K Number
K041164

Validate with FDA (Live)

Device Name
L-TENS
Date Cleared
2004-05-20

(17 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The L-TENS is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

Device Description

The L-TENS is a handheld battery powered TENS device, which is used for pain relief. The device would generate electrical pulses and transmit it to the electrodes, which are attached to the patient's skin. Consequently, the electrical pulses would then pass through the skin to the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain. L-TENS has one output channel and four preset programs. The program mode is displayed on a LCD. The user can adjust the output intensity by 14 steps.

AI/ML Overview

This is a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) device called "L-TENS". This type of device, used for pain relief, is typically cleared based on substantial equivalence to a predicate device rather than extensive clinical studies with specific acceptance criteria that demonstrate performance against a 'ground truth' in the way AI/ML devices often are.

Therefore, the requested information elements related to acceptance criteria, ground truth, expert adjudication, multi-reader multi-case studies, and standalone performance for AI/ML devices are not applicable to this 510(k) submission.

Here's a breakdown of why and what information is available:

1. Table of Acceptance Criteria and Reported Device Performance:

  • Not Applicable. For a TENS device seeking 510(k) clearance, the "acceptance criteria" are primarily based on demonstrating substantial equivalence to a legally marketed predicate device (FDTENS 2010, K994266) regarding its intended use and fundamental technological characteristics, along with compliance to relevant non-clinical standards. The device is not making claims that require specific performance metrics (like sensitivity, specificity, or AUC) against a clinical ground truth for a diagnostic or AI/ML algorithm.

2. Sample size used for the test set and the data provenance:

  • Not Applicable. There was no "test set" in the context of clinical data for performance evaluation as would be done for an AI/ML device. The clearance is based on technological comparison and compliance with non-clinical standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. No ground truth was established for a clinical test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. No clinical test set requiring adjudication was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI/ML device, and no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. The device did not rely on a clinical "ground truth" for its 510(k) clearance, but rather on demonstrating equivalence to an existing device and meeting safety standards through non-clinical testing.

8. The sample size for the training set:

  • Not Applicable. This is not an AI/ML device, so there was no "training set."

9. How the ground truth for the training set was established:

  • Not Applicable. There was no training set or associated ground truth.

Summary of available information from the document:

  • Device Name: L-TENS
  • Predicate Device: FDTENS 2010 (K994266)
  • Intended Use: Symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
  • Technological Comparison: Substantially equivalent to the predicate. Both are battery-powered with adjustable output. The L-TENS has one channel (predicate has dual) and an added "open-circuit detection feature" to prevent output increment in the absence of load. Both comply with "shrouded patient cable connectors" rule.
  • Non-clinical Testing:
    • Compliance to voluntary standards: ANSVAAMI NS4-1986, EN 60601-1, EN 60601-1-1, and EN 60601-1-2.
    • Software verification carried out according to FDA software guidance.
  • Clinical Testing: "Not Applicable as there are no new or innovative aspects that have been introduced."
  • Conclusion: The device is substantially equivalent to the predicate and poses no new questions of safety and effectiveness.

{0}------------------------------------------------

MAY 2 0 2004

10 510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________

10.1 Submitter's Identification

Fuji Dynamics Ltd. Unit 1-3, 23/F., Laws Commercial Plaza, 788 Cheung Sha Wan Road, Kowloon, Hong Kong Tel: (852) 2786 4218 Fax: (852) 2744 6775

Contact Person: Anthony Ah Yin, Shum April 29th 2004 Date Prepared:

10.2 Name of Device:

Proprietary Name:L-TENS
Common or Usual Name:TENS unit
Classification Name:Stimulator, Nerve, Transcutaneous, for Pain Relief(21 CFR 882.5890)
Device Classification:Class II

10.3 Predicate Device Information:

The L-TENS is equivalent to the FDTENS 2010 (K994266).

10.4 Device Description:

The L-TENS is a handheld battery powered TENS device, which is used for pain relief. The device would generate electrical pulses and transmit it to the electrodes, which are attached to the patient's skin. Consequently, the electrical pulses would then pass through the skin to the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain.

L-TENS has one output channel and four preset programs. The program mode is displayed on a LCD. The user can adjust the output intensity by 14 steps.

{1}------------------------------------------------

10.5 Intended Use:

TENS is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

10.6 Technological Comparison to Predicate Device:

The L-TENS has basic technological characteristics that are substantially equivalent to the predicate device. Both devices are battery powered and have adjustable output amplitudes. On the contrary, the legally marketed predicate device has dual channels, while L-TENS is a 1-channel device. The only significant technological difference between the two devices is that L-TENS possesses an open-circuit detection feature. It means that L-TENS could check the continuity between the output terminals, and avoid increment of output in the absence of load.

All units use "shrouded patient cable connectors" to comply with the FDA's Final Rule "Mcdical Devices: Establishment of Performance Standards for Electrode Lead Wires and Patient Cables."

10.7 Non-clinical Testing:

Compliance to applicable voluntary standards includes ANSVAAMI NS4-1986, as well as EN 60601-1, EN 60601-1-1 and EN 60601-1-2 requirements.

In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.

10.8 Clinical Testing

Not Applicable as there are no new or innovative aspects that have been introduced.

10.9 Conclusions:

The L-TENS has the same intended use and similar technical characteristics as the FDTENS 2010 (K994266).

The information supplied in this 510(k) illustrates that the device do not pose any new questions of safety and effectiveness. Therefore, the L-TENS is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 0 2004

Anthony Ah Yin, Shum R & D Department Unit 1-3, Laws Commercial Plaza 788 Cheung Sha Wan Road Hong Kong. China

Re: K041164

Trade/Device Name: L-TENS Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator Regulatory Class: Class II Product Code: GZJ Dated: April 30, 2004 Received: May 3, 2004

Dear Mr. Shum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be adviscd that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Anthony Ah Yin, Shum

This letter will allow you to begin marketing your device as described in your Section 510(k) I mo lotest notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilitics under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
Mark A. Milburn

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: L-TENS

Indications For Use:

The L-TENS is used for the symptomatic rclief and management of chronic intractable pain The 1.- I ENS is used in the symptomation of post-surgical and post-surgical and post-traumatic pain.

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation
for Mark
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

510(k) NumberK091169
------------------------

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).