K Number
K041296

Validate with FDA (Live)

Date Cleared
2004-06-10

(27 days)

Product Code
Regulation Number
882.5550
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HAKIM™ Precision Valve System: The Nonprogrammable Valve System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
HAKIM™ (and HAKIM™ Micro) Programmable Valve Shunt System: The HAKIM™ Programmable Valve Shunt System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
SiphonGuard™ CSF Control Device: The SiphonGuard™ device can be used as a component of hydrocephalus shunt systems designed to shunt CSF from the lateral ventricles of the brain into the peritoneal cavity or right atrium of the heart. The SiphonGuard™ device is designed to reduce the potential hazards of excessive lowering of intraventricular pressure (with respect to atmospheric pressure) when a patient is in an erect position.

Device Description

The devices that are the subject of this Special 510(k): Device Modification are identical to their respective predicate predecessors, with the exception of one material component. The proposed alternative epoxy is the single difference in the devices.

AI/ML Overview

The provided text is a 510(k) summary for a medical device modification, specifically regarding the CODMAN HAKIM™ Precision, Programmable, Micro Programmable Valve Systems and the SiphonGuard™ CSF Control Device. The core of this submission is a material change (alternate epoxy) to existing devices.

Therefore, the study performed is not a typical performance study comparing a device's diagnostic or therapeutic efficacy against a ground truth. Instead, it focuses on demonstrating that the new material does not negatively impact the device's established performance.

Here's an analysis based on the information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Maintain the established performance and safety of the predicate devices for their intended use despite the material change (alternate epoxy)."This submission relied upon appropriate bench and biocompatibility testing necessary to support the device for its intended use." (Implies successful completion of these tests, demonstrating that the new epoxy does not compromise device performance or biocompatibility.)

Explanation: The acceptance criteria here are implicitly that the modified devices (with the new epoxy) demonstrate equivalence to the predicate devices in terms of performance and safety, as assessed through bench and biocompatibility testing. The "reported device performance" is the statement that these tests were performed and were "appropriate... to support the device for its intended use," which signifies that the modifications meet the safety and effectiveness requirements.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size for the test set used in the bench and biocompatibility testing. The data provenance is not explicitly stated. Given that it's a 510(k) submission for a material modification, these would typically be retrospective tests conducted by the manufacturer in a laboratory setting (bench testing) and possibly with in vitro or animal studies for biocompatibility, rather than human clinical trials with geographical data provenance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. For a material change and bench/biocompatibility testing, the "ground truth" is typically defined by established engineering specifications, material science standards, and biocompatibility guidelines. These are assessed by engineers, material scientists, and toxicologists, rather than medical experts establishing a diagnostic ground truth. The document does not specify the number or qualifications of these experts.

4. Adjudication Method for the Test Set

Not applicable. This type of submission does not involve adjudication of expert readings or diagnoses. The "adjudication" would be internal technical review and approval by the manufacturer's quality and regulatory teams to ensure the test results meet the defined criteria, followed by FDA review of the submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data, with and without AI assistance, to assess the AI's impact on diagnostic accuracy. This submission concerns a material change in existing Shunt Systems and a CSF Control Device, which are implantable mechanical devices.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. There is no algorithm or AI component in these mechanical devices.

7. The Type of Ground Truth Used

The ground truth for this type of submission (material change) would be based on:

  • Engineering specifications and performance standards: Ensuring the modified device meets predefined mechanical, flow, and durability specifications.
  • Biocompatibility standards: Demonstrating that the new epoxy material is non-toxic and compatible with the human body as per recognized international standards (e.g., ISO 10993 series).

8. The Sample Size for the Training Set

Not applicable. There is no AI model involved that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no AI model, there is no training set or associated ground truth.

{0}------------------------------------------------

Codman & Shurtleff, Inc. Special 510(k): Device Mouilication Special 310(K). Device Moutheaton
CODMAN® HAKIM™ Precision, Programmable, Micro Programmable Valve Systems; and the SiphonGuard™ CSF Control Device

Page 1 of 2

K041296

(JUN 1 0 2004

510(k) Summary

  • Codman and Shurtleff, Inc. Submitter: 325 Paramount Drive Raynham, MA 02780 Kathryn Wunder Contact Person: Phone Number: (508) 880-8351 Fax Number: (508) 828-3212 May 13, 2004 Date Prepared: Classification Name: Central Nervous System Fluid Shunt and Components Item 1: HAKIM™ Precision Valve System Proprietary Name: Item 2: HAKIM™ Programmable Valve System Item 3: HAKIM™ Micro Programmable Valve System Item 4: SiphonGuardTM CSF Control Device Item 1: HAKIM™ Precision Valve System Predicate Device: Item 2: HAKIM™ Programmable Valve System Item 3: HAKIM™ Micro Programmable Valve System Item 4: SiphonGuard™ CSF Control Device HAKIM™ Precision Valve System: Intended Use: The Nonprogrammable Valve System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. HAKIM™ (and HAKIM™ Micro) Programmable Valve Shunt System: The HAKIM™ Programmable Valve Shunt System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. SiphonGuard™ CSF Control Device: The SiphonGuard™ device can be used as a component of hydrocephalus shunt systems designed to shunt CSF from the lateral ventricles of the brain into the peritoneal cavity or right atrium of the heart.

{1}------------------------------------------------

The SiphonGuard™ device is designed to reduce the potentialhazards of excessive lowering of intraventricular pressure (withrespect to atmospheric pressure) when a patient is in an erectposition.
Materials:All materials remain the same for the indicated devices, with theexception of an alternate epoxy. The current epoxy is Stycast1267. The proposed alternative is Loctite M-31CL.
DeviceDescription:The devices that are the subject of this Special 510(k): DeviceModification are identical to their respective predicatepredecessors, with the exception of one material component. Theproposed alternative epoxy is the single difference in the devices.
PerformanceData:This submission relied upon appropriate bench andbiocompatibility testing necessary to support the device for itsintended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines that resemble a person with outstretched arms. The seal is black and white and appears to be a logo or emblem.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

'JUN 1 0 2004

Ms. Kathryn Wunder Regulatory Affairs Codman & Shurtleff, Inc. 325 Paramount Drive Raynham, Massachusetts 02767

Re: K041296

K041296
Trade/Device Name: HAKIM™ Precision Valve Shunt System
Trade/Device Name: HAKIM™ 20075550 Regulation Number: 21 CFR 882.5550 Regulation Number. 21 CFR 862.9596
Regulation Name: Central nervous system fluid shunt and components Regulatory Class: II Product Code: JXG Dated: May 13, 2004 Received: May 14, 2004

Dear Ms. Wunder:

We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section >10(t) prematic is substantially equivalent (for the indications
referenced above and have determined the device marketed in interstate referenced above and have determined the devices marketed in interstate for use stated in the enclosure) to regally marked the of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment of the Federal Food Drug commerce prior to May 28, 1970, the enaculier and other of the Federal Food, Drug, devices that have been reclassified in accordance was are askets on oval application (PMA).
and Cosmetic Act (Act) that do not requires and control controlsions of the Act. and Cosmetic Act (Act) that do not require approvations provisions of the Act. The
You may, therefore, market the device, subject to the general constructions of the Act. You may, therefore, market the device, subject to the gentlal registration, listing of
general controls provisions of the Act include requirements for annual registranding an general controls provisions of the Act mender requirements of the manufacturing and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) more controls it (spealations affecting your device can
may be subject to such additional controls. Existing major regulations affec may be subject to such additional controis. Existing major toganto and on addion, FDA may
be found in the Code of Federal-Regulations, Title 21, Parts 800 to 898 - 11 additi be found in the Code of rederaritegalations, Friday, Friday, In the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I-DA's issuance of a subscribe only of the requirements of the Act
that FDA has made a determination that your device also also annones. You must that FDA has made a determilation his your arres on other Fecderal agencies. You must
or any Federal statutes and regulations administered by registration and listin or any Federal statutes and regulations administered on registration and listing (21
comply with all the Act's requirements, including, but times assets requirements asst comply with all the Act s requirements, including, out manufacturing practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice the e CFR Part 807); labeling (21 CFR Patt 801); good mananation: and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 2001); 1000-1050 forth in the quality systems (QS) regulanon (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1

{3}------------------------------------------------

Page 2 - Ms. Kathryn Wunder

This letter will allow you to begin marketing your device as described in your Section 510(k)
10 Alegal This letter will allow you to begin marketing your active of your device of your device of your device to a legally
premarket notification. The FDA Innding of substantial eq premarket notification. The FDA inding of substantial equive and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (1 CFR Part 801), please
r the regulation and 1104 - 1001 - 1650 - Also, please note the regulation e If you desire specific advice for your device on our accessing regulation entitled,
contact the Office of Compliance at (301) 594-4659. Also, please note the Office of contact the Office of Compliance at (301) 554-407. The Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). Your may oftai "Misbranding by reference to premarkst nonitication " (2) User from the Division of Small
other general information on your responsibilities under (800) 638-204 other general information on your responsibilities and its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance as and orders and human Manufacturers, International and Consumer 71355stance as no 1
(301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

L. Merle A. Milliken

L. Merle Milliken, Ph.D., M.D.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

Device Name:

Indications For Use:

K041296

SiphonGuard™ CSF Control Device

The SiphonGuard™ device can be used as a The Siphon Oudrocephalus shunt systems designed component of myaross lateral ventricles of the brain to shunt OOF front of the right atrium of the heart.

The SiphonGuard™ device is designed to reduce the me olphonomanas of excessive lowering of potential nazarde of exect to atmospheric intraventhoulan a patient is in an upright position.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Office of Device Evaluation (ODE) Concurrence of CDRH

Mark A M

Page 1 of 1

(Division Sign-On)
Division of General, Restorative,
and Neurological Devices

Re

510(k) Number K041296

{5}------------------------------------------------

K041296

510(k) Number (if known):

Device Name:

Indications For Use:

HAKIM™ Precision Valve Shunt System

The HAKIM™ Precision Valve Shunt System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Qffice of Device Evaluation (ODE) Concurrence of CDRH,

Mark A. Millheiser

Page 1 of 1

Sign-Off) (Division (Division of General, Restorative, Division of Generalive, Division of General Division
and Neurological Devices

510(k) Number

{6}------------------------------------------------

K041296

510(k) Number (if known):

Device Name:

Indications For Use:

HAKIM™ Micro Programmable Valve Shunt System

The HAKIM™ Micro Programmable Valve Shunt System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Qffice of Device Evaluation (ODE) Concurrence of CDRH,

Page 1 of 1

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

K04/296

510(k) Number

{7}------------------------------------------------

510(k) Number (if known):

Device Name:

Indications For Use:

K041296

HAKIM™ Programmable Valve Shunt System

The HAKIM™ Programmable Valve Shunt System is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Office of Device Evaluation (ODE) Concurrence of CDRH,

Mark A. Wilkerson

Page 1 of 1 __

Sign-Off) (Division (Division Sign Oct)
Division of General, Restorative, Division of General Division of Schoological Devices

K041296

510(k) Number

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).