K Number
K041380

Validate with FDA (Live)

Date Cleared
2004-06-08

(14 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ImageChecker CT Software Package with Filling Defect Indicator (FDI) is used during the review of contrast-enhanced computed tomography (CT) images of the chest. This software tool enables the radiologist to view and analyze regions of the image containing low density within vascular structures that may be indicative of filling defects or other intravascular abnormalities.

The software is designed to assist the radiologist in characterization and classification of these suspicious candidate thoracic abnormalities in terms of density, size, dimension, shape and position, thus aiding in the patient management care decision process.

Device Description

The ImageChecker CT Software Package with Filling Defect Indicator (FDI), the subject of this notification, is a thoracic image visualization and analysis tool. This tool offers an extended functionality of the ImageChecker CT Workstation, utilizing segmentation techniques similar to the two above-mentioned predicates. Based on physician request, the FDI tool automatically segments locations in the vasculature of the chest containing low densities that typically correlate with filling defects. The process is based on simple segmentation methodology applied to contrast-opacified CT images of the chest segmentation methouorogy appels below a certain HU (Hounsfield Unit) threshold. fuentifying region is highlighted by placing a triangle on the region of the vascular structure of interest. This extended software functionality is designed for use with the structure of interesent on the R2 ImageChecker CT Workstation (K023003) and on other CT workstations that utilize the ImageChecker CT Workstation software tools. The FDI tool contains modifications to the Workstation with a special workflow based on automated segmentation for the visual identification of possible lung filling defects.

AI/ML Overview

Here's an analysis of the provided text regarding the ImageChecker CT Software Package with Filling Defect Indicator:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) premarket notification does not contain any specific acceptance criteria or quantitative performance data on the device's ability to detect filling defects. The document primarily focuses on demonstrating substantial equivalence to predicate devices based on intended use and technological similarities.

Instead, the document states: "The Filling Defect Indicator tools package will undergo design verification tests for conformance with specifications." This indicates that the performance data was likely to be generated after this document was submitted or in a separate report, and not included in this summary.

2. Sample Size Used for the Test Set and Data Provenance

No information is provided regarding the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective). This information would typically be found in a study report, which is not present in this 510(k) summary.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

No information is provided regarding the number of experts, their qualifications, or how ground truth was established for a test set.

4. Adjudication Method for the Test Set

No information is provided regarding any adjudication method (e.g., 2+1, 3+1, none) for a test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is made. The document does not describe any study comparing human readers with AI assistance versus human readers without AI assistance, nor does it provide any effect sizes.

6. Standalone Performance Study

No information is provided regarding a standalone (algorithm only) performance study. The 510(k) states the device is "an adjunctive device that assists the radiologist in the review of CT images of the chest." This implies it's intended for human-in-the-loop use, but no standalone performance data is presented.

7. Type of Ground Truth Used

No information is provided regarding the type of ground truth used (e.g., expert consensus, pathology, outcomes data).

8. Sample Size for the Training Set

No information is provided regarding the sample size for any training set.

9. How the Ground Truth for the Training Set Was Established

No information is provided on how ground truth was established for any training set.

Summary of the Information Provided in the 510(k):

The provided 510(k) Premarket Notification is a regulatory document focused on demonstrating substantial equivalence to predicate devices. It describes the device's intended use ("adjunctive device that assists the radiologist in the review of CT images of the chest"), its technical functionality (segmentation based on Hounsfield Unit threshold to identify low-density regions), and its relationship to existing R2 Technology and Siemens products.

Crucially, it does not include the details of any performance studies, acceptance criteria, or clinical validation data that would typically be found in a separate study report or a more comprehensive submission. The statement "The Filling Defect Indicator tools package will undergo design verification tests for conformance with specifications" indicates that performance testing was planned or ongoing, but the results are not present in this document. Therefore, based solely on the provided text, the specific information requested about acceptance criteria and study details cannot be extracted.

{0}------------------------------------------------

KO 41380

D2 Technology Image necker C I Software Package with Filling Defect Indicator 510(k) Premarket Notification April, 2004

JUN - 8 2004

510(k) Summary of Safety and Effectiveness Prepared April 16, 2004

Submitted by:R2 Technology, Inc.1195 W. Fremont AvenueSunnyvale, CA 94087
Contact Person:Kathy O'ShaughnessyVice President, Regulatory Affairs
Product Name:ImageChecker CT Software Package with Filling DefectIndicator
Common Name:Medical Image Processing Software
Classification Panel:Radiology
Classification:Class II (CFR 21 892.2050)
Product Code:LLZ

Predicate Devices:

ProductCompany510(k) Number
LungCARE CTsoftware package withextended functionalitySiemens MedicalSolutionsK033374
ImageChecker CTWorkstationR2 Technology, Inc.K023003

Device Description:

The ImageChecker CT Software Package with Filling Defect Indicator (FDI), the subject of this notification, is a thoracic image visualization and analysis tool. This tool offers an extended functionality of the ImageChecker CT Workstation, utilizing segmentation techniques similar to the two above-mentioned predicates. Based on physician request, the FDI tool automatically segments locations in the vasculature of the chest containing

{1}------------------------------------------------

R2 Technology ImageChecker CT Software Package with Hilling Defect Indicator 510(k) Premarket Notification April, 2004

low densities that typically correlate with filling defects. The process is based on simple segmentation methodology applied to contrast-opacified CT images of the chest segmentation methouorogy appels below a certain HU (Hounsfield Unit) threshold. fuentifying region is highlighted by placing a triangle on the region of the vascular structure of interest. This extended software functionality is designed for use with the structure of interesent on the R2 ImageChecker CT Workstation (K023003) and on other CT workstations that utilize the ImageChecker CT Workstation software tools. The FDI tool contains modifications to the Workstation with a special workflow based on automated segmentation for the visual identification of possible lung filling defects.

Summary of Intended Use:

The ImageChecker CT Software Package with Filling Defect Indicator is intended to be used to enable the radiologist to view and analyze regions of the image containing low density within vascular structures that may be indicative of filling defects or other density within vasoular bes. This segmentation tool is an extended functionality of the R2 ImageChecker CT Workstation software, and can be used on that self-standing device as well as other imaging workstations that utilize the ImageChecker CT Workstation woftware. This tool is intended for use as an adjunctive device that assists the radiologist in the review of CT images of the chest.

Comparison with Predicate Devices:

R2 Technology's ImageChecker CT Software Package with Filling Defect Indicator (FDI) software tool, Siemens LungCARE CT software package with extended functionality, and the R2 ImageChecker CT Workstation enable radiologists to view and analyze areas of CT exams. These software packages allow the radiologist to examine segmented candidate regions of interest in the CT volume data sets. The three software packages also enable the radiologist to document and follow up any identified segmented areas of interest.

The Siemens LungCARE CT Software Package with Extended Functionality is a self contained software package containing the specifically developed function of supporting a special workflow based on automated segmentation for the visual identification of possible lesions (Nodule Enhanced Viewing). Likewise, R2's ImageChecker CT Software Package with Filling Defect Indicator is built on the base ImageChecker CT Workstation software with a special workflow based on automated segmentation for the visual identification of possible filling defects of the lung vasculature (FDI).

Studies:

The Filling Defect Indicator tools package will undergo design verification tests for conformance with specifications.

{2}------------------------------------------------

R2 Technology Kz Technology
ImageChecker CT Software Package with Filling Defect Indicator 510(k) Premarket Notification April, 2004

Conclusion:

The ImageChecker CT Software Package with Filling Defect Indicator has the same intended use as the two predicate devices identified in this section. Any technological niterences in the ImageChecker CT Software Package with Filling Defect Indicator and the predicate devices do not raise any new questions regarding safety or effectiveness. thus, R2 Technology's ImageChecker CT Filling Defect Indicator (FDI) software tool is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized caduceus, a symbol often associated with medicine and healthcare.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 1 2009

R2 Technologies, Inc. % Ms. Denise Leung Klinker Reviewer, Medical Device Services Underwriters Laboratories. Inc. 1655 Scott Boulevard SANTA CLARA CA 95050-4169

Re: K041380

Trade/Device Name: ImageChecker CT Software Package with Filling Defect Indicator Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: OEB Dated: May 24, 2004 Received: May 25, 2004

Dear Ms. Klinker:

This letter corrects our substantially equivalent letter of June 8, 2004.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801: good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely vours.

Huld Reemen

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

KO 4138.0 510(k) Number

Device Name: ImageChecker CT Software Package with Filling Defect Indicator

Indications for Use:

The ImageChecker CT Software Package with Filling Defect Indicator (FDI) is used during the review of contrast-enhanced computed tomography (CT) images of the chest. This software tool enables the radiologist to view and analyze regions of the image containing low density within vascular structures that may be indicative of filling defects or other intravascular abnormalities.

The software is designed to assist the radiologist in characterization and classification of these suspicious candidate thoracic abnormalities in terms of density, size, dimension, shape and position, thus aiding in the patient management care decision process.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel A. Lyons

Division Sign-Off) Division of Rentoductive, Abdor Jurimin : 31 Device 51 Jk) Number -

Page 1 of 1

3

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).