, Model UA-7671'

Performance Data:

The device has been certified as meeting the standards of the ISO 9001/ISO 13485 Quality System Standard. Further, it has been certified as complying with the applicable requirements of the European Union Medical Devices Directive (93/42/EEC). The device also passed testing to show conformance with the ANSVAAMI SP10 Standard.

{1}------------------------------------------------

Public Health Service

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three horizontal lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 3 1 2004

Maxi Aids, Inc. c/o Mr. Wayne H. Matelski 1050 Connecticut Avenue Washington, DC 20036-5339

Re: K041778

K041778
Trade Name: Reitzen Talking Blood Pressure Monitor, Model SF860T Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: June 28, 2004 Received: July 1, 2004

Dear Mr. Matelski :

We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your Section 310(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regary manated povice Amendments, or to commerce prof to May 28, 1976, the encenture with the provisions of the Federal Food. Drug, devices that have been icclassified in acceraalso was a proval approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval applica and Cosment Act (Act) that do not require approvinces in controls provisions of the Act. The Y ou may, merelore, market the devices, subject to annual registrator, listing of
general controls provisions of the Act include requirements for annual registration, disco. general controls provisions of the first labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classifica (seculored in the exist is a major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional controller Entroller Controller of the 898. In addition, FDA may be found in the Oous overning your device in the Federal Register.

{2}------------------------------------------------

Page 2 – Mr. Wayne H. Matelski

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that IDA s issualice of a subcements with other requirements of the Act
that FDA has made a determination that your device complies with of the Act that FDA has made a determination alat your as recess by other Federal agencies. You must a or any Federal statules and regulations daminities or registration and listing (21)
comply with all the Act's requirements, including, but not limited to: registration resear comply with all the Act s requirements, more of any cannon of the requirements as setting the closetter CFK Part 807); labeling (21 CFR Part 800); government 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 11 am 1000 1 forin in the quality systems (QD) regalication (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Doctors 597 - vice as described in your Section 510(k)
This letter will allow you to begin marketing your device to logics to 100) I his letter will anow you to ocgin marketing your antial equivalence of your device to a legally premarket nother device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire specific advice for your ac 110 %. Also, please note the regulation entitled, contact the Office of Complians as (Set ) , (21CFR Part 807.97). You may obtain " Misoranding by reserved to premainter to be act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Goress http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Neil R. Opler

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known): K041778

Device Name: Reizen Talking Blood Pressure Monitor, Model SF860T

Indications For Use:

The device measures systolic and diastolic blood pressures and pulse rate. It uses the i he device measures systems and diacters brose prosessor age and older) at home or in an outpatient setting.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

× Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R.P. Ogden

(Division Sign-Off)Division of Cardiovascular Devices
-----------------------------------------------------------

510(k) Number,

Page 1 of 1

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).