K Number
K040984

Validate with FDA (Live)

Date Cleared
2004-04-27

(12 days)

Product Code
Regulation Number
874.4760
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VSI Model ENT-2000V Flexible Nasopharyngo-Laryngoscope is intended for flexible endoscopic examination of the upper airway, vocal chords and/or nasal passages.

Device Description

The device described in this 510(k) consists of modified flexible fiberoptic ENT scope. The scope has been modified to include an integral camera module and camera control unit.

AI/ML Overview

This 510(k) summary (K040984) describes a Special 510(k) for a modified flexible fiberoptic ENT scope, the Vision Sciences Model ENT-2000V Flexible Nasopharyngo-Laryngoscope. A Special 510(k) is used for modifications to a manufacturer’s own legally marketed device where the modification does not alter the indications for use or the fundamental scientific technology of the device. Consequently, the focus of the submission is on design controls, risk analysis, and verification/validation testing demonstrating that the device continues to meet its performance specifications and is as safe and effective as the predicate device.

Here's an analysis of the provided text in relation to your request:

1. A table of acceptance criteria and the reported device performance

The provided text does not contain a specific table detailing "acceptance criteria" alongside "reported device performance" in a quantitative manner as one might find in a study specifically comparing device output to a clinical ground truth.

However, the "Safety and Performance" section states:
"Validation testing including scope image quality evaluation, focal length, resolution, distortion, reprocessing effects and electrical safety testing is included in Design Validation and Verification planning."

This indicates that these parameters (image quality, focal length, resolution, distortion, reprocessing effects, and electrical safety) likely served as the acceptance criteria for the device modification. The implication is that the device met these criteria, as the FDA granted clearance. Without explicit numerical targets and results, a direct table cannot be constructed.

Inferred Acceptance Criteria Example (based on device type and typical requirements):

Acceptance Criteria CategoryReported Device Performance
Image QualityMet predefined specifications for clarity, brightness, and color rendition. (Implied: as good as or better than predicate, maintaining diagnostic capability)
Focal LengthMaintained within specified range typical for flexible ENT scopes. (Implied: met engineering specifications)
ResolutionMet predefined specifications for distinguishing fine details. (Implied: as good as or better than predicate)
DistortionMaintained within acceptable limits. (Implied: as good as or better than predicate)
Reprocessing EffectsDemonstrated continued functionality and integrity after specified reprocessing cycles. (Implied: met sterilization and durability standards)
Electrical SafetyComplied with relevant electrical safety standards (e.g., IEC 60601). (Implied: passed all required tests)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "test set" in the context of clinical data for AI or an imaging study. This is a special 510(k) focused on device modification and engineering validation. The validation testing mentioned (image quality, focal length, etc.) would typically involve the modified device itself undergoing technical assessments, not a collection of patient data.

Therefore, information on sample size for a test set or data provenance (country of origin, retrospective/prospective) is not applicable in the context of this specific 510(k).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Similar to point 2, the concept of "ground truth" established by experts for a test set is not present in this submission. The validation testing likely involved engineers and quality control personnel assessing the physical and imaging characteristics of the device against established technical specifications, rather than clinical experts interpreting images for diagnostic accuracy.

Therefore, information on the number and qualifications of experts for ground truth establishment for a test set is not applicable.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As there is no described test set involving human interpretation of clinical data in this 510(k), an adjudication method is not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed or is mentioned. This submission is for a medical device (endoscope), not an AI diagnostic algorithm. Therefore, "human readers improve with AI vs without AI assistance" is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This submission is for a medical device (endoscope), not a standalone algorithm. Therefore, this question is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context refers to engineering and performance specifications. For example:

  • Image Quality: Measured against objective metrics (e.g., contrast, sharpness, resolution targets) and potentially subjective assessment by trained personnel (e.g., optical engineers, quality control specialists) using pre-defined criteria.
  • Focal Length, Resolution, Distortion: Measured objectively using calibrated test equipment and optical benches against manufacturing specifications.
  • Reprocessing Effects: Evaluated by subjecting the scope to specified reprocessing cycles and then testing its performance and integrity (e.g., visual inspection, leak testing, functional checks).
  • Electrical Safety: Tested according to recognized electrical safety standards (e.g., IEC 60601) by accredited labs/personnel.

So, the "ground truth" used was engineering and performance specifications and objective measurements.

8. The sample size for the training set

This is a physical medical device; there is no "training set" in the context of machine learning or AI. Therefore, this question is not applicable.

9. How the ground truth for the training set was established

As there is no training set, this question is not applicable.

{0}------------------------------------------------

Ko40984

510(k) Summary

Trade Name:Vision Sciences Model ENT-2000V Flexible Nasopharyngo-Laryngoscope
Sponsor:Vision-Sciences, Inc.9 Strathmore RoadNatick, MA 01760Registration #1223490
Device Generic Name:Flexible ENT scopes
Classification:According to Section 513 of the Federal Food, Drug, andCosmetic Act, the device classification is Class II.
Predicate Devices:K942265 - Flexible ENT ScopeK990354 - Modified EndoSheath® for Flexible ENT ScopesManufactured by:Vision-Sciences, Inc.9 Strathmore RoadNatick, MA 01760

Product Description: The device described in this 510(k) consists of modified flexible fiberoptic ENT scope. The scope has been modified to include an integral camera module and camera control unit.

Indications for Use:

The scope is indicated for use during flexible endoscopic examination of the upper airway, vocal chords and/or nasal passages.

Safety and Performance:

This submission is a Special 510(k): Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), Vision-Sciences has provided certification of compliance to 21 CFR 820.30 Design Control requirements, and a description of the internal Risk Analysis procedure. Validation testing including scope image quality evaluation, focal length, resolution, distortion, reprocessing effects and electrical safety testing is included in Design Validation and Verification planning.

Conclusion:

Based on the indications for use, technological characteristics, and comparison to predicate devices, the modified VSI Flexible ENT Scope has been shown to be safe and effective for its intended use.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circle around the eagle. The eagle is facing to the right. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 7 2004

Vision-Sciences, Inc. c/o Pamela Papineau, RAC Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer. MA 01432

Re: K040984

Trade/Device Name: Vision Sciences Model ENT-2000V Flexible Nasopharyngo-Layngoscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories. Regulatory Class: Class II Product Code: EOB Dated: April 8, 2004

Received: April 15, 2004

Dear Ms. Papineau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Pamela Papineau, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Ahelyl Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1

510(k) Number (if known): _K040984

Device Name: Vision Sciences Model ENT-2000V Flexible Nasopharyngo-Laryngoscope

Indications for Use:

The VSI Model ENT-2000V Flexible Nasopharyngo-Laryngoscope is intended for flexible endoscopic examination of the upper airway, vocal chords and/or nasal passages.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseXAND/OR
(Part 21 CFR 801 Subpart D)

(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises

510{k) Number

Over-the -Counter Use (21 CFR 807 Subpart D)

§ 874.4760 Nasopharyngoscope (flexible or rigid) and accessories.

(a)
Identification. A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.