K Number
K041068

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-10-20

(177 days)

Product Code
Regulation Number
866.3940
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative presumptive detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis. Positive results must be confirmed by plaque reduction neutralization test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagnosis of this disease.

Device Description

The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the presumptive laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis.

AI/ML Overview

Acceptance Criteria and Device Performance for PANBIO West Nile Virus IgG Indirect ELISA

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are not explicitly stated as numerical targets in the provided text (e.g., "sensitivity must be > X%"). Rather, the studies aim to demonstrate adequate performance through descriptive statistical measures (sensitivity, specificity, positive and negative presumptive agreement) and reproducibility. The performance is reported with 95% Confidence Intervals. The cross-reactivity study aims to show the device's analytical specificity.

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Study Site 1 (Louisiana, USA)
Negative Presumptive Agreement (Endemic Normal)Demonstrate high agreement with endemic normal specimens not known to have flavivirus related illness.90.5% (181/200) with 95% CI: 85.6 - 94.2%. Adjusted specificity: 95.0% (191/201) with 95% CI: 91.0 - 97.6%.
Serological Sensitivity (PRNT Confirmed WNV)Demonstrate high sensitivity for WNV PRNT confirmed positive specimens.79.0% (79/100) with 95% CI: 69.7 – 86.5%. Adjusted sensitivity: 88.9% (88/99) with 95% CI: 81.0 – 94.3%.
Study Site 2 (Utah, USA)
Positive Presumptive Agreement (Encephalitis/Meningitis Patients, IgG IFA Pos)Demonstrate agreement with IFA positive samples from symptomatic patients.81.3% (26/32) with 95% CI: 63.6 – 92.8%.
Negative Presumptive Agreement (Encephalitis/Meningitis Patients, IgG IFA Neg)Demonstrate agreement with IFA negative samples from symptomatic patients.100.0% (2/2) with 95% CI: 15.8 – 100.0%.
Positive Presumptive Agreement (WNV IFA Positive)Demonstrate agreement with all WNV positive samples confirmed by IFA.88.0% (286/325) with 95% CI: 84.5 - 91.5%.
Negative Presumptive Agreement (WNV IFA Negative)Demonstrate agreement with all WNV negative samples confirmed by IFA.88.1% (140/159) with 95% CI: 83.0 - 93.1%.
Study Site 3 (Ohio, USA)
Negative Presumptive Agreement (Endemic Normal)Demonstrate high agreement with endemic normal specimens not known to have arbovirus infection and negative by IFA.92.3% (180/195) with 95% CI: 87.6 - 95.6%.
Clinical Sensitivity (PRNT) (Encephalitis/Meningitis Patients, PRNT & IgG IFA Pos)Demonstrate sensitivity for symptomatic patients confirmed by PRNT and IFA.80.4% (41/51) with 95% CI: 66.9 - 90.2%.
Positive Presumptive Agreement (Encephalitis/Meningitis Patients, IgG IFA Pos)Demonstrate agreement with symptomatic patients confirmed by IFA.76.3% (29/38) with 95% CI: 59.8 - 88.6%.
ReproducibilityDemonstrate acceptable precision (low variability) across runs, days, and sites for reactive and negative samples.Reactive Sample: Total CV 4.2% (SD 0.24). Negative Sample: Total CV 38.8% (SD 0.07). (Values for other samples ranged from 9.0% to 41.0% total CV).
Cross-ReactivityDemonstrate low cross-reactivity with antibodies to other diseases, especially other flaviviruses.Dengue: 14/15 positive or equivocal (high cross-reactivity). St. Louis encephalitis: 28/35 positive or equivocal (high cross-reactivity). Japanese encephalitis: 3/3 positive (high cross-reactivity). La Crosse encephalitis: 2/26 positive (low cross-reactivity). California encephalitis: 1/11 positive or equivocal (low cross-reactivity). Eastern Equine encephalitis: 0/1 positive. Varicella-Zoster, Cytomegalovirus, Epstein-Barr, Enterovirus, Rheumatoid Factor, Anti-Nuclear Antibody: Showed varying levels of observed reactivity, which was generally low to moderate for most non-flavivirus agents. Ross River virus: 11/39 positive or equivocal (flavivirus positive confirmed by Western blot for 11/11 reactive samples). Barmah Forest virus: 13/36 positive or equivocal (flavivirus positive confirmed by Western blot for 9/13 reactive samples).

2. Sample Sizes and Data Provenance for Test Sets

  • Study Site 1 (Louisiana, USA):

    • Sample Size: 300 retrospective sera.
    • Data Provenance: Louisiana, USA. Retrospective samples collected in 2002-2003.
  • Study Site 2 (Utah, USA):

    • Sample Size: 325 retrospective sera.
    • Data Provenance: Utah, USA. Retrospective samples.
  • Study Site 3 (Ohio, USA):

    • Sample Size: 284 retrospective sera.
    • Data Provenance: Ohio, USA. Retrospective samples collected in 2002.
  • Reproducibility Study:

    • Sample Size: 8 sera, tested 3 times each on 3 different assays (total 72 tests per parameter).
    • Data Provenance: One Australian study site and two study sites in the USA.
  • Cross-Reactivity Study:

    • Sample Size: 314 specimens in total, with varying numbers per disease state.
    • Data Provenance: Multiple sites including PANBIO (Site 5), a state health laboratory in Louisiana (Site 1), a private reference laboratory in Utah (Site 2), a hospital laboratory in Ohio (Site 3), and a private research laboratory in Maryland (Site 6). The exact timing of collection for these cross-reactive samples is not specified beyond being "from patients with confirmed diseases other than WNV."

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the "number of experts" or their specific "qualifications" (e.g., "radiologist with 10 years of experience") for establishing the ground truth. However, the ground truth was established by:

  • Plaque Reduction Neutralization Test (PRNT): This is a gold standard serological test, implying expert virological laboratory practices and interpretation.
  • West Nile Virus IgG IFA (Immunofluorescence Assay): This is another standard serological method, likely performed and interpreted by trained laboratory personnel or experts.
  • Clinical and serological characterization: This suggests a combination of clinical diagnosis and laboratory test results, presumably by medical professionals and laboratory specialists.
  • Western blot analysis: Used for confirmation of flavivirus positivity in cross-reactivity samples, indicating expert serological techniques.

4. Adjudication Method for the Test Set

The document does not describe a formal "adjudication method" involving multiple human readers or a consensus process for discrepant results in the context of device performance evaluation (e.g., 2+1, 3+1 for image interpretation). Instead, the comparison is directly between the PANBIO ELISA results and the established "gold standard" or "characterization" of the specimens (PRNT, IgG IFA, or clinical characterization).

Equivocal results from the PANBIO ELISA were generally "not retested" due to sample unavailability or because the cut-off was modified post-clinical trials, indicating that a formal re-adjudication process for equivocal outcomes was not systematically applied within these studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The device is an in-vitro diagnostic (IVD) assay designed for laboratory testing, not an imaging device requiring human-in-the-loop interpretation. Therefore, the concept of human readers improving with or without AI assistance is not applicable to this device.

6. Standalone Performance

Yes, a standalone performance study was done. The entire evaluation presented (sensitivity, specificity, presumptive agreements, reproducibility, and cross-reactivity) describes the performance of the PANBIO West Nile Virus IgG Indirect ELISA algorithm/assay itself, without human-in-the-loop assistance in its operation or primary interpretation. The results are based on how the assay performs when processing samples and generating qualitative results (Positive, Equivocal, Negative).

7. Type of Ground Truth Used

The ground truth used for the test sets included:

  • Expert Consensus/Reference Methods: Primarily, Plaque Reduction Neutralization Test (PRNT) and West Nile Virus IgG Immunofluorescence Assay (IFA) were used as reference methods to characterize samples as positive or negative for West Nile Virus IgG. These are established laboratory methods considered reliable.
  • Clinical Characterization: Samples from patients with "clinical symptoms consistent with encephalitis/meningitis" were also used, implying correlation with patient presentation alongside serological markers.
  • Disease State Confirmation: For the cross-reactivity study, samples were characterized by their "disease state" (e.g., Dengue virus, St. Louis encephalitis, Cytomegalovirus) and for some flaviviruses, confirmed by Western blot analysis.

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" or "validation set" in the context of machine learning. This is an IVD assay, and the "training" of the assay's cut-off or parameters would typically involve internal development studies and optimization rather than a distinct "training set" like in AI/ML contexts. The performance characteristics described are based on testing against retrospectively collected "test sets" or panels of samples.

9. How the Ground Truth for the Training Set was Established

Since a distinct "training set" for an AI/ML algorithm is not described, the method for establishing its ground truth is not applicable in this context. The assay's performance evaluation relies on comparing its output to established reference methods (PRNT, IFA, Western blot) and clinical characterization using the "test sets" described above. The statement that "cut-off was modified following clinical trials" in Study Site 2 hints at an iterative process of optimizing the assay's interpretive criteria based on initial performance data.

{0}------------------------------------------------

OCT 2 0 2004

1.9 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K041068

Applicant Information:

Submission Date:21st April, 2004
Date Modified:15th October, 2004
Name:PANBIO Limited
Address:116 Lutwyche Road, WindsorQueensland 4030 Australia
Contact Person:Phone Number:Fax Number:Craig Breadmore / Jillianne Keller+61-(0)7-3357-1177+61-(0)7-3357-1222

Device Information:

Trade Name:West Nile Virus IgG Indirect ELISA
Common Name:West Nile Virus IgG Indirect EIA Test
Classification Name:West Nile virus, serological reagents (21 CFR 866.3940)

Equivalent Device:

No marketed EIA device was available at the time that performance studies were conducted. Equivalence shall be demonstrated through gold standard test reference methods. The West Nile Virus IgG Indirect ELISA is a new device that falls under the FDA's classification name "West Nile virus, serological reagents" (21 CFR 866.3940).

Device Description:

The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the presumptive laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis.

Intended Use:

The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative presumptive detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis. Positive results must be confirmed by Plaque Reduction Neutralization Test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagnosis of this disease.

Assay performance characteristics have not been established for testing cord blood, neonate, prenatal screening, general population screening without symptoms of meningioencephalitis or automated instruments. The user is responsible for establishing these assay performance characteristics.

Principle of Procedure:

Serum antibodies combine with purified and inactivated WNV antigen coated on the polystyrene surface of the microwell test strips (assay plate). Residual serum is removed from the assay plate by washing. HRP-conjugated anti-human IgG monoclonal antibody (Mab) is added to the assay plate. After incubation, the microwells are washed and a colorless substrate system, tetramethylbenzidine/hydrogen peroxide (TMB/H₂O₂), is added. The substrate is hydrolyzed by the HRP, if present, and the chromogen changes to a blue color. After stopping the reaction with acid, the TMB becomes yellow. Color development is indicative of the presence of WNV antibodies in the test sample.

{1}------------------------------------------------

PERFORMANCE CHARACTERISTICS Study Site 1:

Three hundred (300) retrospective sera from individuals of various ages and both genders were tested at a state health laboratory in Louisiana, USA. The sera include samples from the following groups: 100 samples characterised as positive for WN virus by PRNT and 200 randomly selected normal specimens from routine laboratory testing not known to have a flavivirus related illness. These samples were masked and tested on the PANBIO West Nile Virus IgG Indirect ELISA and the results were compared to the clinical and serological characterisation of the samples to determine performance of the assay. The data is summarised in Table 1.

Table 1 - Study Site 1 PANBIO West Nile Virus IgG Indirect ELISA Reactivity with Endemic Normal and WNV PRNT Confirmed Specimens

Specimen CharacterisationPANBIO IgG ELISA ResultsPosEqv aNegTotal
Endemic normal specimens(randomly selected) b18 d1181200
West Nile virus Positive c(PRNT confirmed)79021 e100
Total971202300

a Retesting of equivocals was not conducted as samples were unavailable.

b Randomly selected normal specimens from routine laboratory testing from 2002-2003. Not known to have a flavivirus related illness. Further tested by WNV IgG IFA.

& West Nile virus PRNT positive. Collected in 2002. Further tested by WNV IgG IFA.

8 Nine of the 18 endemic specimens that were positive by PANBIÓ ELISA were also positive by IgG IFA. Refer to note below.

e Ten of the 21 PRNT confirmed positive specimens that were negative by PANBIO ELISA were negative by IgG IFA and positive by PANBIO WNV IgM Capture ELISA.

95% Confidence Interval

Endemic normal specimens Negative presumptive agreement = 181/200 90.5% 85.6 - 94.2% Adjusted specificity ' 91.0 - 97.6% = 191/201 95.0%

West Nile virus positive specimens

Serological sensitivity (PRNT)= 79/100 79.0%69.7 – 86.5%
Adjusted sensitivityf= 88/99 88.9%81.0 – 94.3%

Adjusted specificity and sensitivity calculations have been presented to reflect the presumptive redefinition of the specimens identified in " and " above.

Adjusted specificity: (9 endemic / IgG IFA positive specimens excluded and 10 PRNT positive / IgM ELISA positive / IgG IFA negative specimens included).

Adjusted sensitivity: (9 endemic / IgG IFA positive specimens included and 10 PRNT positive / IgM ELISA positive / IgG IFA negative specimens excluded).

{2}------------------------------------------------

Study Site 2:

Three hundred and twenty-five (325) retrospective sera from individuals of various ages and both genders were tested at a private reference laboratory in Utah, USA. The serum panel was comprised of 166 samples that were characterized positive and 159 samples that were characterized negative for WNV by IFA slides (ASR). Thirty-four of these samples were from patients with clinical symptoms consistent with encephalitis / meningitis, of which 32 were characterized positive and 2 were characterized negative for WNV by IFA. The samples were not masked and were tested by the PANBIO West Nile Virus IgG Indirect ELISA. Assay performance was determined by comparing PANBIO West Nile Virus IgG Indirect ELISA results with the clinical and serological characterization of the samples. The data is summarised in Table 2 and Table 3.

Table 2 - Study Site 2 PANBIO West Nile Virus IgG Indirect ELISA Reactivity with Encephalitis / Meningitis Patients

SpecimenPANBIO IgG ELISA Results
CharacterizationPosEqvaNegTotal
Encephalitis/ meningitispatients (IgG IFA positive)262432
Encephalitis/meningitispatients (IgG IFA negative)0022
Total262634

8 Retesting of equivocals was not conducted as cut-off was modified following clinical trials.

Encephalitic symptoms (IgG IFA positive)95% Confidence Interval
Positive Presumptive
Agreement= 26/3281.3%63.6 – 92.8%
WNV (presumptive by (lgG IFA negative ))
Negative Presumptive
Agreement=2/2100.0%15.8 – 100.0%

Table 3 - Study Site 2 PANBIO West Nile Virus IgG Indirect ELISA Reactivity with WNV IFA Characterized Specimens

Specimen CharacterizationPANBIO IgG ELISA Results
PosEqvaNegTotal
WNV positive(IgG IFA positive)146515166
WNV negative(IgG IFA negative)154140159
Total1619155325

8 Retesting of equivocals was not conducted as cut-off was modified following clinical trials.

{3}------------------------------------------------

WNV IFA positive (presumptive)

Positive Presumptive
Agreement= 286/325 88.0%84.5 - 91.5%
Negative Presumptive
Agreement= 140/159 88.1%83.0 - 93.1%

Study Site 3:

Two-hundred and eighty-four (284) retrospective sera collected in 2002 from individuals of various ages and both genders were tested at a hospital laboratory in Ohio, USA. The sera included specimens from 89 patients with symptoms of encephalitis / meningitis (51 samples were confirmed positive for WN virus by PRNT and positive for IFA, and 38 samples were positive by IFA), and 195 randomly selected normal specimens from routine laboratory testing, negative for WN virus by IFA. These samples were masked and tested on the PANBIO West Nile Virus IgG Indirect ELISA and the results were compared to the clinical and serological characterisation of the samples to determine the performance of the assay. The data is summarised in Table 4.

Table 4 - Study Site 3 PANBIO West Nile Virus IqG Indirect ELISA Reactivity with Endemic Normal and Encephalitis / Meningitis Patients

Specimen CharacterizationPANBIO IgG ELISA Results
PosEqvaNegTotal
Endemic normal specimens(randomly selected / IgG IFA negative)b105180195
Encephalitis/meningitis patients(PRNT positive & IgG IFA positive)c411951
Encephalitis/meningitis patients(IgG IFA positive)c292738
Total808196284

Retesting of equivocals was not conducted as cut-off was modified following clinical trials.

· Endemic randomly selected specimens not known to have an arbovirus infection. All specimens

(n=195) further tested negative by PANBIO WNV IgG IFA ASR.
^ Encephalitis/meningitis patients. Specimens collected in 2002. Fifty-one samples were tested by WNV PRNT. All specimens (n=89) further tested positive by PANBIO WNV IgG IFA ASR.

95% Confidence Interval

Endemic normal specimens
Negative Presumptive Agreement = 180/19592.3%87.6 - 95.6%
Encephalitis/meningitis patients
Clinical sensitivity (PRNT) = 41/5180.4%66.9 - 90.2%
Positive Presumptive Agreement = 29/3876.3%59.8 - 88.6%

{4}------------------------------------------------

REPRODUCIBILITY

The reproducibility of the PANBIO West Nile Virus IgG Indirect ELISA was determined by testing 8 sera 3 times each on 3 different assays at one Australian study site and two study sites in the USA. Within-run, between day, between site and total precision were estimated by Analysis of Variance (ANOVA) Type II. The results are presented in Table 5 below.

Samplen*MeanWithinBetween DayBetween SiteTotal
*SDCV*SDCV*SDCV*SDCV
Reactive275.830.254.3%0.000.0%0.000.0%0.244.2%
Negative270.190.028.8%0.000.0%0.0845.5%0.0738.8%
#1272.240.167.0%0.125.1%1.0848.3%0.9241.0%
#2272.940.206.9%0.165.5%0.134.3%0.269.0%
#3274.070.327.9%0.000.0%0.5212.8%0.5313.1%
#4271.480.117.7%0.117.6%0.085.6%0.1611.0%
#5271.530.064.2%0.063.9%0.1811.9%0.1711.2%
#6271.230.097.4%0.032.3%0.1613.3%0.1713.5%
#7270.820.056.2%0.021.9%0.1316.0%0.1214.8%
#8270.840.077.9%0.000.0%0.1214.7%0.1214.4%

Table 5 - Reproducibility Data PANBIO West Nile Virus IgG Indirect ELISA Precision Measures - ANOVA Type II (Using Cut-Off Ratio*)

All values are calculated from Ratios *

SD = Standard Deviation; CV = Coefficient of Variation (%)

Notes:

  • Standard deviation results have been rounded to two decimal places for tabulation purposes.
  • The cut-off ratio * is calculated by dividing the sample absorbance by the cut-off value. The cut-off value is calculated by multiplying the average absorbance of the triplicates of the calibrator by the calibration factor. The calibrator cannot be calculated as a true cut-off ratio and therefore has been excluded from this analysis.

{5}------------------------------------------------

CROSS REACTIVITY

This study consisted of a panel of 314 specimens from patients with confirmed diseases other than WNV. The purpose of this study was to establish the analytical specificity of the PANBIO West Nile Virus IgG Indirect ELISA through the analysis of specimens with diseases that have the potential for cross-reactivity. Each of the specimens included in the study was characterised with respect to disease state prior to analysis of the specimens with the PANBIO West Nile Virus IgG Indirect ELISA. Table 6 below provides a summary of specimens in the disease panel outlined in Table 7.

PANBIO West Nile virus IgG Indirect ELISA
Disease StateTotalSpecimensaPANBIO IgG ELISA ResultsPosEqvPos and Eqv
Dengue virus1514014/15
St. Louis encephalitis3525328/35
Japanese encephalitis3303/3
La Crosse encephalitis26202/26
California encephalitis11011/11
Eastern Equine encephalitis1000/1
Varicella-Zoster virus15000/15
Cytomegalovirus48707/48
Epstein-Barr virus40707/40
Enterovirus15101/15
Ross River virus3910111/39b
Barmah Forest virus3610313/36c
Rheumatoid Factor15404/15
Anti-Nuclear Antibody15202/15
Table 6 - Summary of Cross-reactivity Analysis
PANBIO West Nile Virus laG Indirect ELISA

a Sample testing was conducted at PANBIO (Site 5) except as follows:

Site 1: Testing on 25 Saint Louis encephalitis, 9 California encephalitis and 1 Eastern Equine encephalitis specimens was conducted at a state health laboratory in Louisiana, USA,

Site 2: Testing on 5 Dengue virus, 10 Saint Louis encephalitis and 2 California encephalitis specimens was conducted at a private reference laboratory in Utah, USA.

Site 3: Testing on 26 La Crosse encephalitis specimens was conducted at a hospital laboratory in Ohio, USA.

Site 6: Testing on 25 Epstein-Barr virus and 33 Cytomegalovirus specimens was conducted at a private research laboratory in Maryland, USA.

6 11 of 11 samples that were reactive on PANBIO West Nile virus IgG Indirect ELISA were confirmed flavivirus positive by Western blot.

6 9 of 13 samples that were reactive on PANBIO West Nile virus IgG Indirect ELISA were confirmed flavivirus positive by Western blot.

{6}------------------------------------------------

PANBIO West Nile Virus IgG Indirect ELISA
SampleSite *IgG Antibody TypePANBIO E-WNV01GELISA ResultUnitsResult
12Dengue virus25.6P
22Dengue virus100.3P
32Dengue virus5.1N
42Dengue virus102.3P
52Dengue virus82.7P
65Dengue virus96.1P
75Dengue virus103.6P
85Dengue virus105.7P
95Dengue virus114.7P
105Dengue virus126.7P
115Dengue virus102.5P
125Dengue virus123.5P
135Dengue virus116.6P
145Dengue virus112.3P
155Dengue virus118.6P
161Saint Louis encephalitis15.6E
171Saint Louis encephalitis21.4P
181Saint Louis encephalitis4.2N
191Saint Louis encephalitis8.7N
201Saint Louis encephalitis18.2P
211Saint Louis encephalitis28.9P
221Saint Louis encephalitis35.9P
231Saint Louis encephalitis40.5P
241Saint Louis encephalitis1.2N
251Saint Louis encephalitis28.2P
261Saint Louis encephalitis7.8N
271Saint Louis encephalitis27.0P
281Saint Louis encephalitis47.8P
291Saint Louis encephalitis43.0P
301Saint Louis encephalitis57.2P
311Saint Louis encephalitis47.9P
321Saint Louis encephalitis29.4P
331Saint Louis encephalitis31.0P
341Saint Louis encephalitis16.5P
351Saint Louis encephalitis43.1P
361Saint Louis encephalitis68.5P
371Saint Louis encephalitis59.5P
381Saint Louis encephalitis10.0N
Sample Site *IgG Antibody TypePANBIO E-WNV01GELISA Result
UnitsResult
39Saint Louis encephalitis13.4N
40Saint Louis encephalitis54.6P
41Saint Louis encephalitis15.4E
42Saint Louis encephalitis15.7E
43Saint Louis encephalitis41.6P
44Saint Louis encephalitis6.0N
45Saint Louis encephalitis73.6P
46Saint Louis encephalitis32.6P
47Saint Louis encephalitis79.1P
48Saint Louis encephalitis62.8P
49Saint Louis encephalitis32.8P
50Saint Louis encephalitis25.3P
51Japanese encephalitis66.0P
52Japanese encephalitis41.6P
53Japanese encephalitis103.9P
54La Crosse encephalitis2.1N
55La Crosse encephalitis1.8N
56La Crosse encephalitis3.6N
57La Crosse encephalitis2.1N
58La Crosse encephalitis16.5P
59La Crosse encephalitis4.4N
60La Crosse encephalitis5.8N
61La Crosse encephalitis2.5N
62La Crosse encephalitis1.9N
63La Crosse encephalitis3.2N
64La Crosse encephalitis2.2N
65La Crosse encephalitis2.2N
66La Crosse encephalitis2.8N
67La Crosse encephalitis2.2N
68La Crosse encephalitis1.3N
69La Crosse encephalitis3N
70La Crosse encephalitis5.5N
71La Crosse encephalitis6.4N
72La Crosse encephalitis3.6N
73La Crosse encephalitis3.6N
74La Crosse encephalitis13.2N
75La Crosse encephalitis1.7N
76La Crosse encephalitis5.4N
SampleSite *IgG Antibody TypePANBIO E-WNV01GELISA Result
UnitsResult
773La Crosse encephalitis2.5N
783La Crosse encephalitis6N
793La Crosse encephalitis16.8P
801California encephalitis6.1N
811California encephalitis11.5N
821California encephalitis14.1E
831California encephalitis1.5N
841California encephalitis5.4N
851California encephalitis8.5N
861California encephalitis6.2N
871California encephalitis4N
881California encephalitis4.7N
892California encephalitis13.2N
902California encephalitis4.4N
911Eastern Equine enceph.3.6N
925Varicella-Zoster virus3.7N
935Varicella-Zoster virus2.4N
945Varicella-Zoster virus5.7N
955Varicella-Zoster virus5.3N
965Varicella-Zoster virus3.8N
975Varicella-Zoster virus7.7N
985Varicella-Zoster virus2.6N
995Varicella-Zoster virus3.8N
1005Varicella-Zoster virus10.0N
1015Varicella-Zoster virus1.2N
1025Varicella-Zoster virus2.2N
1035Varicella-Zoster virus2.5N
1045Varicella-Zoster virus6.7N
1055Varicella-Zoster virus2.6N
1065Varicella-Zoster virus3.9N
1075Cytomegalovirus4.6N
1085Cytomegalovirus1.5N
1095Cytomegalovirus1.9N
1105Cytomegalovirus1.8N
1115Cytomegalovirus16.7P
1125Cytomegalovirus4.4N
1135Cytomegalovirus1.9N
1145Cytomegalovirus44.5P
SampleSite *IgG Antibody TypePANBIO E-WNV01GELISA Result
UnitsResult
1155Cytomegalovirus69.4P
1165Cytomegalovirus5.2N
1175Cytomegalovirus3.8N
1185Cytomegalovirus46.9P
1195Cytomegalovirus16.2P
1205Cytomegalovirus3.3N
1215Cytomegalovirus31.0P
1226Cytomegalovirus11.9N
1236Cytomegalovirus0.5N
1246Cytomegalovirus0.3N
1256Cytomegalovirus3.7N
1266Cytomegalovirus1.3N
1276Cytomegalovirus0.6N
1286Cytomegalovirus2.9N
1296Cytomegalovirus1.1N
1306Cytomegalovirus2.1N
1316Cytomegalovirus4.4N
1326Cytomegalovirus4.1N
1336Cytomegalovirus1.4N
1346Cytomegalovirus0.4N
1356Cytomegalovirus7.8N
1366Cytomegalovirus4.9N
1376Cytomegalovirus11.6N
1386Cytomegalovirus20.4P
1396Cytomegalovirus7.5N
1406Cytomegalovirus0.8N
1416Cytomegalovirus2.2N
1426Cytomegalovirus12.1N
1436Cytomegalovirus4.4N
1446Cytomegalovirus9.8N
1456Cytomegalovirus1.3N
1466Cytomegalovirus0.5N
1476Cytomegalovirus2.8N
1486Cytomegalovirus1.3N
1496Cytomegalovirus1.7N
1506Cytomegalovirus0.4N
1516Cytomegalovirus5.5N
1526Cytomegalovirus3.4N
SampleSite *IgG Antibody TypePANBIO E-WNV01G
ELISA Result
UnitsResult
1536Cytomegalovirus0.6N
1546Cytomegalovirus6.2N
1555Epstein-Barr virus3.3N
1565Epstein-Barr virus2.0N
1575Epstein-Barr virus3.8N
1585Epstein-Barr virus2.4N
1595Epstein-Barr virus18.0P
1605Epstein-Barr virus5.5N
1615Epstein-Barr virus1.9N
1625Epstein-Barr virus19.9P
1635Epstein-Barr virus2.3N
1645Epstein-Barr virus6.7N
1655Epstein-Barr virus3.4N
1665Epstein-Barr virus17.9P
1675Epstein-Barr virus6.6N
1685Epstein-Barr virus134.2P
1695Epstein-Barr virus3.2N
1706Epstein-Barr virus10.2N
1716Epstein-Barr virus4.6N
1726Epstein-Barr virus5.8N
1736Epstein-Barr virus5.8N
1746Epstein-Barr virus17.4P
1756Epstein-Barr virus7.4N
1766Epstein-Barr virus5.1N
1776Epstein-Barr virus5.5N
1786Epstein-Barr virus29.0P
1796Epstein-Barr virus6.5N
1806Epstein-Barr virus5.6N
1816Epstein-Barr virus3.4N
1826Epstein-Barr virus4.0N
1836Epstein-Barr virus1.9N
1846Epstein-Barr virus50.0P
1856Epstein-Barr virus2.8N
1866Epstein-Barr virus12.3N
1876Epstein-Barr virus6.0N
1886Epstein-Barr virus8.9N
1896Epstein-Barr virus2.0N
1906Epstein-Barr virus3.3N
SampleSite *IgG Antibody TypePANBIO E-WNV01G
ELISA Result
UnitsResult
1916Epstein-Barr virus12.3N
1926Epstein-Barr virus13.9N
1936Epstein-Barr virus4.8N
1946Epstein-Barr virus6.0N
1955Enterovirus2.6N
1965Enterovirus7.0N
1975Enterovirus1.3N
1985Enterovirus4.7N
1995Enterovirus1.7N
2005Enterovirus6.1N
2015Enterovirus2.9N
2025Enterovirus1.8N
2035Enterovirus2.6N
2045Enterovirus1.2N
2055Enterovirus1.6N
2065Enterovirus3.5N
2075Enterovirus17.7P
2085Enterovirus2.8N
2095Enterovirus8.5N
2105Ross River virus95.8Pᵃ
2115Ross River virus12.5N
2125Ross River virus22.6Pᵃ
2135Ross River virus4.3N
2145Ross River virus8.6N
2155Ross River virus3.2N
2165Ross River virus6.2N
2175Ross River virus5.8N
2185Ross River virus30.4Pᵃ
2195Ross River virus7.3N
2205Ross River virus3.0N
2215Ross River virus4.4N
2225Ross River virus17.9Pᵃ
2235Ross River virus12.7N
2245Ross River virus4.9N
2255Ross River virus19.7Pᵃ
2265Ross River virus5.6N
2275Ross River virus5.0N
2285Ross River virus7.1N
PANBIO E-WNV01G
SampleSite *IgG Antibody TypeELISA Result
UnitsResult
2295Ross River virus38.9Pa
2305Ross River virus12.3N
2315Ross River virus6.5N
2325Ross River virus15.8Ea
2335Ross River virus9.3N
2345Ross River virus16.3Pa
2355Ross River virus12.6N
2365Ross River virus6.1N
2375Ross River virus4.8N
2385Ross River virus5.4N
2395Ross River virus8.1N
2405Ross River virus7.1N
2415Ross River virus16.1Pa
2425Ross River virus8.1N
2435Ross River virus7.9N
2445Ross River virus6.8N
2455Ross River virus60.8Pa
2465Ross River virus10.1N
2475Ross River virus6.9N
2485Ross River virus17.5Pa
2495Barmah forest virus5.9N
2505Barmah forest virus8.1N
2515Barmah forest virus15.7Ea
2525Barmah forest virus9.7N
2535Barmah forest virus11.6N
2545Barmah forest virus5.8N
2555Barmah forest virus4.6N
2565Barmah forest virus4.9N
2575Barmah forest virus7.5N
2585Barmah forest virus2.5N
2595Barmah forest virus12.5N
2605Barmah forest virus34.9Pa
2615Barmah forest virus8.0N
2625Barmah forest virus98.0Pa
2635Barmah forest virus9.3N
2645Barmah forest virus10.9N
2655Barmah forest virus10.6N
2665Barmah forest virus8.8N
SampleSite *IgG Antibody TypePANBIO E-WNV01GELISA Result
UnitsResult
2675Barmah forest virus55.2Pa
2685Barmah forest virus6.4N
2695Barmah forest virus8.8N
2705Barmah forest virus8.8N
2715Barmah forest virus52.4Pa
2725Barmah forest virus10.4N
2735Barmah forest virus19.5Pa
2745Barmah forest virus6.2N
2755Barmah forest virus9.0N
2765Barmah forest virus5.5N
2775Barmah forest virus26.0P
2785Barmah forest virus27.6Pa
2795Barmah forest virus14.2E
2805Barmah forest virus34.3P
2815Barmah forest virus6.2N
2825Barmah forest virus79.1Pa
2835Barmah forest virus49.7P
2845Barmah forest virus14.4Ea
2855Rheumatoid Factor16.3P
2865Rheumatoid Factor2.5N
2875Rheumatoid Factor6.4N
2885Rheumatoid Factor5.0N
2895Rheumatoid Factor21.3P
2905Rheumatoid Factor1.4N
2915Rheumatoid Factor3.6N
2925Rheumatoid Factor3.9N
2935Rheumatoid Factor7.1N
2945Rheumatoid Factor4.6N
2955Rheumatoid Factor45.2P
2965Rheumatoid Factor7.5N
2975Rheumatoid Factor36.5P
2985Rheumatoid Factor4.5N
2995Rheumatoid Factor4.5N
3005Anti-Nuclear Antibody2.88N
3015Anti-Nuclear Antibody1.94N
3025Anti-Nuclear Antibody1.28N
3035Anti-Nuclear Antibody7.04N
3045Anti-Nuclear Antibody4.07N
SampleSite *IgG Antibody TypePANBIO E-WNV01GELISA Result
UnitsResult
3055Anti-Nuclear Antibody5.63N
3065Anti-Nuclear Antibody13.33N
3075Anti-Nuclear Antibody8.2N
3085Anti-Nuclear Antibody48.9P
3095Anti-Nuclear Antibody2.6N
3105Anti-Nuclear Antibody7.2N
3115Anti-Nuclear Antibody3.5N
3125Anti-Nuclear Antibody6.1N
3135Anti-Nuclear Antibody50.0P
3145Anti-Nuclear Antibody2.1N

Table 7 – Results of Cross-reactivity Analysis

PANBIO West Nile Virus IgG Indirect ELISA

{7}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

{8}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

{9}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

{10}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

{11}------------------------------------------------

Table 7 -- Results of Cross-reactivity Analysis -- Continued

{12}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

{13}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

{14}------------------------------------------------

Table 7 – Results of Cross-reactivity Analysis – Continued

a Flavivirus reactivity in these samples has been confirmed by Western blot analysis.

INTERPRETATION

AN R A. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. BBuch BandadeCHANNELE AND LE CHANNEL CALLA BROND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AN
· NDIA ·MANIE------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company of a comments of a collection and1A COLORIO COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTIONﻟ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Canadian Children Children Children Children Children Children Children CharlesWHERE ANNUAL A COLL L A & A
  • Refer to summary table for explanation of sites.

.

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle, there is a stylized symbol that resembles three parallel lines curving upwards, often interpreted as representing human aspiration and progress.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 2 0 2004

Ms. Jillianne Keller Regulatory Affairs Officer PANBIO Limited. 116 Lutwyche Road, Windsor Brisbane, Queensland 4030 Australia

K041068 Re:

Trade/Device Name: West Nile Virus IgG Indirect ELISA Regulation Number: 21 CFR 866.3940 Regulation Name: West Nile Virus Serological Reagents Regulatory Class: Class II Product Code: NOP Dated: September 17, 2004 Received: September 20, 2004

Dear Ms. Keller:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to conniner of the 112) 2011-12-11 in accordance with the provisions of the Federal Food, Drug, de vices may been recial of require approval of a premarket approval application (PMA). and Cosmetic rear (110) market the device, subject to the general controls provisions of the Act. The I ou may, dicrere, mance of the Act include requirements for annual registration, listing of general controls provisions and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may be subject to back adon additions (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be advised that i Drimination that your device complies with other requirements of the Act that I DIT has made a and regulations administered by other Federal agencies. You must or any it cach statutes and regarments, including, but not limited to: registration and listing (21 compry with an the Tre (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{16}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Saartys

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{17}------------------------------------------------

Indications for Use

20 October, 2004

510(k) Number (if known):_K041068

Device Name: West Nile Virus IgG Indirect ELISA

Indications For Use:

The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative presumptive detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis. Positive results must be confirmed by plaque reduction neutralization test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagnosis of this disease.

Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Page 1 of 1

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)________________________________________________________________________________________________________________________________________________________________________ KOFIDER

§ 866.3940 West Nile virus serological reagents.

(a)
Identification. West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.(b)
Classification. Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.